← Product Code [LYZ](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYZ) · K974120

# POWDER-FREE VINYL PATIENT EXAMINATION GLOVES (K974120)

_Shanghai Super Gloves Co., Ltd. · LYZ · Dec 17, 1997 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYZ/K974120

## Device Facts

- **Applicant:** Shanghai Super Gloves Co., Ltd.
- **Product Code:** [LYZ](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYZ.md)
- **Decision Date:** Dec 17, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

## Device Story

Disposable vinyl patient examination glove; worn by healthcare personnel on hands/fingers; acts as physical barrier to prevent contamination between patient and examiner; used in clinical/medical settings; non-sterile; powder-free; manufactured to ASTM D5250-92 standards; provides protection against fluids, waste, and environmental contaminants.

## Clinical Evidence

No clinical data. Bench testing only: physical dimensions, water leak test (1000 ml), and biocompatibility (primary skin irritation and sensitization) performed per ASTM D5250-92.

## Technological Characteristics

Material: Vinyl. Standard: ASTM D5250-92. Form factor: Disposable examination glove. Powder-free (verified via USP Iodine test and particulate testing). Physical testing: Water leak test (AQL 2.5, Inspection Level S-4), dimensions/physical properties (AQL 4.0, Inspection Level S-2).

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Sunmax Enterprise Shanghai Co., Ltd. Powder-Free Vinyl Patient Examination Gloves ([K964964](/device/K964964.md))
- Cheer & Merit Co., Ltd. Powder-Free Vinyl Patient Examination Gloves ([K942042](/device/K942042.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# DEC I 7 1997

### EXHIBIT #1 Page 1 of 2

### 510 (K) SUMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

The assigned 510(k) number is: 974120

#### Submitter's Identification: 1.

Ms. Lee Hung Yun-Shan Shanghai Super Gloves Co., Ltd. No. 4, Tingwai Road, Zhuhang Zhen, Jinshan County, Shanghai, China

Date Summary Prepared: October 29, 1997

#### 2. Name of the Device:

Shanghai Super Gloves Co., Ltd. Powder-Free Vinyl Patient Examination Gloves

#### 3. Predicate Device Information:

Sunmax Enterprise Shanghai Co., Ltd. Powder-Free Vinyl Patient Examination Gloves, K#964964 Cheer & Merit Co., Ltd. Powder-Free Vinyl Patient Examination Gloves, K#942042

#### 4 . Device Description:

Classified by FDA's General and Plastic Surgery Device Panel Class I, 21 CFR 880.6250, Powder-Free Vinyl Patient as Examination Glove, 80LYZ and meets all requirements of ASTM Standard D5250-92.

#### 5. Intended Use:

A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.

#### 6. Comparison to Predicate Devices:

Shanghai Super Gloves Co., Ltd. Powder-Free Vinyl Patient Examination Gloves, is substantially equivalent in safety and

{1}------------------------------------------------

## EXHIBIT #1 Page 2 of 2

effectiveness to the Sunmax Enterprise Shanghai Co., Ltd. Powder-Free Vinyl Patient Examination Gloves and the Cheer & Merit Powder-Free Vinyl Patient Examination Gloves.

### Discussion of Non-Clinical Tests Performed for Determination 7 . of Substantial Equivalence are as follows:

The standards used for Shanghai Super Gloves Co., Ltd. glove production are based on ASTM-D-5250-92. All testing meets requirements for Physical and Dimensions Testing conducted on gloves, Inspection Level S-2, AQL 4.0.

The FDA 1000 ml. Water Fill Test was also conducted with samplings of AQL 2.5, Inspection Level S-4, meeting these requirements. Primary Skin Irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions.

There are no special labeling claims and we do not claim our gloves as hypoallergenic on our labels.

Test for Starch at finished inspection is A USP Iodine conducted to insure that our gloves meet our "powder-free" We adhere to all USP Iodine Test methodogy and testing claim. conducted revealed passing results. We will employ the use of with a "powder-residue by weight"and are contracting laboratory to conduct particulate testing for added assurance. Final release testing consists of a light transmission test using a spectrophotometer.

#### Discussion of Clinical Tests Performed: 8.

Not Applicable - There is no hypoallergenic claim.

#### 9. Conclusions:

Shanghai Super Gloves Co., Ltd. Vinyl Patient Examination Gloves conform fully to ASTM-D-5250-92 standards as well as CFR references, applicable 21 and, meets pinhole IRDA biocompatibility requirements and labeling requirements, claims as shown by data in Section 7. There are no safety/efficacy issues or new claims from the "substantial equivalence" products cited.

{2}------------------------------------------------

Image /page/2/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with its wings spread, with a human profile incorporated into the body of the eagle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle image.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC 17 1997

Shanghai Super Gloves Company, Ltd. C/O Ms. Susan D. Goldstein-Falk Official Correspondent MDI Consultants, Incorporated 55 Northern Boulevard, Suite 410 Great Neck, New York 11021

K974120 Re : Powder-Free Vinyl Patient Examination Gloves Trade Name: Requlatory Class: I LYZ Product Code: Dated: October 29, 1997 Received: October 31, 1997

Dear Ms. Goldstein-Falk:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਸ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531

{3}------------------------------------------------

Page 2 - Ms. Goldstein-Falk

through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a leqally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Timothy A. Whetstone

Timot Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

ATTACHMENT TA

Page_1_of_1

510(k) Number (if known):_ K974120

Device Name:___ Shanghai Super Gloves Co., Ltd. Vinyl Patient Examination Gloves-Powder-Free

Indications For Use:

乐

A patient examination glove is a disposable device intended for

medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

(PLEASE DO NOT WRITE:BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

*Phim S. bin*
(Division Sign-Off)
Division of Dental, Infection Control,
and General Hospital Devices

General Hospital Devices
510(k) Number K974120

Prescription Use (Per 21 CFR 801.109) CR

Over-The-Counter Use X

(Optional Format 1-2-96)

। :

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYZ/K974120](https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYZ/K974120)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
