# Nitrile Examination Gloves (K212924)

_Shanxi Nacosa Medical Technology Co.,Ltd · LZA · Dec 13, 2021 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA/K212924

## Device Facts

- **Applicant:** Shanxi Nacosa Medical Technology Co.,Ltd
- **Product Code:** [LZA](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA.md)
- **Decision Date:** Dec 13, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

The nitrile examination gloves are intended to be worn on the hands of examiners to prevent contamination between patient and examiner. This is a single-use, powder-free, non-sterile device.

## Device Story

Nitrile examination gloves; worn on hands of examiners; prevent cross-contamination between patient and examiner. Single-use; powder-free; non-sterile; blue color; available in S, M, L, XL sizes. Ambidextrous design. Used in clinical settings by healthcare providers. Provides physical barrier protection. Benefits patient and examiner by reducing risk of contamination during examinations.

## Clinical Evidence

No clinical data. Bench testing only. Compliance verified via ASTM D6319-19 (physical dimensions/properties), ASTM D5151-19 (watertightness/holes), and ASTM D6124-06 (residual powder). Biocompatibility testing performed per ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation/sensitization), and ISO 10993-11 (acute systemic toxicity).

## Technological Characteristics

Material: Nitrile. Form factor: Ambidextrous, blue, non-sterile gloves. Sizes: S, M, L, XL. Standards: ASTM D6319-19, ASTM D5151-19, ASTM D6124-06. Biocompatibility: ISO 10993-5, ISO 10993-10, ISO 10993-11. No software or electronic components.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Nitrile examination gloves ([K203593](/device/K203593.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 13, 2021

Shanxi Nacosa Medical Technology Co.,Ltd % Boyle Wang Official Correspondent Shanghai Truthful information Technology Co., Ltd. RM. 1801, No. 161 Lujiazui East Rd.,Pudong Shanghai, Shanghai 200120 China

Re: K212924

Trade/Device Name: Nitrile Examination Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: September 7, 2021 Received: September 14, 2021

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray, III, PhD. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K212924

Device Name Nitrile Examination Gloves

Indications for Use (Describe)

The nitrile examination gloves are intended to be worn on the hands of examiners to prevent contamination between patient and examiner. This is a single-use, powder-free, non-sterile device.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary K212924

This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.

#### 1.0 Submitter's Information

Name: Shanxi Nacosa Medical Technology Co.,Ltd Address: Chengxi Industrial Park, Wenxi Economic and Technological Development Zone, Yuncheng City, Shanxi Province, China. Phone Number: +86 15955123917 Contact: Lu Yingying Date of Preparation: Sept 7, 2021

#### Designated Submission Correspondent

Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 1801, No. 161 East Lujiazui Rd., Pudong,Shanghai 200120 ,China Tel: +86-21-50313932 Email: Info@truthful.com.cn

#### 2.0 Device Information

Trade name: Nitrile Examination Gloves Common name: Patient Examination Gloves Classification name: Non-powdered patient examination glove Model(s): S, M, L, XL

#### 3.0 Classification

Production code: LZA Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital

#### 4.0 Predicate Device Information

Manufacturer: GUANG DONG KINGFA SCI. & TECH.CO.,LTD. Device: Nitrile examination gloves 510(k) number: K203593

{4}------------------------------------------------

#### 5.0 Indication for Use

The disposable medical nitrile examination gloves are intended to be worn on the hands of examiners to prevent contamination between patient and examiner. This is a single-use, powder-free, non-sterile device.

#### 6.0 Device Description

The subject device is powder free nitrile examination gloves. The subject device is blue. It can be available in four specifications: S,M,L and XL. The subject device is non-sterile.

