K241252 · Anqing Pukang Medical Technology Co., Ltd. · LYZ · May 20, 2024 · General Hospital
Device Facts
Record ID
K241252
Device Name
Disposable Vinyl Examination Gloves
Applicant
Anqing Pukang Medical Technology Co., Ltd.
Product Code
LYZ · General Hospital
Decision Date
May 20, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Attributes
3rd-Party Reviewed
Indications for Use
Disposable Vinyl Examination Gloves is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Device Story
Disposable Vinyl Examination Gloves are non-sterile, powder-free, ambidextrous gloves made of poly (vinyl chloride) compound. Available in sizes S, M, L, and XL, they are intended for use by healthcare professionals during medical examinations to provide a barrier against contamination. The gloves are worn on the hands to prevent the transfer of microorganisms or fluids between the patient and the examiner. They are single-use devices.
Clinical Evidence
No clinical data. Bench testing only. Biocompatibility testing (ISO 10993-5, 10, 23) confirmed no cytotoxicity, irritation, or sensitization. Physical performance testing (ASTM D5250-19, D6124-06, D5151-19) confirmed compliance with dimensions, tensile strength, elongation, freedom from holes (AQL 2.5), and residual powder limits (<2mg/glove).
Technological Characteristics
Material: Poly (vinyl chloride) compound. Form factor: Ambidextrous, clear, non-sterile, powder-free. Sizes: S, M, L, XL. Standards: ASTM D5250-19 (physical properties/dimensions), ASTM D5151-19 (freedom from holes), ASTM D6124-06 (residual powder), ISO 10993-5/10/23 (biocompatibility).
Indications for Use
Indicated for use as a disposable medical glove worn on the examiner's hand or finger to prevent cross-contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 20, 2024
Anqing Pukang Medical Technology Co., LTD. % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114
Re: K241252
Trade/Device Name: Disposable Vinyl Examination Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Non-powdered patient examination glove Regulatory Class: Class I, reserved Product Code: LYZ Dated: May 3, 2024 Received: May 3, 2024
Dear Prithul Bom:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{3}------------------------------------------------
Sincerely,
# Allan Guan -S
For Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K241252
Device Name Disposable Vinyl Examination Gloves
Indications for Use (Describe)
Disposable Vinyl Examination Gloves is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
## 510(K) Summary
This summary of 510(k) is being submitted in accordance with 21 CFR 807.92. Date prepared: May 20, 2024 A. Applicant:
Name: Anqing Pukang Medical Technology Co., LTD. Address: No 7 Haian Road, Anqing City, Anhui,China. Contact: Ms. Min Gao Title: QE Tel: +86-18266365220 Email: 474112523@qq.com
Submission Correspondent: Primary contact: Max Li Anqing Pukang Medical Technology Co., LTD. No 7 Haian Road, Anqing City, Anhui,China. Tel: +86-536-6136888 Email: maxli@basicmedical.com
## B. Device:
Trade Name: Disposable Vinyl Examination Gloves Common Name: Non-Powdered Patient Examination Glove
Regulatory Information Classification Name: Vinyl Patient Examination Glove Classification: Class I Product code: LYZ Regulation Number: 21 CFR 880.6250 Review Panel: General Hospital and Personal Use
## C. Predicate device:
K223008 Name of the device: Disposable Vinyl Examination Gloves Owner's Name: JIANGXI HANDSPRO PRODUCTS SOLUTIONS CO., LTD.
## D. Indications for use of the device:
Disposable Vinyl Examination Gloves is a disposable device intended for medical purposes that is worn on the examiners' hand or finger to prevent contamination between patient and examiner.
#### Device Description: نن
The subject device is a patient examination glove made of Poly (vinyl chloride) compound. It is Non-Sterile, Powder free and is clear in colour. The device is ambidextrous and can be worn on either the left or right hand. (Per 21 CFR 880.6250, class I). The device provides four specifications: S, M, L and XL. The device meets all the
{6}------------------------------------------------
specifications in ASTM D5250-19, Standard specification for Vinyl Examination Gloves. Additionally, the gloves have been tested for biocompatibility per ISO 10993-10, ISO 10993-23, ISO 10993-5
## F. Summary of Technological Characteristics
Disposable Vinyl Examination Gloves are compared with the predicate device in terms of intended use, design, material, specifications, and performance.
