Powder Free Vinyl Patient Examination Gloves, Colored (Yellow) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Device Story
Disposable patient examination glove; made of PVC; colored yellow. Worn on examiner's hand or finger; acts as physical barrier to prevent cross-contamination between patient and examiner. Used in clinical or medical settings. Non-sterile; powder-free; single-patient use. Provides protection for healthcare providers during examinations.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including water leak tests (ASTM D5151), residual powder testing (ASTM D6124), and biocompatibility studies (ISO 10993-10).
Indicated for use as a disposable medical glove worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
{0}------------------------------------------------
"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."
"The assigned 510(k) number is: در بہ (applicant leave blank)
#### Premarket Notification [510(k)] Summary
[(a)(1)]. The summary contains on the first page, preferably on your letterhead paper, the submitter's name, address, phone and fax numbers, name of contact person, and date the summary was prepared :
| Submitter's name : | HEBEI LEADER PLASTIC CO., LTD | | |
|--------------------------------|------------------------------------------------------|--|--|
| Submitter's address : | #77, ZHONGSHAN WEST RD.<br>SHIJIAZHUANG,050000 CHINA | | |
| Phone number : | (86) 311-67699886 | | |
| Fax number : | (86) 311-67699916 | | |
| Name of contact person: | Mr. WANG Fei | | |
| Date the summary was prepared: | Sep. 19th, 2012 | | |
[(a)(2)]. The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known
| Device Name: | Powder free Vinyl Patient Examination Gloves, Colored<br>(Yellow) |
|-------------------------|-------------------------------------------------------------------|
| Proprietary/Trade name: | Powder Free Vinyl Patient Examination Gloves |
| Common Name: | Patient examination glove |
| Classification Name: | Patient examination glove |
| Device Classification: | I |
| Regulation Number: | 21 CFR 880.6250 |
| Panel: | General Hospital (80) |
| Product Code: | LYZ |
[(a)(3)]. An identification of the legally marketed device to which your firm is claiming substantial equivalence .
Class I* powder free vinyl patient examination gloves , colored (yellow) that meets all of the requirements of ASTM standard D 5250-06 (Reaffirmation 2011).
Predicate device : Powder free Vinyl Patient Examination Gloves(White, Yellow), SHIJIAZHUANG MANFUL LIGHT INDUSTRIAL PRODUCTS CO., LTD. K051156 .
### [(a)(4)] A description of the device
Device Description : Powder free Vinyl Patient Examination Gloves, Colored (Yellow) that meets all of the requirements of ASTM standard D 5250-06(Reaffirmation 2011).
{1}------------------------------------------------
## [(a)(5)] The summary describes the intended use of the device
Device Intended Use: Powder free Vinyl Patient Examination Gloves, Colored (Yellow) is a. disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
### [(a)(6)] A summary of the technological characteristics of new device compared to the predicate device.
The powder free vinyl patient examination gloves, non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.
| Features &<br>Description | Predicate Device | Medical Glove<br>Guidance<br>Manual(1661) | Subject Device | Result of<br>Comparison |
|------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Company | SHIJIAZHUAN<br>G MANFUL<br>LIGHT<br>INDUSTRIAL<br>PRODUCTS<br>CO., LTD | | HEBEI LEADER<br>PLASTIC CO.,<br>LTD | -- |
| Product name | Powder free<br>Vinyl Patient<br>Examination<br>Gloves(White,<br>Yellow) | | Powder free Vinyl<br>Patient<br>Examination<br>Gloves, Colored<br>(Yellow) | -- |
| Intend for use | Powder free<br>Vinyl Patient<br>Examination<br>Gloves, Colored<br>(White, Yellow)<br>is a disposable<br>device intended<br>for medical<br>purposes that is<br>worn on the<br>examiner's hand<br>or finger to<br>prevent<br>contamination<br>between patient<br>and examiner. | Powder-Free<br>Examination<br>Gloves:<br>A powder-free<br>patient examination<br>glove is a disposable<br>device intended for<br>medical purposes<br>that is worn on the<br>examiner's hand or<br>finger to prevent<br>contamination<br>between patient and<br>examiner. | Powder free Vinyl<br>Patient<br>Examination<br>Gloves, Colored<br>(Yellow) is a<br>disposable device<br>intended for<br>medical purposes<br>that is worn on the<br>examiner's hand or<br>finger to prevent<br>contamination<br>between patient and<br>examiner. | Substantially<br>equivalent |
| Device<br>Description and<br>Specifications | Meets ASTM<br>D5250 | If vinyl:<br>Do the vinyl<br>examination gloves<br>meet all the current<br>specifications listed<br>under ASTM<br>Specification D5250<br>or an equivalent<br>consensus standard? | Meets ASTM<br>D5250 | Substantially<br>equivalent |
| Compare all<br>materials used to<br>fabricate the<br>devices | PVC | If the glove is made<br>of a polymer or<br>other type of<br>material. identify the<br>material. | PVC | Substantially<br>equivalent |
| Dusting or<br>Donning Powder: | PU | If a donning<br>lubricant is used | PU | Substantially<br>equivalent |
