← Product Code [LYZ](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYZ) · K122656

# POWDER-FREE VINYL PATIENT EXAMINATION GLOVES, YELLOW COLOR (K122656)

_Shandong Yifan Plastic Products Co., Ltd. · LYZ · May 6, 2014 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYZ/K122656

## Device Facts

- **Applicant:** Shandong Yifan Plastic Products Co., Ltd.
- **Product Code:** [LYZ](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYZ.md)
- **Decision Date:** May 6, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or finger to prevent contamination between patient and examiner.

## Device Story

Disposable vinyl patient examination glove; worn by healthcare examiners on hands or fingers; acts as physical barrier to body fluids and blood-borne pathogens; prevents cross-contamination between patient and examiner; conforms to hand for procedural movement; non-sterile; yellow color.

## Clinical Evidence

Bench testing only. Testing included physical dimensions, tensile strength, ultimate elongation (before and after aging), residual powder content (ASTM D6124-06), pinhole testing (ASTM D5151-06), and biocompatibility (primary skin irritation and dermal sensitization). All results met ASTM D5250-06 requirements.

## Technological Characteristics

Material: Polyvinyl chloride, polyurethane, and diisononyl phthalate (DINP). Form: Disposable, powder-free, yellow vinyl glove. Standards: ASTM D5250-06, ASTM D6124-06, ASTM D5151-06. Dimensions: 230mm length, 95mm palm width, 0.085mm palm thickness, 0.090mm finger thickness. Non-sterile.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Synthetic Vinyl Patient Examination Gloves - Powder Free ([K992821](/device/K992821.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510 (K) SUMMARY

# MAY 0 6 2014

This summary of 510(K) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21CFR 807.92.

#### 1. Submitter's Identification:

Shandong Yifan Plastic Products Co., Ltd. Building 11, Sunny Industries Park, Shuangyang Village, Zichuan District, Zibo City, Shandong, China 255000 Date summary prepared: April 15, 2014

#### 2. Name of the Device:

Shandong Yifan Plastic Products Co., Ltd. Powder-free Vinyl Patient Examination Gloves, Yellow Color

#### 3. Common name/classification name of the Device:

Regulation Number: 21 CFR 880.6250, Regulation Name: Patient Examination Glove Regulation Class: Class I Product Code: LYZ

#### 4. Contact Person:

Jennifer Guo, Tel: 909-548-4828

#### ડ. Predicate Device Information:

Shijiazhuang Hongxiang Plastic Products Ltd. Synthetic Vinyl Patient Examination Gloves - Powder Free (K992821)

#### 6. Device Description:

Classified by FDA's General and Plastic Surgery Device panel as Class I, and meets all requirement of ASTM Standard D5250-06. The subject device is a garment covering the hand and wrist area. Clovers have separate sheaths or openings for each finger and the thumb. And vinyl films form a barrier to body fluid and blood-borne pathogens, and that the vinyl rubber is water tight under normal conditions of use. Its tensile

{1}------------------------------------------------

properties cause it to conform to the hand, allowing movements necessary for a medical procedure.

#### 7. Intended Use:

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or finger to prevent contamination between patient and examiner.

#### 8. Comparison to Predicate Devices:

Shandong Yifan Plastic Products Co., Ltd. Powder-free Vinyl Patient Examination Gloves, Yellow Color are substantially equivalent in safety and effectiveness to the Shijiazhuang Hongxiang Plastic Products Co., Ltd.(K992821)

## Table 7-2. Side-by-Side Comparison of Intended Use, Design, Material, Physical, Biocompatibility, and Performance Testing

|                                                                    | Proposed Device                                                                                                                                                                                          | Predicate Device (K992821)                                                                                                                                                                               |
|--------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Description                                                        | Shandong Yifan Plastic Products Co.,<br>Ltd Powder-Free Vinyl Patient<br>Examination Gloves, Yellow Color                                                                                                | Shijiazhuang Hongxiang<br>Powder-free Vinyl Patient<br>Examination Gloves                                                                                                                                |
| Indication for Use                                                 | A patient examination glove is a<br>disposable device intended for medical<br>purposes that is worn upon the<br>examiner's hands or fingers to prevent<br>contamination between patient and<br>examiner. | A patient examination glove is a<br>disposable device intended for medical<br>purposes that is worn upon the<br>examiner's hands or fingers to prevent<br>contamination between patient and<br>examiner. |
| Labeling: Instruction<br>for use                                   | A garment covering the hand and wrist<br>area. Clovers have separate sheaths or<br>openings for each finger and the<br>thumb.                                                                            | Substantially equivalent                                                                                                                                                                                 |
| Labeling: Labels on<br>the carton                                  | Labels include: Product name; color;<br>"single use Only" size, piece count, lot<br>number, distributor name, and<br>manufacturer address.                                                               | Substantially equivalent                                                                                                                                                                                 |
| Device Materials                                                   | Poly Vinyl Chloride<br>Polyurethane<br>Diisononyl Phthalate (DINP)                                                                                                                                       | Substantially equivalent                                                                                                                                                                                 |
| Before Aging: Tensile<br>Strength(Mpa) and<br>Ultimate Elongations | Average Tensile Strength (Mpa):<br>16.9<br>Average Ultimate Elongations: 550%                                                                                                                            | Substantially equivalent                                                                                                                                                                                 |
| After Aging: Tensile<br>Strength (Mpa)                             | Average Tensile Strength (Mpa):                                                                                                                                                                          | Substantially equivalent                                                                                                                                                                                 |

