POWDER FREE VINYL PATIENT EXAMINATION GLOVES, CLEAR (NON-COLORED)

K130800 · Kaixuan Plastic Co., Ltd. · LYZ · Feb 18, 2014 · General Hospital

Device Facts

Record IDK130800
Device NamePOWDER FREE VINYL PATIENT EXAMINATION GLOVES, CLEAR (NON-COLORED)
ApplicantKaixuan Plastic Co., Ltd.
Product CodeLYZ · General Hospital
Decision DateFeb 18, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

Powder Free Vinyl Patient Examination Gloved) is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Story

Device is a disposable, non-sterile vinyl patient examination glove. Acts as a physical barrier to body fluids and bloodborne pathogens. Material is PVC; flexible and water-tight to conform to hand during medical procedures. Used in clinical settings by healthcare providers to prevent contamination between patient and examiner. Provides protection through barrier function; no electronic or mechanical components.

Clinical Evidence

No clinical data. Substantial equivalence is supported by non-clinical bench testing, including physical property testing (tensile strength, elongation), water leak testing (ASTM D5151), residual powder testing (ASTM D6124), and biocompatibility testing (ISO 10993-10).

Technological Characteristics

Material: PVC. Surface coating: PU (polyurethane). Standards: ASTM D5250-06 (Reaffirmation 2011), ASTM D5151-06 (Reaffirmation 2011), ASTM D6124-06 (Reaffirmation 2011), ISO 10993-10:2010. Form factor: Disposable, non-sterile, clear, ambidextrous glove. Dimensions: ≥230mm length; thickness ≥0.05mm (finger), ≥0.08mm (palm).

