← Product Code [LYY](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYY) · K032293

# NON-STERILE, POWDER FREE GREEN LATEX EXAMINATION GLOVES WITH OR WITHOUT SOUR APPLE OR PEPPERMINT SCENTING PLUS A PROTEIN (K032293)

_Sgmp Co., Ltd. · LYY · Aug 28, 2003 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYY/K032293

## Device Facts

- **Applicant:** Sgmp Co., Ltd.
- **Product Code:** [LYY](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYY.md)
- **Decision Date:** Aug 28, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

A medical glove to be worn on the hand of the health care and similar personnel to prevent contamination between health care personnel and the patient.

## Device Story

Non-sterile, powder-free green latex examination gloves; optional sour apple or peppermint scenting; protein content labeled <50 µg/g. Used by healthcare personnel in clinical settings as a protective barrier. Device functions as a physical barrier to prevent contamination between wearer and patient. Performance verified against ASTM D3578-01aE2 standards for dimensions, tensile strength, ultimate elongation, and water-tightness (pinhole testing).

## Clinical Evidence

Bench testing only. Testing included physical property evaluation (tensile strength, elongation), water-tightness (1,000 ml leak test), residual powder content (ASTM D6124-01), residual protein levels (ASTM D5712-99), and biocompatibility testing. All results met or exceeded ASTM D3578-01aE2 requirements.

## Technological Characteristics

Latex examination glove; powder-free; green color; optional sour apple or peppermint scent. Dimensions and physical properties comply with ASTM D3578-01aE2. Residual powder <2 mg/g; residual protein <50 µg/g. Non-sterile.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- POWDER FREE PURPLE LATEX EXAM GLOVES WITH AND WITHOUT GRAPE SCENTING ([K011370](/device/K011370.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K032293

#### APPENDIX M

#### 510(k) SUMMARY

#### NON-STERILE POWDER-FREE GREEN LATEX EXAMINATION GLOVES WITH OR WITHOUT SOUR APPLE OR PEPPERMINT SCENTING PLUS A PRQTEIN LABELING CLAIM (<50uG/G) (Contains 50 grams or less of water extractable protein per gram of glove)

Submitted For: SGMP COMPANY LTD.

TUCKER & ASSOCIATES Submitted By: Official Correspondent for SGMP COMPANY LTD. JANNA P. TUCKER, President - CEO 198 Avenue de la D'emerald Sparks, NV 89434-9550 775-342-2612 Phone: 775-342-2613 Fax: E-Mail: Tuckerjan@aol.com

#### Equivalent Predicate Device: POWDER FREE PURPLE LATEX EXAM GLOVES WITH AND WITHOUT GRAPE SCENTING which was approved for marketing as K011370.

This summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.

#### Device Information:

| Trade Name:                                         | NON-STERILE POWDER FREE GREEN LATEX EXAM<br>GLOVES WITH/WITHOUT SOUR APPLE/PEPPERMINT<br>SCENTING AND WITH PROTEIN LABELING (<50uG/G)                                                                    |
|-----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name:                                        | Latex Exam Gloves.                                                                                                                                                                                       |
| Classification Name:                                | Patient Examination Glove, Latex.                                                                                                                                                                        |
| Classification Information<br>& Device Description: | Class I latex patient examination glove, 80LYY, powder free<br>And meeting all the requirements of ASTM D-3578-01aE2,<br>Standard Specification for Latex Examination Gloves for<br>Medical Application. |
| Intended Use of Device:                             | A medical glove to be worn on the hand of the health care and<br>similar personnel to prevent contamination between health<br>care personnel and the patient.                                            |

{1}------------------------------------------------

Technological Characteristics of Device:

| DIMENSION   |        | ASTM D3578-01aE2                | SGMP         |
|-------------|--------|---------------------------------|--------------|
| X-Small     |        | 70 mm +/- 10 mm                 | 70 - 75 mm   |
| Small       |        | 80 mm +/- 10mm                  | 80 - 85 mm   |
| Medium      |        | 95 mm +/- 10mm                  | 90 - 97 mm   |
| Large       |        | 111mm +/- 10mm                  | 105 - 111 mm |
| Length      |        | 230 mm minimum<br>for all sizes | 242mm        |
| Thickness - | Finger | 0.08mm min                      | 0.08 mm min  |
|             | Palm   | 0.08mm min                      | 0.08 mm min  |

