← Product Code [LYZ](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYZ) · K110536

# POWDER-FREE VINYL PATIENT EXAMINATION GLOVES (K110536)

_Zibo Tengyue Plastic Products Co., Ltd. · LYZ · Mar 19, 2012 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYZ/K110536

## Device Facts

- **Applicant:** Zibo Tengyue Plastic Products Co., Ltd.
- **Product Code:** [LYZ](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYZ.md)
- **Decision Date:** Mar 19, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or finger to prevent contamination between patient and examiner (21CFR 880.6250)

## Device Story

Disposable non-sterile vinyl patient examination gloves; worn on hands/fingers by clinicians or users to prevent cross-contamination between patient and examiner. Device functions as a physical barrier. Manufactured to meet ASTM D5250-06 (2011) standards for physical dimensions and residual powder limits (<2mg/pc). Used in clinical or general medical settings.

## Clinical Evidence

Bench testing only. No clinical data provided. Testing included physical dimensions, water fill test for pinholes (AQL 2.5), and biocompatibility (primary skin irritation and sensitization).

## Technological Characteristics

Vinyl material; powder-free (<2mg/pc per ASTM D6124-06). Non-sterile. Dimensions and physical properties conform to ASTM D5250-06 (2011).

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Shijiazhuang Star Plastic Co., Ltd. Vinyl Examination Gloves - Powder Free ([K100699](/device/K100699.md))
- Tangshan Zhonghong Pulin Food Products Co., Ltd. Vinyl Examination Gloves - Powder Free ([K022091](/device/K022091.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# KI10536

MARY 11 9, 2012

## 510 (K) SUMMARY

This summary of 510(K) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21CFR 807.92.

#### Submitter's Identification: 1.

Zibo Tengyue Plastic Products Co., Ltd. 77 Center of Zhangdian, Zhangdian District Zibo, Shandong, 255400

Contact Person: Jiang, Xiyan Product Manager Phone Number: 011-86-13355336122

Date summary prepared: August 18, 2011

#### 2. Name of the Device:

Zibo Tengyue Plastic Products Co., Ltd. Powder-Free Vinyl Patient Examination Gloves, Non-Sterile

#### 3. Predicate Device Informaton:

Shijiazhuang Star Plastic Co., Ltd. Vinyl Examination Gloves - Powder Free (K 100699)

Tangshan Zhonghong Pulin Food Products Co., Ltd. Vinyl Examination Gloves - Powder Free (K022091)

#### 4. Device Description:

A Powder-Free Vinyl Patient Examination Gloves is a disposable device intended for medical purposes that is worn upon the examiner's hands or finger to prevent contamination between patient and examiner (21CFR 880.6250). Our Powder-Free Vinyl Patient Examination Gloves do not contain any UPS powder according to the results of testing conducted on the gloves. The residual powder testing is conducted per standards of ASTM IN-6124-06, which demonstrates that the powder free glove is less than 2mg/pc. Classified by FDA's General and Plastic Surgery Device panel as Class I, 21 CFR 880.6250, Powder-Free Vinyl Patient Examination Glove, 80LtZ, and meets all requirement of ASTM Standard D5250-06 (2011).

{1}------------------------------------------------

#### Intended Use: 5.

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or finger to prevent contamination between patient and examiner (21CFR 880.6250)

#### Comparison to Predicate Devices: ર્.

Zibo Tengyue Plastic Products Co., Ltd. Powder-Free Vinyl Patient Examination Gloves are substantially equivalent in safety and effectiveness to Tangshan Zhonghong Pulin Food Products Co., Ltd. (K022091) and Shijiazhuang Star Plastic., Co., Ltd. (K100699) Powder-Free Viny) Patient Examination Gloves.

#### 7. Discussion of Non-Clinical tests performed for Determination of Substantial Equivalence are as follows:

The standards used for Zibo Tengyue Plastic Products Co., Ltd. glove production are based on ASTM-D-5250-06 (2011). All testing meets requirements for Physical and Dimensions Testing conducted on gloves. Inspection Level S-2, AOL 2.5.

The FDA 1000 ml. Water Fill Test was also conducted with samplings of AOL 2.5, Inspection Level I, meeting these requirements, Primary Skin irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions.

Our gloves meet standard of ASTM D6124-06 (Standard Test Method for Residual Powder on Medical Gloves).

#### Discussion of Clinical Tests Performed: 8.

Not Applicable - There is no hypoallergenic claim.

#### の。 Conclusions:

Zibo Tengyue Plastic Products Co., Ltd. Powder-Free Vinyl Patient Examination Gloves conforms fully to ASTM-D-5250-06 (2011) standard as well as applicable 21 CFR references, and, meets pinhole FDA requirements, biocompatibility requirements and labeling claims as shown by data in Sentic, 7. There are no safety/efficacy issues or new claims from the "substantial equivalence" products cited.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter of the circle.

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Zibo Tengyue Plastic Products Company, Limited C/O Mr. Frank Liu President Basic Medical Industries, Incorporated 12390 East End Avenue Chino, California 91710

MAR 1 9 2012

Re: K110536

Trade/Device Name: Powder-Free Vinyl Patient Examination Gloves, Non-Sterile Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: February 10, 2012 Received: February 24, 2012

Dear Mr. Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

### Page 2- Mr. Liu

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

G

for

Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Anthony D. Watson, B.S., M.S., M.B.A.

Enclosure

{4}------------------------------------------------

# INDICATIONS FOR USE

Zibo Tengyue Plastic Products Co., Ltd. Applicant:

510(k) Number: KI 10236

Powder-Free Vinyl Patient Examination Gloves, Non-Sterile Device Name:

Indications of Use:

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or finger to prevent contamination between patient and examiner (21CFR 880.6250)

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over the Counter Use X (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

= trill = " = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = =

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: ___ KJ1 05 34

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYZ/K110536](https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYZ/K110536)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
