← Product Code [LZA](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA) · K210990

# Non-sterile Powder Free Nitrile Examination Gloves (K210990)

_Rubberex Alliance Sdn Bhd · LZA · Nov 22, 2021 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA/K210990

## Device Facts

- **Applicant:** Rubberex Alliance Sdn Bhd
- **Product Code:** [LZA](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA.md)
- **Decision Date:** Nov 22, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

A non-sterile powder free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. The device is for over-the-counter use.

## Device Story

Disposable, non-sterile, powder-free nitrile examination glove; worn on examiner's hand/finger to prevent cross-contamination between patient and examiner. Used in clinical or home settings; operated by healthcare professionals or lay users. Device acts as a physical barrier. Benefits include protection against contamination. Subject device is blue nitrile; predicate is aqua green nitrile. No electronic or software components.

## Clinical Evidence

Bench testing only. Device evaluated for dimensions, physical properties (tensile strength, elongation), freedom from pinholes (ASTM D5151-19), residual powder (ASTM D6124-06), and biocompatibility (ISO 10993-5, 10, 11). All tests passed, confirming conformance with ASTM D6319-19.

## Technological Characteristics

Material: Nitrile compound. Form factor: Disposable, non-sterile, powder-free glove. Standards: ASTM D6319-19 (dimensions, physical properties), ASTM D412-16 (tension), ASTM D5151-19 (pinholes), ASTM D6124-06 (residual powder). Biocompatibility: ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation/sensitization), ISO 10993-11 (systemic toxicity). No energy source, connectivity, or software.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Powder Free Nitrile Examination Glove (Aqua Green) ([K200326](/device/K200326.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 22, 2021

Rubberex Alliance Sdn Bhd % Aristotle Nafpliotis Regulatory Affairs Consultant/Engineer mdi Consultants. Inc. 55 Northern Blvd Great Neck, New York 11021

Re: K210990

Trade/Device Name: Non-sterile Powder Free Nitrile Examination Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: October 18, 2021 Received: October 19, 2021

Dear Aristotle Nafpliotis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K210990

Device Name

Non-sterile Powder Free Nitrile Examination Gloves

Indications for Use (Describe)

A non-sterile powder free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. The device is for over-the-counter use.

## Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY

# K210990

#### 1. Submitter's Identification:

RUBBEREX ALLIANCE SDN BHD Lot 138201 Off ¾ Mile, Jalan Bercham Kawasan Perindustrian Bercham 31400 Ipoh, Perak, Malaysia

Date Summary Prepared: 17 November 2021

Contact: Sabri Hamid

#### 2. Name of the Device:

Non-sterile Powder Free Nitrile Examination Gloves K#: K210990 Regulation Number: 880.6250 Regulation Name: Polymer Patient Examination Glove Requlatory Class: 1 Product Code: LZA

#### 3. Information for the 510(k) Cleared Device (Predicate Device):

Predicate device: K200326 Trade/Device Name: Powder Free Nitrile Examination Glove (Aqua Green) Device Classification Name: Patient Examination gloves (21 CFR 880.6250) Device Class: Class I Product Code: LZA Applicant name: Riverstone Resources SDN BHD Lot 55, No 13 Jalan Jasmin 2 Kawasan Perindustrian Bukit Beruntung, 48300 My

#### 4. Device Description:

The subject device in this 510(k) Notification is Powder Free Nitrile Examination Glove. The subject device is a patient examination glove made from nitrile compound, powder free and non-sterile (Per 21 CFR 880.6250, class I). The device meets the specifications in ASTM D6319-19, standard specification for Nitrile Examination Gloves.

{4}------------------------------------------------

Specifications: Dimension and Thickness of Gloves (ASTM D6319-10)

| Dimension                       | Size XS  | Size S   | Size M   | Size L   | Size XL  |
|---------------------------------|----------|----------|----------|----------|----------|
| Overall Length (mm)             | min 220  | min 220  | min 230  | min 230  | min 230  |
| Width (± 10mm)                  | 70       | 80       | 95       | 110      | 120      |
| Thickness at Palm<br>(mm)       | min 0.05 | min 0.05 | min 0.05 | min 0.05 | min 0.05 |
| Thickness at Finger<br>Tip (mm) | min 0.05 | min 0.05 | min 0.05 | min 0.05 | min 0.05 |

|  |  |  | Specifications: Gloves Physical Properties and Holes (ASTM D6319-19) |  |
|--|--|--|----------------------------------------------------------------------|--|
|  |  |  |                                                                      |  |

| Measurement             | Before Aging                    | After Aging at 70°C for 168<br>hrs @ 100°C for 22 hrs |
|-------------------------|---------------------------------|-------------------------------------------------------|
| Tensile Strength (MPa)  | min 14                          | min 14                                                |
| Ultimate Elongation (%) | min 500                         | min 400                                               |
| Pin-hole Level          | AQL 2.5<br>Inspection Level G-1 | AQL 2.5<br>Inspection Level G-1                       |

### 5. Indications for Use:

A non-sterile powder free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. The device is for over-thecounter use.

