K243317 · Infinitum Eta, Ltd. · KCT · Feb 11, 2025 · General Hospital
Device Facts
Record ID
K243317
Device Name
NUVENTUS NV.C™ Surgical Cassette and Tray
Applicant
Infinitum Eta, Ltd.
Product Code
KCT · General Hospital
Decision Date
Feb 11, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The NUVENTUS NV.C™ Surgical Cassette and Tray is intended to enclose other medical devices that are to be sterilized by a health care provider. The NUVENTUS NV.C™ Surgical Cassette and Tray is intended to allow sterilization of the enclosed medical devices. The NUVENTUS NV.C™ Surgical Cassette and Tray requires the use of an FDA cleared wrap to maintain the sterility of the enclosed devices. The NUVENTUS NV.C™ Surgical Cassette and Tray should be enclosed in a sterilization wrap that is FDA cleared for the indicated cycle, and moist heat (steam) sterilized using the following cycle: Pre-vacuum steam: 132°C (270°F) for 4 minutes with 30 minutes drying time. The NUVENTUS NV.C™ Surgical Cassette and Tray is intended for sterilization of non-porous loads. The NUVENTUS NV.C™ Surgical Cassette and Tray is recommended not to be stacked during sterilization. The combined weight of the NUVENTUS NV.C™ Surgical Cassette and Tray and the associated instruments is 220 grams. The weight of the empty NUVENTUS NV.C™ Surgical Cassette and Tray is 200 grams.
Device Story
Reusable rigid sterilization container; comprises base, removable inner tray, and lid. Manufactured from polyphenylsulfone (PPSU) with medical-grade silicone instrument holders. Perforated design facilitates steam penetration. Used by healthcare providers to organize, protect, and sterilize non-porous surgical instruments. Requires external FDA-cleared sterilization wrap to maintain sterility post-process. Subject to 225 reprocessing cycles (cleaning, disinfection, sterilization). Benefits include instrument protection and organized sterilization workflow.
Clinical Evidence
No clinical data. Bench testing only: validation of manual and automated cleaning (ISO 17664, AAMI ST98), sterilization validation (ISO 17665, AAMI TIR 12), use-life testing (225 cycles), and biocompatibility (ISO 10993-5).
Technological Characteristics
Rigid container; materials: polyphenylsulfone (PPSU) base/lid/tray, medical-grade silicone holders. Dimensions: 138L x 98W x 62H mm. Perforated for moist heat (steam) sterilization. Reusable (up to 225 cycles).
Indications for Use
Indicated for use by healthcare providers to enclose and sterilize non-porous medical devices via pre-vacuum steam sterilization (132°C for 4 minutes, 30 minutes drying). Requires FDA-cleared sterilization wrap. Not for stacked sterilization.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
February 11, 2025
Infinitum Eta Ltd. Avital Barlev Vice President QA&RA Hatnufa St. 6 Petach Tikva, 4951024 Israel
Re: K243317
Trade/Device Name: NUVENTUS NV.CTM Surgical Cassette and Tray Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: October 22, 2024 Received: October 22, 2024
Dear Avital Barlev:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{2}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{3}------------------------------------------------
Sincerely,
Image /page/2/Picture/3 description: The image contains the text "Stephen A. Anisko -S". The text is arranged in a vertical format, with "Stephen A." on the first line, "Anisko" on the second line, and "-S" on the third line. The text is in a simple, sans-serif font and is black in color. The background of the image is white.
Digitally signed by Stephen A. Anisko -S Date: 2025.02.11 11:03:44 -05'00'
for: Christopher K. Dugard Division Director DHT4C: Division of Infection Control Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known) K243317
#### Device Name
NUVENTUS NV.CTM Surgical Cassette and Tray
Indications for Use (Describe)
The NUVENTUS NV.C™ Surgical Cassette and Tray is intended to enclose other medical devices that are to be sterilized by a health care provider. The NUVENTUS NV.C™ Surgical Cassette and Tray is intended to allow sterilization of the enclosed medical devices.
The NUVENTUS NV.C™ Surgical Cassette and Tray requires the use of an FDA cleared wrap to maintain the sterility of the enclosed devices. The NUVENTUS NV.C™ Surgical Cassette and Tray should be enclosed in a sterlization wrap that is FDA cleared for the indicated cvcle, and moist heat (steam) sterilized using the following cvcle: Pre-vacum steam: 132°C (270°F) for 4 minutes with 30 minutes drying time.
The NUVENTUS NV.CTM Surgical Cassette and Tray is intended for sterilization of non-porous loads.
