← Product Code [LZA](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA) · K223235

# Nitrile Examination Gloves Powder Free (Tested for use with Chemotherapy Drugs) (K223235)

_Pt. Sintong Unigolden Glove · LZA · Oct 26, 2022 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA/K223235

## Device Facts

- **Applicant:** Pt. Sintong Unigolden Glove
- **Product Code:** [LZA](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA.md)
- **Decision Date:** Oct 26, 2022
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

Nitrile Examination Gloves Powder Free (Tested for use with Chemotherapy Drugs) is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs. This is a single-use, powder-free, non-sterile device.

## Device Story

Disposable, non-sterile, powder-free nitrile examination glove; worn by clinicians to prevent cross-contamination between patient and examiner. Manufactured from carboxylated nitrile rubber; textured surface; beaded cuff; ambidextrous design. Tested for permeation resistance against specific chemotherapy drugs per ASTM D6978-05 (2019). Used in clinical settings; provides barrier protection. Output is physical barrier; informs clinical decision-making regarding personal protective equipment selection during chemotherapy handling. Benefits patient and clinician by reducing exposure to contaminants and hazardous drugs.

## Clinical Evidence

Bench testing only. No clinical data provided. Testing included physical properties (tensile strength, elongation), dimensions, freedom from holes (ASTM D5151-19), residual powder (ASTM D6124-06), and chemotherapy drug permeation (ASTM D6978-05). Biocompatibility testing performed per ISO 10993-5, -10, -11, and -23.

## Technological Characteristics

Material: Carboxylated nitrile rubber. Form factor: Textured, ambidextrous, beaded cuff, powder-free. Standards: ASTM D6319-19 (physical specs), ASTM D5151-19 (holes), ASTM D6124-06 (powder), ASTM D6978-05 (chemo permeation). Biocompatibility: ISO 10993 series. Non-sterile. Connectivity: None.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Nitrile Patient Examination Gloves Blue Colored Tested For Use With Chemotherapy Drugs ([K213040](/device/K213040.md))

## Reference Devices

- Non-Sterile Nitrile Powder Free Examination Gloves - Blue, Green and Black color ([K210388](/device/K210388.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

October 26, 2022

PT. Sintong Unigolden Glove % Manoj Zacharias Consultant Liberty Management Group Ltd. 75 Executive Dr. STE 114 Aurora, Illinois 60504

Re: K223235

Trade/Device Name: Nitrile Examination Gloves Powder Free (Tested for use with Chemotherapy Drugs) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC, OPJ Dated: October 19, 2022 Received: October 19, 2022

Dear Manoj Zacharias:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

#### 510(k) Number (if known) K223235

#### Device Name

Nitrile Examination Gloves Powder Free (Tested for use with Chemotherapy Drugs)

#### Indications for Use (Describe)

Nitrile Examination Gloves Powder Free (Tested for use with Chemotherapy Drugs) is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. In addition, these gloves were tested for use with cherapy drugs in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs. This is a single-use, powderfree, non-sterile device.

The tested chemotherapy drugs and their breakthrough detection times are as follows:

| Chemotherapy Drug   | Concentration           | Breakthrough Detection Time in Minutes |
|---------------------|-------------------------|----------------------------------------|
| Carboplatin         | 10 mg/ml (10,000 ppm)   | > 240 Minutes                          |
| Carmustine          | 3.3 mg/ml (3,300 ppm)   | 15.9 Minutes                           |
| Cisplatin           | 1 mg/ml (1,000 ppm)     | > 240 Minutes                          |
| Cyclophosphamide    | 20 mg/ml (20,000 ppm)   | > 240 Minutes                          |
| Cytarabine HCl      | 100 mg/ml (100,000 ppm) | > 240 Minutes                          |
| Dacarbazine         | 10 mg/ml (10,000 ppm)   | > 240 Minutes                          |
| Docetaxel           | 10 mg/ml (10,000 ppm)   | > 240 Minutes                          |
| Doxorubicin HCl     | 2 mg/ml (2,000 ppm)     | > 240 Minutes                          |
| Etoposide           | 20 mg/ml (20,000 ppm)   | > 240 Minutes                          |
| Fluorouracil        | 50 mg/ml (50,000 ppm)   | > 240 Minutes                          |
| Gemcitabine         | 38 mg/ml (38,000 ppm)   | > 240 Minutes                          |
| Ifosfamide          | 50 mg/ml (50,000 ppm)   | > 240 Minutes                          |
| Irinotecan          | 20 mg/ml (20,000 ppm)   | > 240 Minutes                          |
| Mechlorethamine HCl | 1 mg/ml (1,000 ppm)     | > 240 Minutes                          |
| Melphalan           | 5 mg/ml (5,000 ppm)     | > 240 Minutes                          |
| Methotrexate        | 25 mg/ml (25,000 ppm)   | > 240 Minutes                          |
| Mitoxantrone        | 2 mg/ml (2,000 ppm)     | > 240 Minutes                          |
| Paclitaxel          | 6 mg/ml (6,000 ppm)     | > 240 Minutes                          |
| Thiotepa            | 10 mg/ml (10,000 ppm)   | 54.2 Minutes                           |

