← Product Code [KCT](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/KCT) · K150540

# Instru-Safe Instrument Protection System (K150540)

_Summit Medical, Inc. · KCT · Jun 22, 2015 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/KCT/K150540

## Device Facts

- **Applicant:** Summit Medical, Inc.
- **Product Code:** [KCT](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/KCT.md)
- **Decision Date:** Jun 22, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6850
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

Instru-Safe® Instrument Protection System cassettes are used to organize and protect other medical devices that are sterilized by a healthcare provider. Instru-Safe Instrument Protection System cassettes are intended to allow sterilization of the enclosed medical devices during Amsco V-PRO Low Temperature Sterilization Cycles. The Instrument Protection System cassettes are intended to be used in conjunction with a legally marketed wrap or Aesculap rigid container. The Instru-Safe Instrument Protection System cassettes are not intended on their own to maintain sterility. A full list of device models is provided in table 1.

## Device Story

Instrument protection system consisting of perforated cassettes/trays with silicone inserts; used to organize and protect surgical instruments during sterilization, storage, and transport. Cassettes feature rectangular base with latchable cover; perforations allow sterilant penetration. Used in healthcare facilities by sterile processing staff. Device does not maintain sterility independently; must be used with legally marketed wrap or Aesculap rigid container. Validated for specific Amsco V-PRO Low Temperature Sterilization cycles (Standard, Lumen, Non-Lumen). Benefits include organized instrument handling and protection during sterilization cycles.

## Clinical Evidence

Bench testing only. Sterilization validation performed to demonstrate compatibility with Amsco V-PRO Low Temperature Sterilization systems. Biocompatibility testing (MEM Elution Cytotoxicity per ISO 10993-5) confirmed samples are non-cytotoxic.

## Technological Characteristics

Materials: Anodized aluminum, stainless steel, silicone. Form factor: Perforated rectangular base and cover with latching mechanism and internal silicone inserts (hold-it/hold-down). Sterilization: Compatible with Amsco V-PRO Low Temperature Sterilization cycles. No software or electronic components.

## Regulatory Identification

A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.

## Predicate Devices

- Instru-Safe Instrument Protection System ([K133015](/device/K133015.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of three faces in profile, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 22, 2015

Summit Medical Inc. Ms. Nicole Dove Quality Assurance/Regulatory Affairs Manager 815 Northwest Pkwy, Suite 100 St. Paul, MN 55121

Re: K150540

Trade/Device Name: Instru-Safe® Instrument Protection System Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: KCT Dated: May 20, 2015 Received: May 22, 2015

Dear Ms. Dove:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Ms. Dove

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tejashri Purohit-Sheth, M.D.

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K150540

#### Device Name

Instru-Safe® Instrument Protection System

#### Indications for Use (Describe)

Instru-Safe® Instrument Protection System cassettes are used to organize and protect other medical devices that are sterilized by a healthcare provider. Instru-Safe Instrument Protection System cassettes are intended to allow sterilization of the enclosed medical devices during Amsco V-PRO Low Temperature Sterilization Cycles. The Instrument Protection System cassettes are intended to be used in conjunction with a legally marketed wrap or Aesculap rigid container. The Instru-Safe Instrument Protection System cassettes are not intended on their own to maintain sterility. A full list of device models is provided in table 1.

|  |  |  | Amsco V-PRO Low Temperature Sterilization Cycles: |  |  |
|--|--|--|---------------------------------------------------|--|--|
|--|--|--|---------------------------------------------------|--|--|

| Sterilizer   | Standard<br>Cycle | Lumen<br>Cycle | Non Lumen<br>Cycle |
|--------------|-------------------|----------------|--------------------|
| V-PRO 1      | X                 | N/A            | N/A                |
| V-PRO 1 Plus | N/A               | X              | X                  |
| V-PRO maX    | N/A               | X              | X                  |

| Summit Cassette Model | Aesculap Container Model |
|-----------------------|--------------------------|
| IN-8823               | *JM444                   |
| IN-6105               | *JM742                   |

