← Product Code [LZA](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA) · K200326

# Powder Free Nitrile Examination Glove (Aqua Green) (K200326)

_Riverstone Resources Sdn Bhd · LZA · Apr 9, 2020 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA/K200326

## Device Facts

- **Applicant:** Riverstone Resources Sdn Bhd
- **Product Code:** [LZA](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA.md)
- **Decision Date:** Apr 9, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

The Powder Free Nitrile Examination Glove (Aqua Green) patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. It is for over-the-counter use.

## Device Story

Disposable nitrile examination glove; worn on examiner's hands or fingers; prevents contamination between patient and examiner. Used in medical settings; over-the-counter availability. Device acts as physical barrier; protects both patient and clinician from cross-contamination. No electronic or mechanical components.

## Clinical Evidence

Bench testing only. Device conforms to ASTM D6319-10 (dimensions, physical properties), ASTM D5151-06 (pinholes), ASTM D6124-06 (residual powder), and ISO 10993-10/11 (biocompatibility: primary skin irritation, dermal sensitization, acute systemic toxicity).

## Technological Characteristics

Material: Nitrile compound. Form factor: Disposable, non-sterile, powder-free examination glove. Dimensions: 230mm length; palm/finger thickness min 0.05mm. Standards: ASTM D6319-10, ASTM D412-16, ASTM D5151-06, ASTM D6124-06, ISO 10993-10, ISO 10993-11, ISO 28590:2017.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Emg Black Nitrile Medical Examination Glove Powder Free ([K141579](/device/K141579.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo, which features a stylized caduceus symbol. To the right of that is the FDA logo, which is the acronym "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 9, 2020

Riverstone Resources SDN BHD Suresh Kumar Official Correspondent Lot 55. No 13 Jalan Jasmin 2 Kawasan Perindustrian Bukit Beruntung, 48300 My

Re: K200326

Trade/Device Name: Powder Free Nitrile Examination Glove (Aqua Green) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: February 2, 2020 Received: February 10, 2020

Dear Suresh Kumar:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/6 description: The image shows a light blue FDA logo. The letters are large and bold. The background is white.

Elizabeth F. Claverie -S

CAPT Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Section 1A-1

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

#### Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

#### 510(k) Number (if known) K200326

R200326

Device Name

Powder Free Nitrile Examination Glove (Aqua Green)

#### Indications for Use (Describe)

The Powder Free Nitrile Examination Glove (Aqua Green) is a disposable device intended for Medical purpose that is worn on the examiner's hands or finger to prevent contamination patient and examiner.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW,*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(K) Summary

# K200326

## Powder Free Nitrile Examination Glove (Aqua Green)

Preparation Date: April 8th, 2020

## 1. Submitter:

Company Name: Riverstone Resource Sdn Bhd. Company Address: Lot 55,No 13, Jalan Jalan Jasmin 2 Kawasan Perindustrian Bukit Beruntung 48300. Selangor Malaysia.

Contact Person: Mr Suresh Kumar Telephone No: 603-60283033 Email: qa1@riverstone.com.my

## 2. Name of the Device

Trade Name / Proprietary Name: Powder Free Nitrile Examination Glove (Aqua Green) Device Name: Nitrile Patient Examination gloves. Device Classification Name: Patient Examination gloves (21 CFR 880.6250). Device Class: Class I. Product Code: LZA

## 3. Identification of The Legally Marketed Device:

Predicate Device: K141579 Device Name: Emg Black Nitrile Medical Examination Glove Powder Free Device Classification Name: Patient Examination gloves (21 CFR 880.6250). Device Class: Class I. Product Code: LZA

## 4. Device Description

The subject device in this 510(k) Notification is Powder Free Nitrile Examination Glove (Aqua Green). The subject device is a patient examination glove made from nitrile compound, Aqua Green color, powder free and non-sterile (Per 21 CFR 880.6250, class I). The device meets the specifications in ASTM D6319-10, Standard specification for Nitrile Examination Gloves.

