K214041 · Aesculap, Inc. · KCT · Dec 30, 2022 · General Hospital
Device Facts
Record ID
K214041
Device Name
Aesculap AICON Container
Applicant
Aesculap, Inc.
Product Code
KCT · General Hospital
Decision Date
Dec 30, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The AESCULAP Aicon™ Container is a reusable rigid sterilization container intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed device and also maintain sterility of the enclosed device during transport and until used. This container is compatible for use in the following sterilization modalities in the configurations listed below: - Ethylene Oxide (EtO) - STERIS V-PRO maX/maX 2: Lumen, Non-Lumen, Flex - STERIS V-PRO 60: Lumen, Non- Lumen, Flex - STERRAD 100NX: Standard, Express, Flex, Duo - STERRAD NX: Standard, Advanced - STERRAD 100S - STERIZONE VP4 - PreVac Steam
Device Story
Reusable rigid sterilization container system; constructed from anodized aluminum with single-use polypropylene filters. Encloses medical devices for sterilization; maintains sterility during transport and storage. Used in healthcare facilities by sterile processing staff. Compatible with multiple sterilization modalities: EtO, STERIS V-PRO (maX/maX 2, 60), STERRAD (100NX, NX, 100S), STERIZONE VP4, and PreVac Steam. Available in solid bottom or enhanced drying system (EDS) configurations. Healthcare providers load instruments into the container; process through validated sterilization cycles; store until use. Benefits include protection of sterile instruments and maintenance of sterility barrier for up to 365 days.
Clinical Evidence
No clinical data. Bench testing only: sterilization efficacy (SAL 10^-6), dry time, microbial aerosol challenge, simulated use for flexible scopes, material compatibility, cytotoxicity (ISO 10993-5), and sterility maintenance (365 days). All tests passed.
Technological Characteristics
Reusable rigid container; anodized aluminum construction; silicone gaskets; single-use polypropylene filters. Dimensions: Full, Three-Quarter, and Half sizes. Vent-to-volume ratios: 0.04mm^-1 * 10^-3 to 0.26mm^-1 * 10^-3. Sterilization modalities: EtO, V-PRO, STERRAD, STERIZONE, PreVac Steam. No software or electronic components.
Indications for Use
Indicated for use by healthcare providers to enclose medical devices for sterilization and to maintain sterility during transport and storage. Compatible with various sterilization modalities including EtO, specific STERIS V-PRO cycles, STERRAD cycles, STERIZONE VP4, and PreVac Steam. Patient population is not specified; intended for use in clinical settings.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.
December 30, 2022
Aesculap Inc. Tim Stoudt Associate Director- Regulatory Affairs 3773 Corporate Parkway Center Valley, Pennsylvania 18034
Re: K214041
Trade/Device Name: Aesculap AICON™ Container Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: November 30, 2022 Received: December 2, 2022
Dear Tim Stoudt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Clarence W. Murray III -S
Clarence W. Murray III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K214041
Device Name AESCULAP Aicon™ Container
#### Indications for Use (Describe)
The AESCULAP Aicon™ Container is a reusable rigid sterilization container intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed device and also maintain sterility of the enclosed device during transport and until used. This container is compatible for use in the following sterilization modalities in the configurations listed below:
- · Ethylene Oxide (EtO)
- STERIS V-PRO maX/maX 2: Lumen, Non-Lumen, Flex
- · STERIS V-PRO 60: Lumen, Non- Lumen, Flex
- · STERRAD 100NX: Standard, Express, Flex, Duo
- STERRAD NX: Standard, Advanced
- · STERRAD 100S
- STERIZONE VP4
- PreVac Steam
The AESCULAP Aicon™ consists of two different container styles: a solid bottom container and enhanced drying system (EDS) container.
The following tables identify the load configurations.