* Validated by Summit Medical for use in Amso V-PRO Low Temperature Sterilization systems ONLY. When using the Aesculap container as a sterile barrier, the load (Summit tray and contents), should not exceed the load claims for the container in weight or load type.

| Lumen size of instrumentation validated includes: |                         |                |                  |
|---------------------------------------------------|-------------------------|----------------|------------------|
| Summit Cassette Model                             | Minimum Inside Diameter | Maximum Length | Number of Lumens |
| IN-8823                                           | 3 mm                    | 400 mm         | 2                |
| IN-6105                                           | 3 mm                    | 200 mm         | 1                |
| IN-2681                                           | 1 mm                    | 64 mm          | 1                |

The worst case validated load by vent-to-volume calculation is the IN-2681 tray.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|X Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Summit Medical. The logo consists of the word "SUMMIT" in a stylized font, with the word "medical" in a smaller, sans-serif font underneath. Above the word "SUMMIT" are three blue diamonds, arranged horizontally. The logo has a clean and modern design.

## Indications for Use Statement

| Table 1 – Device Models |                          |                                              |
|-------------------------|--------------------------|----------------------------------------------|
| Part Number             | Maximum # of Instruments | Estimated Weight - Tray w/ instruments (lbs) |
| IN-1315                 | 30                       | 3.5                                          |
| IN-2006                 | 8                        | 2                                            |
| IN-2681                 | 13                       | 0.64                                         |
| IN-2682                 | 13                       | 0.89                                         |
| IN-2683                 | 13                       | 1.39                                         |
| IN-2840                 | 36                       | 8.75                                         |
| IN-2842                 | 24                       | 6.5                                          |
| IN-2843                 | 36                       | 8.75                                         |
| IN-2880                 | 56                       | 12.1                                         |
| IN-2900                 | 22                       | 4.18                                         |
| IN-2950                 | 12                       | 4.1                                          |
| IN-3030                 | 34                       | 9.5                                          |
| IN-4000                 | 20                       | 3.2                                          |
| IN-4003                 | 30                       | 3.25                                         |
| IN-4010                 | 10                       | 1.98                                         |
| IN-5009                 | 8                        | 5                                            |
| IN-5401                 | 6                        | 2                                            |
| IN-5401-02              | 2                        | 1                                            |
| IN-5401-03              | 2                        | 1                                            |
| IN-5401-08              | 8                        | 3.2                                          |
| IN-5401-12              | 12                       | 3.25                                         |
| IN-6103                 | 2                        | 2.15                                         |
| IN-6105                 | 2                        | 2.15                                         |
| IN-6110                 | 3                        | 2.15                                         |
| IN-6203                 | 2                        | 2.75                                         |
| IN-6205                 | 2                        | 2.75                                         |
| IN-6210                 | 2                        | 2.75                                         |
| IN-6240                 | 2                        | 2.75                                         |
| IN-6303                 | 2                        | 3.28                                         |
| IN-6305                 | 2                        | 3.28                                         |
| IN-6310                 | 2                        | 3.28                                         |
| IN-6403                 | 2                        | 3.28                                         |
| IN-6405                 | 2                        | 3.28                                         |
| IN-6410                 | 2                        | 3.28                                         |
| IN-6500                 | 30                       | 12.5                                         |
| IN-7010                 | 2                        | 2                                            |
| IN-7012                 | 1                        | 1.07                                         |
| IN-7020                 | 30                       | 8.5                                          |
| IN-7030                 | 30                       | 10                                           |
| IN-7032                 | 2                        | 1.1                                          |
| IN-7073                 | 10                       | 5                                            |
| IN-7120                 | 45                       | 11.25                                        |
| IN-7123                 | 45                       | 12                                           |
| IN-7130                 | 45                       | 13.5                                         |
| IN-7140                 | 45                       | 14.5                                         |
| IN-7150                 | 8                        | 1.9                                          |
| IN-7153                 | 6                        | 1.7                                          |
| IN-7220                 | 45                       | 14.5                                         |
| IN-7223                 | 10                       | 9.2                                          |
| IN-7230                 | 45                       | 14.5                                         |
| IN-7234                 | 45                       | 14.5                                         |
| IN-7240                 | 45                       | 14.5                                         |
| IN-7250                 | 45                       | 14.5                                         |
| IN-7252                 | 25                       | 8                                            |
| IN-7260                 | 45                       | 14.5                                         |
| IN-7273                 | 10                       | 6                                            |
| IN-7274                 | 30                       | 8                                            |
| IN-7322                 | 45                       | 14.5                                         |
| IN-7323                 | 45                       | 14.5                                         |
| IN-7343                 | 45                       | 14.5                                         |
| IN-7344                 | 1                        | 4                                            |
| IN-7360                 | 45                       | 14.5                                         |
| IN-7423                 | 45                       | 14.5                                         |
| IN-7452                 | 10                       | 8                                            |
| IN-7453                 | 10                       | 8                                            |
| IN-7540                 | 45                       | 14.5                                         |
| IN-7560                 | 45                       | 14.5                                         |
| IN-7644                 | 45                       | 14.5                                         |
| IN-7723                 | 15                       | 7.18                                         |
