RELIANCE FAMILY OF WASHER-DISINFECTORS

K983068 · STERIS Corporation · LDS · May 31, 2000 · General Hospital

Device Facts

Record IDK983068
Device NameRELIANCE FAMILY OF WASHER-DISINFECTORS
ApplicantSTERIS Corporation
Product CodeLDS · General Hospital
Decision DateMay 31, 2000
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6991
Device ClassClass 2

Indications for Use

The Amsco Reliance Family of Washer/Disinfectors are indicated for use in the cleaning and low-level disinfection of soiled reusable utensils, trays, glassware, bedpans and urinals, rubber and plastic goods, simple hard-surfaced rigid surgical instruments such as forceps and clamps, and other similar and related articles found in healthcare facilities.

Device Story

Amsco Reliance Washer/Disinfectors are computer-controlled mechanical systems for decontaminating reusable medical items. Items are placed in a stainless steel chamber; pressurized fluids are delivered via a pump-driven circulation system. The device uses pre-programmed cycles—including pre-wash, enzymatic soak, detergent wash, sonic cleaning, hot water rinse, thermal disinfection, and optional lubrication/drying—to remove soil and disinfect loads. Configurations include steam or electrical heating. Accessories include loading racks and transfer carts; some models feature barcode-reading conveyor systems for automated cycle initiation. Operated by healthcare facility personnel, the device provides automated cleaning and thermal disinfection, reducing manual handling of contaminated items and ensuring items are safe for further processing or sterilization.

Clinical Evidence

Bench testing only. Studies included simulated-use type-tests assessing cleaning efficacy, residue removal, and thermal disinfection. Cleaning efficacy verified via side-by-side testing with predicate using blood-soil challenge. Residue testing measured ionic/non-ionic detergent and enzymatic levels against acceptance criteria. Thermal disinfection validated via challenge testing with microorganisms, achieving >8 log reduction. Equipment decontamination validated via 3-week biofilm growth study. Lubrication testing confirmed no interference with subsequent sterilization.

Technological Characteristics

Stainless steel construction; pump-driven fluid circulation with rotating spray arms; steam or electrical heating; microprocessor-based cycle control. Connectivity includes barcode reading for cycle automation. Sterilization method: thermal disinfection (hot water).

Indications for Use

Indicated for cleaning and low-level disinfection of soiled reusable medical items including utensils, trays, glassware, bedpans, urinals, rubber/plastic goods, and simple hard-surfaced rigid surgical instruments (e.g., forceps, clamps) in healthcare facilities.

Regulatory Classification

Identification

A medical washer is a device that is intended for general medical purposes to clean and dry surgical instruments, anesthesia equipment, hollowware, and other medical devices.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Medical Washers and Medical Washer-Disinfectors.” The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 880.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## MAY 3 1 2000 #### XG83068 510(k) SUMMARY This pre-market notification was submitted by the following individual: Raymond Ursick. Vice-President, Regulatory Affairs and Quality Systems STERIS Corporation 5960 Heisley Road Mentor, Ohio 44060 Tel: 440-354-2600 This 510(k) summary was prepared on May 12, 2000. | Device identification | | |-----------------------|------------------------------------------------| | Trade name | Amsco® Reliance® Family of Washer/Disinfectors | | Common name | Washer/Disinfector | | Classification | II (As recommended by classification panel) | | Predicate devices | | |-------------------|-------------------------------| | Name | Model 7550 Washer/Disinfector | | | T-21 Decontaminating Machine | | Manufacturer | MDT/Castle | | | HAMO, USA, Inc. | | Classification | NOT CLASSIFIED | | | NOT CLASSIFIED | | 510(k) number | K911087 | | | K911120 | #### DEVICE DESCRIPTION The Amsco Reliance Family of Washer/Disinfectors are mechanical, computercontrolled Washer/Disinfectors. They are designed with pre-programmed cycles intended specifically for surgical instruments, delicate instrumentation (gentle cycle), utensils, glassware, plastic goods, anesthesia/respiratory goods and equipment decontamination. Additional cycles are available for customized programming to meet specific operating requirements. Essentially, the Amsco Reliance Family of Washer/Disinfectors operate as follows: the load items are placed in the chamber where they are exposed to pressurized fluids delivered through a circulation system driven by a pump and controlled by a programmable computer. Injection pumps allow the automatic delivery of the selected {1}------------------------------------------------ chemicals in the sump water. Available configurations include steam or electrical heating of the sump water; and the drying option (electrically heated on some models. steam heated on others). The cabinet, chamber, and circulation system (including the pump) are primarily constructed of stainless steel. The available cycle phases can be summarized as follows: - 1. cold tap water pre-wash for the removal of gross soil; - soaking in an enzymatic solution to break down proteinaceous materials; 2. - 3. detergent wash to complete the soil removal: - sonic cleaning to remove fine debris in crevices: র্ব - ട്. hot tap water rinse to remove detergent and soil residues: - thermal rinse to disinfect the load items; റ. - 7. pure water rinse and instrument lubrication: - 8. drvina. Note: Phases 2 and 4 are not offered on all models. Phases 2, 4, 7 and 8 are optional features. The available accessories include a variety of single and multi-level loading racks, and transfer carts. Some models offer conveyor systems used to automate the loading of racks into the chamber, to initiate the proper cycle by reading a bar code tag located on the loading rack, and to provide queuing capacity. ### INTENDED USE The Amsco Reliance Family of Washer/Disinfectors are indicated for use in the cleaning and low-level disinfection of soiled reusable utensils, trays, glassware, bedpans and urinals, rubber and plastic goods, simple hard-surfaced rigid surgical instruments such as forceps and clamps, and other similar and related articles found in healthcare facilities. ### SUBSTANTIAL EQUIVALENCE - TECHNOLOGICAL COMPARISON In most respects, including intended use, operating principle, energy sources, construction materials and control technology, the Amsco Reliance Family of Washer/Disinfectors are quite similar to both predicate devices. All use impingement and chemical means to remove soil, contaminants and residues from used medical devices and provide hot water thermal disinfection, in order to render them safe for handling and further processing by healthcare facility personnel. The devices are constructed primarily of stainless steel with rotating spray arms in a closed chamber. Hot water and detergents are circulated in varying cycles, in order to wash and disinfect a variety of medical devices approved for reuse. Microprocessors are used to provide cycle control. {2}------------------------------------------------ There is a notable difference in the cycle parameters structure, which is attibutable to There is a notable difference in the cycle paramoced screenshish the fact of accomplish the the fact that the MDT/Castle predicate device uses the wash phase to accomplish th the fact that the MDT/Castle predicate Election Framily of Washer/ thermal disinfection whereas the HAMO and Alliance Family of Washer/ thermal disimection whereas the fixhe and Antionally, the MDT/Castle device has a Disinfectors uses the themial hise phase. A longer thermal disinfection phase time. high-level disinfection claim which requires a formal disinfection also in the restricted high-level disintection claim which requires a leinfection claim is restricted to The Amsco Reliance Family of Wasileiner that this level is suitable for contact low-level disinfection, based on the understanding that the lover is one. with intact skin, and therefore compatible with the safe handling of processed items by healthcare facility staff. # PERFORMANCE TESTS - TEST SCHEME DESCRIPTION In order to demonstrate the actual performance levels achieved by the Amsco Reliance In order to demonstrate the actual penomismes to simulated-use type-tests have been Family of Washer/Disimectors, a sches of of the cleaning, rinsing of residues and conducted. These included assessment of the clouns of the areas of the (thermal) disintection encles, in ropresentially be colonized by heterotrophic machine and accessones that occated potentially challenge test to prove that the microorganisms have been submitted to a whiles are effective to prevent any recommended equipment decomamination processor inmersion the twas conducted to bacterial growin inside the unit. I Thany, a instrument lubricant on hinged surgical demonstrate that the practice of applying interfere with subsequent sterilization. ### TEST RESULTS - CLEANING Side-by-side testing was performed with the HAMO device to verify the Amsco Reliance Side-by-side testing was penomiculum to clean a representative load. The test method Family of viasneriDisiniector's ability to clour a representative soil, and performing a consisted of covering test nems with blood of the presence of visible residual visual inspection of hems and probecaing to colling of Washer/Disinfectors provided soll. It was shown that the Amoso Rollance Parison and when used as directed. ### TEST RESULTS - RESIDUES Testing was performed to establish the efficacy of the rinse phases to remove detergent Testing was performed to establish the emsels known to be non-toxic to patients and and enzymatic residues on load terns to technologictivity increase of test solution users. The test method consisted of measuring the conductivity increase of test solution users. The test method consisted on measuring the son test tems to determine the caused by the elution