### Technological Gharacteristics Comparison between the predicate and 6. subiect devices:

| Item description                                                                                                                        | Subject Device,<br>Non-sterile Powder<br>Free Nitrile<br>Examination Glove<br>K210990                                                                                                                                                                                             | Predicate Device,<br>Powder Free<br>Examination Glove<br>(Aqua green),<br>K200326                                                                                                                                                                                                 | Similar or Different |
|-----------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| K Number:                                                                                                                               | K210990                                                                                                                                                                                                                                                                           | K200326                                                                                                                                                                                                                                                                           |                      |
| Product Code                                                                                                                            | LZA                                                                                                                                                                                                                                                                               | LZA                                                                                                                                                                                                                                                                               | Similar              |
| Intended use                                                                                                                            | A powder free<br>patient<br>examination glove<br>is a disposable<br>device intended for<br>medical purposes<br>that is worn on the<br>examiner's hand or<br>finger to prevent<br>contamination<br>between patient<br>and examiner. The<br>device is for over-<br>the-counter use. | A powder free<br>patient<br>examination glove<br>is a disposable<br>device intended for<br>medical purposes<br>that is worn on the<br>examiner's hand or<br>finger to prevent<br>contamination<br>between patient<br>and examiner. The<br>device is for over-<br>the-counter use. | Similar              |
| Material use                                                                                                                            | Nitrile compound                                                                                                                                                                                                                                                                  | Nitrile compound                                                                                                                                                                                                                                                                  | Similar              |
| Colour                                                                                                                                  | Blue                                                                                                                                                                                                                                                                              | Aqua Green                                                                                                                                                                                                                                                                        | Different            |
| Sterility                                                                                                                               | Non sterile                                                                                                                                                                                                                                                                       | Non sterile                                                                                                                                                                                                                                                                       | Similar              |
| Single used                                                                                                                             | Single used                                                                                                                                                                                                                                                                       | Single used                                                                                                                                                                                                                                                                       | Similar              |
| Dimensions                                                                                                                              | Overall<br>Length (mm)<br>Min 220mm<br>Width (±<br>10mm) Size<br>XS = 70mm<br>S = 80mm<br>Size M=<br>95mm Size<br>L = 110mm<br>Size XL =<br>120mm<br>Thickness at<br>Palm (mm)<br>Min; 0.05<br>mm<br>Thickness at<br>Finger Tip<br>(mm) Min<br>0.05 mm                            | Overall<br>Length (mm)<br>Min 230mm<br>Width (±<br>5mm) Size S<br>= 85mm<br>Size M=<br>95mm Size<br>L = 105mm<br>Size XL =<br>115mm<br>Thickness at<br>Palm (mm)<br>Min; 0.05<br>mm<br>Thickness at<br>Finger Tip<br>(mm) Min<br>0.05 mm                                          | Similar              |
| Physical properties                                                                                                                     | Before Ageing<br>Tensile Strength<br>(MPa) = 14min<br>Ultimate Elongation<br>(%) = 500min After<br>Aging at 70 +- 2 oC<br>for 166 +- 2 hrs<br>Tensile Strength<br>(MPa) = 14min<br>Ultimate Elongation<br>(%) = 400min                                                            | Before Ageing<br>Tensile Strength<br>(MPa) = 14min<br>Ultimate Elongation<br>(%) = 500min After<br>Aging at 70oC for<br>168 hrs @ 100oC<br>for 22 hrs Tensile<br>Strength (MPa) =<br>14min Ultimate<br>Elongation (%) =<br>400min                                                 | Similar              |
| Freedom from<br>pinholes                                                                                                                | AQL 2.5<br>Inspection<br>Level G-1                                                                                                                                                                                                                                                | AQL 2.5<br>Inspection<br>Level G-1                                                                                                                                                                                                                                                | Similar              |
| Residual Powder                                                                                                                         | < 2.0 mg/pc                                                                                                                                                                                                                                                                       | < 2.0 mg/pc                                                                                                                                                                                                                                                                       | Similar              |
| Biocompatibility<br>ISO 10993-10-<br>Biological<br>Evaluation<br>on Medical Device<br>- Primary Skin<br>Irritation Test                 | Under the<br>conditions of this<br>study, the test<br>article was a non-<br>irritant.                                                                                                                                                                                             | Under the<br>conditions of this<br>study, the test<br>article was a non-<br>irritant.                                                                                                                                                                                             | Similar              |
| Biocompatibility<br>ISO 10993-10-<br>Biological<br>Evaluation<br>on Medical Device<br>- Dermal<br>Sensitization Assay                   | Under the<br>conditions of this<br>study, the test<br>article was a non-<br>sensitizer.                                                                                                                                                                                           | Under the<br>conditions of this<br>study, the test<br>article was a non-<br>sensitizer.                                                                                                                                                                                           | Similar              |
| Biocompatibility<br>ISO 10993-5<br>Biological<br>evaluation of<br>medical<br>devices - Part<br>5: Tests for in<br>vitro<br>cytotoxicity | Under the<br>conditions of this<br>study, the test<br>was found<br>cytotoxic at 100%<br>after an exposure<br>period of 48<br>hours, non-<br>cytotoxic at 10%                                                                                                                      | Predicate did not<br>perform this test<br>as part of this<br>submission.                                                                                                                                                                                                          | Different            |
| Biocompatibility<br>ISO 10993-11-<br>Acute Systemic<br>test                                                                             | Not induce<br>systemic toxicity                                                                                                                                                                                                                                                   | Not induce<br>systemic toxicity                                                                                                                                                                                                                                                   | Similar              |