Please note that the following drugs have low permeation times: Carmustine 3.3 mg/ml 15.9 Minutes Thiotepa 10 mg/m1 54.2 Minutes

Warning: Please do not use with Carmustine and Thiotepa.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

### A. APPLICANT INFORMATION

| 510(K) Owner's Name | PT. Sintong Unigolden Glove                                                                                                   |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Address             | Jl. Lintas Sumatera, Kel. Hessa Perlompongan,<br>Kec. Air Batu, Kab. Asahan, Kota Kisaran,<br>Sumatera Utara, Indonesia 21272 |
| Phone               | +62623533339                                                                                                                  |
| Fax                 | +62623533330                                                                                                                  |
| E-mail              | sug.qualityassurance@gmail.com                                                                                                |
| Contact Person      | Vivekanandan                                                                                                                  |
| Designation         | Production Manager                                                                                                            |
| Contact Number      | +681253206449                                                                                                                 |
| Contact Email       | vivekmp80@gmail.com                                                                                                           |
| Date Submitted      | 13 October 2022                                                                                                               |

### B. DEVICE IDENTIFICATION

| Name of the device                | Nitrile Examination Gloves Powder Free (Tested for use with Chemotherapy Drugs) |
|-----------------------------------|---------------------------------------------------------------------------------|
| Product proprietary or trade name | Nitrile Examination Gloves Powder Free (Tested for use with Chemotherapy Drugs) |
| Common or usual name              | Nitrile Examination Gloves Powder Free (Tested for use with Chemotherapy Drugs) |
| Classification name               | Patient Examination Glove, Specialty                                            |
| Device Classification             | Class-1                                                                         |
| Product Code                      | LZA, LZC, OPJ                                                                   |
| Regulation Number                 | 21 CFR 880.6250                                                                 |
| Review Panel                      | General Hospital                                                                |

### C. PREDICATE DEVICE

| Predicate Device | Nitrile Patient Examination Gloves Blue Colored<br>Tested For Use With Chemotherapy Drugs |
|------------------|-------------------------------------------------------------------------------------------|
| 510(k) Number    | K213040                                                                                   |
| Regulatory Class | Class 1                                                                                   |
| Product code     | LZA, LZC, OPJ                                                                             |

| Reference Device | Non-Sterile Nitrile Powder Free Examination<br>Gloves - Blue, Green and Black color |
|------------------|-------------------------------------------------------------------------------------|
| 510(k) Number    | K210388                                                                             |
| Regulatory Class | Class 1                                                                             |
| Product code     | LZA                                                                                 |

{5}------------------------------------------------

### D. DESCRIPTION OF THE DEVICE:

Nitrile Examination Gloves Powder Free (Tested for use with Chemotherapy Drugs) is a Class I patient examination gloves bearing the product codes LZA, LZC, OPJ (21CFR880.6250). They meet all the current specifications listed under the ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application and also complies with requirements for Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs as per ASTM D6978-05 (2019). They are made from Carboxylated Nitrile. These gloves are blue in color and are powder free. The product is non-sterile, textured, ambidextrous with beaded cuff and single use only.

Nitrile Examination Gloves Powder Free (Tested for use with Chemotherapy Drugs) with sizes X-Small, Small, Medium, Large, X-Large and XX-Large are included in the submission.

### E. INDICATION FOR USE OF THE DEVICE:

Nitrile Examination Gloves Powder Free (Tested for use with Chemotherapy Drugs) is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs. This is a single-use, powder-free, non-sterile device.

