← Product Code [LZA](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA) · K210276

# Disposable Nitrile Gloves (K210276)

_Fujian Ercon Medical Management Co., Ltd. · LZA · Apr 16, 2021 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA/K210276

## Device Facts

- **Applicant:** Fujian Ercon Medical Management Co., Ltd.
- **Product Code:** [LZA](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA.md)
- **Decision Date:** Apr 16, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

The Disposable Nitrile Gloves are intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

## Device Story

Disposable, non-sterile, powder-free nitrile examination gloves; ambidextrous design; blue color. Worn by clinicians during medical examinations to provide a protective barrier against contamination. Device functions as a physical barrier; no electronic or mechanical components. Performance verified through physical property testing (tensile strength, elongation) and biocompatibility assessments. Used in clinical settings; no specialized training required for operation.

## Clinical Evidence

No clinical data. Bench testing only. Compliance verified with ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation/sensitization), ASTM D6124 (residual powder), ASTM D5151 (freedom from holes), and ASTM D6319 (physical properties/dimensions).

## Technological Characteristics

Material: Nitrile. Form factor: Ambidextrous, powder-free, non-sterile gloves. Standards: ASTM D6319-19 (specifications), ASTM D5151-19 (holes), ASTM D6124-06 (residual powder), ISO 10993-5/10 (biocompatibility). No energy source, software, or connectivity.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Disposable Powder Free Nitrile Examination Glove, White/ Blue/ Black/ Pink Color ([K171422](/device/K171422.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 16, 2021

Fujian Ercon Medical Management Co., Ltd. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.608, No.738, Shangcheng Rd., Pudong Shanghai, Shanghai 200120 China

Re: K210276

Trade/Device Name: Disposable Nitrile Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: January 20, 2021 Received: February 1, 2021

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801: medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Ryan Ortega, PhD Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K210276

Device Name Disposable Nitrile Gloves

Indications for Use (Describe)

The Disposable Nitrile Gloves are intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary (K210276)

This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.

#### 1.0 Submitter's Information

Name: FUJIAN ERCON MEDICAL MANAGEMENT CO., LTD. Address: Unit 701, Xinyi International Center, No.3911 Huandao East Road, Siming District, Xiamen City, Fujian Province, China 361000. Tell: +86- 592-5721376 Contact: Lin Ruijin Date of Preparation: Apr. 10,2021

#### Designated Submission Correspondent

Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 608, No. 738 Shangcheng Rd., Pudong Shanghai, 200120 China Tel: +86-21-50313932 Email: Info@truthful.com.cn

#### 2.0 Device Information

Trade name: Disposable Nitrile Gloves Common name: Patient Examination Gloves Classification name: Non-powdered patient examination glove S,M,L,XL Model(s):

#### 3.0 Classification

Production code: LZA Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital

#### 4.0 Predicate Device Information

Manufacturer: Ever Global (Vietnam) Enterprise Corp Disposable Powder Free Nitrile Examination Glove, White/ Device:

{4}------------------------------------------------

#### Blue/ Black/ Pink Color 510(k) number: K171422

### 5.0 Indication for Use

The Disposable Nitrile Gloves are intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

#### 6.0 Device Description

The subject device is powder free nitrile examination gloves. The subject device is blue. The subject device is non-sterile.

#### 7.0 Technological Characteristic Comparison Table

| Table1-General Comparison   |                                                                                                                                                                                        |                                                                                                                                                                                                                                                                              |        |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Item                        | Subject Device<br>(K210276)                                                                                                                                                            | Predicated Device<br>(K171422)                                                                                                                                                                                                                                               | Remark |
| Product Code                | LZA                                                                                                                                                                                    | LZA                                                                                                                                                                                                                                                                          | Same   |
| Regulation No.              | 21CFR880.6250                                                                                                                                                                          | 21CFR880.6250                                                                                                                                                                                                                                                                | Same   |
| Class                       | I                                                                                                                                                                                      | I                                                                                                                                                                                                                                                                            | Same   |
| Intended Use                | The Disposable<br>Nitrile Gloves are<br>intended for<br>medical purposes<br>that is worn on the<br>examiner's hands<br>to prevent<br>contamination<br>between patient<br>and examiner. | The Disposable<br>Powder Free Nitrile<br>Examination Glove,<br>White/ Blue/ Black/<br>Pink Color is a<br>disposable device<br>intended for<br>medical purposes<br>that is worn on the<br>examiner's hands<br>to prevent<br>contamination<br>between patient<br>and examiner. | Same   |
| Powdered or<br>Powered free | Powdered free                                                                                                                                                                          | Powdered free                                                                                                                                                                                                                                                                | Same   |
| Design Feature              | Ambidextrous                                                                                                                                                                           | Ambidextrous                                                                                                                                                                                                                                                                 | Same   |
| Labeling Information        | Single-use<br>indication, powder<br>free, device color,<br>device name,<br>glove size and                                                                                              | Single-use<br>indication, powder<br>free, device color,<br>device name, glove<br>size and quantity                                                                                                                                                                           | Same   |

