← Product Code [LZA](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA) · K212078

# Medical Examination Gloves (K212078)

_Guangdong Gymda Medical Technology Co.,Ltd · LZA · Oct 8, 2021 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA/K212078

## Device Facts

- **Applicant:** Guangdong Gymda Medical Technology Co.,Ltd
- **Product Code:** [LZA](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA.md)
- **Decision Date:** Oct 8, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

The Medical Examination Gloves are disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner.

## Device Story

Disposable, non-sterile, blue nitrile examination gloves; worn by healthcare examiners to provide a barrier against cross-contamination between patient and examiner. Device is ambidextrous and available in sizes S, M, L, and XL. Used in clinical or medical environments. Provides physical barrier protection; performance verified through tensile strength, elongation, and freedom from holes testing per ASTM standards. Biocompatibility assessed via irritation, sensitization, cytotoxicity, and systemic toxicity testing.

## Clinical Evidence

No clinical data. Bench testing only. Biocompatibility testing performed per ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation/sensitization), and ISO 10993-11 (systemic toxicity). Physical performance testing conducted per ASTM D6124, ASTM D5151, ASTM D6319, and ASTM D7160.

## Technological Characteristics

Material: Nitrile. Form factor: Ambidextrous, non-sterile, blue, powder-free. Standards: ASTM D6124 (residual powder), ASTM D5151 (holes), ASTM D6319 (specifications), ASTM D7160 (expiration). Biocompatibility: ISO 10993-5, ISO 10993-10, ISO 10993-11.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Disposable Powder Free Nitrile Examination Glove, White/ Blue/ Black/ Pink Color ([K171422](/device/K171422.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 8, 2021

Guangdong Gymda Medical Technology Co., Ltd % Boyle Wang General Manager Shanghai Truthful Information Technology Co., Ltd. Room 608, No.738, Shangcheng Rd., Pudong Shanghai, Shanghai 200120 China

Re: K212078

Trade/Device Name: Medical Examination Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: September 6, 2021 Received: September 10, 2021

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K212078

Device Name Medical Examination Gloves

Indications for Use (Describe)

The Medical Examination Gloves are disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with 21 CFR 807.92.

#### 1.0 submitter's information

Name: Guanqdong Gymda Medical Technology Co.,Ltd Address: No.13, Quan'an Third Road, Phase 2 of High-tech Zone, Nanxiong City, Shaoguan City, Guangdong Province, 512400, China Phone Number: +86-20-37738661 Contact: Olivia Chen Date of Preparation: 2021.05.27

#### Designated Submission Correspondent

Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Tel: +86-21-50313932 Email: Info@truthful.com.cn

# 2.0 Device information

Trade name: Medical Examination Gloves Common name: Patient Examination Gloves Classification name: Non-powdered patient examination glove Model(s): S, M, L, XL

#### 3.0 Classification

Production code: LZA Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital

#### 4.0 Predicate device information

Manufacturer: Ever Global (Vietnam) Enterprise Corp Device: Disposable Powder Free Nitrile Examination Glove, White/ Blue/ Black/ Pink Color 510(k) number: K171422

{4}------------------------------------------------

#### 5.0 Intended use

The Medical Examination Gloves are disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner.

#### 6.0 Device description

The proposed device is Powder Free Medical Examination Gloves. The proposed device is blue. The design of proposed device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D6319. The proposed device is non-sterile.

#### 7.0 Summary comparing technological characteristics with predicate device

| Table1-General Comparison |                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                  |        |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Item                      | Proposed device                                                                                                                                                                                      | Predicated device                                                                                                                                                                                                                                                | Remark |
| 510(k) number             | Pending                                                                                                                                                                                              | K171422                                                                                                                                                                                                                                                          |        |
| Product Code              | LZA                                                                                                                                                                                                  | LZA                                                                                                                                                                                                                                                              | Same   |
| Regulation No.            | 21CFR880.6250                                                                                                                                                                                        | 21CFR880.6250                                                                                                                                                                                                                                                    | Same   |
| Class                     | I                                                                                                                                                                                                    | I                                                                                                                                                                                                                                                                | Same   |
| Intended Use              | The Medical Examination<br>Gloves is a disposable<br>device intended for<br>medical purposes that is<br>worn on the examiner's<br>hands to prevent<br>contamination between<br>patient and examiner. | The Disposable Powder<br>Free Nitrile Examination<br>Glove, White/ Blue/ Black/<br>Pink Color is a disposable<br>device intended for<br>medical purposes that is<br>worn on the examiner's<br>hands to prevent<br>contamination between<br>patient and examiner. | Same   |
| Powdered or Powered free  | Powdered free                                                                                                                                                                                        | Powdered free                                                                                                                                                                                                                                                    | Same   |
| Design Feature            | ambidextrous                                                                                                                                                                                         | ambidextrous                                                                                                                                                                                                                                                     | Same   |
| Labeling Information      | Single-use indication,<br>powder free, device color,<br>device name, glove size<br>and quantity, Medical<br>Examination Gloves,<br>Non-Sterile                                                       | Single-use indication,<br>powder free, device color,<br>device name, glove size<br>and quantity, Disposable<br>Powder Free Nitrile<br>Examination Glove,<br>Non-Sterile                                                                                          | Same   |