| Chemotherapy Drug   | Concentration           | Breakthrough Detection Time in Minutes |
|---------------------|-------------------------|----------------------------------------|
| Carboplatin         | 10 mg/ml (10,000 ppm)   | > 240 Minutes                          |
| Carmustine          | 3.3 mg/ml (3,300 ppm)   | 15.9 Minutes                           |
| Cisplatin           | 1 mg/ml (1,000 ppm)     | > 240 Minutes                          |
| Cyclophosphamide    | 20 mg/ml (20,000 ppm)   | > 240 Minutes                          |
| Cytarabine HCl      | 100 mg/ml (100,000 ppm) | > 240 Minutes                          |
| Dacarbazine         | 10 mg/ml (10,000 ppm)   | > 240 Minutes                          |
| Docetaxel           | 10 mg/ml (10,000 ppm)   | > 240 Minutes                          |
| Doxorubicin HCl     | 2 mg/ml (2,000 ppm)     | > 240 Minutes                          |
| Etoposide           | 20 mg/ml (20,000 ppm)   | > 240 Minutes                          |
| Fluorouracil        | 50 mg/ml (50,000 ppm)   | > 240 Minutes                          |
| Gemcitabine         | 38 mg/ml (38,000 ppm)   | > 240 Minutes                          |
| Ifosfamide          | 50 mg/ml (50,000 ppm)   | > 240 Minutes                          |
| Irinotecan          | 20 mg/ml (20,000 ppm)   | > 240 Minutes                          |
| Mechlorethamine HCl | 1 mg/ml (1,000 ppm)     | > 240 Minutes                          |
| Melphalan           | 5 mg/ml (5,000 ppm)     | > 240 Minutes                          |
| Methotrexate        | 25 mg/ml (25,000 ppm)   | > 240 Minutes                          |
| Mitoxantrone        | 2 mg/ml (2,000 ppm)     | > 240 Minutes                          |
| Paclitaxel          | 6 mg/ml (6,000 ppm)     | > 240 Minutes                          |
| Thiotepa            | 10 mg/ml (10,000 ppm)   | 54.2 Minutes                           |

The tested chemotherapy drugs and their breakthrough detection times are as follows:

{6}------------------------------------------------

Please note that the following drugs have low permeation times: Carmustine (3.3 mg/ml) 15.9 Minutes Thiotepa (10 mg/ml) 54.2 Minutes Warning: Do not use with Carmustine & Thiotepa

