Jewel Precision Reusable Rigid Sterilization Container System
K162600 · Jewel Precision Sheet Metal & Machine, Inc. · KCT · Dec 9, 2016 · General Hospital
Device Facts
Record ID
K162600
Device Name
Jewel Precision Reusable Rigid Sterilization Container System
Applicant
Jewel Precision Sheet Metal & Machine, Inc.
Product Code
KCT · General Hospital
Decision Date
Dec 9, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The Jewel Precision Reusable Rigid Sterilization Container System is a device intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. It allows sterilization of the enclosed medical device and maintains sterility of the enclosed device until used, for a maximum of 30 days. Jewel Precision Reusable Rigid Sterilization Containers are suitable for dynamic air removal (pre-vacuum) steam sterilization at 270°F (132°C) for 4 minutes with a drying time of 30 minutes. Reusable containers, covers, filter holders, insert trays, and accessory items such as brackets of various configuration, dividers, caddies, and silicone nipple mats are intended to organize and secure the enclosed medical devices during sterilization and storage of the container. Filter media allows ingress and egress of sterilant gas while providing a microbial barrier. Filter media is single use only.
Device Story
Reusable rigid container system for steam sterilization of surgical instruments; comprises aluminum base/cover, stainless steel hardware, silicone gaskets, and single-use paper filters. Used in healthcare facilities to organize, protect, and transport instruments during pre-vacuum steam sterilization (270°F/132°C, 4-min cycle, 30-min dry). Latching mechanism compresses gasket to maintain sterile barrier. Filter media allows sterilant gas ingress/egress while providing microbial barrier. Protects instruments during storage for up to 30 days. No direct patient contact.
Clinical Evidence
No clinical data. Bench testing only. Validation included sterilization efficacy (12-log reduction, SAL 10^-6), dry time validation, and biocompatibility testing (ISO 10993-5) on materials after 100 sterilization cycles.
Technological Characteristics
Materials: Aluminum 5052 H32/6061, Stainless Steel 304/303, USP Class VI Polyamide 11 (nylon coating), USP Class VI silicone elastomer. Filter: Ahlstrom Reliance 335 WL11395D (FDA cleared K800123). Principle: Rigid container with gasketed seal and filter-protected vents for steam penetration. Sterilization: Dynamic air removal (pre-vacuum) steam. Connectivity: None.
Indications for Use
Indicated for healthcare providers to enclose medical devices for dynamic air removal (pre-vacuum) steam sterilization and to maintain sterility during storage for up to 30 days. Suitable for non-porous and porous items.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Predicate Devices
Genesis Reusable Rigid Sterilization Container System (K142529)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 9, 2016
Jewel Precision Sheet Metal and Machine, Inc. c/o Mark Job Regulatory Technology Services, LLC 1394 25th Street, NW Buffalo, MN 55313
Re: K162600
Trade/Device Name: Jewel Precision Reusable Rigid Sterilization Container System Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization wrap Regulatory Class: Class II Product Code: KCT Dated: December 2, 2016 Received: December 5, 2016
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/1/Picture/7 description: The image shows the name "Michael J. Ryan -S" in a large, bold font. The text is black and appears to be centered on a white background. The letters are clearly legible, and the name is presented in a straightforward manner.
for Tina Kiang, Ph.D.
Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162600
Device Name
Jewel Precision Reusable Rigid Sterilization Container System
Indications for Use (Describe)
The Jewel Precision Reusable Rigid Sterilization Container System is a device intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. It allows sterilization of the enclosed medical device and maintains sterility of the enclosed device until used, for a maximum of 30 days.
Jewel Precision Reusable Rigid Sterilization Containers are suitable for dynamic air removal (pre-vacuum) steam sterilization at 270°F (132°C) for 4 minutes with a drying time of 30 minutes.
Reusable containers, covers, filter holders, insert trays, and accessory items such as brackets of various configuration, dividers, caddies, and silicone nipple mats are intended to organize and secure the enclosed medical devices during sterilization and storage of the container.
