The SteriCUBE Multiple Tray Sterilization System (SteriCUBE® System) is indicated for enclosing other medical devices that are to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed materials and maintain sterility for up to 30 days until used. The unit must be used with the SteriCUBE System Transfer Cart, SteriCUBE System Filters and SteriCUBE System Integrity Locks. The SteriCUBE System is intended to be used in prevacuum steam sterilizers with a prevacuum cycle of 270°F and exposure time of 4 minutes. Use no more than 3 trays per shelf or 25 lbs. per tray. The SteriCUBE System was tested and validated with rigid instruments containing lumens with an inner diameter of 0.7mm and an overall length of 500mm as well as lumens with an inner diameter of 1.0mm and an overall length of 850mm. Do not use with instruments containing lumens with an inner diameter smaller than 0.7mm and an overall length longer than 500mm or lumens with an inner diameter smaller than 1.0mm and an overall length longer than 850mm. Use only uncovered, perforated or wire mesh general delivery trays within the SteriCUBE System.
Device Story
SteriCUBE Multiple Tray Sterilization System (MTS225/MTS300) consists of stainless steel sterilization cabinet and adjustable transfer cart; used to house surgical instruments during steam sterilization. Healthcare providers load uncovered, perforated, or wire mesh trays into cabinet; cabinet deployed into prevacuum steam sterilizer via transfer cart. System utilizes disposable paper filters and optional reusable ceramic drain filter to maintain sterility. Output is sterilized, protected instrument set. Benefits include maintained sterility for 30 days post-processing. Device operates in clinical sterilization departments; requires specific filters and integrity locks for function. Clinical decision-making relies on successful sterilization verification via indicators.
Clinical Evidence
Bench testing only. Sterilization efficacy validated per ANSI/AAMI ST77:2013. Testing confirmed 10^-6 SAL for specified lumen sizes (0.7mm x 500mm and 1.0mm x 850mm) and verified sterility maintenance for 30 days with a 5-minute dry cycle.
Indicated for enclosing medical devices for sterilization by healthcare providers and maintaining sterility for up to 30 days. Intended for use in prevacuum steam sterilizers (270°F, 4-minute exposure). Compatible with rigid instruments with minimum lumen ID 0.7mm (max length 500mm) or 1.0mm (max length 850mm). Contraindicated for lumens smaller than these dimensions.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Predicate Devices
MTS225 and MTS300 Multiple Tray Sterilization Systems (K190541)
Related Devices
K152014 — MTS300 Multiple Tray Sterilization System · Pmbs, LLC · Oct 20, 2015
K190541 — MTS225 and MTS300 Multiple Tray Sterilization Systems · Pmbs, LLC · Nov 5, 2019
K233222 — Aygün Sterilization Container Systems · Aygun Surgical Instruments Co., Inc. · Dec 12, 2023
K172850 — SterilContainer with PrimeLine Pro Lid · Aesculap, Inc. · Nov 17, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services - USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a recognizable symbol of the agency's role in regulating and overseeing food and drug products in the United States.
September 28, 2022
SteriCUBE, LLC % Annette Hillring President Hillring & Associates, Inc. 3012 St. Charles Drive Tampa, Florida 33618
Re: K222328
Trade/Device Name: SteriCUBE® Multiple Tray Sterilization Systems Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: July 29, 2022 Received: August 2, 2022
Dear Annette Hillring:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K222328
Device Name SteriCUBE(R) Multiple Tray Sterilization Systems
#### Indications for Use (Describe)
The SteriCUBE Multiple Tray Sterilization System) is indicated for enclosing other medical devices that are to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed materials and maintain sterility for up to 30 days until used. The unit must be used with the SteriCUBE System Transfer Cart, SteriCUBE System Filters and SteriCUBE System Integrity Locks.
The SteriCUBE System is intended to be used in prevacuum steam sterilizers with a prevacuum cycle of 270°F and exposure time of 4 minutes. Use no more than 3 trays per shelf or 25 lbs. per tray.
The SteriCUBE System was tested and validated with rigid instruments containing lumens with an inner diameter of 0.7mm and an overall length of 500mm as well as lumens with an inner diameter of 1.0mm and an overall length of 850mm. Do not use with instruments containing lumens with an inner diameter smaller than 0.7mm and an overall length longer than 500mm or lumens with an inner diameter smaller than 1.0mm and an overall length longer than 850mm.
Use only uncovered, perforated or wire mesh general delivery trays within the SteriCUBE System.
