← Product Code [LYZ](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYZ) · K211262

# Disposable Vinyl Nitrile Synthetic Gloves Powder Free (K211262)

_Tangshan Hongyun Healthcare Products Co., Ltd. · LYZ · Jul 24, 2021 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYZ/K211262

## Device Facts

- **Applicant:** Tangshan Hongyun Healthcare Products Co., Ltd.
- **Product Code:** [LYZ](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYZ.md)
- **Decision Date:** Jul 24, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands to prevent contamination between patient and examiner.

## Device Story

Disposable, non-sterile, ambidextrous vinyl nitrile synthetic patient examination gloves; blue color. Worn by healthcare examiners to provide a barrier against contamination between patient and examiner. Device functions as a physical barrier; no electronic or active components. Used in clinical or medical settings. Benefits include reduction of cross-contamination risk during patient examinations.

## Clinical Evidence

No clinical data. Bench testing only. Compliance verified against ASTM D5250 (physical dimensions, tensile strength, elongation), ASTM D5151 (watertightness/holes), ASTM D6124 (residual powder), ISO 10993-5 (cytotoxicity), and ISO 10993-10 (irritation and sensitization). All tests met acceptance criteria.

## Technological Characteristics

Materials: Polyvinyl chloride, polyurethane, nitrile, and di-(2-ethylhexyl) terephthalate (DOTP). Non-sterile, powder-free, ambidextrous. Dimensions: 230mm length; 85-115mm width; ≥0.05mm thickness. Standards: ASTM D5250, ASTM D5151, ASTM D6124, ISO 10993-5, ISO 10993-10.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Synmax Synthetic Patient Examination Vinyl Gloves, Powder Free, Blue ([K153028](/device/K153028.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

July 24, 2021

Tangshan Hongyun Healthcare Products Co., Ltd. Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.608, No.738, Shangcheng Rd., Pudong Shanghai, Shanghai 200120 China

Re: K211262

Trade/Device Name: Disposable Vinyl Nitrile Synthetic Gloves Powder Free Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LYZ Dated: April 19, 2021 Received: April 26, 2021

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K211262

Device Name

Disposable Vinyl Nitrile Synthetic Gloves Powder Free

Indications for Use (Describe)

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

# K211262

This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.

### 1.0 Submitter's Information

Name: Tangshan Hongyun Healthcare Products Co.,Ltd. Address: South Shenggezhuang Village, Pachigang Town, Luannan, Tangshan, Hebei 063500, China. Phone Number: +86-13933365259 Contact: Suying Le Date of Preparation: 07/23/2021

#### Designated Submission Correspondent

Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 608, No. 738 Shangcheng Rd., Pudong Shanghai, 200120 China Tel: +86-21-50313932 Email: Info@truthful.com.cn

## 2.0 Device Information

Trade name: Disposable Vinyl Nitrile Synthetic Gloves Powder Free Common name: Vinyl Patient Examination Glove Classification name: Non-powdered Patient Examination Glove Model(s): S, M, L, XL

## 3.0 Classification

Production code: LYZ Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital

#### 4.0 Predicate Device Information

Manufacturer: Zibo Huiying Medical Products, Co. Ltd. Device: Synmax Synthetic Patient Examination Vinyl Gloves, Powder Free, Blue 510(k) number: K153028

#### 5.0 Indication for Use

{4}------------------------------------------------

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands to prevent contamination between patient and examiner.

#### 6.0 Device Description

The subject device is powder free vinyl synthetic patient examination gloves. The subject device is blue. The design of subject device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D5250. The subject device is non-sterile.

