← Product Code [LZA](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA) · K234097

# Disposable Powder Free Nitrile Examination Glove, Natural Blue Color, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (K234097)

_Ever Growth (Vietnam) Corporation Limited · LZA · Mar 27, 2024 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA/K234097

## Device Facts

- **Applicant:** Ever Growth (Vietnam) Corporation Limited
- **Product Code:** [LZA](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA.md)
- **Decision Date:** Mar 27, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. In addition, these gloves were tested respectively for use with chemotherapy drugs and Fentanyl citrate in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

## Device Story

Disposable, non-sterile, ambidextrous, five-fingered nitrile examination glove; made from NBR, accelerator, bulking agent, and pigment; powder-free. Used by healthcare professionals in clinical settings to provide a barrier against contamination. Tested for permeation resistance against 36 chemotherapy drugs and Fentanyl citrate per ASTM D6978-05. Provides protection during patient examination and handling of hazardous substances; helps prevent cross-contamination between patient and examiner.

## Clinical Evidence

No clinical data. Bench testing only, including physical dimensions, tensile strength, elongation, freedom from holes, residual powder, biocompatibility (cytotoxicity, skin sensitization, skin irritation), and chemotherapy/Fentanyl citrate permeation resistance per ASTM D6978-05.

## Technological Characteristics

Nitrile Butadiene Rubber (NBR) construction; powder-free; non-sterile; ambidextrous. Complies with ASTM D6319-19 for physical properties and dimensions. Biocompatibility per ISO 10993-5 and 10993-10. Permeation resistance tested per ASTM D6978-05.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested for Use with Chemotherapy Drugs ([K190736](/device/K190736.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 27, 2024

Ever Growth (Vietnam) Corporation Limited Deng Elizabeth U.S. Agent Long Khanh Industrial Zone, Binh Loc Commune. Long Khanh City, Dong Nai 76461 Vietnam

Re: K234097

Trade/Device Name: Disposable Powder Free Nitrile Examination Glove, Natural Blue Color, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC, OPJ, QDO Dated: December 1, 2023 Received: December 26, 2023

Dear Deng Elizabeth:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

# Bifeng Qian -S

Bifeng Qian, M.D, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

### Indications for Use

510(k) Number (if known) K234097

#### Device Name

Disposable Powder Free Nitrile Examination Glove, Natural Blue Color, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate

#### Indications for Use (Describe)

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. In addition, these gloves were tested respectively for use with chemotherapy drugs and Fentanyl citrate in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

| No. | Test Chemotherapy Drug and Concentration | Minimum Breakthrough Detection Time (Min.) |
|-----|------------------------------------------|--------------------------------------------|
| 1   | Bleomycin Sulfate 15.0 mg/ml             | > 240                                      |
| 2   | Busulfan 6.0 mg/ml                       | > 240                                      |
| 3   | Carboplatin 10.0 mg/ml                   | > 240                                      |
| 4   | Carmustine (BCNU) 3.3 mg/ml              | 59.4                                       |
| 5   | Chloroquine 50.0 mg/ml                   | > 240                                      |
| 6   | Cisplatin 1 mg/ml                        | > 240                                      |
| 7   | Cyclophosphamide 20 mg/ml                | > 240                                      |
| 8   | Cyclosporin A 100.0 mg/ml                | > 240                                      |
| 9   | Cytarabine 100.0 mg/ml                   | > 240                                      |
| 10  | Dacarbazine (DTIC), 10.0 mg/ml           | > 240                                      |
| 11  | Daunorubicin 5.0 mg/ml                   | > 240                                      |
| 12  | Docetaxel 10.0 mg/ml                     | > 240                                      |
| 13  | Doxorubicin Hydrochloride 2.0 mg/ml      | > 240                                      |
| 14  | Epirubicin 2.0 mg/ml                     | > 240                                      |
| 15  | Etoposide (Toposar) 20.0 mg/ml           | > 240                                      |
| 16  | Fludarabine 25.0 mg/ml                   | > 240                                      |
| 17  | Fluorouracil 50.0 mg/ml                  | > 240                                      |
| 18  | Gemcitabine 38.0 mg/ml                   | > 240                                      |
| 19  | Idaribicin 1.0 mg/ml                     | > 240                                      |
| 20  | Ifosfamide 50.0 mg/ml                    | > 240                                      |
| 21  | Irinotecan 20.0 mg/ml                    | > 240                                      |
| 22  | Mechlorethamine HCl 1.0 mg/ml            | > 240                                      |
| 23  | Melphalan 5.0 mg/ml                      | > 240                                      |
| 24  | Methotrexate 25 mg/ml                    | > 240                                      |
| 25  | Mitomycin C 0.5 mg/ml                    | > 240                                      |
| 26  | Mitoxantrone 2.0 mg/ml                   | > 240                                      |
| 27  | Oxaliplatin 2.0 mg/ml                    | > 240                                      |
| 28  | Paclitaxel (Taxol), 6.0 mg/ml            | > 240                                      |
| 29  | Paraplatin 10 mg/ml                      | > 240                                      |
| 30  | Retrovir 10 mg/ml                        | > 240                                      |
| 31  | Rituximab 10 mg/ml                       | > 240                                      |
| 32  | Thiotepa, 10.0 mg/ml                     | 118.5                                      |
| 33  | Topotecan HCl 1.0 mg/ml                  | > 240                                      |
| 34  | Trisonex 1.0 mg/ml                       | > 240                                      |
| 35  | Velcade (Bortezomib) 1.0 mg/ml           | > 240                                      |
| 36  | Vincristine 1.0 mg/ml                    | > 240                                      |