### F. SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICE

| CHARACTERISTICS     | STANDARDS        | DEVICE PERFORMANCE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Comparison                                                                                                                                         |
|---------------------|------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
|                     |                  | PREDICATE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | REFERENCE                                                                                                                                                                                                                                                | SUBJECT                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                    |
| 510(K) Number       | ---              | K213040                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | K210388                                                                                                                                                                                                                                                  | K223235                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                    |
| Name of device      | ---              | Nitrile Patient<br>Examination Gloves<br>Blue Colored Tested<br>For Use With<br>Chemotherapy Drugs                                                                                                                                                                                                                                                                                                                                                                                                    | Non Sterile<br>Nitrile<br>Powder Free<br>Examination<br>Gloves –<br>Blue, Green<br>And Black<br>color                                                                                                                                                    | Nitrile Examination<br>Gloves Powder Free<br>(Tested for use with<br>Chemotherapy Drugs)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Similar<br>to<br>predicate<br>device                                                                                                               |
| Product Code        | ---              | LZA, LZC, OPJ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | LZA                                                                                                                                                                                                                                                      | LZA, LZC, OPJ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Similar to<br>predicate<br>device                                                                                                                  |
| Indication for use  | ---              | The blue colored nitrile<br>examination glove is<br>intended to be worn on<br>the hands of examiner's<br>to prevent<br>contamination<br>between patient and<br>examiner. This is a<br>single-use, powder-free,<br>non-sterile device.<br>The Nitrile Patient<br>Examination Gloves<br>Blue Colored were<br>tested for use with<br>chemotherapy drugs per<br>ASTM D6978-05<br>(Reapproved 2019)<br>Standard Practice for<br>Assessment of Medical<br>Gloves to Permeation<br>by Chemotherapy<br>Drugs. | A powder<br>free patient<br>examination<br>glove is a<br>disposable<br>device<br>intended for<br>medical<br>purposes that<br>is worn on<br>the<br>examiner's<br>hand or<br>finger to<br>prevent<br>contaminatio<br>n between<br>patient and<br>examiner. | Nitrile Examination<br>Gloves Powder Free<br>(Tested for use with<br>Chemotherapy Drugs) is<br>a disposable device<br>intended for medical<br>purpose that is worn on<br>the examiner's hand to<br>prevent contamination<br>between patient and<br>examiner. In addition,<br>these gloves were tested<br>for use with<br>chemotherapy drugs in<br>accordance with ASTM<br>D6978-05 (2019)<br>Standard Practice for<br>Assessment of<br>Resistance of Medical<br>Gloves to Permeation<br>by Chemotherapy<br>Drugs. This is a single-<br>use, powder-free, non-<br>sterile device. | Similar to<br>predicate<br>device                                                                                                                  |
| Regulation Number   | ---              | 21 CFR 880.6250                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 21 CFR<br>880.6250                                                                                                                                                                                                                                       | 21 CFR 880.6250                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                                               |
| CHARACTERISTICS     | STANDARDS        | DEVICE PERFORMANCE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Comparison                                                                                                                                         |
|                     |                  | PREDICATE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | REFERENCE                                                                                                                                                                                                                                                | SUBJECT                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                    |
| 510(K) Number       | ---              | K213040                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | K210388                                                                                                                                                                                                                                                  | K223235                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                    |
| Material            |                  | Powder-Free Nitrile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Carboxylated<br>Butadiene<br>Acrylonitrile as<br>base material                                                                                                                                                                                           | Carboxylated Nitrile<br>Rubber                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Same as the<br>reference<br>device                                                                                                                 |
| Color               | ---              | Blue                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Blue, Green and<br>Black                                                                                                                                                                                                                                 | Blue                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same as<br>predicate<br>device                                                                                                                     |
| Size                | ---              | X-Small, Small,<br>Medium, Large, X-<br>Large and<br>XX-Large                                                                                                                                                                                                                                                                                                                                                                                                                                         | Small, Medium,<br>Large and<br>X-Large                                                                                                                                                                                                                   | X-Small, Small,<br>Medium, Large, X-<br>Large and<br>XX-Large                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same as<br>predicate<br>device                                                                                                                     |
| Single Use          | ---              | Single Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Single Use                                                                                                                                                                                                                                               | Single Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Same                                                                                                                                               |
| Sterile/non sterile | ---              | non sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | non sterile                                                                                                                                                                                                                                              | non sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                                                                                               |
| Rx Only or OTC      | ---              | Over the Counter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Over the<br>Counter                                                                                                                                                                                                                                      | Over the Counter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                               |
| Dimensions - Length | ASTM<br>D6319-19 | XS (220mm min)<br>S (220mm min)<br>M (230mm min)<br>L (230mm min)<br>XL (230mm min)<br>XXL (230mm min)                                                                                                                                                                                                                                                                                                                                                                                                | 240-246 mm<br>(Medium)                                                                                                                                                                                                                                   | XS (220mm min)<br>S (220mm min)<br>M (230mm min)<br>L (230mm min)<br>XL (230mm min)<br>XXL (230mm min)                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same as<br>predicate<br>device<br><br>Size Average<br>value<br>X- Small 270<br>Small 273<br>Medium 274<br>Large 276<br>X-Large 278<br>XX-Large 276 |
| Dimensions - Width  | ASTM<br>D6319-19 | XS (70±10mm)<br>S (80±10mm)<br>M (95±10mm)<br>L(110±10mm)<br>XL (120±10mm)<br>XXL (≥120mm)                                                                                                                                                                                                                                                                                                                                                                                                            | 95-98 mm<br>(Medium)                                                                                                                                                                                                                                     | XS (70±10mm)<br>S (80±10mm)<br>M (95±10mm)<br>L(110±10mm)<br>XL (120±10mm)<br>XXL (130±10mm)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Similar<br><br>Size Average<br>value<br>X-Small 72<br>Small 82<br>Medium 97<br>Large 103<br>X-Large 112<br>XX-Large 126                            |