| Sterilization Cycle | Container Sizes | Compatible Lumens |
|-----------------------------------------------------------------------------------------|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Ethylene Oxide (EtO)<br>(130°F, 60 minute exposure<br>50-70% RH, concentration 735mg/L) | Full Size | (2) Stainless Steel Lumens (≥3.0mm ID x ≤ 400mm L) |
| | Three-Quarter Size | (2) Stainless Steel Lumens (≥3.0mm ID x ≤ 400mm L) |
| | Half Size | (2) Stainless Steel Lumens (≥3.0mm ID x ≤ 400mm L) |
| STERIS V-PRO maX /maX 2<br>Lumen Cycle | Full Size | (6) Stainless Steel Lumens (≥1mm ID x ≤ 400mm L) |
| | Three-Quarter Size | (6) Stainless Steel Lumens (≥0.77mm ID x ≤ 250mm L) |
| | Half Size | (6) Stainless Steel Lumens (≥1mm ID x ≤ 400mm L)<br>(6) Stainless Steel Lumens (≥0.77mm ID x ≤ 250mm L) |
| | | (12) Stainless Steel Lumens (≥1mm ID x ≤ 150mm L) |
| STERIS V-PRO Max/ Max 2<br>Non-Lumen Cycle | Full Size | Non lumened stainless steel instruments |
| | Three-Quarter Size | Non lumened stainless steel instruments |
| | Half Size | Non lumened stainless steel instruments |
| STERIS V-PRO maX / maX 2<br>Flex Cycle | Full Size | (1) Flexible endoscope with a light cord (if not integral to<br>endoscope) without any additional load. The flexible endoscope may be single lumen<br>that is ≥1mm ID x ≤1050mm L |
| | Three-Quarter Size | (1) Flexible endoscope with a light cord (if not integral to<br>endoscope) without any additional load. The flexible endoscope may be single lumen<br>that<br>is ≥1mm ID x ≤1050mm L |
| | Half Size | (2) Flexible endoscopes with a light cord (if not integral to |
#### Table 1. AESCULAP Aicon™ Container Lumen Configurations
{3}------------------------------------------------
| | | endoscope) without any additional load. The flexible endoscope may be single lumen<br>that is ≥1mm ID x ≤1050mm L or two lumens with one ≥ 1 mm ID and ≤ 990 mm L<br>and the other ≥ 1 mm ID and ≤ 850 mm L<br>OR<br>(1) flexible endoscope with a light cord (if not integral<br>to endoscope) and additional non-lumened instruments. The flexible endoscope can<br>have:<br>a single lumen that is ≥ 1mm ID and ≤ 1050 L or two lumens with one ≥ 1 mm ID and<br>< 990 L and the other ≥ 1 mm ID and ≤ 850 mm L | | | |
|-------------------------------------|--------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|
| | Full Size | (6) Stainless Steel Lumens (≥1mm ID x ≤ 400mm L) | | | |
| | | (6) Stainless Steel Lumens (≥0.77mm ID x ≤ 250mm L) | | | |
| STERIS VPRO 60 | Three-Quarter Size | (6) Stainless Steel Lumens (≥1mm ID x ≤ 400mm L) | | | |
| Lumen Cycle | | (6) Stainless Steel Lumens (≥0.77mm ID x ≤ 250mm L) | | | |
| | Half Size | (12) Stainless Steel Lumens (≥1mm ID x ≤ 150mm L) | | | |
| STERIS VPRO 60 | Full Size | Non lumened stainless steel instruments | | | |
| Non-Lumen Cycle | Three-Quarter Size | Non lumened stainless steel instruments | | | |
| | Half Size | Non lumened stainless steel instruments | | | |
| | Full Size | (1) flexible surgical endoscope or bronchoscope with a<br>light cord (if not integral to scope) and single lumen (≥1mm ID x ≤990mm L) | | | |
| | | | | | |
| STERIS VPRO 60 | Three-Quarter Size | (1) flexible surgical endoscope or bronchoscope with a<br>light cord (if not integral to scope) and single lumen (≥1mm ID x ≤990mm L) | | | |
| Flex Cycle | | | | | |
| | Half Size | (1) flexible surgical endoscope or bronchoscope with a<br>light cord (if not integral to scope) and single lumen or dual lumens<br>(≥1mm ID x ≤990mm L) | | | |
| | Full Size | (5) Stainless-Steel Lumens (≥1.0mm ID x ≤500mm L) | | | |