{4}------------------------------------------------

Fentanyl Citrate Injection 100 mcg/2ml

Thiotepa and Carmustine have low permeation times

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

区 Over-The-Counter Use (21 CFR 801 Subpart C)

>240

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

#### 0.0 Summary Preparation Date: March 23, 2024

#### 1.0 Submitter:

Submitter's name:

Submitter's address:

Phone number: Fax number: Name of contact person:

#### 2.0 US Agent:

US Representative Name: Company Address:

Telephone Number: Contact Email Address:

#### 3.0 Name of the Device

Proprietary/Trade name:

Common Name:

Classification Name: Device Classification: Regulation Number: Product Code:

#### 4.0 Predicate device

Device Name:

Company name:

510(K) Number:

EVER GROWTH (VIETNAM) CORPORATION LIMITED Long Khanh Industrial Zone, Binh Loc Commune, Long Khanh City, Dong Nai, VN 76461 84-61-3514022 84-61-3514023 Jerry Lin

Elizabeth Deng 5748 Eaglewood Place Rancho Cucamonga, California Rancho Cucamonga, CA 91739 909 4659188 baxianunited48(@yahoo.com

Disposable Powder Free Nitrile Examination Glove, Natural Blue Color, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate Nitrile Examination Gloves, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate Non-powdered Patient Examination Glove Class I 21 CFR 880.6250 LZA, LZC, OPJ, QDO

Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested for Use with Chemotherapy Drugs EVER GROWTH (VIETNAM) CORPORATION LIMITED K190736

{6}------------------------------------------------

#### 5.0 Device Description:

This device, Disposable Powder Free Nitrile Examination Glove, Natural Blue Color, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate, is a nitrile non-sterile medical examination glove available in various sizes (from extra-small to extra-large) and packaging configuration. The gloves are non-sterile medical examination gloves, seamless, ambidextrous, five fingered. They are made from proprietary formulations composed of NBR (Nitrile Butadiene Rubber), accelerator, bulking agent, and pigment. In addition, no former release powder or chemical is used in its manufacturing process.

