POWDER-FREE VINYL PATIENT EXAMINATION GLOVES, YELLOW COLOR
K100486 · Zibo Yinghao Medical Products Co., Ltd. · LYZ · May 5, 2010 · General Hospital
Device Facts
Record ID
K100486
Device Name
POWDER-FREE VINYL PATIENT EXAMINATION GLOVES, YELLOW COLOR
Applicant
Zibo Yinghao Medical Products Co., Ltd.
Product Code
LYZ · General Hospital
Decision Date
May 5, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.
Device Story
Disposable synthetic vinyl patient examination gloves; non-sterile; yellow color. Worn by healthcare personnel during patient examinations to provide a barrier against contamination from body fluids, waste, or environmental factors. Manufactured to meet ASTM D5250-06 standards. Used in clinical or medical settings. Provides protection for both patient and examiner.
Clinical Evidence
Bench testing only. Testing included physical dimensions, 1000 ml water fill test for pinholes (AQL 2.5), and biocompatibility (primary skin irritation and sensitization). Results confirmed compliance with ASTM D5250-06 and FDA requirements.
Technological Characteristics
Synthetic vinyl material; powder-free (≤2 mg powder/glove); non-sterile; yellow color. Complies with ASTM D5250-06. Physical testing conducted per Inspection Level S-2/AQL 2.5. Water fill testing conducted per Inspection Level I/AQL 2.5.
Indications for Use
Indicated for use as a disposable medical glove worn on the hands or fingers of healthcare personnel to prevent cross-contamination between the patient and the examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
{0}------------------------------------------------
K100486
## 510 (K) SUMMARY
MAY - 5 2010
This summary of 510(K) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21CFR 807.92.
#### 1. Submitter's Identification:
Zibo Yingbo Medical Products Co., Ltd. Qilu Chemcial Industry Park Zhangdian District, Zibo, Shandong, China
Date summary prepared: Feb. 10, 2010
#### 2. Name of the Device:
Zibo Yingbo Medical Products Co., Ltd. Synthetic Vinyl Patient Examination Gloves -- Powder Free, Yellow Color
#### 3. Predicate Device Informaton:
Shijiazhuang Hongxiang Plastic Products Ltd. Synthetic Vinyl Patient Examination Gloves - Powder Free (K992821)
#### 4. Device Description:
Classified by FDA's General and Plastic Surgery Device panel as Class I, 21 CFR 880.6250, Powder-Free Vinyl Patient Examination Glove, 80LYZ, and meets all requirement of ASTM Standard D5250-06.
#### 5. Intended Use:
A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.
#### 6. Comparison to Predicate Devices:
Zibo Yingbo Medical Products Co., Ltd. Synthetic Vinyl Patient Examination Gloves, Powder-Free, yellow color are substantially equivalent in safety and effectiveness to the Shijiazhuang Hongxiang Plastic Products Co., Ltd. and Sunmax Enterprise Shanghai Co., Ltd. Powder-Free Vinyl Patient Examination Gloves, Powder free.
{1}------------------------------------------------
### 7. Discussion of Non-Clinical tests performed for Determination of Substantial Equivalence are as follows:
The standards used for Zibo Yingbo Medical Products Co., Ltd. glove production are based on ASTM-D-5250-06. All testing meets requirements for Physical and Dimensions Testing conducted on gloves, Inspection Level S-2, AOL 2.5.
The FDA 1000 ml. Water Fill Test was also conducted with samplings of AOL 2.5, Inspection Level I, meeting these requirements, Primary Skin irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitize was reactions.
There are no special labeling claims and we do not claim our gloves as hypoallergenic is conducted to insure that our gloves meet our "overder-free" claims (contain no more than 2 mg powder per glove).
#### Discussion of Clinical Tests Performed: 8.
Not Applicable - There is no hypoallergenic claim.
#### 9. Conclusions:
Zibo Yingbo Medical Products Co., Ltd. Synthetic Vinyl Patient Examination Gloves, Powder-Free, yellow color conform fully to ASTM-D-5250-06 standard as well as applicable 21 CFR references, and, meets pinhole FDA requirements, biocompatibility requirements and labeling claims as shown by data in Section 7. There are no safety/efficacy issues or new claims from the "substantial equivalence" products cited.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three curved lines representing its wings. The eagle is enclosed in a circle, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference of the circle. The text is in all caps and is written in a simple, sans-serif font.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Zibo Yingbo Medical Products Company, Limited C/O Mr. Frank Liu President Basic Medical Industries, Incorporated 12390 East End Avenue Chino, California 91710
MAY - 5 2010
Re: K100486
Trade/Device Name: Patient Vinyl Examination Gloves, Powder Free, Non-Sterile Yellow Color Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: April 21, 2010 Received: April 23, 2010
Dear Mr. Liu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm1115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Susan Runner
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K100486
# Zibo Yingbo Medical Products Co., Ltd.
Qilu Chemical Industry Park Zhangdian District, Zibo, Shandong, China
## INDICATIONS FOR USE
Zibo Yingbo Medical Products Co., Ltd. Applicant:
510(k) Number: K | 004 86
Device Name:
Patient Vinyl Examination Gloves, Powder free, Non-Sterile, Yellow Color
Indications of Use:
A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or finger to prevent contamination between patient and examiner (21CFR 880.6250)
Prescription Use
Over the Counter Use _____X
Factory Initials
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
K100486 510(k) Number: