← Product Code [KZH](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/KZH) · K021700

# ANIMUS EZSERTER INFUSION SET INSERTER (K021700)

_Animas Corp. · KZH · Jun 4, 2002 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/KZH/K021700

## Device Facts

- **Applicant:** Animas Corp.
- **Product Code:** [KZH](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/KZH.md)
- **Decision Date:** Jun 4, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6920
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

The Animas ezSerter™ Infusion Set Inserter is intended to aid in the insertion of Animas ezSet™ infusion sets. This device is intended for home use and is a prescription device.

## Device Story

Animas ezSerter is a spring-loaded mechanical inserter for subcutaneous placement of Animas ezSet angled infusion sets. Device consists of molded plastic housing, internal spring-loaded carrier, and stainless steel springs. User loads sterile infusion set into carrier; cocks spring mechanism; positions device at 30-degree angle to skin fold; presses trigger button to release carrier, driving needle and cannula into subcutaneous tissue. User then removes introducer needle, secures adhesive pad, and attaches tubing. Used in home setting by patients or caregivers. Simplifies insertion process; ensures consistent needle angle and depth; reduces manual handling of needles.

## Clinical Evidence

No clinical data; bench testing only.

## Technological Characteristics

Molded plastic housing and carrier; stainless steel springs. Mechanical spring-loaded actuation. Manual operation. Non-sterile device (used with sterile infusion sets).

## Regulatory Identification

A syringe needle introducer is a device that uses a spring-loaded mechanism to drive a hypodermic needle into a patient to a predetermined depth below the skin surface.

## Predicate Devices

- MiniMed Sil-SerterTM Infusion Set Insertion System

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K021700

# Section 1 - Information Required by the Safe Medical Devices Act of 1990

| 1.1 510(k) Summary   |                                                                                                                                                         |  |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Date:                | March 13, 2002                                                                                                                                          |  |
| Submitter:           | Animas Corporation, 590 E. Lancaster Avenue, Frazer, PA 19355                                                                                           |  |
| Contact:             | Richard R. Michelin, Vice President of Quality Assurance,<br>Telephone: (610)-644-8990, ext 185, Fax: (610)-644-8717,<br>Email: richardm@animascorp.com |  |
| Name of Device:      | Animas ezSerter         TM Infusion Set Inserter                                                                                                        |  |
| Common Name:         | Inserter                                                                                                                                                |  |
| Classification Name: | Introducer, Syringe Needle                                                                                                                              |  |
| Predicate Device:    | MiniMed Sil-SerterTM Infusion Set Insertion System                                                                                                      |  |

Description of the New Device: The Animas ezSerter™ Infusion Set Inserter is a simple device for automatically inserting the cannula of the Animas ezSef™ angled infusion sets subcutaneously. The inserter is composed of several molded plastic pieces and stainless steel springs. An elongated molded plastic housing acts as a handle and a quide for an internally mounted spring-loaded carrier. An opening in the proximal end of the carrier accepts and retains the handle of the introducer needle hub.

The inserter may be held in either hand. To operate the inserter, the user places a sterile infusor base into the carrier. The spring-loaded retainer locks the base in place, automatically aligning and retaining the insertion needle axis. The user then folds the front flap of the adhesive pad up under the retainer trigger to expose the introducer needle and cannula. The user draws back the carrier until it locks in the cocked position. At this stage, the sliding safety may be switched to the "locked" position to prevent the accidental release of the carrier while the inserter is positioned or the site is cleaned. The protective sleeve over the cannula is then removed.

A small fold is formed at the infusion site by pinching the skin between the thumb and finger of the free hand. After releasing the safety, the inserter is held at approximately a 30 degree angle to the surface of the skin. The proximal edge of the inserter is positioned at the edge of the skin fold with the needle is pointing towards the middle of the fold. When the release button is pressed, the carrier

{1}------------------------------------------------

springs forward, quickly and smoothly inserting the needle and cannula subcutaneously. The skin fold is released and the backing paper is removed from the front flap of the adhesive pad. The adhesive pad is then smoothed over the skin, covering the point of entry with the transparent window of the pad. The sides of the introducer needle hub are pinched together to release the introducer needle from the cannula housing; the inserter and attached needle are withdrawn. The inserter is set aside while the rear backing paper is removed and the pad is smoothed to the skin. The tubing set may then be attached to the cannula housing to allow the insulin to flow into the infusion site. The retainer button is pressed to release the introducer needle from the inserter for disposal.

Intended Use of the New Device: The Animas ezSerter™ Infusion Set Inserter is intended to aid in the insertion of Animas ezSet™ infusion sets.

This device is intended for home use and is a prescription device.

Comparision of the Technological Features of the New Device and the Predicate Device: The new device and the predicate devices are substantially similar in terms of technology. Both are spring loaded devices. The user must load the infusion set into the device, compress a spring and activate the device by pressing a trigger button.

The differences between the new device and the predicate devices do not affect the safety or effectiveness of the device.

{2}------------------------------------------------

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized graphic of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a sense of depth. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular fashion around the graphic.

### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN 0 4 2002

Animas Company Mr. Robert Mosenkis Responsible Third Party Official CITECH 5200 Butler Pike Plymouth Meetings, Pennsylvania 19462-1298

Re: K021700

Trade/Device Name: Animus ezSerter™ Infusion Set Inserter Regulation Number: 880.6920 Regulation Name: Syringe Needle Introducer Regulatory Class: II Product Code: KZH Dated: May 22, 2002 Received: May 23, 2002

Dear Mr. Mosenkis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements

{3}------------------------------------------------

### Page 2 - Mr. Mosenkis

of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours.

Ous L. for
Timothy A. Ulatowski

Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

#### 1.3 Indications for Use Statement

510(k) Number:

Animas ezSerter™ Infusion Set Inserter Device Name:

Indications for Use: The Animas ezSerter™ Infusion Set Inserter is intended to aid in the insertion of the Animas ezSet™ infusion sets.

This device is intended for home use and is a prescription device.

Concurrence of CDRH, Office of Device Evaluation (ODE)

Patricia Cuenca

(Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices 021700 510(k) Number ________________________________________________________________________________________________________________________________________________________________

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/KZH/K021700](https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/KZH/K021700)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