#### 6.0 Device Indications for use:

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. In addition, these gloves were tested respectively for use with chemotherapy drugs and Fentanyl citrate in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by

| No. | Test Chemotherapy Drug and Concentration | Minimum Breakthrough Detection Time (Min.)* |
|-----|------------------------------------------|---------------------------------------------|
| 1   | Bleomycin Sulfate 15.0 mg/ml             | > 240                                       |
| 2   | Busulfan 6.0 mg/ml                       | > 240                                       |
| 3   | Carboplatin 10.0 mg/ml                   | > 240                                       |
| 4   | Carmustine (BCNU) 3.3 mg/ml              | 59.4                                        |
| 5   | Chloroquine 50.0 mg/ml                   | > 240                                       |
| 6   | Cisplatin 1 mg/ml                        | > 240                                       |
| 7   | Cyclophosphamide 20 mg/ml                | > 240                                       |
| 8   | Cyclosporin A 100.0 mg/ml                | > 240                                       |
| 9   | Cytarabine 100.0 mg/ml                   | > 240                                       |
| 10  | Dacarbazine (DTIC), 10.0 mg/ml           | > 240                                       |
| 11  | Daunorubicin 5.0 mg/ml                   | > 240                                       |
| 12  | Docetaxel 10.0 mg/ml                     | > 240                                       |
| 13  | Doxorubicin Hydrochloride 2.0 mg/ml      | > 240                                       |
| 14  | Epirubicin 2.0 mg/ml                     | > 240                                       |
| 15  | Etoposide (Toposar), 20.0 mg/ml          | > 240                                       |
| 16  | Fludarabine 25.0 mg/ml                   | > 240                                       |
| 17  | Fluorouracil 50.0 mg/ml                  | > 240                                       |
| 18  | Gemcitabine 38.0 mg/ml                   | > 240                                       |

{7}------------------------------------------------

| No. | Test Chemotherapy Drug and Concentration  | Minimum Breakthrough<br>Detection Time (Min.)* |
|-----|-------------------------------------------|------------------------------------------------|
| 19  | Idaribicin 1.0 mg/ml                      | > 240                                          |
| 20  | Ifosfamide 50.0 mg/ml                     | > 240                                          |
| 21  | Irinotecan 20.0 mg/ml                     | > 240                                          |
| 22  | Mechlorethamine HCl 1.0 mg/ml             | > 240                                          |
| 23  | Melphalan 5.0 mg/ml                       | > 240                                          |
| 24  | Methotrexate 25 mg/ml                     | > 240                                          |
| 25  | Mitomycin C 0.5 mg/ml                     | > 240                                          |
| 26  | Mitoxantrone 2.0 mg/ml                    | > 240                                          |
| 27  | Oxaliplatin 2.0 mg/ml                     | > 240                                          |
| 28  | Paclitaxel (Taxol) 6.0 mg/ml              | > 240                                          |
| 29  | Paraplatin 10 mg/ml                       | > 240                                          |
| 30  | Retrovir 10 mg/ml                         | > 240                                          |
| 31  | Rituximab 10 mg/ml                        | > 240                                          |
| 32  | Thiotepa 10.0 mg/ml                       | 118.5                                          |
| 33  | Topotecan HCl 1.0 mg/ml                   | > 240                                          |
| 34  | Trisonex 1.0 mg/ml                        | > 240                                          |
| 35  | Velcade (Bortezomib) 1.0 mg/ml            | > 240                                          |
| 36  | Vincristine 1.0 mg/ml                     | > 240                                          |
|     | Fentanyl Citrate Test                     |                                                |
|     | Fentanyl Citrate Injection<br>100 mcg/2ml | >240                                           |

*The maximum testing time is 240 minutes

Please note that the following drug has an extremely low permeation time:

| No. | Test Chemotherapy Drug and Concentration | Minimum Breakthrough<br>Detection Time (Min.) |
|-----|------------------------------------------|-----------------------------------------------|
| 4   | Carmustine (BCNU) 3.3 mg/ml              | 59.4                                          |
| 32  | Thiotepa, 10.0 mg/ml                     | 118.5                                         |