{7}------------------------------------------------

{8}------------------------------------------------

| CHARACTERISTICS                         | STANDARDS       | DEVICE PERFORMANCE                                  |                                                                  |                                                                                                          | Comparison                     |  |  |  |  |  |  |  |  |  |  |  |  |  |  |
|-----------------------------------------|-----------------|-----------------------------------------------------|------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|--------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|
|                                         |                 | PREDICATE                                           | REFERENCE                                                        | SUBJECT                                                                                                  |                                |  |  |  |  |  |  |  |  |  |  |  |  |  |  |
| 510(K) Number                           | ---             | K213040                                             | K210388                                                          | K223235                                                                                                  |                                |  |  |  |  |  |  |  |  |  |  |  |  |  |  |
| Physical Properties-Tensile Strength    | ASTM D6319-2019 | Before Ageing<br>Tensile Strength<br>≥14MPa, min    | Before Ageing<br>Tensile Strength<br>25.9-32.0 MPa<br>(Medium)   | Before Ageing<br>Tensile Strength<br>14MPa, min                                                          | Similar                        |  |  |  |  |  |  |  |  |  |  |  |  |  |  |
|                                         |                 |                                                     |                                                                  | Size Average value   X-Small 17.7   Small 18.1   Medium 18.3   Large 18.4   X-Large 18.2   XX-Large 18.5 |                                |  |  |  |  |  |  |  |  |  |  |  |  |  |  |
|                                         |                 | After Ageing<br>Tensile Strength<br>≥14MPa, min     | After Ageing<br>Tensile Strength<br>25.4-34.0 MPa<br>(Medium)    | After Ageing<br>Tensile Strength<br>14MPa, min                                                           | Similar                        |  |  |  |  |  |  |  |  |  |  |  |  |  |  |
|                                         |                 |                                                     |                                                                  | Size Average value   X-Small 16.8   Small 17.3   Medium 17.2   Large 17.6   X-Large 17.5   XX-Large 17.6 |                                |  |  |  |  |  |  |  |  |  |  |  |  |  |  |
| Physical Properties-Ultimate Elongation | ASTM D6319-2019 | Before Ageing<br>Ultimate<br>Elongation<br>500% min | Before Ageing<br>Ultimate<br>Elongation<br>500-540 %<br>(Medium) | Before Ageing<br>Ultimate Elongation<br>500% min                                                         | Same as<br>predicate<br>device |  |  |  |  |  |  |  |  |  |  |  |  |  |  |
|                                         |                 |                                                     |                                                                  | Size Average value   X-Small 637   Small 634   Medium 647   Large 630   X-Large 644   XX-Large 644       |                                |  |  |  |  |  |  |  |  |  |  |  |  |  |  |
|                                         |                 | After Ageing<br>Ultimate<br>Elongation<br>400% min  | After Ageing<br>Ultimate<br>Elongation<br>480-520 %<br>(Medium)  | After Ageing<br>Ultimate Elongation<br>400% min                                                          | Same as<br>predicate<br>device |  |  |  |  |  |  |  |  |  |  |  |  |  |  |
|                                         |                 |                                                     |                                                                  | Size Average value   X-Small 532   Small 539   Medium 542   Large 536   X-Large 529   XX-Large 533       |                                |  |  |  |  |  |  |  |  |  |  |  |  |  |  |

{9}------------------------------------------------

| CHARACTERISTICS                                                             | STANDARDS                | DEVICE PERFORMANCE                                                                                                                                                                                                                                                                                                                                                                      |                                     |                                                                                                                                                                                                                                                                                                                                                                                                                       | Comparison                                                                        |  |  |  |  |  |  |  |  |  |  |  |  |  |         |  |  |  |  |  |  |  |
|-----------------------------------------------------------------------------|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|---------|--|--|--|--|--|--|--|
|                                                                             |                          | PREDICATE                                                                                                                                                                                                                                                                                                                                                                               | REFERENCE                           | SUBJECT                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                   |  |  |  |  |  |  |  |  |  |  |  |  |  |         |  |  |  |  |  |  |  |