| | | (5) Stainless-Steel Lumens (≥0.7mm ID x ≤400mm L) | | | |
| STERRAD 100NX | Three-Quarter Size | (5) Stainless-Steel Lumens (≥1.0mm ID x ≤500mm L) | | | |
| Standard Cycle | | (5) Stainless-Steel Lumens (≥0.7mm ID x ≤400mm L) | | | |
| | Half Size | (5) Stainless-Steel Lumens (≥0.7mm ID x ≤250mm L) | | | |
| | | (5) Stainless-Steel Lumens (≥0.7mm ID x ≤200mm L) | | | |
| | Full Size | (5) Stainless Steel Lumens (≥3.0mm I.D. x ≤ 400mm L) | | | |
| STERRAD 100S | Three-Quarter Size | (5) Stainless Steel Lumens (≥3.0mm I.D. x ≤ 400mm L) | | | |
| | Half Size | (5) Stainless Steel Lumens (≥2.0mm ID x ≤ 250mm L) | | | |
| | Full Size | (10)Stainless-steel lumens (≥2.5mm ID x ≤ 400mm L) | | | |
| STERRAD NX | Three-Quarter Size | (10)Stainless-steel lumens (≥2.5mm ID x ≤ 400mm L) | | | |
| Standard Cycle | | | | | |
| | Half Size | (10) Stainless-steel lumens (≥2mm ID x ≤ 200mm L) | | | |
| | Full Size | (5) Stainless-steel lumens (≥2.5mm ID x ≤500mm L) | | | |
| | | (5) Flexible Lumens (≥1mm ID x ≤850mm L) | | | |
| STERRAD NX<br>Advanced Cycle | Three-Quarter Size | (5) Stainless-steel lumens (≥2.5mm ID x ≤500mm L) | | | |
| | Half Size | (5) Flexible Lumens (≥1mm ID x ≤850mm L)<br>(10) Stainless-steel lumens (≥2mm ID x ≤ 200mm L) | | | |
| | | | | | |
| STERRAD 100NX | Full Size | Non lumened stainless steel instruments | | | |
| Express Cycle | Three-Quarter Size | Non lumened stainless steel instruments | | | |
| | Full Size | Non lumened stainless steel instruments | | | |
| | | | | | |
| STERRAD 100NX | Full Size | (1) Flexible Lumen (≥1mm ID x ≤ 850mm L) | | | |
| DUO Cycle | Three-Quarter Size | (1) Flexible Lumen (≥1mm ID x ≤ 850mm L) | | | |
| | Half Size | (1) Flexible Lumen (≥1.mm ID x ≤ 650mm L) | | | |
| | Full Size | (1) Flexible Lumen(≥1.5mm ID x ≤ 650mm L) | | | |
| STERRAD 100NX | Three-Quarter Size | (1) Flexible Lumen (≥1.5mm ID x ≤ 650mm L) | | | |
| Flex Cycle | | | | | |
| | Half Size | (1) Flexible lumen (≥1.mm ID x ≤ 650mm L) | | | |
| | Full Size | (4) Stainless Steel Lumens (≥1.mm ID x ≤200mm L) | | | |
| PreVac Steam with dry time | | | | | |
| (270°F, 4 min Exposure) | Three-Quarter | (4) Stainless Steel Lumens (≥1.0mm ID x ≤200mm L) | | | |
| | Half Size | (4) Stainless Steel Lumens (≥1.0mm ID x ≤200mm L) | | | |
| PreVac Steam without dry time | Full Size | (4) Stainless Steel Lumens (≥1mm ID x ≤ 125 mm L) | | | |
| (270°F, 4 min Exposure) | Three-Quarter Size | (4) Stainless Steel Lumens (≥1mm ID x ≤ 125 mm L) | | | |
| Porous | Half Size | (4) Stainless Steel Lumens (≥1mm ID x ≤ 125 mm L) | | | |
| | | | | | |
| Prevac cycles without dry times are | Prevac cycles without dry times are<br>Immediate Use Sterilization Cycles<br>(IUSS). | | | | |
{4}------------------------------------------------
| (270°F, 3 min Exposure) | Full Size | Non-lumened instruments |
|--------------------------------------------------------------------------------------|--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Non-Porous | Three-Quarter Size | Non-lumened instruments |
| Prevac cycles without dry times are<br>Immediate Use Sterilization Cycles<br>(IUSS). | Half Size | Non-lumened instruments |
| STERIZONE VP4<br>Load 8<br>(Aicon Validated Load 1) | Full Size | (1) Dual channel flexible scope (≥1.5mm ID x ≤850mm L) and<br>(≥1.5mm ID x ≤989mm L)<br>(1) Single channel flexible scope (≥1.5mm ID x ≤850mm L) |
| | Three-Quarter Size | (1) Dual channel flexible scope (≥1.5mm ID x ≤850mm L) and<br>(≥1.5mm ID x ≤989mm L)<br>(1) Single channel flexible scope (≥1.5mm ID x ≤850mm L) |