{8}------------------------------------------------

#### 7.0 Comparison of device technological characteristics:

| Device<br>Characteristic   | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                          | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                    | Comparison           |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Product Name               | Disposable Powder Free Nitrile<br>Examination Glove, Blue Color,<br>Tested for Use with<br>Chemotherapy Drugs                                                                                                                                                                                                                                                                                                             | Disposable Powder Free Nitrile<br>Examination Glove, Natural<br>Blue Color, Tested for Use with<br>Chemotherapy Drugs and<br>Fentanyl Citrate                                                                                                                                                                                                                                                                                                     | same except<br>color |
| 510(K) No.                 | K190736                                                                                                                                                                                                                                                                                                                                                                                                                   | K234097                                                                                                                                                                                                                                                                                                                                                                                                                                           | n/a                  |
| Product Owner              | Ever Growth (Vietnam)<br>Corporation Limited                                                                                                                                                                                                                                                                                                                                                                              | Ever Growth (Vietnam)<br>Corporation Limited                                                                                                                                                                                                                                                                                                                                                                                                      | same                 |
| Product Code               | LZA & LZC                                                                                                                                                                                                                                                                                                                                                                                                                 | LZA & LZC, OPJ, QDO                                                                                                                                                                                                                                                                                                                                                                                                                               | same                 |
| Regulation                 | 21 CFR 880.6250                                                                                                                                                                                                                                                                                                                                                                                                           | 21 CFR 880.6250                                                                                                                                                                                                                                                                                                                                                                                                                                   | same                 |
| Class                      | I                                                                                                                                                                                                                                                                                                                                                                                                                         | I                                                                                                                                                                                                                                                                                                                                                                                                                                                 | same                 |
| Indications for<br>Use     | A patient examination gloves is a<br>disposable device intended for<br>medical purpose that is worn on<br>the examiner's hand or fingers to<br>prevent contamination between<br>patient and examiner. In<br>addition, these gloves were<br>tested for use with chemotherapy<br>drugs in accordance with ASTM<br>D6978-05 Standard Practice for<br>Assessment of Medical gloves to<br>Permeation by Chemotherapy<br>Drugs. | A patient examination gloves is a<br>disposable device intended for<br>medical purpose that is worn on<br>the examiner's hand or fingers to<br>prevent contamination between<br>patient and examiner. In<br>addition, these gloves were<br>tested for use with<br>chemotherapy drugs and<br>Fentanyl citrate in accordance<br>with ASTM D6978-05 Standard<br>Practice for Assessment of<br>Medical gloves to Permeation<br>by Chemotherapy Drugs. | similar              |
| Power free                 | Yes                                                                                                                                                                                                                                                                                                                                                                                                                       | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                               | same                 |
| Size                       | XS/S/M/L/XL                                                                                                                                                                                                                                                                                                                                                                                                               | XS/S/M/L/XL                                                                                                                                                                                                                                                                                                                                                                                                                                       | same                 |
| Over-The-<br>Counter Use   | Yes (21 CFR 801 Subpart C)                                                                                                                                                                                                                                                                                                                                                                                                | Yes (21 CFR 801 Subpart C)                                                                                                                                                                                                                                                                                                                                                                                                                        | same                 |
| Single Use                 | YES                                                                                                                                                                                                                                                                                                                                                                                                                       | YES                                                                                                                                                                                                                                                                                                                                                                                                                                               | same                 |
| Non-Sterile                | YES                                                                                                                                                                                                                                                                                                                                                                                                                       | YES                                                                                                                                                                                                                                                                                                                                                                                                                                               | same                 |
| Dimensions-<br>Length      | Complies with ASTM D6319-10<br>≥ 230 mm                                                                                                                                                                                                                                                                                                                                                                                   | Complies with ASTM D6319-19<br>XS&S ≥ 220 mm;<br>M&L&XL ≥ 230 mm                                                                                                                                                                                                                                                                                                                                                                                  | similar              |
| Dimensions -<br>Palm Width | Complies with ASTM D6319-10<br>XS: 70 ± 10 mm;<br>S: 80 ± 10 mm;<br>M: 95 ± 10 mm;<br>L: 110 ± 10 mm;<br>XL: 120 ± 10 mm                                                                                                                                                                                                                                                                                                  | Complies with ASTM D6319-19<br>XS: 70 ± 10 mm;<br>S: 80 ± 10 mm;<br>M: 95 ± 10 mm;<br>L: 110 ± 10 mm;<br>XL: 120 ± 10 mm                                                                                                                                                                                                                                                                                                                          | same                 |
| Dimensions -<br>Thickness  | Complies with ASTM D6319-10<br>Palm ≥0.05 mm;<br>Finger ≥0.05 mm                                                                                                                                                                                                                                                                                                                                                          | Complies with ASTM D6319-19<br>Palm ≥0.05 mm;<br>Finger ≥0.05 mm                                                                                                                                                                                                                                                                                                                                                                                  | same                 |