The NUVENTUS NV.C™ Surgical Cassette and Tray is recommended not to be stacked during sterilization. The combined weight of the NUVENTUS NV.C™ Surgical Cassette and Tray and the associated instruments is 220 grams. The weight of the empty NUVENTUS NV.C™ Surgical Cassette and Tray is 200 grams.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
## INFIN
#### 510(K) Summary (K243317) for Infineta's NUVENTUS NV.CTM Surgical Cassette and Tray
| Date Prepared: | 11 February 2025 |
|-------------------------|-----------------------------------------------------------|
| 1. Submitter | |
| Company Name: | Infinitum Eta Ltd |
| Address: | Hatnufa St. 6, Petach Tikva, 4951024, Israel |
| Phone and E-mail: | avitalb@infinetamed.com; +972-54-2272773 |
| Contact Person: | Avital Barlev, Vice President QA&RA |
| 2. Device: | |
| Proprietary/Trade Name: | NUVENTUS NV.CTM Surgical Cassette and Tray |
| Common Name: | Instrument Sterilization Trays |
| Regulation Number: | 21 CFR 880.6850 |
| Regulation Name: | Containers, Trays, Surgical Cassettes & Other Accessories |
| Regulatory Class: | Class II |
| Product Code: | KCT |
3. Predicate Device: Z-Systems AG's ST-Z5, cleared under K201878.
#### 4. Device Description
The subject device is a reusable rigid container, comprising a base (bottom), a removable inner tray, and a lid (cover). It is designed to organize and protect NUVENTUS NV.CTM instruments that fall within the cleaning and sterilization validation scope and are sterilized in the tray by a healthcare provider. The base, inner tray, and lid components are designed to be integrated into a single unit which contains and protects the interior contents during sterilization. The tray is perforated to allow for penetration of the sterilant, are to be used with moist heat (steam), and require the use of an FDA cleared wrap to maintain sterility. The subject device components are manufactured from injection molded polyphenylsulfone (PPSU), and holders of various geometries to position instruments in the trays are manufactured from silicone.
#### 5. Indications for Use
The NUVENTUS NV.C™ Surgical Cassette and Tray is intended to enclose other medical devices that are to be sterilized by a health care provider. The NUVENTUS NV.C™ Surgical Cassette and Tray is intended to allow sterilization of the enclosed medical devices.
The NUVENTUS NV.C™ Surgical Cassette and Tray requires the use of an FDA cleared wrap to maintain the sterility of the enclosed devices. The NUVENTUS NV.C™ Surgical Cassette and Tray should be enclosed in a sterilization wrap that is FDA cleared for the indicated cycle, and moist heat (steam) sterilized using the following cycle: Pre-vacuum steam: 132°C (270°F) for 4 minutes with 30 minutes drying time.
The NUVENTUS NV.C™ Surgical Cassette and Tray is intended for sterilization of non-porous loads.
{6}------------------------------------------------
## INFINETA
The NUVENTUS NV.C™ Surgical Cassette and Tray is recommended not to be stacked during sterilization. The combined weight of the NUVENTUS NV.C™ Surgical Cassette and Tray and the associated instruments is 220 grams. The weight of the empty NUVENTUS NV.C™ Surgical Cassette and Tray is 200 grams.