| IN-7724                 | 15                       | 7.2                                          |
| IN-7725                 | 10                       | 9.5                                          |
| IN-7781                 | 45                       | 14.5                                         |
| IN-7823                 | 45                       | 14.5                                         |
| IN-7830                 | 45                       | 14.5                                         |
| IN-7840                 | 45                       | 13.5                                         |
| IN-7940                 | 20                       | 13.25                                        |
| IN-8240                 | 20                       | 13.5                                         |
| IN-8610                 | 2                        | 6.65                                         |
| IN-8612                 | 2                        | 6.8                                          |
| IN-8613                 | 2                        | 6.1                                          |
| IN-8615                 | 2                        | 5.8                                          |
| IN-8616                 | 2                        | 5.8                                          |
| IN-8620                 | 3                        | 7.2                                          |
| IN-8621                 | 4                        | 7.18                                         |
| IN-8622                 | 4                        | 7.18                                         |
| IN-8630                 | 3                        | 6.5                                          |
| IN-8632                 | 3                        | 6.45                                         |
| IN-8633                 | 3                        | 6.8                                          |
| IN-8640                 | 4                        | 5.35                                         |
| IN-8642                 | 4                        | 5.35                                         |
| IN-8643                 | 5                        | 5.35                                         |
| IN-8645                 | 4                        | 5.35                                         |
| IN-8650                 | 4                        | 5.85                                         |
| IN-8660                 | 4                        | 5.35                                         |
| IN-8662                 | 4                        | 5.35                                         |
| IN-8663                 | 3                        | 5.5                                          |
| IN-8700                 | 40                       | 14                                           |
| IN-8810                 | 20                       | 13.5                                         |
| IN-8820                 | 15                       | 8.75                                         |
| IN-8823                 | 45                       | 14                                           |
| IN-8830                 | 15                       | 8.75                                         |
| IN-8833                 | 45                       | 14                                           |
| IN-8840                 | 20                       | 13.75                                        |
| IN-8850                 | 15                       | 8.75                                         |
| IN-8853                 | 45                       | 14                                           |
| IN-8860                 | 15                       | 8.75                                         |
| IN-8862                 | 30                       | 10.5                                         |
| IN-8863                 | 45                       | 14                                           |
| IN-8880                 | 2                        | 3.28                                         |
| IN_8882                 | 16                       | 12.1                                         |
| IN-8883                 | 2                        | 3.28                                         |
| IN-8884                 | 4                        | 5.35                                         |
| IN-8885                 | 1                        | 2.25                                         |
| IN-8886                 | 6                        | 12.1                                         |
| IN-8889                 | 6                        | 12.1                                         |
| IN-8891-S               | 1                        | 2                                            |
| IN-8891-SI-12-S         | 1                        | 2                                            |
| IN-8891-SI-85-S         | 1                        | 2                                            |
| IN-8892                 | 12                       | 12.1                                         |
| IN-8893                 | 9                        | 7.5                                          |
| IN-8894                 | 5                        | 16.1                                         |
| IN-8897                 | 8                        | 6                                            |
| IN-8898                 | 10                       | 10.25                                        |
| IN-8899                 | 7                        | 6.5                                          |
| IN-8901                 | 1                        | 2.25                                         |
| IN-8902-G2              | 22                       | 17                                           |
| IN-8903                 | 15                       | 13.25                                        |
| IN-8904                 | 22                       | 17                                           |
| IN-8907                 | 7                        | 12.5                                         |
| IN-8931                 | 1                        | 2.4                                          |
| IN-8932                 | 9                        | 9.5                                          |
| IN-8933                 | 3                        | 3.75                                         |
| IN-8936                 | 6                        | 11.5                                         |
| IN-8937                 | 16                       | 14.5                                         |
| IN-8938                 | 8                        | 12.5                                         |
| IN-8939                 | 10                       | 11.6                                         |
| IN-8940                 | 5                        | 5.18                                         |
| IN-8942                 | 11                       | 10                                           |
| IN-8943                 | 1                        | 2.7                                          |
| IN-8944                 | 6                        | 4.7                                          |
| IN-8945                 | 2                        | 5.18                                         |
| IN-8946                 | 9                        | 6.1                                          |
| IN-8980                 | 20                       | 9.5                                          |
| IN-8982-01              | 17                       | 9.5                                          |
| IN-8983-01              | 16                       | 9.5                                          |
| IN-8984-01              | 15                       | 9.5                                          |
| IN-8986-S               | 2                        | 6.5                                          |
| IN-8987-S               | 2                        | 6.5                                          |
| IN-8988-S               | 2                        | 6                                            |
| IN-8989-S               | 2                        | 6                                            |
| IN-9999-160             | 6                        | 12.1                                         |
| IN-9999-162-S           | 2                        | 5.8                                          |
| IN-9999-168-S           | 2                        | 5.8                                          |
| IN-9999-172-S           | 2                        | 5.8                                          |
| IN-9999-178-S           | 2                        | 5.8                                          |