| | Half Size | (1) Dual channel flexible scope (≥1.5mm ID x ≤850mm L) and<br>(≥1.5mm ID x ≤989mm L)<br>(1) Single channel flexible scope (≥1.5mm ID x ≤850mm L) |
| STERIZONE VP4<br>Load 4<br>(Aicon Validated Load 2) | Full Size | (1) Dual channel semi rigid scope (≥0.7mm ID x 500mm L; ≥1.1mm ID x ≤500mm L)<br>(1) Dual channel semi rigid scope (≥1.1 mm ID x 500mm L; ≥1.1 mm ID x ≤500mm L)<br>(2) Stainless Steel Lumens (≥2mm ID x ≤250mm L)<br>(2) Stainless Steel Lumens (≥3mm ID x ≤400mm L) |
| | Three-Quarter Size | (1) Dual channel semi rigid scope (≥0.7mm ID x 500mm L; ≥1.1mm ID x ≤500mm L)<br>(1) Dual channel semi rigid scope (≥1.1 mm ID x 500mm L; ≥1.1 mm ID x ≤500mm L)<br>(2) Stainless Steel Lumens (≥2mm ID x ≤250mm L)<br>(2) Stainless Steel Lumens (≥3mm ID x ≤400mm L) |
| | Half Size | (1) Stainless Steel Lumen (≥0.7 mm ID x ≤300mm L)<br>(1) Stainless Steel Lumen (≥1.0 mm ID x ≤300mm L)<br>(2) Stainless Steel Lumens (≥2.0 mm ID x ≤250mm L)<br>(2) Stainless Steel Lumens (≥3.0 mm ID x ≤300mm L) |
| STERIZONE VP4<br>Load 7<br>(Aicon Validated Load 3) | Full Size | Non-Lumened Stainless Steel Instruments (including hinges, luerlocks) |
| | Three-Quarter Size | Non-Lumened Stainless Steel Instruments (including hinges, luerlocks) |
| | Half Size | Non-Lumened Stainless Steel Instruments (including hinges, luerlocks) |
## Table 2. AESCULAP Aicon™ Container Configurations
| Sterilization<br>Method | Container Size | Standard<br>Container | EDS<br>Container | Container<br>Lid Part # | Total Loaded<br>Container (lbs) |
|--------------------------------------------------------------------------------------|------------------------------|-----------------------|------------------|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| EtO<br>(130°F, 60 minute exposure<br>50-70% RH, concentration 735mg/L) | Full Size 4" | JJ110 | JJ111 | | |
| | Full Size 6" | JJ120 | JJ121 | | |
| | Full Size 8" | JJ130 | JJ131 | JJ410 | 25 |
| | Full Size 10" | JJ140 | JJ141 | | |
| | Three-Quarter Size 4" | JJ310 | JJ311 | | |
| | Three-Quarter Size 6" | JJ320 | JJ321 | | |
| | Three-Quarter Size 8" | JJ330 | JJ331 | JJ430 | 25 |
| | Three-Quarter Size 10" | JJ340 | JJ341 | | |
| | Half Size 4" | JJ210 | JJ211 | | |
| | Half Size 6" | JJ220 | JJ221 | | |
| | Half Size 8" | JJ230 | JJ231 | JJ420 | 25 |
| | Half Size 10" | JJ240 | JJ241 | | |
| STERIS VPRO<br>maX/ maX 2 Lumen | Full Size 4" | JJ110 | JJ111 | | |
| | Full Size 6" | JJ120 | JJ121 | | |
| | Full Size 8" | JJ130 | JJ131 | JJ410 | 19.65 |
| | Full Size 10" | JJ140 | JJ141 | | |
| | Three-Quarter Size 4" | JJ310 | JJ311 | | |
| | Three-Quarter Size 6" | JJ320 | JJ321 | | |
| | Three-Quarter Size 8" | JJ330 | JJ331 | JJ430 | 19.65 |
| | Three-Quarter Size 10" | JJ340 | JJ341 | | |
| | Half Size 4" | JJ210 | JJ211 | | |
| | Half Size 6" | JJ220 | JJ221 | JJ420 | 13.8 |
| | Half Size 8" | JJ230 | JJ231 | | |
| | Half Size 10" | JJ240 | JJ241 | | |
| | Full Size 4" | JJ110 | JJ111 | | 25 |
| STERIS VPRO<br>maX / maX 2 Non-Lumen | Full Size 6" | JJ120 | JJ121 | JJ410 | |
| | Full Size 8" | JJ130 | JJ131 | | |
| | Full Size 10" | JJ140 | JJ141 | | |
| | Three-Quarter Size 4" | JJ310 | JJ311 | | 25 |
| | Three-Quarter Size 6" | JJ320 | JJ321 | JJ430 | |
| | Three-Quarter Size 8" | JJ330 | JJ331 | | |
| | Three-Quarter Size 10" | JJ340 | JJ341 | | |
| | Half Size 4" | JJ210 | JJ211 | | |
| | Half Size 6" | JJ220 | JJ221 | | JJ420 |
| | Half Size 8" | JJ230 | JJ231 | | |
| | Half Size 10" | JJ240 | JJ241 | |…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.