## Table1-General Comparison

{5}------------------------------------------------

|  | quantity, Nitrile<br>Glove Powder<br>Free Blue, Non-<br>Sterile | Disposable Powder<br>Free Nitrile<br>Examination Glove,<br>Non-Sterile |
|--|-----------------------------------------------------------------|------------------------------------------------------------------------|
|--|-----------------------------------------------------------------|------------------------------------------------------------------------|

### Table2 Device Dimensions Comparison

| Predicate<br>Device(K171422) | Designation    | Size |     |     |     | Tolerance |     |
|------------------------------|----------------|------|-----|-----|-----|-----------|-----|
|                              |                | XS   | S   | M   | L   | XL        |     |
|                              | Length, mm     | 230  | 230 | 230 | 230 | 230       | min |
|                              | Width, mm      | 75   | 85  | 95  | 105 | 115       | ±5  |
|                              | Thickness, mm: |      |     |     |     |           |     |
|                              | Finger         | 0.05 |     |     |     |           | min |
|                              | Palm           | 0.05 |     |     |     |           | min |
| Subject<br>Device(K210276)   | Designation    | Size |     |     |     | Tolerance |     |
|                              |                | S    | M   | L   | XL  |           |     |
|                              | Length, mm     | 230  | 230 | 230 | 230 | ±10       |     |
|                              | Width, mm      | 85   | 95  | 110 | 115 | ±10       |     |
|                              | Thickness, mm: |      |     |     |     |           |     |
|                              | Finger         | 0.05 |     |     |     |           | min |
|                              | Palm           | 0.05 |     |     |     |           | min |
| Remark                       | Similar        |      |     |     |     |           |     |

Analysis: The tolerance of length is different with that of the predicate, but they all meet the requirements of ASTM D6319-19,so the differences do not raise any new safety or performance questions.

| Item                   |                 | Subject device<br>(K210276)     | Predicated device<br>(K171422) | Remark |
|------------------------|-----------------|---------------------------------|--------------------------------|--------|
| Colorant               |                 | Blue                            | White/ Blue/ Black/ Pink       | Same   |
| Physical<br>Properties | Before<br>Aging | Tensile Strength<br>14MPa, min  | 14MPa, min                     | Same   |
|                        |                 | Ultimate Elongation<br>500% min | 500% min                       | Same   |
|                        | After<br>Aging  | Tensile Strength<br>14MPa, min  | 14MPa, min                     | Same   |
|                        |                 | Ultimate Elongation<br>400%min  | 400%min                        | Same   |
| Freedom from Holes     |                 | Comply with ASTM D6319          | Comply with ASTM<br>D6319      | Same   |
|                        |                 | Be free from                    | Be free from                   | Same   |

#### Table3 Performance Comparison

{6}------------------------------------------------

|                | holes when<br>tested in<br>accordance with<br>ASTMD5151<br>AQL=2.5 | holes when<br>tested in<br>accordance with<br>ASTMD5151<br>AQL=2.5 |      |
|----------------|--------------------------------------------------------------------|--------------------------------------------------------------------|------|
| Powder Content | $0.01 \text{ mg per glove}$                                        | Meet the<br>requirements<br>of ASTM<br>D6124                       | Same |

#### Table4 Safety Comparison

| Table4 Safety Comparison |                                                                                                                                                            |                                                                               |                                |         |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|--------------------------------|---------|
| Item                     | Subject<br>device<br>(K210276)                                                                                                                             | Predicated<br>device<br>(K171422)                                             | Remark                         |         |
|                          | Material                                                                                                                                                   | Nitrile                                                                       |                                | Nitrile |
| Biocompatibility         | Irritation (ISO<br>10993-10:2010<br>Biological<br>Evaluation of<br>Medical Devices -<br>Part 10: Tests For<br>Irritation And Skin<br>Sensitization)        | Under the<br>conditions of<br>the study, not<br>an irritant                   | Comply with<br>ISO10993-<br>10 | Same    |
|                          | Sensitization<br>(ISO 10993-<br>10:2010<br>Biological<br>Evaluation of<br>Medical Devices -<br>Part 10: Tests For<br>Irritation And Skin<br>Sensitization) | Under<br>conditions of<br>the study, not<br>a sensitizer.                     |                                |         |
|                          | Cytotoxicity (ISO<br>10993-5:2009<br>Biological<br>Evaluation of<br>Medical Devices -<br>Part 5: Tests For<br>In Vitro<br>Cytotoxicity)                    | Under<br>conditions of<br>the study,<br>device<br>extract is not<br>cytotoxic | /                              | Similar |

# 8.0 Discussion of Non-clinical and Performance Testing

{7}------------------------------------------------

Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.

ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity

ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves

ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.

ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.

### 9.0 Discussion of Clinical and Performance Testing

Clinical testing is not needed for this device.

#### 10.0 Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device, Disposable Nitrile Gloves ,is as safe, as effective, and performs as well as or better than the legally marketed predicated device under K171422.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA/K210276](https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA/K210276)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).