Type of Use *(Select one or both, as applicable)*
| <span style="display:inline-block;">☐</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------------------------------------|
| <span style="display:inline-block;">❌</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K222328 510(k) Summary
| Submitter | SteriCUBE, LLC<br>2701 Beach Boulevard South<br>Gulfport, FL 33707<br>Contact person: Michele E. Mauzerall, CEO<br>Phone: 908-300-7093<br>Email: mem.pmbs@gmail.com<br>Date prepared: September 27, 2022 |
|--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device | SteriCUBE® Multiple Tray Sterilization Systems |
| Predicate Device | MTS225 and MTS300 Multiple Tray Sterilization Systems (“MTS Systems”) cleared most recently November 5, 2019, via K190541 |
| Device Description | The SteriCUBE® Multiple Tray Sterilization Systems are also referred to as the “SteriCUBE Systems” and include the SteriCUBE MTS225 System and the SteriCUBE MTS300 System. The SteriCUBE System is comprised of the stainless steel sterilization cabinet and adjustable transfer cart. An optional reusable ceramic drain filter and single use filters and integrity locks with sterilization indicator dots are used with the system for each sterilization cycle. The optional ceramic drain filter should be replaced every 6 months or after 160 uses, whichever comes first. An optional STEAMPlus™ tray record card with STEAMPlus sterilization integrator (SPS Medical) may be utilized by the healthcare provider. Adjustable shelves within the cabinet can hold up to three trays (uncovered, perforated or wire mesh) of devices and/or surgical instruments intended for a single patient surgery. No more than 25 pounds can be loaded per tray for a maximum of 75 pounds per shelf. The MTS300 Cabinet (with four shelves) can hold up to 300 pounds per load and the MTS225 Cabinet (with three shelves) can hold up to 225 pounds per load. The loaded cabinet is deployed from the adjustable transfer cart and processed in pre-vacuum steam sterilizers. Processing includes the standard conditioning phase of the autoclave followed by a sterilization cycle time of 4 minutes and a dry cycle of a minimum of 5 minutes. Sterility is maintained for up to 30 days. |
Continued on next page
{4}------------------------------------------------
# Section 5: 510(k) Summary, continued
| Indications for<br>Use | The SteriCUBE Multiple Tray Sterilization System (SteriCUBE® System) is<br>indicated for enclosing other medical devices that are to be sterilized by a<br>healthcare provider. It is intended to allow sterilization of the enclosed<br>materials and maintain sterility for up to 30 days until used. The unit must be<br>used with the SteriCUBE System Transfer Cart, SteriCUBE System Filters<br>and SteriCUBE System Integrity Locks. | |
|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The SteriCUBE System is intended to be used in prevacuum steam sterilizers<br>with a prevacuum cycle of 270°F and exposure time of 4 minutes. Use no<br>more than 3 trays per shelf or 25 lbs. per tray. | |
| | The SteriCUBE System was tested and validated with rigid instruments<br>containing lumens with an inner diameter of 0.7mm and an overall length of<br>500mm as well as lumens with an inner diameter of 1.0mm and an overall<br>length of 850mm. Do not use with instruments containing lumens with an<br>inner diameter smaller than 0.7mm and an overall length longer than 500mm<br>or lumens with an inner diameter smaller than 1.0mm and an overall length<br>longer than 850mm. | |
| | Use only uncovered, perforated or wire mesh general delivery trays within the<br>SteriCUBE System. | |
| Comparison of<br>Technological<br>Characteristics<br>with the<br>Predicate<br>Device | There are no new technological characteristics associated with the modified<br>SteriCUBE System. A comparison of the technological characteristics of the<br>subject device, the SteriCUBE System (SteriCUBE MTS225 and SteriCUBE<br>MTS300 Systems) to the predicate, the currently marketed MTS225 and<br>MTS300 Systems, is provided in the table on the following pages. | |
| | Continued on next page | |
| Characteristic | Subject Device: | Predicate Device: |
| | SteriCUBE® Systems (SteriCUBE MTS225 and | MTS225 and MTS300 Systems |
| | SteriCUBE MTS300 Systems) | (K190541) |
| Indications for Use | The SteriCUBE Multiple Tray Sterilization System is indicated<br>for enclosing other medical devices that are to be sterilized by a<br>healthcare provider. It is intended to allow sterilization of the<br>enclosed materials and maintain sterility for up to 30 days until<br>used. The unit must be used with the SteriCUBE System Transfer<br>Cart, SteriCUBE System Filters and SteriCUBE System Integrity<br>Locks. | The MTS System (Multiple Tray Sterilization System) is indicated<br>for enclosing other medical devices that are to be sterilized by a<br>healthcare provider. It is intended to allow sterilization of the<br>enclosed materials and maintain sterility for up to 30 days until<br>used. The unit must be used with the MTS System Transfer Cart,<br>MTS System Filters and MTS System Integrity Locks. |
