← Product Code [LYY](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYY) · K032016

# DERMAGRIP-X POWDER FREE LATEX EXAMINATION GLOVES COATED WITH ALPHA LIPOIC ACID AND VITAMIN E, AND WITH PROTEIN CONTENT (K032016)

_Wrp Specialty Products Sdn. Bhd. · LYY · May 11, 2004 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYY/K032016

## Device Facts

- **Applicant:** Wrp Specialty Products Sdn. Bhd.
- **Product Code:** [LYY](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYY.md)
- **Decision Date:** May 11, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

The Powder Free Latex Examination Gloves Coated with Vitamin E, and with Protein Content Labeling Claim are disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

## Device Story

Disposable natural latex examination glove; coated with Vitamin E; powder-free; protein content labeled at 50 micrograms or less per gram. Used in clinical settings by healthcare providers to prevent cross-contamination between patient and examiner. Acts as a physical barrier. No electronic or software components.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by non-clinical bench testing, including physical property testing, pinhole leak testing, and biocompatibility assessments (primary skin irritation and dermal sensitization) per ASTM standards.

## Technological Characteristics

Natural latex material; powder-free (< 2 mg/glove); Vitamin E coating; protein content < 50 µg/g. Meets ASTM D 3578-01aE2 (dimensions, physical properties, pinholes), ASTM D 6124-01 (powder-free), and ASTM D 5712-95 (protein level). Biocompatible (non-irritant, non-sensitizer).

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

MAY 11 2004

K032016

# 510(k) Summary

#### 1.0 Submitter:

| Name:      | WRP Specialty Products Sdn Bhd.                                                                              |
|------------|--------------------------------------------------------------------------------------------------------------|
| Address:   | Lot 11, Jalan 2, Kawasan Perushaan Bandar Baru Salak<br>Tinggi, 43900 Sepang, Selangor Darul Ehsan, MALAYSIA |
| Phone No.: | +60 3 8706 9788                                                                                              |
| Fax No.:   | +60 3 8706 5020                                                                                              |

Date of Summary Prepared:

#### 2.0 Contact Person:

| Name:      | V. Nadarajan    |
|------------|-----------------|
| Phone No.: | +60 3 8706 9788 |
| Fax No.:   | +60 3 8706 5020 |

#### 3.0 Name of the device:

| Proprietary or Trade Name: | 1) DermaGrip-X, and                                |
|----------------------------|----------------------------------------------------|
|                            | 2) Multiple or Customer's Trade Name               |
| Classification Name :      | Patient Examination Gloves (per 21 CFR 880.6250)   |
| Device Name :              | Powder Free Latex Examination Gloves Coated with   |
|                            | Vitamin E, and with Protein Content Labeling Claim |
|                            | (50 Micrograms or less)                            |

#### Identification of The Legally Marketed Device: 4.0

Class I patient examination gloves, 80LYY, powder free, that meets all the
requirements of ASTM standard D 3578 – 01a 22 and FDA 21 CFR 800.20.

#### Description of The Device: 5.0

:

The Powder Free Latex Examination Gloves, Non Sterile with Protein Content Labeling Claim (50 Micrograms or less) meets all the requirements of ASTM standard D 3578 - 01aE2 and FDA 21 CFR 800.20.

{1}------------------------------------------------

#### Intended Use of the Device: 6.0

The Powder Free Latex Examination Gloves Coated with Vitamin E, and with Protein Content Labeling Claim are disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

#### Summary of The Technological Characteristics of The Device: 7.0

The Powder Free Latex Examination Gloves Coated with Vitamin E, and with Protein Content Labeling Claim are summarized with the following technological characteristics compared to ASTM or equivalent standards.