#### Table1-General Comparison

# Table2 Device Dimensions Comparison

{5}------------------------------------------------

| Predicate       | Designation    | Size |     |     |     |     | Tolerance |
|-----------------|----------------|------|-----|-----|-----|-----|-----------|
| Device(K171422) |                | XS   | S   | M   | L   | XL  |           |
|                 | Length, mm     | 230  | 230 | 230 | 230 | 230 | min       |
|                 | Width, mm      | 75   | 85  | 95  | 105 | 115 | $\pm 5$   |
|                 | Thickness, mm: |      |     |     |     |     |           |
|                 | Finger         | 0.05 |     |     |     |     | min       |
|                 | Palm           | 0.05 |     |     |     |     | min       |
| Proposed Device | Designation    | Size |     |     |     |     | Tolerance |
|                 |                | S    | M   | L   | XL  |     |           |
|                 | Length, mm     | 220  | 230 | 230 | 230 |     | min       |
|                 | Width, mm      | 80   | 95  | 110 | 120 |     | $\pm 10$  |
|                 | Thickness, mm: |      |     |     |     |     |           |
|                 | Finger         | 0.05 |     |     |     |     | min       |
|                 | Palm           | 0.05 |     |     |     |     | min       |
| Remark          | Analysis 1     |      |     |     |     |     |           |

Analysis1: The sizes and tolerances of proposed device are different with those of the predicate, but they all meet the requirements of ASTM D6319-19, so the differences do not raise any new safety or performance questions.

| Item                   | Proposed device                                                     | Predicated device                                                   | Remark     |      |
|------------------------|---------------------------------------------------------------------|---------------------------------------------------------------------|------------|------|
| Colorant               | blue                                                                | White/ Blue/ Black/ Pink                                            | Analysis2  |      |
| Physical<br>Properties | Before<br>Aging<br>Tensile<br>Strength                              | 14MPa, min                                                          | 14MPa, min | SAME |
|                        | Ultimate<br>Elongation                                              | 500%min                                                             | 500%min    | SAME |
|                        | After<br>Aging<br>Tensile<br>Strength                               | 14MPa, min                                                          | 14MPa, min | SAME |
|                        | Ultimate<br>Elongation                                              | 400%min                                                             | 400%min    | SAME |
| Comply with ASTM D6319 |                                                                     | Comply with ASTM D6319                                              | SAME       |      |
| Freedom from Holes     | Be free from holes when tested in accordance with ASTMD5151 AQL=2.5 | Be free from holes when tested in accordance with ASTMD5151 AQL=2.5 | SAME       |      |
| Powder Content         | 0.15                                                                | Meet the requirements of ASTM D6124                                 | SIMILAR    |      |

#### Table3 Performance Comparison

Analysis 2: The proposed device has different color to the predicate device, but all proposed devices are conducted the biocompatibility test, the test results shown that the color difference do not effect the safety of proposed device

{6}------------------------------------------------

| Item               | Proposed device        | Predicated device                                                                                               | Remark                                           |           |
|--------------------|------------------------|-----------------------------------------------------------------------------------------------------------------|--------------------------------------------------|-----------|
| Material           | Nitrile                | Nitrile                                                                                                         | SAME                                             |           |
| Biocompatibility   | Irritation             | Under the conditions of the study, not an irritant                                                              | Comply with ISO10993-10                          | SAME      |
|                    | Sensitization          | Under conditions of the study, not a sensitizer.                                                                |                                                  |           |
|                    | Cytotoxicity           | Under the conditions of the study, the device is potentially cytotoxic                                          | Comply with ISO10993-5                           | Analysis3 |
|                    | Systemic toxicity      | Under the conditions of the study, the device does not elicit a systemic toxicity response in the model animal. | Complies with ISO 10993-11 Third edition 2017-09 |           |
| Label and Labeling | Meet FDA's Requirement | Meet Requirement                                                                                                | FDA's SAME                                       |           |

#### Table4 Safety Comparison

Analysis3: The proposed device is potentially cytotoxic, but all proposed devices are conducted the systemic toxicity test, the test results show that the proposed device is safe.

#### 8.0 Discussion of non-clinical testing

The following tests were performed to evaluate the biocompatibility of the proposed device, the tests' results to show substantial equivalence to the predicate device:

ISO 10993-10:2010 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.

ISO 10993-5:2009 Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity

ISO 10993-11 Third edition 2017-09, Biological evaluation of medical devices -Part 11: Tests for systemic toxicity.

Clinical testing is not needed for this device.

#### 9.0 Discussion of performance testing (Bench)

In summary, the performance testing of the subject device was conducted to adequately demonstrate the effectiveness of the device in accordance with the relevant test methods cited below:

{7}------------------------------------------------

ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves

ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.

ASTM D6319-19,Standard Specification For Medical Examination Gloves For Medical Application.

ASTM D7160-16,Standard Practice for Determination of Expiration Dating for Medical Gloves

#### Conclusion 10.0

The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as the legally marketed predicated device.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA/K212078](https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA/K212078)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