of lonic detergent roomatic residue levels were measured by quantitles of residues. The non-lonic enzymatic roclase 10 relieve the following table shows {3}------------------------------------------------ # K98.3068 | Chemical | Items | Acceptance Criteria (mg/item) | Results (mg/item)1 | |--------------------------------------|---------------------------|-------------------------------|--------------------| | STERIS Liqui Jet | Surgical Instruments | 0.79 | 0.023 | | | Micro Surgery Instruments | 0.79 | 0.069 | | | Suction Tips | 6.34 | 0.08 | | | Respiratory Equipment | 12.66 | 4.1 | | Enzy Care II<br>(detergent fraction) | Surgical Instruments | 4.8 | 0.11 | | | Micro Surgery Instruments | 4.8 | 0.4 | | | Suction Tips | 9.5 | 0.31 | | Enzy Care II<br>(enzyme fraction) | Surgical Instruments | 14.5 | <0.8 | | | Micro Surgery Instruments | 14.5 | <3.9 | | | Suction Tips | 29 | <3.7 | | Descaler | Surgical Instruments | 2.9 | 0.016 | | | Micro Surgery Instruments | 2.9 | 0.016 | | | Suction Tips | 5.8 | 0.016 | | | Respiratory Equipment | 46.4 | 0.016 | | NpH-Klenz® | Rigid MIS Instruments | 2.8 | 0.262 | | Klenzyme®<br>(detergent Fraction) | Rigid MIS Instruments | 7.7 | 0.632 | | Klenzyme®<br>(enzyme fraction) | Rigid MIS Instruments | 15.5 | <0.272 | #### the acceptance criteria and results for each item family and chemical tested. Note 1: Test results as obtained with the Amsco Reliance 444 Single-Chamber Washer/Disinfector, and are considered as representative of the results obtained for the Family of Washer/Disinfectors. Note 2: The MIS rack only applies to the Amsco Reliance 444 Single-Chamber Washer/Disinfector. ## TEST RESULTS - THERMAL LOW-LEVEL DISINFECTION Challenge testing was performed to verify the ability of the Amsco Reliance Family of Washer/Disinfectors to achieve low-level disinfection of commonly processed items. The test method consisted of placing vials containing known concentrations of microorganisms in the coldest areas of the chamber and measuring logarithmic reductions after processing through the thermal rinse treatment. A sample test conducted with the MIS rack on the Amsco Reliance 444 Single-Chamber Washer/Disinfector, using microorganisms recognized as suitable to test lowlevel disinfection, demonstrated that greater than 8 logarithmic reductions of the test organisms was achieved. #### TEST RESULTS - EQUIPMENT DECONTAMINATION Testing was performed to validate that the equipment decontamination procedures in the device labeling are effective to prevent the formation of biofilm or growth of heterotrophic microorganisms in critical areas of the device. The test method consisted of operating the Washer/Disinfector with various loading racks for 3 consecutive weeks and, on a weekly basis, swabbing the surfaces of critical areas of the machine and loading racks, and culturing to detect the presence of biofilm {4}------------------------------------------------ KG83068 formation. The test reports conclude that in its current design, the Washer/Disinfector and its accessories, when submitted to the recommended decontamination procedure, are not likely to develop biofilm on identified critical areas. ### TEST RESULTS - INSTRUMENT LUBRICATION Testing was performed to verify that the process of lubricating an item prior to terminal processing did not affect the item's ability to be sterilized. The test method consisted of inoculating hinged instruments with bacterial spores, dipping them into full strength instrument lubricant before sterilization and testing them for sterility. The test report shows that no growth could be detected on any of the lubricated as well as unlubricated instruments, demonstrating that the practice of lubricating instruments in the Washer/Disinfector at any concentration will not impede the ability to sterilize them subsequently. #### CONCLUSIONS Whereas, the test conditions were designed to be representative conditions within the intended use of the Washer/Disinfector; Whereas, these tests have been conducted in keeping with applicable quality systems requlations (Good Laboratory Practice, 21CFR §58), whether conducted in STERIS facilities or subcontracted; Whereas, per STERIS Corporation's intent, the above mentioned test conditions were designed to be adequate, complete and challenging to all aspects relating to the performance, safety and effectiveness of the device for its intended use; STERIS Corporation considers the Amsco Reliance Family of Washer/Disinfectors to be safe, effective, and is substantially equivalent to the predicate devices. {5}------------------------------------------------ Image /page/5/Picture/2 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an emblem that appears to be an abstract representation of an eagle or bird-like figure. The emblem is composed of three curved lines that form the shape of the bird's head and body. ### MAY 3 1 2000 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Raymond Ursick Vice-President, Requlatory Affairs and Quality Systems STERIS Corporation 5960 Heisley Road Mentor, Ohio 44060 Re : K983068 Trade Name: Amsco® Reliance Family of Washer-Disinfectors Regulatory Class: Unclassified Product Code: LDS Dated: March 20, 2000 Received: March 21, 2000 Dear Mr. Ursick: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual reqistration, listing of devices, qood manufacturing practice, labelinq, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਬੇ substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {6}------------------------------------------------ Page 2 - Mr. Ursick This letter will allow you to begin marketing your device as Inis lecter will arrow your market notification. The FDA described in four tial equivalence of your device to a legally Finding of subscancial equires. In a classification for your marketed predicate actro your device to proceed to the market. If you desire specific advice for your device on our labeling II you debire bpoorle and additionally 809.10 for in regulacion (ir devices), please contact the Office of Villy diagnobere at (301) 594-4692. Additionally, for questions on Compilance at (301) 534 4022. Indias device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general premation on your responsibilities under the Act may be Information on your responsible suall Manufacturers Assistance obtained from cho bring (800) 638-2041 or (301) 443-6597 or at at feb corr rreo hamber to://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Susan Burn Timothy A. Ulatowski Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ ### K983068 STERIS Amsco Reliance Family of Washer / Disinfectors ### INDICATIONS FOR USE STATEMENT ### DEVICE NAME: STERIS Amsco Reliance Family of Washer / Disinfectors #### INDICATIONS FOR USE: The STERIS Amsco Reliance Family of Washer / Disinfectors are indicated for use in the cleaning and low-level disinfection of soiled reusable utensils, trays, glassware, bedpans and urinals, rubber and plastic goods, simple hard-surfaced rigid instruments such as forceps and clamp, and other similar and related articles found in healthcare facilities. (Please Do Not Write Below This Line - Continue on Another Page If Needed) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional Format 1-2-96) Qun S. Lin (Division Sign-Off) Division of Dental, Infection Control and General Hospital Devices 510(k) Number
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...