| 510(K) Number                                                               |                          | K213040                                                                                                                                                                                                                                                                                                                                                                                 | K210388                             | K223235                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                   |  |  |  |  |  |  |  |  |  |  |  |  |  |         |  |  |  |  |  |  |  |
| Thickness                                                                   | ASTM<br>D6319-19         | Palm<br>0.05mm min                                                                                                                                                                                                                                                                                                                                                                      | Palm<br>0.06-0.06 mm<br>(Medium)    | Palm 0.05 mm min;<br>Finger 0.05 mm min                                                                                                                                                                                                                                                                                                                                                                               | Similar                                                                           |  |  |  |  |  |  |  |  |  |  |  |  |  |         |  |  |  |  |  |  |  |
|                                                                             |                          | Finger<br>0.11 mm min                                                                                                                                                                                                                                                                                                                                                                   | Finger<br>0.09-0.10 mm<br>(Medium)  | Size Palm<br>(Avg value) Finger<br>(Avg value)   X-Small 0.12 0.14   Small 0.12 0.14   Medium 0.12 0.14   Large 0.12 0.14   X-Large 0.12 0.14   XX-Large 0.12 0.14                                                                                                                                                                                                                                                    |                                                                                   |  |  |  |  |  |  |  |  |  |  |  |  |  |         |  |  |  |  |  |  |  |
| Powder Free Residue                                                         | ASTM D6319-<br>19        | <2mg per glove                                                                                                                                                                                                                                                                                                                                                                          | 0.70<br>mg/glove<br>(Medium)        | ≤2 mg/glove   Size Average value   X-Small 0.21   Small 0.18   Medium 0.18   Large 0.19   X-Large 0.21   XX-Large 0.20                                                                                                                                                                                                                                                                                                |                                                                                   |  |  |  |  |  |  |  |  |  |  |  |  |  | Similar |  |  |  |  |  |  |  |
| Freedom from holes                                                          | ASTM D5151-<br>2019      | Complies with ASTM<br>D6319-19 and ASTM<br>D5151-19 G-1,<br>AQL 1.5                                                                                                                                                                                                                                                                                                                     | Inspection<br>Level G-1;<br>AQL=2.5 | Complies with ASTM<br>D6319-19 and ASTM<br>D5151-19 G-1, AQL 2.5                                                                                                                                                                                                                                                                                                                                                      | Same as<br>reference<br>device                                                    |  |  |  |  |  |  |  |  |  |  |  |  |  |         |  |  |  |  |  |  |  |
| Chemotherapy Drugs<br>Tested with Minimum<br>Breakthrough<br>Detection Time | ASTM D6978-<br>05 (2019) | Bleomycin Sulfate<br>15 mg/ml >240 min.<br>Carboplatin 10 mg/ml<br>>240 min.<br>Carmustine (BCNU)<br>3.3 mg/ml 17.2 min.<br>Cisplatin 1.0 mg/ml<br>>240 min.<br>Cyclophosphamide<br>(Cytoxan)<br>20.0 mg/ml >240 min.<br>Cytarabine HCl<br>100 mg/ml >240 min.<br>Dacarbazine (DTIC)<br>10.0 mg/ml >240 min.<br>Daunorubicin<br>5.0 mg/ml >240 min.<br>Docetaxel 10.0 mg/ml<br>>240 min | NA                                  | Not tested<br>Carboplatin 10 mg/ml<br>(10,000 ppm) > 240 Minutes<br>Carmustine 3.3 mg/ml<br>(3,300 ppm) 15.9 Minutes<br>Cisplatin 1 mg/ml<br>(1,000 ppm) > 240 Minutes<br>Cyclophosphamide 20 mg/ml<br>(20,000 ppm) > 240 Minutes<br>Cytarabine HCl<br>100 mg/ml (100,000 ppm)<br>>240 Minutes<br>Dacarbazine 10 mg/ml<br>(10,000 ppm) > 240 Minutes<br>Not tested<br>Docetaxel 10 mg/ml<br>(10,000 ppm) >240 Minutes | Optional*<br>Same<br>Similar<br>Same<br>Same<br>Same<br>Same<br>Optional*<br>Same |  |  |  |  |  |  |  |  |  |  |  |  |  |         |  |  |  |  |  |  |  |