{9}------------------------------------------------

| Device<br>Characteristic                                             | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Subject Device                                                                                                                                                            | Comparison |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |         |
|----------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|---------|
| Physical<br>Properties                                               | Complies with ASTM D6319-<br>10<br>Tensile Strength<br>Before aging≥14MPa<br>After aging ≥14MPa                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Complies with ASTM D6319-<br>19<br>Tensile Strength<br>Before aging≥14MPa<br>After aging ≥14MPa                                                                           | same       |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |         |
|                                                                      | Complies with ASTM D6319-<br>10<br>Elongation:<br>Before Aging: 500% min.<br>After Aging: 400% min.                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Complies with ASTM D6319-<br>19<br>Elongation:<br>Before Aging: 500% min.<br>After Aging: 400% min.                                                                       | same       |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |         |
| Residual<br>powder                                                   | Complies with ASTM D6319-<br>10<br>< 2mg per glove                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Complies with ASTM D6319-<br>19<br>< 2mg per glove                                                                                                                        | same       |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |         |
| Freedom from<br>Holes                                                | In accordance with ASTM D<br>D6319-10 (G-1 with AQL 2.5)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | In accordance with ASTM<br>D6319-19 (G-1 with AQL 2.5)                                                                                                                    | same       |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |         |
| Biocompatibili<br>ty                                                 | AAMI/ ANSI/ ISO 10993-5, In<br>vitro cytotoxicity test: Passes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | ISO 10993-5: 2009, In vitro<br>cytotoxicity test: Pass                                                                                                                    | same       |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |         |
|                                                                      | AAMI/ANSI/ISO 10993-10,<br>Skin sensitization test & Skin<br>Irritation test: Passes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 10993-10:2021, Tests for skin<br>sensitization: Pass<br>CPSC standards regulated under<br>the FHSA Act set forth in 16<br>CFR 1500.41,<br>Irritation/Corrosion Test: Pass | similar    |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |         |
| Resistance of<br>Gloves to<br>Permeation by<br>Chemotherapy<br>Drugs | no. Test Chemotherapy<br>Drug and<br>Concentration Minimum<br>Breakthro<br>ugh<br>Detection<br>Time<br>(Min.)   1 Bleomycin Sulfate<br>15.0 mg/ml > 240   2 Busulfan 6.0 mg/ml > 240   3 Carboplatin 10.0<br>mg/ml > 240   4 Carmustine (BCNU),<br>3.3 mg/ml 59.4   5 Chloroquine 50.0<br>mg/ml > 240   6 Cisplatin 1.0 mg/ml > 240   7 Cyclophosphamide<br>20 mg/ml > 240   8 Cyclosporin A 100.0<br>mg/ml > 240   9 Cytarabine 100.0<br>mg/ml > 240   10 Dacarbazine (DTIC),<br>10.0 mg/ml > 240   11 Daunorubicin 5.0<br>mg/ml > 240   12 Docetaxel 10.0<br>mg/ml > 240 |                                                                                                                                                                           |            |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  | no. Test Chemotherapy<br>Drug and<br>Concentration Minimum<br>Breakthro<br>ugh<br>Detection<br>Time<br>(Min.)   1 Bleomycin Sulfate<br>15.0 mg/ml > 240   2 Busulfan 6.0 mg/ml > 240   3 Carboplatin 10.0<br>mg/ml > 240   4 Carmustine (BCNU),<br>3.3 mg/ml 59.4   5 Chloroquine 50.0<br>mg/ml > 240   6 Cisplatin 1.0 mg/ml > 240   7 Cyclophosphamide<br>20 mg/ml > 240   8 Cyclosporin A 100.0<br>mg/ml > 240   9 Cytarabine 100.0<br>mg/ml > 240   10 Dacarbazine (DTIC),<br>10.0 mg/ml > 240   11 Daunorubicin 5.0<br>mg/ml > 240   12 Docetaxel 10.0<br>mg/ml > 240 |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  | similar |