{5}------------------------------------------------

{6}------------------------------------------------

### Summarv of Non-Clinical Tests Performed 7.

 Non-sterile Powder Free Nitrile Examination Glove was tested and found in conformance with the following standards:

| ASTM D6319-19 | Standard Specification for Nitrile<br>Examination Gloves for Medical<br>Application      |
|---------------|------------------------------------------------------------------------------------------|
| ASTM D412-16  | Standards test method for Vulcanized<br>Rubber and Thermoplastics Elastomer -<br>Tension |
| ASTM D5151-19 | Standard Test Method for detection of<br>Holes in Medical Gloves                         |

{7}------------------------------------------------

| ASTM D6124-06  | Standard Test Method for Residual<br>Powder on Medical Gloves                                                                                         |
|----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 10993-5    | Biological evaluation of medical devices<br>Part 5: Tests for in vitro cytotoxicity                                                                   |
| ISO 10993-10   | Biological evaluation on medical device<br>Part 10: Test for irritation and Skin<br>Irritation                                                        |
| ISO 10993-11   | Biological evaluation on medical device<br>Part 11: Test for acute systemic toxicity                                                                  |
| ISO 28590:2017 | Sampling Procedure for Inspection by<br>Attributes: Introduction to the ISO 2859<br>series of standards for sampling for<br>inspection by attributes. |

Summary of non-clinical performance test

| Test item             | Test standard                      | Acceptance Criteria                                                                                                                                                                                                            | Conclusion |
|-----------------------|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Dimensions            | ASTM D6319-19                      | Overall Length (mm)<br>Min 230mm Width (±<br>5mm) Size S = 85mm<br>Size M= 95mm Size L<br>= 105mm Size XL =<br>115mm Thickness at<br>Palm (mm) Min; 0.05<br>mm Thickness at<br>Finger Tip (mm) Min<br>0.05 mm                  | Passed     |
| Physical properties   | ASTM D6319-19                      | Before Ageing Tensile<br>Strength (MPa) =<br>14min Ultimate<br>Elongation (%) =<br>500min After Aging at<br>70°C for 168 hrs @<br>100°C for 22 hrs<br>Tensile Strength (MPa)<br>= 14min Ultimate<br>Elongation (%) =<br>400min | Passed     |
| Freedom from pinholes | ASTM D6319-19                      | AQL 2.5<br>Inspection Level G-1                                                                                                                                                                                                | Passed     |
| Residual Powder       | ASTM D6124-06<br>(Reapproved 2017) | < 2.0 mg/pc                                                                                                                                                                                                                    | Passed     |

#### 8. Conclusions:

The Conclusion drawn from the Non-Clinical test demonstrates that the subject device, Non-sterile Powder Free Nitrile Examination Glove is as safe, as effective, and performs as well as or better than the legally marketed Predicate device cleared under K200326.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA/K210990](https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA/K210990)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).