| | The SteriCUBE System is intended to be used in prevacuum<br>steam sterilizers with a prevacuum cycle of 270°F and exposure<br>time of 4 minutes. Use no more than 3 trays per shelf or 25 lbs.<br>per tray. | The MTS System is intended to be used in prevacuum steam<br>sterilizers with a prevacuum cycle of 270°F and exposure time of 4<br>minutes. Use no more than 3 trays per shelf or 25 lbs. per tray. |
| | The SteriCUBE System was tested and validated with rigid<br>instruments containing lumens with an inner diameter of 0.7mm<br>and an overall length of 500mm as well as lumens with an inner<br>diameter of 1.0mm and an overall length of 850mm. Do not use<br>with instruments containing lumens with an inner diameter<br>smaller than 0.7mm and an overall length longer than 500mm or<br>lumens with an inner diameter smaller than 1.0mm and an overall<br>length longer than 850mm. | The MTS System was tested and validated with rigid instruments<br>containing lumens with an inner diameter of 0.7mm and an overall<br>length of 500mm as well as lumens with an inner diameter of<br>1.0mm and an overall length of 850mm. Do not use with<br> |
| | Use only uncovered, perforated or wire mesh general delivery<br>trays within the SteriCUBE System. | Use only uncovered, perforated or wire mesh general delivery trays<br>within the MTS System. |
| | Sterilization | Prevacuum cycle of 270°F and exposure time of 4 |
| Parameters | minutes | minutes |
| Minimum Drying | 5 minutes | 10 minutes |
| Time | | |
| Characteristic | Subject Device:<br>SteriCUBE Systems (SteriCUBE MTS225 and<br>SteriCUBE MTS300 Systems) | Predicate Device:<br>MTS225 and MTS300 Systems<br>(K190541) |
| Sterility<br>Maintenance | 30 days | 30 days |
| Cabinet Material | 16 gauge stainless steel | 16 gauge stainless steel |
| Weight w/ Shelves | 212 lbs (MTS300) and 185 lbs (MTS225) | 212 lbs (MTS300) and 185 lbs (MTS225) |
| Filter Material | Heavy duty sterilization wrap by SPS Medical<br>Optional ceramic drain filter<br>Optional sterilization wrap filter separators | Heavy duty sterilization wrap by SPS Medical |
| Volume-to-Vent<br>Ratio | 129.131 (MTS300) and 94.725 (MTS225) | 129.131 (MTS300) and 94.725 (MTS225) |
| Deployment into<br>Sterilization<br>Chamber | Adjustable transfer cart | Adjustable transfer cart |
| Recommended<br>Sterilization Trays | All manufacturers' trays – uncovered, perforated or<br>wire mesh general delivery trays | All manufacturers' trays – uncovered, perforated or wire<br>mesh general delivery trays |
| Recommended<br>Sterilizers | All makes and models with dimensions compatible<br>with the SteriCUBE System | All makes and models with dimensions compatible with<br>the MTS System |
{5}------------------------------------------------
## Section 5: 510(k) Summary, continued
### Comparison of Technological Characteristics with the Predicate Device, continued
{6}------------------------------------------------
Continued on next page
## Section 5: 510(k) Summary, continued
### Comparison of Technological Characteristics with the Predicate Device, continued
Continued on next page
{7}------------------------------------------------
# Section 5: 510(k) Summary, continued
| Comparison of<br>Technological<br>Characteristics<br>with the<br>Predicate<br>Device,<br>continued | Modifications introduced into the SteriCUBE® Multiple Tray Sterilization<br>Systems do not affect the technological characteristics. The cabinet material<br>remains stainless steel and a filtration system utilizing disposable paper filters<br>of the same material from the same supplier is the method by which<br>sterilization is achieved and maintained. The optional reusable drain ceramic<br>filter and paper filter separators are available for user convenience and<br>performance data confirms their effectiveness. The cabinet continues to be<br>deployed into the sterilization chamber with an adjustable transfer cart. The<br>sterilization parameters and sterility maintenance specifications are identical<br>to the predicate device. | |
|----------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| | The SteriCUBE Systems (both the SteriCUBE MTS225 and MTS300<br>Systems) have demonstrated the ability to:<br>achieve a 10-6 SAL when sterilizing longer and smaller lumen sized<br>instruments (rigid instruments containing lumens with an inner<br>diameter of 0.7mm and an overall length of 500mm as well as lumens<br>with an inner diameter of 1.0mm and an overall length of 850mm), meet the shorter minimum dry time of 5 minutes, and maintain sterility for 30 days. | |
| Performance<br>Data | Nonclinical data was generated to ensure the SteriCUBE System continues to<br>meet the intended use. Sterilization efficacy verification studies were<br>performed in accordance with the appropriate requirements of ANSI/AAMI<br>ST77:2013 Containment devices for reusable medical device sterilization. | |
| Conclusions | In accordance with the Federal Food, Drug and Cosmetic Act and 21 CFR<br>807, and based upon the information and scientifically valid data provided in<br>this premarket notification, SteriCUBE, LLC, concludes that the subject<br>device, the SteriCUBE Systems (SteriCUBE MTS225 and MTS300 Systems)<br>is as safe, as effective and perform as well as the predicate device, the<br>MTS225 and MTS300 Multiple Tray Sterilization Systems (K190541). | |
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.