Filter media allows ingress and egress of sterilant gas while providing a microbial barrier. Filter media is single use only.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K162600 510(k)
## Summary
per 21 CFR 807.92
| Submitter Information | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | Jewel Precision Sheet Metal and Machine, Inc. |
| Address | 200 Commerce Road<br>Cedar Grove, NJ 07009 |
| Phone Number | 973-857-5545 |
| Fax Number | 973-857-5548 |
| Establishment Registration<br>Number | 3004112448 |
| Name of Contact Person | Gary Schneberger, Manager Quality Assurance & Regulatory Affairs |
| Date Prepared | 9-December-2016 |
| Device Information | |
| Trade or Proprietary Name | Jewel Precision Reusable Rigid Sterilization Container System |
| Common or Usual Name | Sterilization Container |
| Classification Name | Sterilization Wrap, Containers, Trays, Cassettes, and Other Accessories |
| Classification Panel | 80 |
| Regulation | Class II per 21CFR 880.6850 |
| Product Code(s) | KCT |
| Legally Marketed Device to<br>which equivalence is claimed | Genesis Reusable Rigid Sterilization Container System<br>K142529 |
| Reason for 510(k) submission | To obtain clearance to market for a new medical device |
| Device Description | The Jewel Precision Reusable Rigid Sterilization Container System is a device<br>intended to be used to enclose another medical device that is to be sterilized by<br>a healthcare provider. It allows sterilization of the enclosed medical device and<br>maintains sterility of the enclosed device until used, for a maximum of 30 days.<br><br>The containers are suitable for dynamic air removal (pre-vacuum) steam<br>sterilization when used as described in the instructions for use.<br><br>Reusable trays and accessory items such as brackets of various configuration,<br>dividers, caddies, and silicone nipple mats are intended to organize and secure<br>the enclosed medical devices during sterilization and storage of the container.<br><br>The Jewel Precision Reusable Rigid Sterilization Container System comprises a<br>sealed sterilization container consisting of a base, cover, filter retainers, single<br>use replaceable paper filters, and accessories consisting of inner trays, and<br>support brackets. Support brackets are configured to securely support container<br>contents, and are secured to inner trays with stainless steel screws and cap<br>nuts. The container is designed to allow efficient exposure of the tray's contents<br>to sterilant gas during the sterilization process, and to contain and protect<br>reusable surgical instruments during sterilization, transport, and storage. The<br>cover incorporates a sealing gasket, and is secured to the base with latches<br>which lock in place. The latching mechanism places a load on the gasket, which<br>is held in compression against the top surface of the container base, maintaining<br>sterile integrity and preventing unwanted separation. The cover also<br>incorporates an alignment flange along its perimeter, which ensures correct<br>alignment of the cover with the container bottom. Provisions are made on the<br>external container surfaces for the use of user provided Sterilizer Load Data<br>Cards and user provided Tamper Evident Locks, through the incorporation of<br>stainless steel Load Data Card Holders and stainless steel basp and pin stude |
| are designed to fit standard autoclaves, and are manufactured of materials<br>capable of withstanding repeated steam sterilization cycles. The containers and<br>accessories are manufactured from materials which are suitable for use with<br>Pulsed Pre-Vacuum Steam Sterilization methods. | |
| Filter media is Ahlstrom Reliance 335 WL11359D paper, FDA cleared under<br>K800123, which allows ingress and egress of sterilant gas while providing a<br>microbial barrier. The Filter media is single use only. | |
| The Jewel Precision Reusable Rigid Sterilization Container System is used to<br>organize, protect, and transport surgical instruments during steam sterilization<br>and subsequent storage. The Jewel Precision Reusable Rigid Sterilization<br>Container System maintains sterility of enclosed instruments, having been<br>validated for a period of 30 days post sterilization, using containers previously<br>subjected to more than 100 sterilization usage cycles. The Jewel Precision<br>Reusable Rigid Sterilization Container System is intended for sterilization of non-<br>porous and porous items, e.g. scissors, hooks, probes, extractors, suction<br>instruments, drivers, rasps, obturators, etc. The container does not come into<br>direct patient contact while in use. | |
{4}------------------------------------------------
The containers, insert trays, covers, brackets, and filters are manufactured from the following materials:
| Material | Specification |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Aluminum Sheet Metal | Alloy 5052 H32 |
| Aluminum Bar Stock | Alloy 6061 |
| Stainless Steel Sheet Metal | SAE 304 |
| Stainless Steel Rod Stock | AISI 303 |
| Nylon Powder Coating | USP Class VI Polyamide 11 |
| Silicone Elastomer | USP Class VI Vinyl Methyl Siloxane,<br>Platinum Cured |
| Filter Paper | Ahlstrom Reliance 335 WL11395D, a Wet<br>Laid Non-Woven Fabric comprised of Natural<br>Wood Pulp Fibers bonded with a Synthetic<br>Resin Binder, having a Basis Weight of 52.2<br>g/m². Cleared by FDA for use as a<br>Sterilization Barrier System under 510(k)<br>Number K800123. |
### Indications For Use
The Jewel Precision Reusable Rigid Sterilization Container System is a device intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. It allows sterilization of the enclosed medical device and maintains sterility of the enclosed device until used, for a maximum of 30 days.
Jewel Precision Reusable Rigid Sterilization Containers are suitable for dynamic air removal (pre-vacuum) steam sterilization at 270 °F (132 °C) for 4 minutes with a drying time of 30 minutes.
Reusable containers, covers, filter holders, insert trays, and accessory items such as brackets of various configuration, dividers, caddles, and silicone nipple mats are intended to organize and secure the enclosed medical devices during sterilization and storage of the container.
Filter media allows ingress and egress of sterilant gas while providing a microbial barrier. Filter media is single use only.
{5}------------------------------------------------
| Summary of the Technological Characteristics of the Device Compared to the Predicate Device | | | |
|--------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| Characteristic | Predicate Device - K142529…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.