#### Comparison of Technological Characteristics 6.
A tabular comparison of technological characteristics is presented herein:
## Table 1: Indications for Use Comparison
| Device Name | Indications for Use | |
|--------------------------------|------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject<br>Device | NUVENTUS<br>NV.CTM Surgical<br>Cassette and<br>Tray<br>(K243317) | The NUVENTUS NV.CTM Surgical Cassette and Tray is intended to enclose other medical<br>devices that are to be sterilized by a health care provider. The NUVENTUS NV.CTM Surgical<br>Cassette and Tray is intended to allow sterilization of the enclosed medical devices.<br>The NUVENTUS NV.CTM Surgical Cassette and Tray requires the use of an FDA cleared wrap<br>to maintain the sterility of the enclosed devices. The NUVENTUS NV.CTM Surgical Cassette<br>and Tray should be enclosed in a sterilization wrap that is FDA cleared for the indicated cycle,<br>and moist heat (steam) sterilized using the following cycle: Pre-vacuum steam: 132°C (270°F)<br>for 4 minutes with 30 minutes drying time.<br>The NUVENTUS NV.CTM Surgical Cassette and Tray is intended for sterilization of non-<br>porous loads.<br>The NUVENTUS NV.CTM Surgical Cassette and Tray is recommended not to be stacked<br>during sterilization. The combined weight of the NUVENTUS NV.CTM Surgical Cassette and<br>Tray and the associated instruments is 220 grams. The weight of the empty NUVENTUS<br>NV.CTM Surgical Cassette and Tray is 200 grams. |
| Primary<br>Predicate<br>Device | ST-Z5<br>(K201878) | ST-Z5 surgical trays are intended to be used to enclose other medical devices that are to be<br>sterilized by a health care provider. ST-Z5 surgical trays are intended to allow sterilization of<br>the enclosed medical devices. ST-Z5 surgical trays require the use of an FDA cleared wrap to<br>maintain the sterility of the enclosed devices. The trays are to be enclosed in a sterilization<br>wrap that is FDA cleared for the indicated cycle, and moist heat (steam) sterilized using the<br>following cycle: Pre-vacuum steam: 132 °C (270 °F) for 4 minutes with 20 minutes drying<br>time. ST-Z5 surgical trays are intended for sterilization of non-porous loads. ST-Z5 surgical<br>trays are recommended not to be stacked during sterilization. The combined weight of the ST-<br>Z5_c_m_c(t)_m(t) tray and the associated instruments is 441 grams. The weight of the empty<br>ST-Z5_c_m_c(t)_m(t) tray is 361 grams. The combined weight of the Z5-BL/TL tray and the<br>associated instruments is 427 grams. The weight of the empty Z5-BL/TL tray is 363 grams. |
## Table 2: Technological Characteristics Comparison
| Feature | Subject Device | Primary Predicate Device | Technological<br>Characteristics<br>Comparison |
|--------------|---------------------------------------------------------|------------------------------------------------------|------------------------------------------------|
| | NUVENTUS NV.CTM Surgical<br>Cassette and Tray (K243317) | ST-Z5 (K201878) | |
| Product Code | KCT | KCT | Identical |
| Design | Rigid polymer base, lid, and removable<br>inner tray | Rigid polymer base, lid, and removable<br>inner tray | Identical |
| Materials | Polyphenylsulfone [lid, base, tray] | Polyphenylsulfone [lid, base, tray] | Identical |
{7}------------------------------------------------
# INFINETA
| | Subject Device | Primary Predicate Device | Technological<br>Characteristics<br>Comparison |
|------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|
| Feature | NUVENTUS NV.CTM Surgical<br>Cassette and Tray (K243317) | ST-Z5 (K201878) | |
| | Medical grade silicone<br>[grommets/holders] | Medical grade silicone<br>[grommets/holders] | |
| Materials<br>compatible<br>with<br>Sterilization<br>Method | Yes | Yes | Identical |
| Perforated | Yes; allows moist heat (steam)<br>penetration to achieve sterilization | Yes; allows moist heat (steam)<br>penetration to achieve sterilization | Identical |
| Number of<br>Overall Sizes | 1 | 1 | Identical |
| Number of<br>Configurations | 1 | 2 | Similar |
| Overall<br>Dimensions | 138L x 98W x 62H [mm] | 185.1L x 133.6W x 61.5H [mm] | Similar |
| Vent to<br>Volume Ratio | 0.0172 cm2 / cm3 | 0.01596 cm2 / cm3 | Similar |
| Reusable | Yes | Yes | Identical |
| Number of<br>Uses | 225 | 101 | Similar |
| Use Life<br>Testing | Reusable up to 225 cycles<br>Assembled, cleaned, disinfected and<br>sterilized<br>Visual inspection<br>Defect and deformation<br>Functional performance (Cassette and<br>Tray) | Reusable up to 101 cycles<br>Assembled and sterilized<br>Visual inspection<br>Component dimensional fit verification<br>Functional closure (lid-base latch)<br>verification | Similar |
| Sterilization<br>Method | Moist Heat | Moist Heat | Identical |
| Sterilization<br>Cycles | Fractionated vacuum (pre-vacuum)<br>Exposure at 132 °C (270 °F) for 4<br>minutes with 30 minutes drying time | Fractionated vacuum (pre-vacuum)<br>Exposure at 132 °C (270 °F) for 4<br>minutes with 20 minutes drying time | Similar |
| Sterile<br>Barrier | Sterilization wrap, FDA cleared for<br>indicated method and cycle | Sterilization wrap, FDA cleared for<br>indicated method and cycle | Identical |
#### 7. Technological Comparison Discussion:
The subject device is similar in indications and design principles to the predicate device.
The subject device and the predicate device have the same intended use, the same product classification and product code (KCT), and have similar Indications for Use statements.
