← Product Code [LZA](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA) · K212485

# Disposable Nitrile Examination gloves (K212485)

_Fujian Yuanxin Safety Protection Technology Co., Ltd. · LZA · Nov 5, 2021 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA/K212485

## Device Facts

- **Applicant:** Fujian Yuanxin Safety Protection Technology Co., Ltd.
- **Product Code:** [LZA](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA.md)
- **Decision Date:** Nov 5, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

The Disposable Nitrile Examination gloves are disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner.

## Device Story

Disposable nitrile examination gloves; non-sterile; ambidextrous design. Used by clinicians or patients to provide a protective barrier against contamination during medical examinations or procedures. Device acts as a physical barrier; prevents cross-contamination between examiner and patient. Manufactured from nitrile rubber; blue color. Performance verified through physical property testing (tensile strength, elongation), biocompatibility (irritation, sensitization, cytotoxicity), and leak testing. Used in clinical or home settings; no specialized operation required.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified via ASTM D6124 (residual powder), ASTM D5151 (freedom from holes), and ASTM D6319 (physical properties). Biocompatibility verified via ISO 10993-10 (irritation/sensitization) and ISO 10993-5 (cytotoxicity).

## Technological Characteristics

Material: Nitrile rubber. Form factor: Ambidextrous, non-sterile, blue. Standards: ASTM D6124, ASTM D5151, ASTM D6319. Biocompatibility: ISO 10993-5, ISO 10993-10. Physical properties: Tensile strength >= 14 MPa, Ultimate elongation >= 500% (before aging).

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Disposable Powder Free Nitrile Examination Glove, White/ Blue/ Black/ Pink Color ([K171422](/device/K171422.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Nov 5, 2021

Fujian Yuanxin Safety Protection Technology Co., Ltd. Boyle Wang General Manager Shanghai Truthful Information Technology Co., Ltd. Room608.No.738.Shangcheng Rd.,Pudong Shanghai, Shanghai 200120 China

Re: K212485

Trade/Device Name: Disposable Nitrile Examination gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: July 30, 2021 Received: August 9, 2021

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K212485

Device Name Disposable Nitrile Examination gloves

Indications for Use (Describe)

The Disposable Nitrile Examination gloves are disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# K212485 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with 21 CFR 807.92.

#### 1.0 submitter's information

Name: Fujian Yuanxin Safety Protection Technology Co., Ltd. Address: Xiacun Medical Industry Zone,Zherong County,Ningde,Fujian, 355300, China Phone Number: +86-593-8311666 Contact: Alan Wu Date of Preparation: 2021.07.30

#### Designated Submission Correspondent

Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Tel: +86-21-50313932 Email: Info@truthful.com.cn

#### 2.0 Device information

Trade name: Disposable Nitrile Examination gloves Common name: Patient Examination Gloves Classification name: Non-powdered patient examination glove Model(s): XS, S, M, L, XL

#### 3.0 Classification

Production code: LZA Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital

#### 4.0 Predicate device information

Manufacturer: Ever Global (Vietnam) Enterprise Corp Device: Disposable Powder Free Nitrile Examination Glove, White/ Blue/ Black/ Pink Color 510(k) number: K171422

FROM THE PIT

{4}------------------------------------------------

#### 5.0 Intended use

The Disposable Nitrile Examination gloves are disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner.

#### 6.0 Device description

The proposed device is Powder Free Disposable Nitrile Examination gloves. The proposed device is blue. The design of proposed device is addressing the standards as ASTM D6124,ASTM D5151, and ASTM D6319. The proposed device is non-sterile.

#### 7.0 Summary comparing technological characteristics with predicate device

| Item                     | Proposed device                                                                                                                                                                            | Predicated device                                                                                                                                                                                                                     | Remark |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| 510(k) number            | Pending                                                                                                                                                                                    | K171422                                                                                                                                                                                                                               |        |
| Product Code             | LZA                                                                                                                                                                                        | LZA                                                                                                                                                                                                                                   | Same   |
| Regulation No.           | 21CFR880.6250                                                                                                                                                                              | 21CFR880.6250                                                                                                                                                                                                                         | Same   |
| Class                    | I                                                                                                                                                                                          | I                                                                                                                                                                                                                                     | Same   |
| Intended Use             | The Disposable Nitrile Examination gloves is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner. | The Disposable Powder Free Nitrile Examination Glove, White/ Blue/ Black/ Pink Color is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner. | Same   |
| Powdered or Powered free | Powdered free                                                                                                                                                                              | Powdered free                                                                                                                                                                                                                         | Same   |
| Design Feature           | ambidextrous                                                                                                                                                                               | ambidextrous                                                                                                                                                                                                                          | Same   |
| Labeling Information     | Single-use indication, powder free, device color, device name, glove size and quantity, Disposable Nitrile Examination gloves, Non-Sterile                                                 | Single-use indication, powder free, device color, device name, glove size and quantity, Disposable Powder Free Nitrile Examination Glove, Non-Sterile                                                                                 | Same   |