{10}------------------------------------------------

| 13 | Doxorubicin<br>Hydrochloride, 2.0<br>mg/ml   | > 240 | 13 | Doxorubicin<br>Hydrochloride, 2.0<br>mg/ml   | > 240 |  |
|----|----------------------------------------------|-------|----|----------------------------------------------|-------|--|
| 14 | Epirubicin 2.0 mg/ml                         | > 240 | 14 | Epirubicin 2.0 mg/ml                         | > 240 |  |
| 15 | Etoposide (Toposar),<br>20.0 mg/ml           | > 240 | 15 | Etoposide (Toposar),<br>20.0 mg/ml           | > 240 |  |
| 16 | Fludarabine 25.0<br>mg/ml                    | > 240 | 16 | Fludarabine 25.0<br>mg/ml                    | > 240 |  |
| 17 | Fluorouracil, 50.0<br>mg/ml                  | > 240 | 17 | Fluorouracil, 50.0<br>mg/ml                  | > 240 |  |
| 18 | Gemcitabine 38.0<br>mg/ml                    | > 240 | 18 | Gemcitabine 38.0<br>mg/ml                    | > 240 |  |
| 19 | Idaribicin 1.0 mg/ml                         | > 240 | 19 | Idaribicin 1.0 mg/ml                         | > 240 |  |
| 20 | Ifosfamide 50.0<br>mg/ml                     | > 240 | 20 | Ifosfamide 50.0<br>mg/ml                     | > 240 |  |
| 21 | Irinotecan 20.0<br>mg/ml                     | > 240 | 21 | Irinotecan 20.0<br>mg/ml                     | > 240 |  |
| 22 | Mechlorethamine<br>HCl 1.0 mg/ml             | > 240 | 22 | Mechlorethamine<br>HCl 1.0 mg/ml             | > 240 |  |
| 23 | Melphalan 5.0 mg/ml                          | > 240 | 23 | Melphalan 5.0 mg/ml                          | > 240 |  |
| 24 | Methotrexate 25<br>mg/ml                     | > 240 | 24 | Methotrexate 25<br>mg/ml                     | > 240 |  |
| 25 | Mitomycin C 0.5<br>mg/ml                     | > 240 | 25 | Mitomycin C 0.5<br>mg/ml                     | > 240 |  |
| 26 | Mitoxantrone 2.0<br>mg/ml                    | > 240 | 26 | Mitoxantrone 2.0<br>mg/ml                    | > 240 |  |
| 27 | Oxaliplatin 2.0 mg/ml                        | > 240 | 27 | Oxaliplatin 2.0 mg/ml                        | > 240 |  |
| 28 | Paclitaxel (Taxol),<br>6.0 mg/ml             | > 240 | 28 | Paclitaxel (Taxol),<br>6.0 mg/ml             | > 240 |  |
| 29 | Paraplatin 10 mg/ml                          | > 240 | 29 | Paraplatin 10 mg/ml                          | > 240 |  |
| 30 | Retrovir 10 mg/ml                            | > 240 | 30 | Retrovir 10 mg/ml                            | > 240 |  |
| 31 | Rituximab 10 mg/ml                           | > 240 | 31 | Rituximab 10 mg/ml                           | > 240 |  |
| 32 | Thiotepa, 10.0 mg/ml                         | 118.5 | 32 | Thiotepa, 10.0 mg/ml                         | 118.5 |  |
| 33 | Topotecan HCl 1.0<br>mg/ml                   | > 240 | 33 | Topotecan HCl 1.0<br>mg/ml                   | > 240 |  |
| 34 | Trisonex 1.0 mg/ml                           | > 240 | 34 | Trisonex 1.0 mg/ml                           | > 240 |  |
| 35 | Velcade<br>(Bortezomib) 1.0<br>mg/ml         | > 240 | 35 | Velcade<br>(Bortezomib) 1.0<br>mg/ml         | > 240 |  |
| 36 | Vincristine 1.0 mg/ml                        | > 240 | 36 | Vincristine 1.0 mg/ml                        | > 240 |  |
| 37 | Fentanyl Citrate<br>Injection<br>100 mcg/2ml | n/a   | 37 | Fentanyl Citrate<br>Injection<br>100 mcg/2ml | >240  |  |