Except for the details concerning the weights of the empty and loaded trays, the indications for use statement for the subject device is identical to that of the predicate device K201878.
{8}------------------------------------------------
## INFINETA
The subject device and the predicate device are reusable rigid sterilization trays intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Components of the subject device and the predicate device are perforated to allow for penetration of the sterilant, are to be used with the same moist heat (steam) sterilization cycle and require the use of an FDA-cleared wrap to maintain sterility.
The subject device and the predicate device include components manufactured from polyphenylsulfone (PPSU) and silicone, and both devices are provided in one overall size. While the predicate device is provided in two configurations, the subject device is provided in one configuration. The overall dimensions of the subject device are smaller than the overall dimensions for the predicate device and the vent to volume ratio of the subject device is larger than the vent to volume ratio of the predicate device.
There are minor differences in the configurations, sizes, or designs between the subject device and the predicate device. The differences are related to the specific design features and system components and are supported by the cleaning and sterilization validation and use life testing performed.
The subject device and the predicate device have different use life, 225 compared to 101, respectively.
#### 8. Non-clinical Testing Summary
Provided below is a summary of the non-clinical performance testing included in this submission, including the test methods, purpose, acceptance criteria, and results.
| Test Methodology | Purpose | Acceptance Criteria | Results |
|----------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Validation of the Manual<br>Cleaning Process<br>According to the ISO<br>17664 and the AAMI<br>ST98 Standards | Evaluate and validate the<br>manual cleaning of the<br>subject devices | Extraction Recovery - The extraction efficiency for<br>each marker should be no less than 70%. The amounts<br>of each marker residual detected in the fourth extract<br>should be less than 10% of that detected in the first<br>extraction.<br>Residual marker testing - Protein Assay Acceptance<br>criterion: less than 6.4 µg/cm2 Assay quantitation<br>limits: 2.5 µg/mL Hemoglobin assay Acceptance<br>criterion: less than 2.2 µg/cm2 Assay quantitation<br>limits: 10.0 µg/mL | Pass |
| Validation of the<br>Automated Cleaning<br>Process According to the<br>ISO 17664 and the AAMI<br>ST98 Standards | Evaluate and validate the<br>automated cleaning of<br>the subject devices | Extraction Recovery - The extraction efficiency for<br>each marker should be no less than 70%. The amounts<br>of each marker residual detected in the fourth extract<br>should be less than 10% of that detected in the first<br>extraction.<br>Residual marker testing - Protein Assay Acceptance<br>criterion: less than 6.4 µg/cm2 Assay quantitation<br>limits: 2.5 µg/mL Hemoglobin assay Acceptance<br>criterion: less than 2.2 µg/cm2 Assay quantitation<br>limits: 10.0 µg/mL | Pass |
| Sterilization Validation<br>According to the ISO<br>17664 - ISO 17665<br>Standards and the AAMI<br>TIR 12 Technical Report | Validate a sterilization<br>cycle and drying time of<br>the subject devices | Acceptance criterion: 3 consecutive half-cycles will<br>demonstrate complete inactivation of all biologic<br>indicators; A minimum SAL of 10-6 is achieved if the<br>Instructions for Use are followed | Pass |
## Table 3: Non-Clinical Testing Summary
{9}------------------------------------------------
## I N
| Test Methodology | Purpose | Acceptance Criteria | Results |
|----------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|---------|
| | | Acceptance criterion: Using pre-cycle and post-cycle<br>weights, the weight gain after drying will not exceed<br>+3% | |
| Use Life Testing<br>FDA Guidance<br>Reprocessing Medical<br>Devices in Health Care<br>Settings: Validation<br>Methods and Labeling and<br>ISO 13402:1995 | Life Cycle (simulated<br>use) testing after 225<br>cleaning, disinfection<br>and sterilization cycles | Acceptance criteria: Visual inspection, functional<br>verification for 225 use cycles | Pass |
| Biocompatibility of<br>Subject Device (by<br>cytotoxicity testing)<br>ISO 10993-5<br>ISO 10993-12 | The purpose of this test<br>is to evaluate the<br>cytotoxicity potential of<br>the test article using an in<br>vitro cell culture assay. | Acceptance criterion: ≤ 2 reactivity grade after<br>exposure to extract of the device | Pass |
In summary, the nonclinical testing provided for these devices met the acceptance criteria for each standard and test methodology used to evaluate the devices as shown in the table above.
No clinical data were included in this submission.
#### 9. Conclusion
To conclude, the conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K201878.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.