#### Table1-General Comparison

#### Table2 Device Dimensions Comparison

| Predicate       | Designation | Size |   |   |   |    | Tolerance |
|-----------------|-------------|------|---|---|---|----|-----------|
| Device(K171422) |             | XS   | S | M | L | XL |           |

{5}------------------------------------------------

|                 | Length, mm     | 230  | 230 | 230 | 230 | 230 | min       |
|-----------------|----------------|------|-----|-----|-----|-----|-----------|
|                 | Width, mm      | 75   | 85  | 95  | 105 | 115 | $\pm 5$   |
|                 | Thickness, mm: |      |     |     |     |     |           |
|                 | Finger         | 0.05 |     |     |     |     | min       |
|                 | Palm           | 0.05 |     |     |     |     | min       |
| Proposed Device | Designation    | Size |     |     |     |     | Tolerance |
|                 |                | XS   | S   | M   | L   | XL  |           |
|                 | Length, mm     | 220  | 220 | 230 | 230 | 230 | min       |
|                 | Width, mm      | 70   | 80  | 95  | 110 | 120 | $\pm 10$  |
|                 | Thickness, mm: |      |     |     |     |     |           |
|                 | Finger         | 0.05 |     |     |     |     | min       |
|                 | Palm           | 0.05 |     |     |     |     | min       |
| Remark          | Analysis1      |      |     |     |     |     |           |

#### Table3 Performance Comparison

| Item                   |                 |                        | Proposed device                                                     | Predicated device                                                   | Remark    |
|------------------------|-----------------|------------------------|---------------------------------------------------------------------|---------------------------------------------------------------------|-----------|
| Colorant               |                 |                        | blue                                                                | White/ Blue/ Black/ Pink                                            | Analysis2 |
| Physical<br>Properties | Before<br>Aging | Tensile<br>Strength    | 14MPa, min                                                          | 14MPa, min                                                          | SAME      |
|                        |                 | Ultimate<br>Elongation | 500%min                                                             | 500%min                                                             | SAME      |
|                        | After<br>Aging  | Tensile<br>Strength    | 14MPa, min                                                          | 14MPa, min                                                          | SAME      |
|                        |                 | Ultimate<br>Elongation | 400%min                                                             | 400%min                                                             | SAME      |
| Comply with ASTM D6319 |                 |                        | Comply with ASTM D6319                                              |                                                                     | SAME      |
| Freedom from Holes     |                 |                        | Be free from holes when tested in accordance with ASTMD5151 AQL=2.5 | Be free from holes when tested in accordance with ASTMD5151 AQL=2.5 | SAME      |
| Powder Content         |                 |                        | 0.16                                                                | Meet the requirements of ASTM D6124                                 | SIMILAR   |

Analysis1: The proposed device has different sizes to the predicate device, but all proposed devices are conducted the properties test, the test results shown that the sizes comply with the requirements of standard ASTM D6319-19, Standard Specification For Disposable Nitrile Examination gloves For Medical Application.

Analysis2: The proposed device has different color to the predicate device, but all proposed devices are conducted the biocompatibility and performance tests, the test results shown that the color difference does not affect the safety and efficacy of proposed device.

{6}------------------------------------------------

| Item               | Proposed device | Predicated device                                                              | Remark                  |         |
|--------------------|-----------------|--------------------------------------------------------------------------------|-------------------------|---------|
| Material           | Nitrile         | Nitrile                                                                        | SAME                    |         |
| Biocompatibility   | Irritation      | Under the conditions of the study, not an irritant                             | Comply with ISO10993-10 | SAME    |
|                    | Sensitization   | Under conditions of the study, not a sensitizer.                               |                         |         |
|                    | Cytotoxicity    | Under conditions of the study, did not show potential toxicity to L-929 cells. | Comply with ISO10993-5  | SIMILAR |
| Label and Labeling |                 | Meet FDA's Requirement                                                         | Meet FDA's Requirement  | SAME    |

#### Biocompatibility Testing Comparison Table4

# 8.0 Summary of Non-Clinical Performance Testing

The following performance data has been provided to demonstrate that the subject device meets the acceptance criteria in the standard.