#### 7.0 Technological Characteristic Comparison Table

| Table1-General Comparison   |                                                                                                                                                                                                                      |                                                                                                                                                                                                                      |            |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Item                        | Subject Device<br>(K212924)                                                                                                                                                                                          | Predicated Device<br>(K203593)                                                                                                                                                                                       | Comparison |
| Product Code                | LZA                                                                                                                                                                                                                  | LZA                                                                                                                                                                                                                  | Same       |
| Regulation No.              | 21CFR880.6250                                                                                                                                                                                                        | 21CFR880.6250                                                                                                                                                                                                        | Same       |
| Class                       | I                                                                                                                                                                                                                    | I                                                                                                                                                                                                                    | Same       |
| Intended Use                | The nitrile examination<br>glove is intended to be<br>worn on the hands of<br>examiners to<br>prevent contamination<br>between patient and<br>examiner. This is a<br>single-use, powder-free,<br>non-sterile device. | The nitrile examination<br>glove is intended to be<br>worn on the hands of<br>examiners to<br>prevent contamination<br>between patient and<br>examiner. This is a<br>single-use, powder-free,<br>non-sterile device. | Same       |
| Material                    | Nitrile                                                                                                                                                                                                              | Nitrile                                                                                                                                                                                                              | Same       |
| Powdered or<br>Powered free | Powdered free                                                                                                                                                                                                        | Powdered free                                                                                                                                                                                                        | Same       |
| Design Feature              | Ambidextrous                                                                                                                                                                                                         | Ambidextrous                                                                                                                                                                                                         | Same       |
| Colorant                    | Blue                                                                                                                                                                                                                 | Blue                                                                                                                                                                                                                 | Same       |
| Labeling<br>Information     | Single-use indication,<br>powder free, device<br>color, device name,<br>glove size and quantity,<br>Non-Sterile                                                                                                      | Single-use indication,<br>powder free, device<br>color, device name,<br>glove size and quantity,<br>Non-Sterile                                                                                                      | Same       |
| Dimensions(mm)              | Length(mm):<br>>230;<br>Width(mm):<br>S: Average 84mm<br>M: Average 95mm                                                                                                                                             | Length:<br>S (220mm min)<br>M (230mm min)<br>L (230mm min)<br>XL (230mm min)                                                                                                                                         | Similar    |

#### Table1-General Comparison

{5}------------------------------------------------

|                        |                 | L: Average 111mm<br>XL: Average 115mm                                                                                                               | Width:<br>Small (80 $\pm$ 10mm)<br>Medium (95 $\pm$ 10mm)<br>Large (110 $\pm$ 10mm)<br>X large (120 $\pm$ 10mm) |                                                                                                                                                                               |            |         |
|------------------------|-----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|---------|
|                        |                 | Meet the requirements<br>of ASTM D6319-19                                                                                                           | meet the requirements<br>of ASTM D6319-19                                                                       |                                                                                                                                                                               |            |         |
| Thickness(mm)          |                 | Finger: 0.12-0.15<br>Palm: 0.08-0.10                                                                                                                | Palm: 0.05mm min<br>Finger: 0.05mm min                                                                          | Similar                                                                                                                                                                       |            |         |
| Physical<br>Properties | Before<br>Aging | Tensile<br>Strength                                                                                                                                 | 17-38MPa                                                                                                        | Tensile<br>Strength                                                                                                                                                           | 14MPa, min | Similar |
|                        |                 | Ultimate<br>Elongation                                                                                                                              | 501-565%                                                                                                        | Ultimate<br>Elongation                                                                                                                                                        | 500% min   | Similar |
|                        | After<br>Aging  | Tensile<br>Strength                                                                                                                                 | 18-43MPa                                                                                                        | Tensile<br>Strength                                                                                                                                                           | 14MPa, min | Similar |
|                        |                 | Ultimate<br>Elongation                                                                                                                              | 500-564%                                                                                                        | Ultimate<br>Elongation                                                                                                                                                        | 400%min    | Similar |
| Freedom from<br>Holes  |                 | Be free from holes when<br>tested in accordance<br>with ASTMD5151<br>AQL=2.5                                                                        |                                                                                                                 | Be free from holes when<br>tested in accordance<br>with ASTMD5151<br>AQL=2.5                                                                                                  |            | Same    |
| Powder Content         |                 | 0.1-0.3mg<br>Meet the requirements of<br>ASTM D6124<br><2.0 mg/gloves                                                                               |                                                                                                                 | Meet the requirements of<br>ASTM D6124<br><2.0 mg/gloves                                                                                                                      |            | Similar |
| Biocompatibility       |                 | ISO 10993-10;                                                                                                                                       |                                                                                                                 | ISO 10993-10;                                                                                                                                                                 |            |         |
|                        |                 | Under the conditions of<br>the study, not an irritant<br>or a sensitizer                                                                            |                                                                                                                 | Under the test condition<br>of study not a sensitizer.<br>Under the test condition<br>of study not an irritant.                                                               |            | Same    |
|                        |                 | ISO 10993-11;<br>Under the<br>condition of acute<br>systemic toxicity test,<br>the test article did not<br>show acute systemic<br>toxicity in vivo. |                                                                                                                 | Cytotoxicity is assessed<br>via rationale. Under the<br>condition of acute<br>systemic toxicity test, the<br>test article did not show<br>acute systemic toxicity in<br>vivo. |            | Same    |
|                        |                 | ISO 10993-5<br>Under conditions of the<br>study, device extract is<br>cytotoxic                                                                     |                                                                                                                 | N.A.                                                                                                                                                                          |            | /       |