#### 1. Dimension

## 2. Physical Properties (ASTM-D3578-01aE2 Standard Specification for Latex Exam Gloves) on Lot# 0320

|                                                     | TENSILE STRENGTH |                                     | ULTIMATE ELONGATION |                               |
|-----------------------------------------------------|------------------|-------------------------------------|---------------------|-------------------------------|
|                                                     | ASTM-D3578-01aE2 | SGMP                                | ASTM-D3578-01aE2    | SGMP                          |
| Before Aging<br>X-Small<br>Small<br>Medium<br>Large | Mpa<br>14.0      | Mpa<br>26.0<br>27.3<br>27.9<br>27.4 | %<br>700            | %<br>990<br>880<br>860<br>870 |
| After Aging<br>X-Small<br>Small<br>Medium<br>Large  | 14.0             | 24.5<br>28.5<br>28.0<br>21.5        | 500                 | %<br>950<br>950<br>950<br>920 |

{2}------------------------------------------------

## 3. Water Tight Test

| BATCH # | SIZE    | SAMPLE SIZE | LEAK STATUS | NUMBER<br>LEAKED |
|---------|---------|-------------|-------------|------------------|
| UN-AGED |         |             |             |                  |
| 0320    | X-Small | 125         | Yes         | 1                |
| 0320    | Small   | 125         | No          | 0                |
| 0320    | Medium  | 125         | No          | 0                |
| 0320    | Large   | 125         | Yes         | 1                |
| AGED    |         |             |             |                  |
| 0320    | X-Small | 125         | No          | 0                |
| 0320    | Small   | 125         | No          | 0                |
| 0320    | Medium  | 125         | Yes         | 1                |
| 0320    | Large   | 125         | Yes         | 1                |

Using the FDA specified 1,000 ml water leak test, 125 pieces of each size of the gloves were tested and our results are as given below:

The above figures are within the ASTM D3578-01aE2 requirements for latex exam gloves of 2.5% AQL.

#### Biocompatibility 4.

The bio-compatibility test results are as per attached in APPENDIX L and show that the gloves passed the tests for examination gloves.

## 5. Total Residual Powder Content & Presence of Cornstarch

| TESTS                                          | FDA INTERNAL<br>REQUIREMENT | SGMP's                                         |
|------------------------------------------------|-----------------------------|------------------------------------------------|
| Residual Powder<br>Content<br>(ASTM D 6124-01) | 2 mg glove max              | Range: 0.5-0.8mg glove<br>Mean : 0.63 mg glove |
| Presence of Cornstarch                         | Negative                    | Negative                                       |

{3}------------------------------------------------

### 6. Residual Protein Level

| TESTS          | FDA ALLOWABLE LEVEL | CLAIMED LEVEL |
|----------------|---------------------|---------------|
| ASTM D 5712-99 | < 50 µg/g           | < 50 µg/g     |

## Conclusion:-

The data presented indicate that the Non-sterile Powder Free Green latex examination glove with Sour Apple/Peppermint Scenting

- Standard Specifications For Latex Examination 1. meets/exceeds ASTM- D3578-01aE2 Glove,
- 2. meets FDA pinhole requirements,
- 3. meets FDA claim criterion of a powder free glove.
- 4. meets the protein labeling claim level at <50 µg/g.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the department's mission to promote health, well-being, and human services. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 2 8 2003

SGMP Company Limited C/O Ms. Janna P. Tucker Official Correspondent Tucker & Associates 198 Avenue De La D' emerald Sparks, Nevada 89434-9550

Re: K032293

Trade/Device Name: Non-Sterile Powder Free Green Latex Exam Latex Examination Gloves with or without Sour Apple or Peppermint Scenting Plus a Protein Labeling Claim (<50uG/G) Regulation Number: 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYY Dated: July 23, 2003 Received: July 25, 2003

Dear Ms. Tucker:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 -Ms. Tucker

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally narketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Susan Runse

Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

# INDICATION FOR USE STATEMENT

Applicant : SGMP Company Limited

K032293 510K NUMBER :

Device Name : Non-sterile Powder Free Green Latex Examination Gloves with on FINS 5D microgram orless of Total Vealer expression Indication For Use :

This is a medical glove to be worn on the hand of health care and similar personnel to prevent contamination between health care personnel and the patient.

Qhin S. Lim

(Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental D

510(k) Number: K032293

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Concurrence of CDRH Office of Device Evaluation (ODE)

| Prescription Use .........................<br>Per 21 CFR 801.109 | OR | Over-The-Counter......................... |
|------------------------------------------------------------------|----|-------------------------------------------|
|------------------------------------------------------------------|----|-------------------------------------------|

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYY/K032293](https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYY/K032293)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