# Table 5 Summary of Non-Clinical Performance Testing

| No. | Name of the Test<br>Methodology / Standard                                                                                     | Purpose                                                                                                                                                                                     | Acceptance Criteria                                                                                                                                                                                                                                                                                                                                                                 | Results                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|-----|--------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1   | ISO 10993-10:2010<br>Biological Evaluation Of<br>Medical Devices - Part<br>10: Tests For Irritation<br>And Skin Sensitization. | This part of ISO 10993 assesses possible contact hazards from chemicals released from medical devices, which may produce skin and mucosal irritation, eye irritation or skin sensitization. | Skin Sensitization Test:<br>provided grades less than 1, otherwise sensitization.                                                                                                                                                                                                                                                                                                   | All grades are 0.<br><br>All animals were survived and no abnormal signs were observed during the study.                                                                                                                                                                                                                                                                                                                                           |
| 2   |                                                                                                                                |                                                                                                                                                                                             | Skin Irritation Test:<br>If the primary irritation index is 0-0,4, the response category is Negligible.<br>0,5-1,9 means slight<br>2-4,9 means moderate<br>5-8 means severe                                                                                                                                                                                                         | The primary irritation index is 0.<br><br>The response of the proposed device was categorized as negligible under the test condition                                                                                                                                                                                                                                                                                                               |
| 3   | ISO 10993-5:2009<br>Biological Evaluation Of<br>Medical Devices - Part<br>5: Tests For In Vitro<br>Cytotoxicity                | This part of ISO 10993 describes<br>test methods to<br>assess the in vitro<br>cytotoxicity of<br>medical devices.                                                                           | The viab.% of the<br>100% extract of the<br>test article is the final<br>result, and if viability is<br>reduced to $<70%$ of<br>the blank, it has<br>cytotoxic potential.                                                                                                                                                                                                           | Viab.% of 100% test article<br>extract is 86.5%<br>It means the proposed device<br>have no potential toxicity to<br>L-929 in the MTT method                                                                                                                                                                                                                                                                                                        |
| 4   | ASTM D6124-06<br>(Reapproved<br>2017),<br>Standard Test Method<br>for Residual Powder on<br>Medical Gloves                     | This standard is<br>designed to<br>determine the<br>amount ofresidual<br>powder (or<br>filter-retained<br>mass) found<br>on medical gloves                                                  | powder residue limit of<br>2.0 mg                                                                                                                                                                                                                                                                                                                                                   | 0.16mg                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| 5   | ASTM<br>D5151-06(Reapproved2<br>015), Standard Test<br>Method for Detection of<br>Holes in Medical Gloves.                     | This test method<br>covers the<br>detection of holes<br>in<br>medical gloves.                                                                                                               | Samples number: 125<br>gloves<br>AQL: 2.5 (ISO 2859)<br>Criterion $≤$ 7 gloves<br>for water leakage                                                                                                                                                                                                                                                                                 | no glove water leakage found                                                                                                                                                                                                                                                                                                                                                                                                                       |
| 6   | ASTM<br>D6319-10(Reapproved<br>2015),Standard<br>Specification For Nitrile<br>Examination Gloves For<br>Medical Application.   | This specification<br>covers certain<br>requirements for<br>nitrile rubber<br>gloves used in<br>conducting<br>medical<br>examinations and<br>diagnostic and<br>therapeutic<br>procedures.   | Sterility: no need<br>Freedom from holes:<br>pl. Refer to No. 5 in<br>table 5<br>Dimensions:<br>S: width 80 $±$ 10mm<br>Length $≥$ 220 mm<br>M: width 95 $±$ 10mm<br>Length $≥$ 230 mm<br>L: width 110 $±$ 10mm<br>Length $≥$ 230 mm<br>XL: width 120 $±$ 10mm<br>Length $≥$ 230 mm<br>Thickness:<br>Finger $≥$ 0.05 mm<br>Palm $≥$ 0.05 mm<br>Physical properties:<br>Before aging | N.A.<br>Please refer to No. 5 in table 5<br>Dimensions:<br>S: width: 83-86 mm<br>Length 253-266 mm<br>M: width 96-98 mm<br>Length 243-262 mm<br>L: width 106-110 mm<br>Length 247-254 mm<br>XL: width 112-118 mm<br>Length 245-252 mm<br>Thickness:<br>Finger 0.09-0.11 mm<br>Palm 0.08-0.11 mm<br>Physical properties:<br>Before aging<br>Tensile strength 15.7-17.7 MPa<br>Ultimate Elongation 532.284% -<br>552.072%<br>After Accelerated Aging |

{7}------------------------------------------------

{8}------------------------------------------------

|  | Ultimate Elongation $\ge$ 500%<br>After Accelerated Aging<br>Tensile strength $\ge$ 14MPa<br>Ultimate Elongation $\ge$ 400%<br>Powder-free Residue:<br>pl. Refer to No. 4 in table 5 | Tensile strength 15.2-17.8 MPa<br>Ultimate Elongation 525.947% - 548.352%<br>Powder-free Residue:<br>pl. Refer to No. 4 in table 5 |
|--|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
|--|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|

### 9. Summary of Clinical Performance Test

No clinical study is included in this submission.

### 10.0 Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicated device.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA/K212485](https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA/K212485)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
