← Product Code [LZA](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA) · K213187

# Disposable Nitrile Medical Powder-free Glove (K213187)

_Fujian Chunhui Medical Supply Co., Ltd. · LZA · Jan 3, 2022 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA/K213187

## Device Facts

- **Applicant:** Fujian Chunhui Medical Supply Co., Ltd.
- **Product Code:** [LZA](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA.md)
- **Decision Date:** Jan 3, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

The Disposable Nitrile Medical Powder-free Glove is intended to be worn on the hands of examiners to prevent contamination between patient and examiner. This is a single-use, powder-free, non-sterile device.

## Device Story

Disposable nitrile examination glove; worn on hands of examiners to prevent cross-contamination between patient and examiner. Single-use, non-sterile, powder-free device. Available in sizes S, M, L, XL. Provides physical barrier protection during clinical examinations. No electronic, software, or algorithmic components.

## Clinical Evidence

No clinical data. Bench testing only, including physical dimensions, watertightness (ASTM D5151), powder content (ASTM D6124), physical properties (ASTM D412), and biocompatibility (ISO 10993-5, 10, 11).

## Technological Characteristics

Nitrile material; ambidextrous design; blue color; non-sterile. Complies with ASTM D6319, ASTM D5151, ASTM D6124, ASTM D412, ISO 10993-5, ISO 10993-10, and ISO 10993-11. No software or electronic components.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Nitrile examination gloves ([K203593](/device/K203593.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 3, 2022

Fujian Chunhui Medical Supply Co., Ltd. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM. 1801, No. 161 Lujiazui East Rd.,Pudong Shanghai, Shanghai 200120 China

Re: K213187

Trade/Device Name: Disposable Nitrile Medical Powder-free Glove Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: September 18, 2021 Received: September 29, 2021

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K213187

Device Name Disposable Nitrile Medical Powder-free Glove

Indications for Use (Describe)

The Disposable Nitrile Medical Powder-free Glove is intended to be worn on the hands of examiners to prevent contamination between patient and examiner. This is a single-use, powder-free, non-sterile device.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary K213187

This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.

#### 1.0 Submitter's Information

Name: Fujian Chunhui Medical Supply Co., Ltd. Address: No.179, Shanggeteng Ping, Shifang Village, Shifang County, Wuping Town, Longyan City, Fujian Province, China. Phone Number: +86-18060003170 Contact: Han Cai Date of Preparation: Sept 18, 2021

#### Designated Submission Correspondent

Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 1801, No. 161 East Lujiazui Rd., Pudong,Shanghai 200120 ,China Tel: +86-21-50313932 Email: Info@truthful.com.cn

#### 2.0 Device Information

Trade name: Disposable Nitrile Medical Powder-free Glove Common name: Patient Examination Gloves Classification name: Non-powdered patient examination glove Model(s): S. M. L. XL

#### 3.0 Classification

Production code: LZA Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital

#### 4.0 Predicate Device Information

Manufacturer: GUANG DONG KINGFA SCI. & TECH.CO., LTD. Device: Nitrile examination gloves 510(k) number: K203593

#### 5.0 Indication for Use

{4}------------------------------------------------

The Disposable Nitrile Medical Powder-free Glove is intended to be worn on the hands of examiners to prevent contamination between patient and examiner. This is a single-use, powder-free, non-sterile device.

#### 6.0 Device Description

The subject device is powder free nitrile examination gloves. The subject device is blue. It can be available in four specifications: S,M,L and XL. The subject device is non-sterile.

## 7.0 Technological Characteristic Comparison Table

| Item                        | Subject Device                                                                                                                                                                                                                          | Predicated Device<br>(K203593)                                                                                                                                                                                       | Comparison |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Product Code                | LZA                                                                                                                                                                                                                                     | LZA                                                                                                                                                                                                                  | Same       |
| Regulation No.              | 21CFR880.6250                                                                                                                                                                                                                           | 21CFR880.6250                                                                                                                                                                                                        | Same       |
| Class                       | I                                                                                                                                                                                                                                       | I                                                                                                                                                                                                                    | Same       |
| Intended Use                | The Disposable Nitrile<br>Medical Powder-free Glove<br>is intended to be worn on<br>the hands of examiners to<br>prevent contamination<br>between patient and<br>examiner. This is a<br>single-use, powder-free,<br>non-sterile device. | The nitrile examination<br>glove is intended to be<br>worn on the hands of<br>examiners to<br>prevent contamination<br>between patient and<br>examiner. This is a<br>single-use, powder-free,<br>non-sterile device. | Same       |
| Material                    | Nitrile                                                                                                                                                                                                                                 | Nitrile                                                                                                                                                                                                              | Same       |
| Powdered or<br>Powered free | Powdered free                                                                                                                                                                                                                           | Powdered free                                                                                                                                                                                                        | Same       |
| Design Feature              | Ambidextrous                                                                                                                                                                                                                            | Ambidextrous                                                                                                                                                                                                         | Same       |
| Colorant                    | Blue                                                                                                                                                                                                                                    | Blue                                                                                                                                                                                                                 | Same       |
| Labeling<br>Information     | Single-use indication,<br>powder free, device color,<br>device name, glove size and<br>quantity, Non-Sterile                                                                                                                            | Single-use indication,<br>powder free, device color,<br>device name, glove size<br>and quantity, Non-Sterile                                                                                                         | Same       |
| Dimensions(mm<br>)          | Length:<br>S:≥220;<br>M/L/XL: ≥230;<br>Width:<br>S: 80±10;<br>M: 95±10;<br>L: 110±10;<br>XL: 120±10                                                                                                                                     | Length:<br>S:≥220;<br>M/L/XL: ≥230;<br>Width:<br>S: 80±10;<br>M: 95±10;<br>L: 110±10;<br>XL: 120±10                                                                                                                  | Same       |