The following table shows similarities and differences of use, design, and material between the proposed device and the predicate device.
| Device | Proposed Device | Predicate Device | Result |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| 510K # | K241252 | K223008 | - |
| Product Name | Disposable Vinyl Examination Gloves | Disposable Vinyl Examination Gloves | - |
| Product Code | LYZ | LYZ | Same |
| Regulation Number | 21 CFR 880.6250 | 21 CFR 880.6250 | Same |
| Indications for use | Disposable Vinyl Examination Gloves is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. | Disposable Vinyl Examination Gloves is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. | Same |
| Powder free | Yes | Yes | Same |
| Design feature | Ambidextrous. | Ambidextrous. | Same |
| Material | Vinyl | Vinyl | Same |
| Color | Clear | Clear | Same |
| Size | S, M, L, XL | S, M, L, XL | Same |
| Sterile | Non-Sterile | Non-Sterile | Same |
| Use | Singe use | Single use | Same |
Table 1 General Comparison of Proposed and Predicate Devices
Table 2 Comparison of Physical, Biocompatibility and Performance Testing
| CHARACTERSTICS | STANDARDS | DEVICE PERFORMANCE | | Remarks | |
|---------------------|-------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|----------|--|
| | | SUBJECT | PREDICATE | | |
| Name of device | | Disposable Vinyl<br>Examination Gloves | Disposable Vinyl<br>Examination Gloves | --- | |
| Dimensions | ASTMD5250- 19 | Length:<br>S/M/L/XL: ≥230; | Length:<br>S/M/L/XL: ≥230; | Same | |
| | | | | | |
| Physical Properties | ASTMD5250- 19 | Width: | Width: | Same | |
| | | S: 85±5<br>M: 95±5<br>L: 105±5<br>XL: 115±5 | S: 85±5<br>M: 95±5<br>L: 105±5<br>XL: 115±5 | | |
| | | Thickness:<br>Palm: ≥0.08<br>Finger: ≥0.08 | Thickness:<br>Palm: ≥0.08<br>Finger: ≥0.08 | | |
| | | Before Aging<br>Tensile Strength<br>min 11 Mpa<br>Ultimate Elongation<br>Min 300% | Before Aging<br>Tensile Strength<br>min 11 Mpa<br>Ultimate Elongation<br>Min 300% | | |
| | | After Aging<br>Tensile Strength<br>min 11 Mpa<br>Ultimate Elongation<br>Min 300% | After Aging<br>Tensile Strength<br>min 11 Mpa<br>Ultimate Elongation<br>Min 300% | | |
| Powder Free | ASTMD5250- 19 | ≤2 mg/glove | ≤2 mg/glove | Same | |
| Freedom from Holes | ASTM D5250- 19<br>ASTM D5151- 19 | Complies with<br>ASTM D5250- 19<br>and ASTM D5151- 19 G-<br>1, AQL 2.5 | Complies with<br>ASTM D5250- 19<br>and ASTM D5151- 19 G-<br>1, AQL 2.5 | Same | |
| Biocompatibility | Primary Skin<br>Irritation-<br>ISO 10993-<br>23:2021; | Under the conditions of this<br>study, the polar and<br>nonpolar extracts of the test<br>article did not induce skin<br>irritation. | Under the conditions of<br>the study, not an irritant | Similar① | |
| | Dermal<br>Sensitization-<br>ISO 10993-<br>10:2010 | Under the conditions of<br>this test, the polar and<br>non-polar extracts of the<br>test article did not cause<br>skin sensitization<br>reaction. | Under the conditions of<br>the study, not a sensitizer | Same | |
| | In vitro<br>cytotoxicity<br>ISO10993-<br>5:2009 | Under the conditions of<br>this study, the extract of<br>the test article did not<br>show potential toxicity to<br>L-929 cells. | Under conditions of the<br>study, did not show<br>potential toxicity to L-<br>929 cells. Complies with | Same | |
{7}------------------------------------------------
{8}------------------------------------------------
(1) The new version of standard for Skin Irritation we refer is ISO 10993-23:2021, the test method and process is the same as the old version ISO 10993-10
Analysis: The proposed device has the Skin Sensitization test, Primary Skin Irritation test and the In vitro cytotoxicity test, the same as the predicate device.
