K191566 · Avinent Implant System S.L.U. · KCT · Mar 6, 2020 · General Hospital
Device Facts
Record ID
K191566
Device Name
Avinent Surgical Tray
Applicant
Avinent Implant System S.L.U.
Product Code
KCT · General Hospital
Decision Date
Mar 6, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The Avinent Surgical Tray is designed to hold various dental surgical drills and tools in order to organize, steam sterilize, and transport the instruments between uses. The tray is to be enclosed in an FDA cleared steam sterilizable wrap and sterilized in an FDA cleared sterilizer for the following cycle: Prevaccum Steam - At 132° for 4 minutes with a 30 minute dry time. -The trays are intended for sterilization of non-porous loads. -The trays are recommended not to be stacked during sterilization. -The Biomimetic Coral Sterilization Cassette Guided Surgery represents the worst case validated load due to the number of components (32 instruments) and the weight of 599.6 grams. -A vinent Implant System SLU does not make any lumen claims for the Avinent Surgical Tray. -The tray will be marketed in the variations found in the following table.
Device Story
Reusable rigid sterilization container/organizing tray; holds dental surgical drills and tools. Composed of base, lid, and internal insert tray; all components perforated for steam penetration. Features silicone grommets/cradles to secure instruments. Used in healthcare facilities; operated by clinical staff. Requires enclosure in FDA-cleared steam sterilizable wrap. Subjected to prevacuum steam sterilization (132°C, 4 min, 30 min dry). Protects instruments during transport, sterilization, and storage. Benefits include organized, sterile instrument delivery for dental procedures.
Clinical Evidence
Bench testing only. Validated cleaning cycle (no visible soil, hemoglobin <2.2μg/cm², protein ≤6.4 μg/cm²). Steam sterilization efficacy confirmed via half-cycle testing with no growth and successful chemical indicator performance. Durability verified through 100 sterilization cycles with no functional or visual degradation. Biocompatibility (cytotoxicity) per ISO 10993-5/12 confirmed with score of 0.
Technological Characteristics
Materials: Radel R-5000 polymer resin, medical-grade silicone (FDA 21 CFR 177.2600). Dimensions: 207.5 x 157.5 x 71.7mm. Design: Perforated base, lid, and insert tray with locking mechanism. Sterilization: Prevacuum steam (132°C). Reusable. No electronics or software.
Indications for Use
Indicated for holding, organizing, steam sterilizing, and transporting non-porous dental surgical drills and tools in healthcare facilities. Not for use with lumen instruments. Not to be stacked during sterilization.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Predicate Devices
Implant Direct Interactive Complete Surgical Tray (K142519)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
March 6, 2020
AVINENT Implant System, S.L.U. % Angela Blackwell Senior Consultant Blackwell Device Consulting P.O. Box 718 Gresham, Oregon 97030-0172
Re: K191566
Trade/Device Name: Avinent Surgical Tray Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: January 28, 2020 Received: February 4, 2020
Dear Angela Blackwell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Sreekanth Gutala, Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
## K191566
Device Name Avinent Surgical Tray
Indications for Use (Describe)
The Avinent Surgical Tray is designed to hold various dental surgical drills and tools in order to organize, steam sterilize, and transport the instruments between uses. The tray is to be enclosed in an FDA cleared steam sterilizable wrap and sterilized in an FDA cleared sterilizer for the following cycle:
Prevaccum Steam - At 132° for 4 minutes with a 30 minute dry time.
-The trays are intended for sterilization of non-porous loads.
-The trays are recommended not to be stacked during sterilization.
-The Biomimetic Coral Sterilization Cassette Guided Surgery represents the worst case validated load due to the number of components (32 instruments) and the weight of 599.6 grams.
-A vinent Implant System SLU does not make any lumen claims for the Avinent Surgical Tray.