| Characteristics       | Standards                                                  | Device Performance                                                        |
|-----------------------|------------------------------------------------------------|---------------------------------------------------------------------------|
| Dimensions            | ASTM D 3578 - 01aE2                                        | Meets                                                                     |
| Physical Properties   | ASTM D 3578 - 01aE2                                        | Meets                                                                     |
| Freedom from pinholes | ASTM D 3578 - 01aE2<br>FDA 21 CFR 800.20                   | Meets                                                                     |
| Powder-Free           | ASTM D 6124 - 01                                           | < 2 mg/glove                                                              |
| Protein Level         | ASTM D 5712 - 95                                           | < 50 $\mu$ g/g                                                            |
| Biocompatability      | Primary Skin Irritation in Rabbits<br>Dermal Sensitization | Passes (Not a primary skin irritant)<br>Passes (Not a contact sensitizer) |

#### Substantial Equivalent Based on Assessment of Non-Clinical Performance Data 8.0

The performance test data of the non-clinical tests that support a determination of substantial equivalence is the same as mentioned immediately above.

#### Substantial Equivalent Based on Assessment of Clinical Performance Data 9.0

Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.

#### 10.0 Conclusion

It can be concluded that the Powder Free Latex Examination Gloves Coated with Vitamin E, and with Protein Content Labeling Claim will perform according to the glove performance standards referenced in Section_7 above and meet ASTM standards, and FDA requirements for water leak test on pinhole AQL. Consequently, this device is substantially equivalent to currently marketed devices.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. Inside the circle is a symbol that resembles a stylized caduceus, with three lines representing the snake and a single line representing the staff.

:

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# MAY 11 2004

Mr. Nadarajan Venugopal Manager, Quality Assurance Regulatory Affairs WRP Specialty Products Sdn Bhd Lot 11, Jalan 2, Kawasan Perusahaan Bandar Baru Salak Tinggi, 43900 Sepang, Selangor Darul Ehsan, MALAYSIA

### Re: K032016

Trade/Device Name: Dermagrip- X Powder Free Latex Examination Gloves Coated with Vitamin E, and with Protein Content Labeling Claim (Contains 50 Micrograms or Less of Total Water Extractable Protein Per GRAM) Regulation Number: 880.6250 Regulation Name: Patient Examination Glove Regulatory Class; I Product Code: LY Y Dated: April 6, 2004 Received: April 12, 2004

## Dear Mr. Venugopal:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting (1 NE ), it has obe found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In 

{3}------------------------------------------------

Page 2 - Mr. Venugopal

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements incall that I Dr mai made statutes and regulations administered by other Federal agencies. or the rice of any I edenal the Act's requirements, including, but not limited to: registration 1 od intist comply with and 807); labeling (21 CFR Part 801); good manufacturing practice and ifsime (21 CF CF results in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) Premarket notification. The FDA finding of substantial equivalence of your device to a premairer notificate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), If you desire specific at Compliance at (301) 594-4618. Also, please note the regulation contitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Susan Runno
Chiu Lin, Ph.D.

Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# INDICATIONS FOR USE

:

| Applicant:                | WRP Specialty Products Sdn Bhd                                                                                                                                                                  |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number (if known): | K032016                                                                                                                                                                                         |
| Device Name:              | POWDER FREE LATEX EXAMINATION<br>GLOVES COATED WITH VITAMIN E, AND<br>WITH PROTEIN CONTENT LABELING CLAIM<br>(CONTAINS 50 MICROGRAMS OR LESS OF<br>TOTAL WATER EXTRACTABLE PROTEIN<br>PER GRAM) |

Indications For Use:

The patient examination glove is a disposable device and is made of natural The patient exammation grove is a 'displace' worn on the examiner's hand of furoer to prevent contamination between patient and examiner.

Concurrence of CDRH, Office of Device Evaluation (ODF)

| Prescription Use<br>(Per 21 CFR 801.109) | OR Over-The-Counter <div style="display:inline-block; width:20px;">✓</div>         |
|------------------------------------------|------------------------------------------------------------------------------------|
|                                          | <i>Kei Mulry</i>                                                                   |
|                                          | (Division Sign-Off)                                                                |
|                                          | Division of Anesthesiology, General Hospital,<br>Infection Control, Dental Devices |
| 510(k) Number:                           | K032016                                                                            |

Page I of 1

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYY/K032016](https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LYY/K032016)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