{10}------------------------------------------------

| CHARACTERISTICS                                                             | STANDARDS                                                                   | DEVICE PERFORMANCE                                                                                             |                                                         |                                                                                              | Comparison  |
|-----------------------------------------------------------------------------|-----------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|----------------------------------------------------------------------------------------------|-------------|
|                                                                             |                                                                             | PREDICATE                                                                                                      | REFERENCE                                               | SUBJECT                                                                                      |             |
| 510(K) Number                                                               |                                                                             | K213040                                                                                                        | K210388                                                 | K223235                                                                                      |             |
| Chemotherapy Drugs<br>Tested with Minimum<br>Breakthrough<br>Detection Time | ASTM<br>D6978-05<br>(2019)                                                  | Doxorubicin HCl<br>2.0 mg/ml >240 min.                                                                         | NA                                                      | Doxorubicin HCl 2 mg/ml<br>(2,000 ppm) > 240 Minutes                                         | Same        |
|                                                                             |                                                                             | Etoposide (Toposar)<br>20.0 mg/ml >240 min.                                                                    | NA                                                      | Etoposide 20 mg/ml<br>(20,000 ppm) > 240 Minutes                                             | Same        |
|                                                                             |                                                                             | Fluorouracil<br>50.0 mg/ml >240 min.                                                                           | NA                                                      | Fluorouracil 50 mg/ml<br>(50,000 ppm) > 240 Minutes                                          | Same        |
|                                                                             |                                                                             | Gemcitabine<br>38 mg/ml >240 min.                                                                              | NA                                                      | Gemcitabine 38 mg/ml<br>(38,000 ppm) >240 Minutes                                            | Same        |
|                                                                             |                                                                             | Idarubicin 1 mg/ml<br>>240 min.                                                                                | NA                                                      | Not tested                                                                                   | Optional*   |
|                                                                             |                                                                             | Ifosfamide 50.0 mg/ml<br>>240 min.                                                                             | NA                                                      | Ifosfamide 50 mg/ml<br>(50,000 ppm) >240 Minutes                                             | Same        |
|                                                                             |                                                                             | Irinotecan 20.0 mg/ml<br>>240 min.                                                                             | NA                                                      | Irinotecan 20 mg/ml<br>(20,000 ppm) >240 Minutes                                             | Same        |
|                                                                             |                                                                             | Mechlorethamine HCl<br>1.0 mg/ml >240 min.                                                                     | NA                                                      | Mechlorethamine HCl<br>1 mg/ml (1,000 ppm)<br>>240 Minutes                                   | Same        |
|                                                                             |                                                                             | Melphalan 5 mg/ml<br>>240 min.                                                                                 | NA                                                      | Melphalan 5 mg/ml<br>(5,000 ppm) >240 Minutes                                                | Same        |
|                                                                             |                                                                             | Methotrexate<br>25 mg/ml >240 min.                                                                             | NA                                                      | Methotrexate 25 mg/ml<br>(25,000 ppm) >240 Minutes                                           | Same        |
|                                                                             |                                                                             | Mitromycin C.<br>0.5 mg/ml >240 min.                                                                           | NA                                                      | Not tested                                                                                   | Optional*   |
|                                                                             |                                                                             | Mitoxantrone<br>2.0 mg/ml >240 min.                                                                            | NA                                                      | Mitoxantrone 2 mg/ml<br>(2,000ppm) > 240 Minutes                                             | Same        |
|                                                                             |                                                                             | Paclitaxel (Taxol)<br>6.0 mg/ml >240 min.                                                                      | NA                                                      | Paclitaxel 6 mg/ml<br>(6,000 ppm) > 240 Minutes                                              | Same        |
|                                                                             |                                                                             | Thiotepa 10.0 mg/ml<br>13.9 min.                                                                               | NA                                                      | Thiotepa 10 mg/ml<br>(10,000 ppm) 54.2 Minutes                                               | Similar     |
|                                                                             |                                                                             | Vincristine Sulfate<br>1.0 mg/ml >240 min.                                                                     | NA                                                      | Not tested                                                                                   | Optional*   |
| Contact Durations                                                           | ---                                                                         | Limited <24 hours                                                                                              | NA                                                      | Limited <24 hours                                                                            | Same        |
| Biocompatibility                                                            | Primary Skin<br>Irritation- ISO<br>10993-23:<br>First Edition<br>2021-01    | Not a skin irritant                                                                                            | Non-irritant<br>(Response<br>Category is<br>Negligible) | Under the condition of study<br>not an irritant                                              | Same        |
|                                                                             | Dermal<br>Sensitization-<br>ISO 10993-10:<br>Fourth Edition<br>2021-11      | Not a skin sensitizer                                                                                          | Non-sensitizer<br>(No<br>sensitization)                 | Under the conditions of the<br>study not a sensitizer                                        | Same        |
| CHARACTERISTICS                                                             | STANDARDS                                                                   | DEVICE PERFORMANCE                                                                                             |                                                         |                                                                                              | Comparison  |
|                                                                             |                                                                             | PREDICATE                                                                                                      | REFERENCE                                               | SUBJECT                                                                                      |             |
| 510(K) Number                                                               | ---                                                                         | K213040                                                                                                        | K210388                                                 | K223235                                                                                      |             |
| Biocompatibility                                                            | In vitro<br>cytotoxicity-<br>ISO 10993-5:<br>Third Edition<br>2009-06-01    | At the neat extraction,<br>the test article is<br>considered cytotoxic                                         | ---                                                     | Under the conditions<br>of the study, non-<br>cytotoxic                                      | Different** |
|                                                                             | Acute<br>Systemic<br>Toxicity- ISO<br>10993-11:<br>Third Edition<br>2017-09 | The acute systemic<br>toxicity results<br>demonstrate the<br>device<br>will not cause a<br>systemic<br>effect. | No<br>toxic effects                                     | Under the conditions<br>of the study, the test<br>item did not produce<br>any adverse effect | Same        |

{11}------------------------------------------------

* Predicate device perform additional Chemotherapy drug test.

** The difference does not raise any issue regarding the safety or effectiveness of the subject glove is non-cytotoxic.

There are no significant differences between the products other than In vitro cytotoxicity study and are identical in terms of intended use, materials, design and manufacturing methods. The devices meet the ASTM standard D6319-19 and D6978-05 (2019).

### G. NON-CLINICAL TESTING SUMMARY PERFORMANCE DATA

### BENCH TEST DATA

| TEST METHOD                                                                                           | PURPOSE                                  | ACCEPTANCE CRITERIA                                                                                                                     | RESULT                                                                                                           |
|-------------------------------------------------------------------------------------------------------|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| ASTM D6319-19<br>Standard Specification for<br>Nitrile Examination Gloves<br>for Medical Application. | To determine the<br>length of the gloves | X-Small : 220 mm min<br>Small : 220 mm min<br>Medium: 230 mm min<br>Large : 230 mm min<br>X-Large: 230 mm min<br>XX-Large: 230 mm min   | X-Small : 270 mm<br>Small : 273 mm<br>Medium : 274 mm<br>Large : 276 mm<br>X-Large : 278 mm<br>XX-Large : 276 mm |
| ASTM D6319-19<br>Standard Specification for<br>Nitrile Examination Gloves<br>for Medical Application. | To determine the<br>width of the gloves  | X-Small : 70+/-10 mm<br>Small : 80+/-10 mm<br>Medium: 95+/-10 mm<br>Large: 110+/-10 mm<br>X-Large: 120+/-10 mm<br>XX-Large: 130+/-10 mm | X-Small : 72 mm<br>Small : 82 mm<br>Medium : 97 mm<br>Large : 103 mm<br>X-Large : 112 mm<br>XX-Large : 126 mm    |