Indications for Use

Indicated for use as a non-sterile, disposable patient examination glove worn by medical examiners to prevent cross-contamination between patient and examiner. Suitable for over-the-counter use.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ # Section C ## 510(k) Summary "This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92." "The assigned 510(k) number is: _ K130800 " (applicant leave blank) Premarket Notification [510(k)] Summary [(a)(1)]. The summary contains on the first page, preferably on your letterhead paper, the submitter's name, address, phone and fax numbers, name of contact person, and date the summary was prepared : | Submitter's name : | Kaixuan Plastic Co., Ltd. | |--------------------------------|-------------------------------------------------| | Submitter's address : | No.44, Zhonghua St. Shijiazhuang, 050000, China | | Phone number : | (86) 311-67699828 | | Fax number : | (86) 311-67699906 | | Name of contact person: | Mr. Zhou Rui | | Date the summary was prepared: | 2014-02-07 | [(a)(2)]. The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known | Device Name: | Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) | |-------------------------|-------------------------------------------------------------------| | Proprietary/Trade name: | Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) | | Common Name: | Patient examination glove | | Classification Name: | Patient examination glove | | Device Classification: | I | | Regulation Number: | 21 CFR 880.6250 | | Panel: | General Hospital (80) | | Product Code: | LYZ | [(a)(3)]. An identification of the legally marketed device to which your firm is claiming substantial equivalence . Class I* Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) that meets all of the requirements of ASTM standard D 5250-06 (Reaffirmation 2011). Predicate device : Powder free Vinyl Patient Examination Gloves, Clear(Non-colored), Shijiazhuang Fuguan Plastic Products Co., Ltd. K032908 . {2}------------------------------------------------ ### [(a)(4)] A description of the device Device Description : Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) that meets all of the requirements of ASTM standard D 5250-06(Reaffirmation 2011). -- How the device functions: PVC films form a barrier to body fluids and bloodborne Pathogens -- Scientific concepts that form the basis for the device The PVC rubber is water tight under normal conditions of use. Its tensile properties cause it to conform to the hand, allowing movements necessary for a medical procedure. -- Physical and performance characteristics such as design, materials and physical properties: PVC gloves are known to create a barrier to bloodborne pathogens and body fluids. ASTM conforming tensile properties create a glove that is strong and flexible. The glove is manufactured in accordance with the requirements of ASTM D5250 and ASTM D5151 requirements. #### [(a)(5)] The summary describes the intended use of the device Device Intended Use: Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. ### [(a)(6)] A summary of the technological characteristics of new device compared to the predicate device. The Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard. | Features<br>&<br>Description | Predicate Device | Medical<br>Glove<br>Guidance<br>Manual(1661) | Subject Device | Result<br>of<br>Comparison | |---------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------| | Company | Shijiazhuang Fuguan<br>Plastic Products Co., Ltd. | | Kaixuan Plastic Co.,<br>Ltd. | -- | | 510(K) Number | K032908 | | K130800 | | | Product name | Powder Free Vinyl Patient<br>Examination Gloves,<br>Clear<br>(Non-colored) | | Powder Free Vinyl<br>Patient Examination<br>Gloves, Clear<br>(non-colored) | Same | | Product Code | LYZ | LYZ | LYZ | Same | | Size | Small/ Medium/<br>Large/X large | | Small/ Medium/<br>Large/X large | Substantially<br>equivalent | | Intend for use | Powder free Vinyl Patient<br>Examination Gloves,<br>Clear(Non-colored)is a<br>disposable device<br>intended for medical<br>purposes that is worn on<br>the examiner's hand or<br>finger to prevent<br>contamination between<br>patient and examiner. | Powder Free<br>Examination<br>Gloves:<br>A powder-free<br>patient examination<br>glove is a<br>disposable device<br>intended for<br>medical purposes<br>that is worn on the<br>examiner's hand or<br>finger to prevent<br>contamination<br>between patient and<br>examiner. | Powder free Vinyl<br>Patient Examination<br>Gloves, Clear<br>(Non-colored) is a<br>disposable device<br>intended for medical<br>purposes that is worn on<br>the examiner's hand or<br>finger to prevent<br>contamination between<br>patient and examiner. | Substantially<br>equivalent | | Device<br>Description and<br>Specifications | Meets ASTM D5250<br>-06 (Reapproved 2011) | If vinyl:<br>Do the vinyl<br>examination gloves<br>meet all the current<br>specifications listed<br>under ASTM<br>Specification D5250 | Meets ASTM D5250<br>-06 (Reapproved 2011) | Substantially<br>equivalent | | | | or an equivalent<br>consensus standard? | | | | Dimensions<br>-- Length | Meets ASTM D5250<br>-06 (Reapproved 2011) | ASTM D5250 | 230mm min for all sizes | Substantially<br>equivalent | | | ≥230mm min. | | | | | Dimensions<br>-- Width | Meets ASTM D5250<br>-06 (Reapproved 2011) | | | Substantially<br>equivalent | | | Small 80-90 mmr<br>Medium 90-100mm<br>Large 100-110mm<br>X large 110-120 mm | | Small 80-85 mm<br>Medium 95-97 mm<br>Large 102-108mm<br>X large 114-118 mm | | | Dimensions<br>-- Thickness | Meets ASTM D5250<br>-06 (Reapproved 2011) | | | | | | Finger 0.05mm min.<br>Palm 0.08mm min. | | Finger 0.05mm min.