#### 7.0 Technological Characteristic Comparison Table

| Item                                  | Subject device                                                                                                                                                                 | Predicated device                                                                                                                                                                         | Comparison |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| 510(k) number                         | K211262                                                                                                                                                                        | K153028                                                                                                                                                                                   |            |
| Product Code                          | LYZ                                                                                                                                                                            | LYZ                                                                                                                                                                                       | Same       |
| Regulation No.                        | 21CFR880.6250                                                                                                                                                                  | 21CFR880.6250                                                                                                                                                                             | Same       |
| Class                                 |                                                                                                                                                                                |                                                                                                                                                                                           | Same       |
| Intended Use/<br>Indications foru Use | A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands to prevent contamination between patient and examiner. | A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or fingers to prevent contamination between patient and examiner. | Same       |
| Powdered or Powered free              | Powdered free                                                                                                                                                                  | Powdered free                                                                                                                                                                             | Same       |
| Design Feature                        | Ambidextrous                                                                                                                                                                   | Ambidextrous                                                                                                                                                                              | Same       |
| Labeling Information                  | Single use, powder free, device color, device name, glove size and quantity, product name, Non-Sterile                                                                         | Single use, powder free, device color, device name, glove size and quantity, product name, Non-Sterile                                                                                    | Similar    |

#### Table1-General Comparison

#### Table2 Device Dimensions Comparison

| Predicate<br>Device(K153028) | Designation    | Size                       | Tolerance |
|------------------------------|----------------|----------------------------|-----------|
|                              | Length, mm     | Average over 234 on M size | -         |
|                              | Width, mm      | Average over 96 on M size  | -         |
|                              | Thickness, mm: |                            |           |

{5}------------------------------------------------

|                | Finger         | Average 0.98  |     |     | -         |     |
|----------------|----------------|---------------|-----|-----|-----------|-----|
|                | Palm           | Average 0.096 |     |     | -         |     |
| Subject Device | Designation    | Size          |     |     | Tolerance |     |
|                |                | S             | M   | L   | XL        |     |
|                | Length, mm     | 230           | 230 | 230 | 230       | min |
|                | Width, mm      | 85            | 95  | 105 | 115       | +5  |
|                | Thickness, mm: |               |     |     |           |     |
|                | Finger         | 0.05          |     |     |           | min |
|                | Palm           | 0.05          |     |     |           | min |
| Remark         | Similar        |               |     |     |           |     |

Analysis: The physical dimensions are different with that of the predicate, but they all meet the requirements of ASTM D5250.

| Item                   |                        |                        | Subject device                                                       | Predicated device                                                    | Comparison |
|------------------------|------------------------|------------------------|----------------------------------------------------------------------|----------------------------------------------------------------------|------------|
| Colorant               |                        |                        | Blue                                                                 | Blue                                                                 | Same       |
| Physical<br>Properties | Before<br>Aging        | Tensile<br>Strength    | 11MPa, min                                                           | Average 16.9MPa                                                      | Analysis   |
|                        |                        | Ultimate<br>Elongation | 300%min                                                              | Average 550%                                                         | Analysis   |
|                        | After<br>Aging         | Tensile<br>Strength    | 11MPa, min                                                           | Average 14.4MPa, min                                                 | Analysis   |
|                        |                        | Ultimate<br>Elongation | 300%min                                                              | Average 500%                                                         | Analysis   |
|                        | Comply with ASTM D5250 |                        | Comply with ASTM D5250                                               | Comply with ASTM D5250                                               | Same       |
| Freedom from Holes     |                        |                        | Be free from holes when tested in accordance with ASTM D5151 AQL=2.5 | Be free from holes when tested in accordance with ASTM D5151 AQL=2.5 | Same       |
| Powder Content         |                        |                        | 0.07 mg per glove.<br>Meet the requirements of<br>ASTM D6124         | Meet the requirements of<br>ASTM D6124                               | Similar    |

#### Table3 Performance Comparison

Analysis: The tensile strength and ultimate elongation are different with that of the predicate, but they all meet the requirements of ASTM D5250

## Table4 Safety Comparison

| Item     | Subject device                      | Predicated device                   | Comparison |
|----------|-------------------------------------|-------------------------------------|------------|
| Material | Poly Vinyl Chloride<br>Polyurethane | Poly Vinyl Chloride<br>Polyurethane | Similar    |