| | | state the composition and include biocompatibility data for the lubricant in an identified attachment; also state the name, manufacturer, and address below | | |
| Dusting or<br>Donning Powder:<br>name | PU | Lubricant<br>Generic Name/<br>Lubricant<br>Brand Name | Surface Coating<br>Agent | Substantially<br>equivalent |
| Compare product<br>specifications | Meets ASTM<br>D5250 | We recommend you certify that your finished "powder-free" gloves meet the following:<br>ASTM D 5250 standard or an equivalent standard for vinyl | Meets ASTM<br>D5250 | Substantially<br>equivalent |
| Compare<br>performance data<br>supporting<br>substantial<br>equivalence | Meets<br>ASTM D5151<br>ASTM D5250<br>ASTM D6124 | At this time FDA recognizes the following standards:<br>Patient Examination Gloves(PVC)<br>ASTM D5151(Detection of Holes in Medical Gloves)<br>ASTM D6124(Residual Powder on Medical Gloves)<br>ASTM D5250(Poly(vinyl chloride) Gloves) | Meets<br>ASTM D5151<br>ASTM D5250<br>ASTM D6124 | Substantially<br>equivalent |
| Single Patient Use | Single Patient<br>Use | Single Patient Use | Single Patient Use | Substantially<br>equivalent |
| Biocompatibility | SKIN<br>IRRITATION<br>DERMAL and<br>SENSITIZATION<br>STUDIES<br>Meets ISO<br>10993-10 | SKIN IRRITATION<br>DERMAL and<br>SENSITIZATION<br>STUDIES<br>ISO 10993-10 | SKIN<br>IRRITATION<br>DERMAL and<br>SENSITIZATION<br>STUDIES Meets<br>ISO 10993-10 | Substantially<br>equivalent |
| Labeling for the<br>legally marketed<br>device to which<br>substantial<br>equivalence is<br>claimed. | -Powder-Free<br>-Patient<br>Examination<br>Glove<br>-Single Use Only<br>- Manufactured<br>For:<br>- Lot | Chapter 4 | -Powder-Free<br>-Patient<br>Examination Glove<br>-Single Use Only<br>- Manufactured<br>For:<br>- Lot | Substantially<br>equivalent |
{2}------------------------------------------------
[(b)(1)] A brief discussion of the nonclinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .
Powder free Vinyl Patient Examination Gloves, Colored (Yellow) meet requirements per ASTM D5250-06(Reaffirmation 2011), per ASTM D6124-06(Reaffirmation 2011), per 21 CFR 800.20 and ISO 10993-10: 2006 Third Edition 2010-08-01.
{3}------------------------------------------------
[(b)(2)] A brief discussion of the clinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .
Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.
[(b)(3)] The conclusions drawn from the nonclinical tests that demonstrate that the device is as safe,as effective, and performed as well or better than the legally marketed device identified in (a)(3).
It can be concluded that the Powder Free Vinyl Patient Examination Gloves meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL., meet labeling claims .
{4}------------------------------------------------
## Section 15
## Substantial Equivalence Discussion
The Powder free Vinyl Patient Examination Gloves, Colored (Yellow) made by HEBEI LEADER PLASTIC CO., LTD, have the same intended use and technological characteristics as the Predicate devices, Powder free Vinyl Patient Examination Gloves(White, Yellow) made by SHIJIAZHUANG MANFUL LIGHT INDUSTRIAL PRODUCTS CO., LTD. K051156.
The Subject Device is substantially equivalent to the predicate devices with regard to technologies, in design and very similar construction, material, function, and application. More detail information on the Table 1.
Based on the information in following Table I, II & III , the Subject Device (Powder free Vinyl Patient Examination Gloves, Colored (Yellow)) has the same technological characteristics as the predicate device. The Powder free Vinyl Patient Examination Gloves, Colored (Yellow) made by HEBEI LEADER PLASTIC CO., LTD, are substantially equivalent to the predicate devices.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo for the Department of Health and Human Services - USA. The logo includes a symbol with three curved lines, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the symbol. The text is in a circular format.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
### November 30, 2012
Hebei Leader Plastic Company, Limited C/O Mr. Chu Xiaoan Room 1606 Building 1, Jianxiang Yuan No. 209 Bei Si Huan Zhong Road, Haidian District Beijing, China 100083
Re: K122920
Trade/Device Name: Powder Free Vinyl Patient Examination Gloves, Colored (Yellow) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: October 19, 2012 Received: November 1, 2012
Dear Mr. Xiaoan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
#### Page 2 - Mr. Xiaoan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Heath's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours,
Anthong Lo min
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{7}------------------------------------------------
#### INDICATIONS FOR USE
Applicant: HEBEI LEADER PLASTIC CO., LTD
# 510(k) Number (if known): * 122920
Device Name: Powder Free Vinyl Patient Examination Gloves, Colored (Yellow)
Indications For Use:
Powder Free Vinyl Patient Examination Gloves, Colored (Yellow) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) Maria Cruz-ISher Cruz-Fisher Scorminaria Cruzen. Maria Cruzen. Maria Cruzena. Maria 1920200.00.11.1.
Data 113112 - Data 113.000 1130113011301130113011131114910 (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K122420
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.