{2}------------------------------------------------

| Strength(Mpa) and<br>Ultimate Elongations              | 14.4<br>Average Ultimate Elongations: 500%                                                                                                                                                                                                                                                                                                                                    |                          |
|--------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|
| Overall Length on<br>Medium Size                       | Average over 230mm                                                                                                                                                                                                                                                                                                                                                            | Substantially equivalent |
| Width of Palm on<br>Medium Size                        | Average 95mm                                                                                                                                                                                                                                                                                                                                                                  | Substantially equivalent |
| Palm Thickness                                         | Average 0.085 mm                                                                                                                                                                                                                                                                                                                                                              | Substantially equivalent |
| Figure Thickness                                       | Average 0.090 mm                                                                                                                                                                                                                                                                                                                                                              | Substantially equivalent |
| Residual Powder                                        | Residual powder content ≤ 2 mg per<br>glove per ASTM D6124-06.                                                                                                                                                                                                                                                                                                                | Substantially equivalent |
| Pinhole Results                                        | According to ASTM D5151-06,<br>Testing result indicates pinhole were<br>found less than two pieces gloves out<br>of 125 pieces gloves. AQL 2.5 is met.                                                                                                                                                                                                                        | Substantially equivalent |
| Biocompatibility<br>Result: Primary Skin<br>Irritation | Not a irritant                                                                                                                                                                                                                                                                                                                                                                | Substantially equivalent |
| Dermal Sensitization                                   | Not a sensitizer                                                                                                                                                                                                                                                                                                                                                              | Substantially equivalent |
| Summary of<br>comparison                               | Shandong Yifan Plastic Products Co., Ltd Powder-Free Vinyl Patient<br>Examination Gloves, Non-Sterile, Yellow Color (subject device) and<br>Shijiazhuang Hongxiang Powder-free Vinyl Patient Examination Gloves<br>(predicate device) are substantially equivalent in all technological<br>characteristics, including tensile strength, ultimate elongations size, thickness, |                          |

### Discussion of Non-Clinical tests performed for Determination of 9. Substantial Equivalence are as follows:

The standards used for Shandong Yifan Plastic Products Co., Ltd. glove production are based on ASTM-D-5250-06. All testing meets requirements for Physical and Dimensions Testing conducted on gloves, Inspection Level S-2, AQL 2.5.

The FDA 1000 ml. Water Fill Test was also conducted with samplings of AQL 2.5, Inspection Level I, meeting these requirements, Primary Skin irritation and Skin Sensitization testing was conducted with results showing no primary skin irritant or sensitization reactions.

The subject gloves are powder free.

י

{3}------------------------------------------------

#### Discussion of Clinical Tests Performed: 10.

Not Applicable

#### Conclusions: 11.

The subject device, Shandong Yifan Plastic Products Co., Ltd. Powder-Free Vinyl Patient Examination Gloves, Yellow Color, Non-Sterile, is substantially equivalent to the predicate device (K992821) Shijiazhuang Hongxiang Powder-free Vinyl Patient Examination Gloves, in intended use as well as technological characteristics, including tensile strength, ultimate elongations size, thickness, residual powder and pinhole.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circle around the eagle.

ood and Drug Administration 0903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 6, 2014

Shandong Yifan Plastic Products Company Limited C/O Ms. Ling Zhu Official Correspondent 12390 East End Avenue Chino, CA 91710

Re: K122656

Trade/Device Name: Powder-Free Vinyl Patient Examination Gloves, Yellow Color Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: April 23, 2014 Received: April 25, 2014

### Dear Ms. Zhu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Mr. Zhu

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Tejashri Purohit-Sheth, M.D.
Clinical Deputy Director
DAGRID/ODE/CDRH

Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

### Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on tast page.

### 510(k) Number (il known) K 122656

### Device Name

Powder-free Vinyl Patient Examination Gloves, Yellow Color

### Indications for Use (Describe)

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands of fingers to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable) [] Prescription Use (Part 21 CFR 801 Subpart D)

Do Over-The-Counter Use (21 CFR 801 Subpart C)

# Please do not write below this line – continue on a separate page if needed.

### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

beth F. Clave ou=FDA, ou=Poople.
cn=Eizabeth F. Clav 12 12 12 03.03 STATE OF ST ਾ ਸਿੱਖ ਮੁਹੰਮ

FORM FDA 3881 (1/14)

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYZ/K122656](https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYZ/K122656)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