## G. Summary of Non-Clinical Testing
#### A Biocompatibility
Biocompatibility Testing according to ISO 10993-1:2018, the nature of body contact for the subject device is Surface Device category, Skin Contact and duration of contact is A-Limited (≤24h). The following tests for the subject device were conducted to evaluated the biocompatibility of Disposable Vinyl Examination Gloves:
- ISO 10993-10: Skin Sensitization test
- ISO 10993-23:Skin Irritation test
- · ISO 10993-5: In vitro cytotoxicity test
#### A Performance Testing
Physical performance testing of the proposed device was conducted as per ASTM D5250-19 Standard Specification for Poly (vinyl chloride) Examination Gloves.
To summarize, the performance testing of the subject device was conducted to adequately demonstrate the effectiveness of the device in accordance with the relevant test methods cited below:
- ASTM D5250-19 Standard Specification for Poly (vinyl chloride) Gloves for Medical Application
- ASTM D6124-06 (Reapproved 2017) Standard Test Method for Residual Powder on Medical Gloves
- ASTM D5151-19 Standard Test Method for Detection of Holes in Medical Gloves
| Test Method | Purpose | Acceptance Criteria | Results |
|--------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| Dimensions<br>(length)<br>(width)<br>(thickness) | The purpose of the<br>test is to evaluate the<br>physical dimension of<br>the glove | Length<br>230 mm min | Pass<br>240 mm min |
| | | Width(mm) | Pass |
| | | S: $85\pm5$<br>M: $95\pm5$<br>L: $105\pm5$<br>XL: $115\pm5$ | S: 89-90mm<br>M: 99 mm<br>L:105-106 mm<br>XL:119-120 mm |
| | | Thickness(mm):<br>Palm: Min 0.08<br>Finger: Min 0.08 | Pass<br>Palm – 0.080mm Min<br>Finger–0.080mm Min |
| Physical<br>properties | The purpose of the<br>test is to evaluate the<br>tensile strength and<br>ultimate elongation<br>before and after<br>aging | Before Aging:<br>Tensile Strength: 11 MPa, min.<br>Elongation: 300%, min. | Pass<br>Before Aging:<br>Tensile Strength: 16.0MPa, min.<br>Elongation: 304%, min. |
| | | After Aging:<br>Tensile Strength: 11 MPa, min.<br>Elongation:300%, min. | After Aging:<br>Tensile Strength:16.7MPa, min.<br>Elongation: 301%, min. |
| | | | |
| Freedom from<br>holes | The purpose of the<br>test is to detect holes<br>in the gloves | No leakage at sampling level of<br>G-1, AQL 2.5 | Pass<br>No leakage, 80 of 80 passed |
{9}------------------------------------------------
| Residual<br>Powder | The purpose of the<br>test is to detect the<br>powder residue in<br>the glove | <2mg per glove | Pass<br>average 0.1 mg per glove |
|--------------------|-------------------------------------------------------------------------------|----------------|----------------------------------|
|--------------------|-------------------------------------------------------------------------------|----------------|----------------------------------|
## H. Clinical Test Conclusion
No clinical study is included in this submission.
#### l. Conclusion
The conclusion drawn from the nonclinical tests demonstrates that the subject device, Disposable Vinyl Examination Gloves, are as safe, as effective, and perform as well as or better than the legally marketed predicate device cleared under K223008.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.