{11}------------------------------------------------

#### 8.0 Assessment of Non-Clinical Performance Data:

The following bench testing was conducted for the Disposable Powder Free Nitrile Examination Glove, Natural Blue Color, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate.:

| Test                                                                 | Test Method                                                                         | Purpose                                                                                                  | Acceptance Criteria                                                                                                                                                                                                            | Results |
|----------------------------------------------------------------------|-------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Dimension                                                            | ASTM D3767                                                                          | Determine the<br>geometrical<br>dimension of gloves                                                      | Length: XS&S ≥ 220 mm; M&L&XL ≥230 mm.<br>Thickness:<br>Palm - 0.05 mm min.<br>Finger - 0.05 mm min.<br>Palm Width:<br>X-small 70 ± 10 mm<br>Small 85 ± 10 mm<br>Medium 95 ± 10 mm<br>Large 110 ± 10 mm<br>X-Large 120 ± 10 mm | Pass    |
| Freedom from<br>holes (Water leak)                                   | 21 CFR 800.20.<br>&<br>ASTM D5151-19                                                | Detect the holes on<br>the gloves.                                                                       | G-I/AQL 2.5                                                                                                                                                                                                                    | Pass    |
| Tensile strength<br>(Before<br>aging/After aging)                    | ASTM D412-16<br>& ASTM D573-<br>04                                                  | Evaluate the tensile<br>(tenson) properties of<br>the gloves. In<br>addition, it also<br>determines the  | Before Aging: 14 MPa,<br>min.<br>After Aging: 14 MPa,<br>min                                                                                                                                                                   | Pass    |
| Elongation (Before<br>aging/After aging)                             | ASTM D412-16<br>& ASTM D573-<br>04                                                  | influence of elevated<br>temperature on the<br>physical properties of<br>gloves.                         | Before Aging: 500%<br>min.<br>After Aging: 400% min.                                                                                                                                                                           | Pass    |
| Powder Residual                                                      | ASTM D6124-06                                                                       | Determine the<br>average powder mass<br>found on the gloves                                              | < 2mg per glove                                                                                                                                                                                                                | Pass    |
| Biocompatibility-<br>cytotoxicity                                    | ISO 10993-<br>5:2009                                                                | determine the<br>cytotoxicity potential<br>of glove                                                      | No in vitro cytotoxic as<br>described in<br>ISO 10993-5                                                                                                                                                                        | Pass    |
| Biocompatibility-<br>Skin Sensitization                              | ISO 10993-<br>10:2021                                                               | determine the<br>potential of glove to<br>promote skin<br>sensitization after<br>repeated applications   | No dermal reactions<br>indicative of<br>delayed contact<br>hypersensitivity                                                                                                                                                    | Pass    |
| Biocompatibility-<br>Skin Irritation                                 | CPSC standards<br>regulated under<br>the FHSA Act set<br>forth in 16 CFR<br>1500.41 | determine the<br>potential of glove to<br>promote irritation<br>reactions after<br>repeated applications | Not dermal irritation<br>reactions                                                                                                                                                                                             | Pass    |
| Test                                                                 | Test Method                                                                         | Purpose                                                                                                  | Acceptance Criteria                                                                                                                                                                                                            | Results |
| Resistance of<br>Gloves to<br>Permeation by<br>Chemotherapy<br>Drugs | ASTM D6978-05                                                                       | Assessment of<br>medical gloves to<br>permeation by<br>chemotherapy drugs.                               | The resistance of our<br>device to permeation<br>were challenged against<br>37 chemotherapy drugs<br>and Fentanyl Citrate                                                                                                      | Pass    |

{12}------------------------------------------------

#### 9.0 Assessment of Clinical Performance Data:

Clinical data is not needed for this type of device.

#### 10.0 Conclusion:

The conclusions drawn from the nonclinical tests demonstrate that the subject device Disposable Powder Free Nitrile Examination Glove, Natural Blue Color, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate is as safe, as effective, and performs as well as or better than the legally marketed device.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA/K234097](https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA/K234097)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
