← Product Code [LYZ](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYZ) · K051723

# POWDER-FREE VINYL PATIENT EXAMINATION GLOVES (WHITE, YELLOW) (K051723)

_Zibo Sinocare Plastic Products Co., Ltd. · LYZ · Jul 15, 2005 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYZ/K051723

## Device Facts

- **Applicant:** Zibo Sinocare Plastic Products Co., Ltd.
- **Product Code:** [LYZ](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYZ.md)
- **Decision Date:** Jul 15, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

Powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

## Device Story

Disposable vinyl patient examination gloves; colored white or yellow; worn by healthcare examiners on hands or fingers. Function: barrier protection to prevent contamination between patient and examiner. Used in clinical settings. Benefit: reduces risk of cross-contamination during medical examinations.

## Clinical Evidence

No clinical data. Bench testing only. Device meets ASTM D 5250-00e4 standards for physical properties and dimensions, 21 CFR 800.20 for pinhole AQL, and ISO 10993-10 for biocompatibility.

## Technological Characteristics

Material: Vinyl. Form factor: Disposable patient examination glove (white/yellow). Standards: ASTM D 5250-00e4 (dimensions, physical properties, powder residual), 21 CFR 800.20 (pinhole), ISO 10993-10 (biocompatibility). Non-sterile.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Powder-Free Vinyl Patient Examination Gloves, SHIJIAZHUANG MANFUL LIGHT INDUSTRIAL PRODUCTS CO., LTD. ([K015156](/device/K015156.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# C Summary

"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."

Premarket Notification [510(k)] Summary

[(a)(1)]. The summary contains on the first page, preferably on your letterhead paper, the submitter's name, address, phone and fax numbers, name of contact person, and date the summary was prepared :

| Submitter's name :             | Zibo SinoCare Plastic Products Co., Ltd                                  |
|--------------------------------|--------------------------------------------------------------------------|
| Submitter's address :          | NO. 208 YiXi Road, Linzi District, Zibo, Shandong,<br>P.R. China, 255411 |
| Phone number :                 | (86) 533-7488136                                                         |
| Fax number :                   | (86) 533-7524666                                                         |
| Name of contact person:        | Chen Yan                                                                 |
| Date the summary was prepared: | 20. May 2005                                                             |

[(a)(2)]. The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known

| Device Name:            | Powder Free Vinyl Patient Examination Gloves, Colored (White, Yellow) |  |
|-------------------------|-----------------------------------------------------------------------|--|
| Proprietary/Trade name: | Powder Free Vinyl Patient Examination Gloves                          |  |
|                         | Other clients private labeling                                        |  |
| Common Name:            | Patient examination glove                                             |  |
| Classification Name:    | Patient examination glove                                             |  |
| Device Classification:  | 1                                                                     |  |
| Regulation Number:      | 21 CFR 880.6250                                                       |  |
| Panel:                  | General Hospital (80)                                                 |  |
| Product Code:           | LYZ                                                                   |  |

[(a)(3)]. An identification of the legally marketed device to which your firm is claiming substantial equivalence .

Class I* powder frec vinyl patient examination gloves (white, yellow) that meets all of the requirements of ASTM standard D 5250-0064.

Predicate device : Powder-Free Vinyl Patient Examination Gloves,. SHIJIAZHUANG MANFUL LIGHT INDUSTRIAL PRODUCTS CO., LTD.,K015156

{(a}(4)] A description of the device

Device Description : powder free vinyl patient examination gloves (white, yellow) that meets all of the requirements of ASTM standard D 5250-0084.

{1}------------------------------------------------

K051723

Device Description : powder free vinyl patient examination gloves (white, yellow) that meets all of the requirements of ASTM standard D 5250-004

## [(a)(5)] The summary describes the intended use of the device

Device Intended Use: powder free vinyl patient examination glove colored (white, yellow) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

### J(a)(6)] A summary of the technological characteristics of new device compared to the predicate device.

The powder free vinyl patient examination gloves, non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.

| Characteristics       | Standard                                  | Device performance                   |
|-----------------------|-------------------------------------------|--------------------------------------|
| Dimension             | ASTM standard D 5250-00e4                 | Meets                                |
| Physical Properties   | ASTM standard D 5250-00e4                 | Meets                                |
| Freedom from pinholes | 21 CFR 800.20                             | Meets                                |
| Powder Residual       | ASTM standard D 5250-00e4                 | Meets                                |
| Biocompatability      | and D6124-01                              | <2mg/glove                           |
|                       | Primary Skin Irritation in<br>rabbits     | Passes                               |
|                       |                                           | Not a Primary Skin Irritation        |
|                       | Dermal sensitization in the<br>guinea pig | Passes<br>Not a Dermal sensitization |

## [(b)(1)] A brief discussion of the nonclinical submitted, reference, or relied on in the ((0)(1) A brig. discussion for a determination of substantial equivalence .

Powder free vinyl patient examination gloves (white, yellow) meet requirements per ASTM Powder free Thry Patient 6, 24-01, per 21 CFR 800.20 and ISO10993-10.

## [(b)(2)] A brief discussion of the clinical submitted, reference, or relied on in the premarket ((v)(2)) it bridge for a determination of substantial equivalence .

Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.

#### [(b)(3)] The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe, as effective, and performed as well or better than the legally marketed device identified in (a)(3).

It can be concluded that the Powder Free Vinyl Patient Examination Gloves meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL., meet labeling claims .

{2}------------------------------------------------

Public Health Service

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" around the perimeter. In the center of the logo is a stylized image of three human profiles facing to the right, with three lines above them that appear to represent flowing hair or energy.

JUL 1 5 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Zibo Sinocare Plastic Products Company c/o Mr. Chu Xiaoan 209 Bei Si Huan Zhong Road Haidian District, Bldg. 1 Beijing, CHINA 100083

Re: K051723

Trade/Device Name: Powder Free Vinyl Patient Examination Glove Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: May 20, 2005 Received: June 27, 2005

Dear Mr. Xiaoan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

#### Page 2 - Mr. Xiaoan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that I Drederal statutes and regulations administered by other Federal agencies. You must or any 1 with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.

Sincerely yours,

Sytite y. Michael Ows

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

3.0 Indications for Use Statement: Include the following or equivalent Indications for Use page. The information, data and labeling claims in the entire 510(k) submission must support and agree with the Indications for Use statement.

#### INDICATIONS FOR USE

Zibo SinoCare Plastic Products Co., Ltd Applicant:____________________________________________________________________________________________________________________________________________________________________

K051723 510(k) Number (if known):___

Device Name: Powder Free Vinyl Patient Examination Gloves, Colored (White, Yellow) 1

Indications For Use:

Powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

Shula H. Munder, M.D. 2/14/05

OW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

510(k) Number: _A 311

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYZ/K051723](https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYZ/K051723)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).