#### Table1-General Comparison

{5}------------------------------------------------

|                                | Thickness(mm)         | Finger: ≥0.05;<br>Palm: ≥0.05                                                                                                                                                                                             | Finger: ≥0.05;<br>Palm: ≥0.05 | Same                                                                                                                                                                                                                      |            |      |
|--------------------------------|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|------|
| Physic<br>al<br>Proper<br>ties | Before<br>Aging       | Tensile<br>Strength                                                                                                                                                                                                       | 14MPa, min                    | Tensile<br>Strength                                                                                                                                                                                                       | 14MPa, min | Same |
|                                |                       | Ultimate<br>Elongation                                                                                                                                                                                                    | 500% min                      | Ultimate<br>Elongation                                                                                                                                                                                                    | 500% min   | Same |
|                                | After<br>Aging        | Tensile<br>Strength                                                                                                                                                                                                       | 14MPa, min                    | Tensile<br>Strength                                                                                                                                                                                                       | 14MPa, min | Same |
|                                |                       | Ultimate<br>Elongation                                                                                                                                                                                                    | 400%min                       | Ultimate<br>Elongation                                                                                                                                                                                                    | 400%min    | Same |
|                                | Freedom from<br>Holes | Be free from holes when<br>tested in accordance with<br>ASTMD5151 AQL=2.5                                                                                                                                                 |                               | Be free from holes when<br>tested in accordance with<br>ASTMD5151 AQL=2.5                                                                                                                                                 |            | Same |
|                                | Powder Content        | Meet the requirements of<br>ASTM D6124                                                                                                                                                                                    |                               | Meet the requirements of<br>ASTM D6124                                                                                                                                                                                    |            | Same |
|                                | Biocompatibility      | ISO 10993-10;<br>Under the conditions of the<br>study, not an irritant or a<br>sensitizer                                                                                                                                 |                               | ISO 10993-10;<br>Under the conditions of the<br>study, not an irritant or a<br>sensitizer                                                                                                                                 |            | Same |
|                                |                       | ISO 10993-11;<br>Under the<br>condition of acute<br>systemic toxicity test,<br>the test article did not<br>show acute systemic<br>toxicity in vivo.<br>ISO 10993-5<br>Under conditions of the<br>study, device extract is |                               | ISO 10993-11;<br>Under the<br>condition of acute<br>systemic toxicity test,<br>the test article did not<br>show acute systemic<br>toxicity in vivo.<br>ISO 10993-5<br>Under conditions of the<br>study, device extract is |            | Same |

#### 8.0 Summary of Non-clinical Testing

Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

ISO 10993-5 Third edition 2009-06-01, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity.

ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.

ISO 10993-11:2017, Biological evaluation of medical devices - Part 11: Tests for

{6}------------------------------------------------

systemic toxicity.

ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves

ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.

ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.

| Test<br>Method  | Purpose                                             | Acceptance Criteria                                                                                                                                           | Results                                                                                                                                                                              |
|-----------------|-----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ASTM<br>D6319   | Physical<br>Dimensions<br>Test                      | Length(mm):<br>S:≥220;<br>M/L/XL:≥230;<br>Width(mm):<br>S: 80±10;<br>M: 95±10;<br>L: 110±10;<br>XL: 120±10<br>Thickness (mm):<br>Finger: ≥0.05<br>Palm: ≥0.05 | Length(mm):<br>> 230/Pass;<br>Width(mm):<br>S: 83-86 /Pass<br>M: 94-98/ Pass<br>L: 101-103/ Pass<br>XL:113-115/ Pass<br>Thickness (mm):<br>Finger: 0.13-0.15/Pass<br>Palm: 0.09/Pass |
| ASTM<br>D5151   | Watertightness<br>Test for<br>Detection of<br>Holes | Meet the requirements of ASTM D5151<br>AQL 2.5                                                                                                                | 0/125/Pass                                                                                                                                                                           |
| ASTM<br>D6124   | Powder<br>Content                                   | Meet the requirements of ASTM D6124 <<br>2.0mg                                                                                                                | 0.21-0.24mg/Pass;                                                                                                                                                                    |
| ASTM<br>D412    | Physical<br>properties                              | Before Aging<br>Tensile Strength ≥14MPa<br>Ultimate Elongation ≥500%<br>After Aging<br>Tensile Strength ≥14MPa<br>Ultimate Elongation ≥400%                   | 14.3-20.7MPa/Pass;<br>500-699%/Pass;<br>14.0-17.8MPa/Pass;<br>490-569%/Pass;                                                                                                         |
| ISO<br>10993-11 | Cytotoxicity                                        | Non- acute systemic<br>toxicity                                                                                                                               | Under conditions of<br>the study, did not<br>show acute systemic                                                                                                                     |
|                 |                                                     |                                                                                                                                                               | toxicity in vivo / Pass                                                                                                                                                              |
| ISO<br>10993-10 | Irritation                                          | Non-irritating                                                                                                                                                | Under the conditions<br>of the study, not an<br>irritant/ Pass                                                                                                                       |
| ISO<br>10993-10 | Sensitization                                       | Non-sensitizing                                                                                                                                               | Under conditions of<br>the study, not a<br>sensitizer./ Pass                                                                                                                         |

Table 2 - Summary of non-clinical performance testing

{7}------------------------------------------------

## 9.0 _Summary of Clinical Testing

Clinical testing is not needed for this device.

#### 10.0 Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device Nitrile Examination Gloves is as safe, as effective, and performs as well as or better than the legally marketed predicated device K203593.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA/K213187](https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA/K213187)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).