## 5. Intended use of the Device

Powder Free Nitrile Examination Glove (Aqua Green) patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. It is for over-the-counter use.

{4}------------------------------------------------

| Characteristics          | Acceptance Criteria                                                                                                                                                                                                                                                      | Subject device,<br>Powder Free<br>Examination Glove<br>(Aqua green),<br>K200326                                                                                                                                                                                           | Predicate device,<br>EMG Blue Nitrile<br>Medical Examination<br>Glove, K141579                                                                                                                                                                                             | Remarks   |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Product Code             | LZA.                                                                                                                                                                                                                                                                     | LZA.                                                                                                                                                                                                                                                                      | LZA.                                                                                                                                                                                                                                                                       | same      |
| Intended use             | A powder free patient<br>examination glove is a<br>disposable device<br>intended for medical<br>purposes that is worn<br>on the examiner's hand<br>or finger to prevent<br>contamination between<br>patient and examiner.<br>The device is for over-<br>the-counter use. | A powder free patient<br>examination glove is a<br>disposable device<br>intended for medical<br>purposes that is worn<br>on the examiner's hand<br>or finger to prevent<br>contamination between<br>patient and examiner.<br>The device is for over-<br>the- counter use. | A powder free patient<br>examination glove is a<br>disposable device<br>intended for medical<br>purposes that is worn<br>on the examiner's<br>hand or finger to<br>prevent contamination<br>between patient and<br>examiner. The device<br>is for over-the-counter<br>use. | same      |
| Material use             | Nitrile compound                                                                                                                                                                                                                                                         | Nitrile compound                                                                                                                                                                                                                                                          | Nitrile compound                                                                                                                                                                                                                                                           | same      |
| Colour                   | N/A                                                                                                                                                                                                                                                                      | Aqua Green                                                                                                                                                                                                                                                                | Blue                                                                                                                                                                                                                                                                       | Different |
| Sterility                | N/A                                                                                                                                                                                                                                                                      | Non sterile                                                                                                                                                                                                                                                               | Non sterile                                                                                                                                                                                                                                                                | same      |
| Single used              | Single used                                                                                                                                                                                                                                                              | Single used                                                                                                                                                                                                                                                               | Single used                                                                                                                                                                                                                                                                | same      |
| Dimensions               | Overall Length (mm)<br>Min 230mm Width (±<br>5mm) Size S = 85mm<br>Size M = 95mm Size L<br>= 105mm Size<br>XL = 115mm<br>Thickness at Palm<br>(mm)<br>Min; 0.05 mm<br>Thickness at Finger Tip<br>(mm) Min 0.05 mm                                                        | Meets ASTM D6319-<br>10                                                                                                                                                                                                                                                   | Meets ASTM D6319-<br>10                                                                                                                                                                                                                                                    | same      |
| Physical<br>properties   | Before Ageing Tensile<br>Strength (MPa)<br>= 14min Ultimate<br>Elongation (%)<br>= 500min After Aging<br>at 70oC for<br>168 hrs @ 100oC for<br>22 hrs Tensile Strength<br>(MPa)<br>= 14min Ultimate<br>Elongation (%)                                                    | Meets ASTM D6319-<br>10                                                                                                                                                                                                                                                   | Meets ASTM D6319-<br>10                                                                                                                                                                                                                                                    | same      |
| Freedom from<br>pinholes | AQL 2.5<br>Inspection Level G-1                                                                                                                                                                                                                                          | Meets ASTM D5151-<br>06                                                                                                                                                                                                                                                   | Meets ASTM D5151-<br>06                                                                                                                                                                                                                                                    | same      |
| Residual Powder          | < 2.0 mg/pc                                                                                                                                                                                                                                                              | Meets ASTM D6124-<br>06                                                                                                                                                                                                                                                   | Meets ASTM D6124-<br>06                                                                                                                                                                                                                                                    | same      |
| Biocompatibility         | ISO 10993-10-<br>Biological Evaluation<br>on Medical Device -<br>-Primary Skin Irritation<br>Test                                                                                                                                                                        | Under the conditions of<br>this study, the test<br>article was a non-<br>irritant.                                                                                                                                                                                        | Under the conditions<br>of this study, the test<br>article was a non-<br>irritant.                                                                                                                                                                                         | same      |
|                          | ISO 10993-10-<br>Biological Evaluation<br>on Medical Device -<br>Dermal Sensitization<br>Assay                                                                                                                                                                           | Under the conditions of<br>this study, the test<br>article was a non-<br>sensitizer.                                                                                                                                                                                      | Under the conditions<br>of this study, the test<br>article was a non-<br>sensitizer.                                                                                                                                                                                       | same      |
|                          | ISO 10993-11-<br>Acute Systemic test                                                                                                                                                                                                                                     | Not induce systemic<br>toxicity                                                                                                                                                                                                                                           | N/A                                                                                                                                                                                                                                                                        | Same      |