-The tray will be marketed in the variations found in the following table.
| DEVICE MODEL NAME | MAX no. of<br>INSTRUMENTS | WEIGHT of each<br>tray FULL (g) | WEIGHT of each<br>tray EMPTY (g) |
|-----------------------------------------------------------|---------------------------|---------------------------------|----------------------------------|
| Biomimetic Coral Sterilization<br>Cassette Guided Surgery | 32 | 599.6 | 510.0 |
| Biomimetic Ocean Sterilization<br>Cassette Guided Surgery | 26 | 592.8 | 510.0 |
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR.801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### Avinent Surgical Tray 510k Summary K191566
#### AVINENT Implant System S.L.U.
#### Pol. Ind. Santa Anna I - 08251 Santpedor Barcelona, Spain
#### +34 902 38 38 48
March 3, 2020
Submission Contact: Angela Blackwell Blackwell Device Consulting P.O. Box 718 Gresham, OR 97030-0172 (704) 450-9934 angela@blackwelldevice.com
Classification Name: Sterilization Wrap Containers, Trays, Cassettes, and other accessories
Common Name: Instrument Sterilization Tray
Regulation Number: 21 CFR 880.6850
Product Code: KCT
Class: II
Predicate Device: Implant Direct Interactive Complete Surgical Tray K142519
#### Device Description:
The Avinent Surgical Tray is a reusable rigid sterilization container or organizing tray intended for use in health care facilities for the purpose of containing reusable medical devices for sterilization. It is composed of multiple pieces, designed to be integrated into a single unit which contains and protects the interior components during sterilization. Each tray consists of three components – a base tray, a lid, and an internal individualized insert tray. All of the three components are perforated for steam penetration. The internal tray has the ability to hold individualized pieces and accessories which include dental tools, drills, and torque wrench. The tray is available in one size, 207.5 x 157.5 x 71.7mm. The tray comes in two variations, Coral and Ocean.
The rigid multi-piece tray holds dental device apparatus and accessories before, during and after the sterilization process. The tray set has a locking lid to contain the products. The trays are designed to fit any standard autoclave, which allows them to be effective for sterilization and be able to withstand the environment of repeated steam sterilization cycles in normal operating room. Since the trays are perforated, an FDA cleared wrap must be used for sterilization purposes to maintain the sterility of contents.
{4}------------------------------------------------
The trays are reusable and the tray material allows repeated sterilization cycles. The lid, base, and insert are made of Radel R-5000. This material is a polymer resin produced by Solvay Advanced Polymers, LLC and is identical to the Radel used in predicate device cleared under K142519. Radel R 5000 CL 301 is used for the lid, Radel R-5100 BK # 937 is used for the base and Radel R-5100 GY #1037 is used for the insert tray where all the drills and tools are mounted. The small circular brackets (grommets) throughout the insert tray which are used to contain drills and the tool holder with cradle on the base tray are made of medical grade silicone material which has been manufactured to meet FDA 21 CFR 177.2600. These brackets and holders are used to secure the instruments during transport, sterilization, and storage. Although these trays are reusable they will not be serviced or repaired.
The instruments to be sterilized in the proposed trays are all non-porous devices and included dental surgical drills and tools. The tools and drills are all class I exempt.
## Indications for Use:
The Avinent Surgical Tray is designed to hold various dental surgical drills and tools in order to organize, steam sterilize, and transport the instruments between uses. The tray is to be enclosed in an FDA cleared steam sterilizable wrap and sterilized in an FDA cleared sterilizer for the following cycle:
Prevaccum Steam - At 132° for 4 minutes with a 30 minute dry time.
-The trays are intended for sterilization of non-porous loads.
-The trays are recommended not to be stacked during sterilization.
-The Biomimetic Coral Sterilization Cassette Guided Surgery represents the worst case validated load due to the number of components (32 instruments) and the weight of 599.6 grams.
-Avinent Implant System SLU does not make any lumen claims for the Avinent Surgical Tray.