{12}------------------------------------------------

| TEST METHOD                                                                                           | PURPOSE                                                         | ACCEPTANCE<br>CRITERIA                                                                                                          | RESULT                                                                                                                                                                                                    |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |
|-------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|
| ASTM D6319-19<br>Standard Specification for<br>Nitrile Examination Gloves<br>for Medical Application. | To determine the<br>thickness of the<br>gloves                  | Palm: 0.05 mm min<br>for all sizes<br>Finger: 0.05 mm min<br>for all sizes                                                      | Size Palm Finger   X-Small 0.12 mm 0.14 mm   Small 0.12 mm 0.14 mm   Medium 0.12 mm 0.14 mm   Large 0.12 mm 0.14 mm   X-Large 0.12 mm 0.14 mm   XX-Large 0.12 mm 0.14 mm                                  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |
| ASTM D6319-19<br>Standard Specification for<br>Nitrile Examination Gloves<br>for Medical Application. | To determine the<br>physical properties-<br>Tensile strength    | Before Ageing<br>Tensile Strength<br>14MPa min for all sizes<br>After Ageing<br>Tensile Strength<br>14MPa min for all sizes     | Size Before<br>ageing After<br>ageing   X-Small 17.7 MPa 16.8 MPa   Small 18.1 MPa 17.3 MPa   Medium 18.3 MPa 17.2 MPa   Large 18.4 MPa 17.6 MPa   X-Large 18.2 MPa 17.5 MPa   XX-Large 18.5 MPa 17.6 MPa |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |
|                                                                                                       | To determine the<br>physical properties-<br>Ultimate Elongation | Before Ageing<br>Ultimate Elongation<br>500% min for all sizes<br>After Ageing<br>Ultimate Elongation<br>400% min for all sizes | Size Before<br>ageing After<br>ageing   X-Small 637% 532%   Small 634% 539%   Medium 647% 542%   Large 630% 536%   X-Large 644% 529%   XX-Large 644% 533%                                                 |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |
| ASTM D5151-19 Standard<br>Test Method for Detection<br>of Holes in Medical Gloves                     | To determine the<br>holes in the gloves                         | AQL 2.5                                                                                                                         | Gloves Passes AQL 2.5                                                                                                                                                                                     |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |
| ASTM D6124-06<br>(Reapproved 2017) Standard<br>Test Method for Residual<br>Powder on Medical Gloves   | To determine the<br>residual powder in<br>the gloves            | $≤$ 2 mg/glove                                                                                                                  | X-Small : 0.21 mg/glove<br>Small : 0.18 mg/glove<br>Medium : 0.18 mg/glove<br>Large : 0.19 mg/glove<br>X-Large : 0.21 mg/glove<br>XX-Large : 0.20 mg/glove                                                |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |

| TEST METHOD                                                                                                                                                   | PURPOSE                                                                        | ACCEPTANCE CRITERIA                                     | RESULT                                                  |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|---------------------------------------------------------|---------------------------------------------------------|
| ASTM D6978-05<br>(Reapproved 2019)<br>Standard Practice for<br>Assessment of<br>Resistance of<br>Medical Gloves to<br>Permeation by<br>Chemotherapy<br>Drugs. | To determine the<br>breakthrough<br>detection time of<br>chemotherapy<br>drugs | Carboplatin 10 mg/ml<br>(10,000 ppm) > 240 Minutes      | Carboplatin 10 mg/ml<br>(10,000 ppm) > 240 Minutes      |
|                                                                                                                                                               |                                                                                | Cisplatin 1 mg/ml<br>(1,000 ppm) > 240 Minutes          | Cisplatin 1 mg/ml<br>(1,000 ppm) > 240 Minutes          |
|                                                                                                                                                               |                                                                                | Cyclophosphamide 20 mg/ml<br>(20,000 ppm) > 240 Minutes | Cyclophosphamide 20 mg/ml<br>(20,000 ppm) > 240 Minutes |
|                                                                                                                                                               |                                                                                | Cytarabine HCl 100 mg/ml<br>(100,000 ppm) >240 Minutes  | Cytarabine HCl 100 mg/ml<br>(100,000 ppm) >240 Minutes  |
|…

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA/K223235](https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA/K223235)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