Analysis:

{6}------------------------------------------------

The physical dimensions, physical properties and powder content are different with that of the predicate, but they all meet the requirements of ASTM D6319-19.

## 8.0 Summary of Non-clinical Testing

Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

ISO 10993-5 Third edition 2009-06-01, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity

ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.

ISO 10993-11:2017, Biological evaluation of medical devices - Part 11: Tests

for systemic toxicity.

ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves

ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.

ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.

| Test<br>Method | Purpose                        | Acceptance Criteria                            | Results                                   |
|----------------|--------------------------------|------------------------------------------------|-------------------------------------------|
| ASTM<br>D6319  | Physical<br>Dimensions<br>Test | Length(mm):<br>S: ≥220;                        | Length(mm):<br>> 230/Pass;                |
|                |                                | M/L/XL: ≥230;                                  | Width(mm):                                |
|                |                                | Width(mm):<br>S: 80±10;                        | S: 83-85 /Pass                            |
|                |                                | M: 95±10;                                      | M: 94-96/Pass                             |
|                |                                | L: 110±10;                                     | L: 110-112/Pass                           |
|                |                                | XL: 120±10                                     | XL:113-117/ Pass                          |
|                |                                | Thickness (mm):<br>Finger: ≥0.05               | Thickness (mm):<br>Finger: 0.12-0.15/Pass |
|                |                                | Palm: ≥0.05                                    | Palm: 0.08-0.10/Pass                      |
|                |                                |                                                |                                           |
| ASTM<br>D5151  | Watertightness<br>Test<br>for  | Meet the requirements of ASTM D5151<br>AQL 2.5 | 0/200/Pass                                |

Table 2 - Summary of non-clinical performance testing

{7}------------------------------------------------

|                 | Detection of<br>Holes      |                                             |                        |                                                                   |                                                                                             |
|-----------------|----------------------------|---------------------------------------------|------------------------|-------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| ASTM<br>D6124   | Powder<br>Content          | Meet the requirements of ASTM D6124 < 2.0mg |                        |                                                                   | 0.1-0.3mg/Pass;                                                                             |
| ASTM<br>D412    | Physical<br>properties     | Before<br>Aging                             | Tensile<br>Strength    | ≥14MPa                                                            | 17-38MPa/Pass;                                                                              |
|                 |                            |                                             | Ultimate<br>Elongation | ≥500%                                                             | 501-565%/Pass;                                                                              |
|                 |                            | After<br>Aging                              | Tensile<br>Strength    | ≥14MPa                                                            | 18-43MPa/Pass;                                                                              |
|                 |                            |                                             | Ultimate<br>Elongation | ≥400%                                                             | 500-564%/Pass;                                                                              |
| ISO<br>10993-5  | Cytotoxicity               | toxicity                                    |                        | Under conditions of<br>the study, device<br>extract is cytotoxic. |                                                                                             |
| ISO<br>10993-11 | Acute Systemic<br>Toxicity | Non- acute systemic<br>toxicity             |                        |                                                                   | Under conditions of<br>the study, did not<br>show acute systemic<br>toxicity in vivo / Pass |
| ISO<br>10993-10 | Irritation                 | Non-irritating                              |                        | Under the conditions<br>of the study, not an<br>irritant/ Pass    |                                                                                             |
| ISO<br>10993-10 | Sensitization              | Non-sensitizing                             |                        | Under conditions of<br>the study, not a<br>sensitizer./ Pass      |                                                                                             |

### 9.0 Summarv of Clinical Testing

Clinical testing is not needed for this device.

#### 10.0 Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device Nitrile Examination Gloves is as safe, as effective, and performs as well as or better than the legally marketed predicated device K203593.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA/K212924](https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA/K212924)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).