VINYL CO-POLYMER POWDER FREE EXAMINATION GLOVES, BLUE COLOR

K070852 · Shijiazhang Hongxiang Plastic Products Co., Ltd. · LYZ · Jul 13, 2007 · General Hospital

Device Facts

Record IDK070852
Device NameVINYL CO-POLYMER POWDER FREE EXAMINATION GLOVES, BLUE COLOR
ApplicantShijiazhang Hongxiang Plastic Products Co., Ltd.
Product CodeLYZ · General Hospital
Decision DateJul 13, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.

Device Story

Disposable vinyl co-polymer patient examination glove; blue color. Worn by healthcare personnel on hands/fingers to provide barrier protection against contamination from patient body fluids, waste, or environment. Device is non-sterile and powder-free. Used in clinical settings to reduce risk of cross-contamination between examiner and patient.

Clinical Evidence

Bench testing only. No clinical data provided. Testing included physical dimensions, water-tightness (AQL 2.5), residual powder content (<2 mg/glove), and biocompatibility (primary skin irritation and sensitization).

Technological Characteristics

Vinyl co-polymer material. Powder-free. Complies with ASTM D-5250-00E4. Water-tightness tested per ASTM D-5151-99. Residual powder tested per ASTM D-6124-06. Class I device, product code LYZ.

Indications for Use

Indicated for use as a disposable patient examination glove worn on the examiner's hand or finger to prevent cross-contamination between patient and examiner.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # JUL 1 3 2007 ### 510(K) SUMMARY This summary of 510(K) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR \$807.92. The assigned 510(K) number is: KO 90 852 ### l . Submitter's Identification: Mr. Zhanmin Lu Shijiazhuang Hongxiang Plastic Products Co., Ltd. New high technologies industrial zone, Luquan city, Hebei, China 050200 Date Summary Prepared: Dec 30, 2006 ### 2. Name of the Device: Shijiazhuang Hongxiang Plastic Products Co., Ltd. Vinyl Co-Polymer Powder Free Examination Gloves, Blue Color ### 3. Predicate Device Information: Grand Work Plastic Products Co., Ltd. Vinyl Co-Polymer Powder Free Examination Gloves, Blue Color (K061562) #### 4. Device Description: Classified by FDA's General and Plastic Surgery Device panel as Class I. 21 CFR 880.6250, Power-Free Vinyl Patient Examination Glove, 80 LYZ, and meets all requirements of ASTM standard D-5250-00E4 equirements of ASTM Standard D-5230-00 ### 5. Intended Use: A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment. {1}------------------------------------------------ ## 6. Comparison to Predicate Devices: Shijiazhuang Hongxiang Plastic Products Co., Ltd.'s Vinyl Co-Polymer Powder Free Examination Gloves (Blue Color) is substantially equivalent in safety and effectiveness to the Grand Work Plastic Products Co., Ltd.'s Vinyl Co-Polymer Powder Free Examination Gloves (Blue Color) # 7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as Follows: The standards used for Shijiazhuang Hongxiang Plastic Products Co., Ltd.'s glove product are bassed on ASTM D-5250-00€4. All testing meets requirements for physical and dimensions testing conducted on gloves. Inspection level S-2, AQL 4.0. The FDA 1000 ml. Water Fill Test based on ASTM D-5151-99 was also conducted with samplings of AQL 2.5, inspection level G-1, meeting these requirements. Primary Skin Irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions. There are no special labeling claims and we do not claim our gloves as hypoallergenic on our labels. A Residual Powder Test that based on ASTM D-6124-06 for Starch at finished inspection is conducted to insure that our gloves meet our "powder-free" claims (contain no more than 2 mg powder per glove). ### 8. Discussion of Clinical Tests Performed: Not Applicable - There is no hypoallergenic Claim. # 9. Conclusions: Shijiazhuang Hongxiang Plastic Products Co., Ltd.'s Vinyl Co-Polymer Powder Free Examination Gloves (Blue Color) conform fully to ASTM D-5250-0054 standard as well as applicable 21 CFR references, and meets pinhole FDA requirements. biocompatibility requirements and labeling claims as shown by data in Section 7. There are no safety/efficacy issues or new claims from the "substantial equivalence" products cited. {2}------------------------------------------------ ood and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUL 1 3 2007 Shijiazhuang Hongxiang Plastic Products Company Limited C/O Ms. Kathy Liu Official Correspondent Gloveco, Incorporated 3973 Schaefer Avenue Re: K070852 Chino, California 91710 Trade/Device Name: Vinyl Co-Polymer Powder Free Exam Gloves, Blue Color Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: June 22, 2007 Received: June 25, 2007 Dear Ms. Liu: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Ms. Liu Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Clues Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ### Attachment A Page 1 of 1 510(K) NUMBER (IF UNKNOW): KO70852 Shijiazhuang Hongxiang Plastic Products Co., Ltd. APPLICANT: Vinyl Co-Polymer Powder Free Exam Gloves, Blue Color DEVICE NAME: ### INDICATIONS FOR USE: A patient examination glove is disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. Prescription Use (Per 21 CFR 801.109) OR Over-The-Count Use (Optional Format 1-2-96) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.) | Concurrent of CDRH, Office of Device Evaluation (ODE) | |-------------------------------------------------------| | (Division Sign-Off) | | Division of Anesthesiology, General Hospital | | Infection Control, Dental Devices | 510(k) Number: K070852
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...