-The tray will be marketed in the variations found in the following table.
| DEVICE MODEL NAME | MAX no. of<br>INSTRUMENTS | WEIGHT of each<br>tray FULL (g) | WEIGHT of each<br>tray EMPTY (g) |
|-----------------------------------------------------------|---------------------------|---------------------------------|----------------------------------|
| Biomimetic Coral Sterilization<br>Cassette Guided Surgery | 32 | 599.6 | 510.0 |
| Biomimetic Ocean Sterilization<br>Cassette Guided Surgery | 26 | 592.8 | 510.0 |
## Technological Characteristics Comparison Table:
The table below provide a technological characteristics comparison of the Avinent Surgical Tray to the predicate device the InterActive Complete Surgical Tray.
| Device | Avinent Surgical Tray<br>(K191566) | InterActive Complete Surgical<br>Tray (K142519) |
|--------------|------------------------------------|-------------------------------------------------|
| Product Code | KCT | KCT |
{5}------------------------------------------------
| Indications for Use | The Avinent Surgical Tray is designed to hold various dental surgical drills and tools in order to organize, steam sterilize, and transport the instruments between uses. | The InterActive Complete Surgical Tray is designed to hold various dental surgical drills and tools in order to organize, steam sterilize, and transport the instruments between uses. |
|---------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Material Composition | Polymer Resin Radel-5000, biomedical grade silicone | Polymer Resin Radel-5000, biomedical grade silicone |
| Design | Plastic tray with locking lid and silicone containment brackets. | Plastic tray with locking lid and silicone containment brackets. |
| Sterilization by<br>1 Gravity Steam | No | Yes |
| 2 Pre Vaccuum Steam | Yes | Yes |
| Air Permeance | Yes | Yes |
| Vent to Volume Ratio | $0.011 in²/in³$ | $.184 in²/in³$ |
| Locking System to hold lid in place | Yes | Yes |
| Reusable | Yes | Yes |
| Stackable | No | No |
| Material compatibility with sterilization process | Yes | Yes |
| Sterilant Penetration Studies | Yes | Yes |
| Biocompatibility | Cytotoxicity tests conducted according to ISO 10993-5 and – 12. | Cytotoxicity tests conducted according to ISO 10993-5 and – 12. |
| Cleaning Instructions for Reusable Devices Validated | Yes | Yes |
| Repeat Sterilization to show material compatibility over time | Yes | Yes |
| Drying Time | Yes | Yes |
| Method to Preserve Sterility of the Tray | FDA cleared wrap | FDA cleared wrap |
## Summary of Non-Clinical Testing:
Provided below are the non-clinical testing performed with the subject device to demonstrate that the device met the acceptance criteria of the test method or the standard.
| Name of Testing | Purpose of Testing | Acceptance Criteria | Results |
|---------------------|---------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|
| Cleaning Validation | Validate the cleaning cycle in the instructions for use | No visible soil.<br>Hemoglobin less than 2.2μg/cm². Protein level ≤6.4 μg/cm².<br>MEM reactivity grade of 2 or less. | No visible soil. All acceptance criteria met. |
{6}------------------------------------------------
| Steam Sterilization<br>Validation | Validate the half-cycle<br>of the cycle in the<br>instructions for use | No growth on half<br>cycles. No moisture on<br>full drying cycle. Pass<br>on Chemical Indicators<br>half and full cycle. | Pass on chemical<br>indicators half and full<br>cycle. No moisture on<br>full drying cycle. No<br>growth on half cycle. |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| 100 cycles sterilization | Show the tray will last<br>at least 100<br>sterilization cycles | Pass visual inspection<br>and test of<br>functionality. | At all cycle counts<br>passed visual<br>inspection and test of<br>functionality. |
| Cytotoxicity | Sterilized instruments<br>were tested to<br>demonstrate their<br>were no residuals on<br>them which would<br>change their<br>biocompatibility. | Pass is a score of no<br>greater than 2 | Pass with score of 0 |
Cleaning validation using a manual method was done following simulated use according to the cleaning method in the instructions for use. Steam sterilization (half cycle) based on the cycle listed in the indications for use was completed. A second steam sterilization based on this cycle was done with thermal profiling. Sterilization of the tray was done to verify the tray safely last for 100 cycles. No changes were noted in the test articles after 100 cycles. Cytotoxicity (ISO 10993-5) of the instruments placed in the tray for sterilization was done after a sterilization cycle was completed to show there were no residuals from the tray on the instruments which would change their biocompatibility.
Conclusion: The conclusions drawn from the nonclinical test demonstrate that the Avinent Surgical Tray is as safe, as effective, and performs as well as or better than the legally marketed predicate device, Interactive Complete Surgical Tray.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.