<br>Palm 0.08mm min. | | | Physical<br>Properties | Meets ASTM D5250<br>-06 (Reapproved 2011) | ASTM D5250 | Before aging/after aging | Substantially<br>equivalent | | | Before aging/after aging<br>Elongation ≥300%<br>Tensile Strength ≥14MPa | | Elongation ≥300%<br>Tensile Strength> 14MPa | | | Freedom<br>from<br>Pinholes | Meets<br>• 21 CFR 800.20<br>• ASTM D5250-06<br>(Reapproved 2011)<br>• ASTM D 5151-06<br>(Reapproved 2011) | 21 CFR 800.20<br>ASTM D5250<br>ASTM D 5151 | Meets ASTM D5151<br>-06(Reapproved 2011)<br><br>Holes<br>Inspection Level I<br>AQL2.5 | Substantially<br>equivalent | | Residual<br>Powder | Meets ASTM D 6124<br>-06(Reapproved 2011) | ASTM D 6124 | Meets ASTM D 6124<br>-06(Reapproved 2011)<br><br>Results generated<br>values below 2mg of<br>residual powder | Substantially<br>equivalent | | Compare all<br>materials used<br>to fabricate the<br>devices | PVC | If the glove is made<br>of a polymer or<br>other type of<br>material, identify<br>the material. | PVC | Substantially<br>equivalent | | Dusting or<br>Donning<br>Powder: | PU | If a donning<br>lubricant is used,<br>state the<br>composition<br>and include<br>biocompatibility<br>data for the<br>lubricant in an<br>identified<br>attachment; also<br>state the name,<br>manufacturer, and<br>address below | PU | Substantially<br>equivalent | | Dusting or<br>Donning<br>Powder: name | PU | Lubricant<br>Generic Name/<br>Lubricant<br>Brand Name | Surface Coating Agent | Substantially<br>equivalent | | Compare<br>performance<br>data supporting<br>substantial<br>equivalence | Meets<br>ASTM D5151-06<br>(Reapproved 2011)<br>ASTM D5250-06<br>(Reapproved 2011)<br>ASTM D6124-06<br>(Reapproved 2011) | At this time FDA<br>recognizes the<br>following standards:<br>Patient Examination<br>Gloves(PVC)<br>ASTM<br>D5151(Detection of<br>Holes in Medical<br>Gloves) | Meets<br>ASTM D5151-06<br>(Reapproved 2011)<br>ASTM D5250-06<br>(Reapproved 2011)<br>ASTM D6124-06<br>(Reapproved 2011) | Substantially<br>equivalent | | Single<br>Patient<br>Use | Single Patient Use | D6124(Residual<br>Powder on Medical<br>Gloves)<br>ASTM<br>D5250(Poly(vinyl<br>chloride) Gloves)<br>Single Patient Use | Single Patient Use | Substantially<br>equivalent | | Biocompatibilit<br>y | SKIN IRRITATION<br>DERMAL and<br>SENSITIZATION<br>STUDIES Meets ISO<br>10993-10 :2002/Amd.1:<br>2006 | SKIN IRRITATION<br>DERMAL and<br>SENSITIZATION<br>STUDIES<br>ISO 10993-10 | The test article was a<br>non-irritant or<br>non-sensitizer<br>SKIN IRRITATION<br>DERMAL and<br>SENSITIZATION<br>STUDIES Meets ISO<br>ISO 10993-10 Third<br>Edition 2010-08-01 | Very similar<br>only ISO<br>10993-10<br>version<br>updated<br>Substantially<br>equivalent | | Labeling for the<br>legally<br>marketed<br>device to which<br>substantial<br>equivalence is<br>claimed. | -Powder Free<br>-Patient Examination<br>Glove<br>-glove color, clear<br>(non-colored)<br>- non sterile<br>-Single Use Only<br>- Manufactured For:<br>- Lot | Chapter 4 | -Powder Free<br>-Patient Examination<br>Glove<br>-glove color, clear<br>(non-colored)<br>- non sterile<br>-Single Use Only<br>- Manufactured For:<br>- Lot | Substantially<br>equivalent | {3}------------------------------------------------ Section C(rev.03) {4}------------------------------------------------ [(b)(1)] A brief discussion of the non-clinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence . Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)meet requirements per ASTM D5250-06(Reaffirmation 2011), per ASTM D6124-06(Reaffirmation 2011), per 21 CFR 800.20 and ISO 10993-10 Third Edition 2010-08-01. [(b)(2)] A brief discussion of the clinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence . Clinical data is not needed for gloves or for most devices cleared by the 510(k) process. [(b)(3)] The conclusions drawn from the non-clinical tests that demonstrate that the device is as safe,as effective, and performed as well or better than the legally marketed device identified in (a)(3). It can be concluded that the Powder Free Vinyl Patient Examination Gloves meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL., meet labeling claims and the Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is as safe, as effective, and performs as well as the predicate device, Powder free Vinyl Patient Examination Gloves, Clear(Non-colored), Shijiazhuang Fuguan Plastic Products Co., Ltd. K032908 . {5}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring a symbol that resembles an abstract eagle or bird-like figure with flowing lines. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 18, 2014 Kaixuan Plastic Company, Limited C/O Mr. Chu Xiaoan Official Correspondent Room 1606 Bldg 1 Jianxiang Yuan #209 Bei Si Huan Zhong Road, Haidian District Beijing 100083 CHINA Re: K130800 Trade/Device Name: Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: January 2, 2014 Received: January 8, 2014 Dear Mr. Xiaoan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {6}------------------------------------------------ Page 2 - Mr. Xiaoan Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address > http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Image /page/5/Picture/7 description: The image shows a signature and some text. The text includes "Sincerely yours," followed by a signature. Below the signature, the text reads "Tejashri Purohit-Sheth, M.D., Clinical Deputy Director, DAGRID," and the word "FOR" is at the bottom right. . Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ ### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ### Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page. 510(k) Number (if known) K 1 30800 ### Device Name Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored) ### Indications for Use (Describe) Powder Free Vinyl Patient Examination Gloved) is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. Type of Use (Select one or both, as applicable) [] Prescription Use (Part 21 CFR 801 Subpart D) [X] Over-The-Counter Use (21 CFR 801 Subpart C) ## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. FOR FOR FOR FOA USE ONLY . . 17:1 : 1 :25 Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) Image /page/6/Picture/12 description: The image shows the name "Elizabeth F. Claverie" in bold font. Below the name, the date and time "2014.02.15 00:27:45 -05'00'" is displayed. The text appears to be extracted from a document or record, possibly indicating a timestamp associated with the individual's name. {8}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...