Summit Medical, Inc. | 815 Northwest Parkway, Suite 100 | St. Paul, MN 55121 USA phone: 651.789.3939 | fax: 651.789.3979 www.summitmedicalusa.com

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Summit Medical. The logo consists of the word "SUMMIT" in a stylized font, with the word "medical" in a smaller font below it. Above the word "SUMMIT" are three blue diamonds arranged horizontally. The logo is simple and modern, and the blue color gives it a professional look.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Summit Medical. The logo consists of the word "SUMMIT" in a stylized font, with the word "medical" in a smaller font underneath. Above the word "SUMMIT" are three blue diamonds arranged horizontally. The logo is simple and modern, and the blue color gives it a professional look.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Summit Medical. The logo consists of the word "SUMMIT" in a sans-serif font, with the word "medical" in a smaller font below it. Above the word "SUMMIT" are three blue diamonds, arranged horizontally. The logo is simple and modern, and the blue color gives it a professional look.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Summit Medical. The logo consists of the word "SUMMIT" in a sans-serif font, with the word "medical" in a smaller font below it. Above the word "SUMMIT" are three blue diamonds, arranged horizontally. The logo is simple and modern, and the use of blue gives it a professional and trustworthy feel.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Summit Medical. The logo consists of the word "SUMMIT" in a bold, sans-serif font, with the word "medical" in a smaller font below it. Above the word "SUMMIT" are three blue diamonds, arranged horizontally. The logo is simple and modern, and the use of blue suggests trust and reliability.