{6}------------------------------------------------

|                    |               | Nitrile<br>Di-(2-ethylhexyl)<br>Terephthalate(DOTP)                                  | Diisononyl Phthalate<br>(DINP) |           |      |
|--------------------|---------------|--------------------------------------------------------------------------------------|--------------------------------|-----------|------|
| Biocompatibility   | Irritation    | Under the conditions of the<br>study, not an irritant                                | Comply with<br>ISO10993-10     | SAME      |      |
|                    | Sensitization | Under conditions of the<br>study, not a sensitizer.                                  |                                |           |      |
|                    | Cytotoxicity  | Under conditions of the<br>study, did not show potential<br>toxicity to L-929 cells. | /                              | Different |      |
| Label and Labeling |               | Meet FDA's Requirement                                                               | Meet<br>Requirement            | FDA's     | SAME |

Analysis: The materials of the subject device are little different with that of the predicate, but they all meet the performance requirements of ASTM D5250,also biocompatibility test has been performed on subject device and the test result can meet the requirements of ISO 10993 standards.

#### 8.0 Discussion of Non-clinical and Clinical Test Performed

Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

ISO 10993-10:2010 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.

ISO 10993-5:2009 Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity

ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves

ASTMD5151-19, Standard Test Method for Detection of Holes in Medical Gloves.

ASTM D5250-19, Standard Specification for Poly (vinyl chloride) Gloves for Medical Application.

| Test<br>Methodology | Purpose                        | Acceptance Criteria                                                               | Results                                                                             |
|---------------------|--------------------------------|-----------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| ASTM D5250          | Physical<br>Dimensions<br>Test | Length(mm):≥230;<br>Width(mm):<br>S: 85±5;<br>M: 95±5;<br>L: 105±5;<br>XL: 115±5; | Length: >230<br>Width:<br>S: 84-89<br>M: 93-97<br>L: 105-107<br>XL: 115-116<br>Pass |

{7}------------------------------------------------

|              |                                                     | Thickness (mm):<br>Finger: ≥0.05<br>Palm: ≥0.05 | Finger: 0.08-0.2<br>Palm: 0.07-0.08<br>Pass |        |                                                                                                    |
|--------------|-----------------------------------------------------|-------------------------------------------------|---------------------------------------------|--------|----------------------------------------------------------------------------------------------------|
| ASTM D5151   | Watertightness<br>Test for<br>Detection of<br>Holes | Meet the requirements of<br>ASTM D5151 AQL 2.5  | 0/125 leaks<br>Pass                         |        |                                                                                                    |
| ASTM D6124   | Powder<br>Content                                   | Meet the requirements of<br>ASTM D6124 < 2.0mg  | 0.07 mg;<br>Pass                            |        |                                                                                                    |
| ASTM D412    | Physical<br>properties                              | Before<br>Aging                                 | Tensile<br>Strength                         | ≥11MPa | 11.6-16.2<br>Pass                                                                                  |
|              |                                                     |                                                 | Ultimate<br>Elongation                      | ≥300%  | 308-391<br>Pass                                                                                    |
|              |                                                     | After<br>Aging                                  | Tensile<br>Strength                         | ≥11MPa | 11.0-14.0<br>Pass                                                                                  |
|              |                                                     |                                                 | Ultimate<br>Elongation                      | ≥300%  | 323-337<br>Pass                                                                                    |
| ISO 10993-5  | Cytotoxicity                                        | Non-cytotoxic                                   |                                             |        | Under conditions<br>of the study, did<br>not show potential<br>toxicity to L-929<br>cells.<br>Pass |
| ISO 10993-10 | Irritation                                          | Non-irritating                                  |                                             |        | Under the<br>conditions of the<br>study, not an<br>irritant.<br>Pass                               |
| ISO 10993-10 | Sensitization                                       | Non-sensitizing                                 |                                             |        | Under conditions<br>of the study, not a<br>sensitizer.<br>Pass                                     |

#### 9.0 Clinical Test Conclusion

Clinical testing is not needed for this device.

## 10.0 Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicated device in K153028.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYZ/K211262](https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYZ/K211262)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).