# 6. Technological characteristics Comparison for the proposed and predicate devices

{5}------------------------------------------------

# 7. Summary of non-clinical testing results

Powder Free Nitrile Examination Glove (Aqua Green) was tested and found in conformance with the following standards:

| ASTM D6319-10  | Standard Specification for Nitrile Examination Gloves for Medical<br>Application                                                                |
|----------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| ASTM D412-16   | Standards test method for Vulcanized Rubber and Thermoplastics<br>Elastomer - Tension                                                           |
| ASTM D5151-06  | Standard Test Method for detection of Holes in Medical Gloves                                                                                   |
| ASTM D6124-06  | Standard Test Method for Residual Powder on Medical Gloves                                                                                      |
| 16 CFR:1500.41 | Method of Testing Primary Irritant Substance                                                                                                    |
| ISO 10993-10   | Biological evaluation on medical device Part 10: Test for irritation and<br>Skin Irritation                                                     |
| ISO 10993-11   | Biological evaluation of medical devices Part 11: Tests for systemic<br>toxicity                                                                |
| ISO 28590:2017 | Sampling Procedure for Inspection by Attributes: Introduction to the<br>ISO 2859 series of standards for sampling for inspection by attributes. |

| Dimension and Thickness of Gloves (ASTM D6319-10) |         |         |         |         |
|---------------------------------------------------|---------|---------|---------|---------|
| Dimension                                         | Size S  | Size M  | Size L  | Size XL |
| Overall Length (mm)                               | 230min  | 230min  | 230min  | 230min  |
| Width (± 5mm)                                     | 85      | 95      | 105     | 115     |
| Thickness at Palm (mm)                            | 0.05min | 0.05min | 0.05min | 0.05min |
| Thickness at Finger Tip (mm)                      | 0.05min | 0.05min | 0.05min | 0.05min |

#### (ASTM D6310 10) d Thick f City

{6}------------------------------------------------

| Measurement             | Before Ageing                   | After Aging at 70ºC for<br>168 hrs @ 100ºC for 22<br>hrs |
|-------------------------|---------------------------------|----------------------------------------------------------|
| Tensile Strength (MPa)  | 14min                           | 14 Min                                                   |
| Ultimate Elongation (%) | 500min                          | 400min                                                   |
| Pin-hole Level          | AQL 2.5<br>Inspection Level G-1 | AQL 2.5<br>Inspection Level G-1                          |

Gloves Physical Properties and Holes (ASTM D6319-10)

## Residual Powder (ASTM D6124-06)

| Sample No.   | Residual Powder (mg/glove) |
|--------------|----------------------------|
| 1            | 0.49                       |
| 2            | 0.38                       |
| 3            | 0.43                       |
| 4            | 0.51                       |
| 5            | 0.22                       |
| Requirements | <2.0                       |

## 8. Conclusion

The Conclusion drawn from the Non-Clinical test demonstrates that the subject device, Powder Free Nitrile Examination Glove (Aqua Green) is as safe, as effective, and performs as well as or better than the legally marketed Predicate device cleared under K141579.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA/K200326](https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZA/K200326)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
