← Product Code [KCT](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/KCT) · K172292

# SALUS - Hygiene Instrument Reprocessing Container (K172292)

_Scican , Ltd. · KCT · Nov 24, 2017 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/KCT/K172292

## Device Facts

- **Applicant:** Scican , Ltd.
- **Product Code:** [KCT](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/KCT.md)
- **Decision Date:** Nov 24, 2017
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 880.6850
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The SALUS® - Hygiene is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used. The SALUS® - Hygiene is a reusable instrument reprocessing container intended to enclose and organize instruments that are designed to withstand steam sterilization and to maintain sterility of the reusable instruments during storage. The SALUS - Hygiene is suitable for use in dynamic air removal steam sterilizers.

## Device Story

SALUS® - Hygiene is a reusable instrument reprocessing container; designed to enclose and organize hygiene instruments during steam sterilization; maintains sterility of instruments during storage. Used by healthcare providers in clinical settings. Device facilitates steam penetration and drying within dynamic air removal steam sterilizers; protects instruments from contamination post-sterilization. Healthcare providers load instruments into container; process in validated steam sterilization cycles; store until use. Benefits include organized instrument handling and maintained sterility of reusable medical devices.

## Clinical Evidence

No clinical data. Bench testing only; validated sterilization cycles provided for dynamic-air-removal (pre-vacuum and SFPP) steam sterilizers at 270°F (132°C) and 273°F (134°C) with specified exposure and drying times for solid and lumen hygiene instruments.

## Technological Characteristics

Reusable instrument reprocessing container; designed for dynamic air removal steam sterilization. Materials and dimensions not specified. No software or electronic components.

## Regulatory Identification

A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the logo for the Department of Health & Human Services, which features a stylized human figure. To the right of that is the logo for the U.S. Food & Drug Administration (FDA). The FDA logo has the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.

November 24, 2017

SciCan Ltd. Teresa Boyce Director of Regulatory Affairs 1440 Don Mills Road Toronto, M3B 3P9 Ca

Re: K172292

Trade/Device Name: SALUS® - Hygiene Instrument Reprocessing Container Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: October 27, 2017 Received: October 31, 2017

Dear Teresa Boyce:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Michael J. Ryan -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) k172292

#### Device Name

SALUS® - Hygiene Instrument Reprocessing Container

#### Indications for Use (Describe)

The SALUS® - Hygiene is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used. The SALUS® - Hygiene is a reusable instrument reprocessing container intended to enclose and organize instruments that are designed to withstand steam sterilization and to maintain sterility of the reusable instruments during storage. The SALUS - Hygiene is suitable for use in dynamic air removal steam sterilizers.

Please refer the chart below for validated sterilization cycles.

| Steam Sterilizer Type                                      | Cycle   | Sterilization<br>Temperature<br>-0/+6°F (-0/+3°C) | Exposure<br>Time | Recommended<br>Drying Time | Load<br>Type                                   | Maximum<br>Instrument<br>Load Per<br>Container |
|------------------------------------------------------------|---------|---------------------------------------------------|------------------|----------------------------|------------------------------------------------|------------------------------------------------|
| Dynamic-air-removal<br>(pre-vacuum)                        | Wrapped | 270°F (132°C)                                     | 4 minutes        | 16 minutes                 | Solid Hygiene<br>Instruments &                 | 0.487lb (221g)                                 |
| Dynamic-air-removal<br>(steam-flush, pressure-pulse, SFPP) | Wrapped | 270°F (132°C)                                     | 4 minutes        | 30 minutes                 | Lumen Hygiene<br>Instruments                   | 0.478lb (217g)                                 |
| Dynamic-air-removal<br>(pre-vacuum)                        | Wrapped | 273°F (134°C)                                     | 4 minutes        | 30 minutes                 | (Ø .028" [0.7mm] x L 3.35"<br>[85 mm], Qty. 1) | 0.473lb (215g)                                 |

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|                                                 |  |

| | Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/KCT/K172292](https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/KCT/K172292)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