# 510(k) Summary

Following is a summary of 510(k) safety and effectiveness information in accordance with 21 CFR 807.92

| Submitter:             | Summit Medical Inc.<br>815 Northwest Parkway, Suite 100<br>St. Paul, MN 55121<br>Tel: (651) 789-3939                                                                                                                                                                                                                                                                       |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ER Number:             | 3008719017                                                                                                                                                                                                                                                                                                                                                                 |
| Contact<br>Person:     | Nicole Dove<br>QA/RA Manager Tel: (651)<br>789-3921<br>ndove@summitmedicalusa.com                                                                                                                                                                                                                                                                                          |
| Date Prepared:         | June 22, 2015                                                                                                                                                                                                                                                                                                                                                              |
| Subject<br>Device:     | Trade Name(s):<br>Instru-Safe® Instrument Protection System<br>Classification Name:<br>Sterilization wrap containers, trays, cassettes & other accessory (21 CFR 880.6850<br>Common Name:<br>Instrument Tray, Sterilization Tray, Sterilization Cassettes, Instrument Delivery<br>System<br>Device Class:<br>Class II<br>Device Code:<br>KCT<br>Panel:<br>General Hospital |
| Predicate<br>Device:   | Tradename: Instru-Safe Instrument Protection System<br>510(k) Holder: Summit Medical Inc.<br>510(k) #: K133015                                                                                                                                                                                                                                                             |
| Device<br>Description: | Summit Medical Inc. Instru-Safe Instrument Protection System are cassettes / trays used to enclose and hold surgical instruments and accessories in an organized manner during the sterilization process and subsequent storage and transportation. The cassettes / trays by themselves do not maintain sterility.                                                         |

Page 1 of 11

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for Summit Medical. The logo consists of three blue diamonds arranged horizontally above the word "SUMMIT" in a stylized font. Below "SUMMIT" is the word "medical" in a smaller, sans-serif font. The overall design is clean and professional, suggesting a healthcare or medical-related organization.

| The cassettes / trays are different sizes of the same basic configuration: a rectangular base with latchable cover. The cassettes / trays have perforations to allow sterilant penetration. The cassettes / trays contain silicone inserts in the base and/or cover to hold, organize and protect the surgical instruments within the cassette / tray.                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                |                  |                     |  |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|------------------|---------------------|--|
| Intended Use:                                                                                                                                                                                                                                                                                                                                                                                                    | Instru-Safe ® Instrument Protection System cassettes are used to organize and protect other medical devices that are sterilized by a healthcare provider. Instru-Safe Instrument Protection System cassettes are intended to allow sterilization of the enclosed medical devices during Amsco V-PRO Low Temperature Sterilization Cycles. The Instru-Safe Instrument Protection System cassettes are intended to be used in conjunction with legally marketed wrap or Aesculap rigid container. The Instru-Safe Instrument Protection System cassettes are not intended on their own to maintain sterility. A full list of device models is provided in table 1. |                |                  |                     |  |
| Amsco V-PRO Low Temperature Sterilization Systems                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                |                  |                     |  |
| Sterilizer                                                                                                                                                                                                                                                                                                                                                                                                       | Standard Cycle                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Lumen Cycle    | Non Lumen Cycle  |                     |  |
| V-PRO 1                                                                                                                                                                                                                                                                                                                                                                                                          | X                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | N/A            | N/A              |                     |  |
| V-PRO 1 Plus                                                                                                                                                                                                                                                                                                                                                                                                     | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | X              | X                |                     |  |
| V-PRO maX                                                                                                                                                                                                                                                                                                                                                                                                        | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | X              | X                |                     |  |
| Summit Cassette Model                                                                                                                                                                                                                                                                                                                                                                                            | Aesculap Container Model                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                |                  |                     |  |
| IN-8823                                                                                                                                                                                                                                                                                                                                                                                                          | *JM444                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                |                  |                     |  |
| IN-6105                                                                                                                                                                                                                                                                                                                                                                                                          | *JM742                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                |                  |                     |  |
| *Validated by Summit Medical for use in Amsco V-PRO Low Temperature Sterilization Systems ONLY. When using the Aesculap container as a sterile barrier, the load (Summit tray and contents), should not exceed the load claims for the container in weight or load type.                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                |                  |                     |  |
| Lumen size of instrumentation validated includes:                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                |                  |                     |  |
| Summit Cassette Model                                                                                                                                                                                                                                                                                                                                                                                            | Minimum Inside Diameter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Maximum Length | Number of Lumens |                     |  |
| IN-8823                                                                                                                                                                                                                                                                                                                                                                                                          | 3mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 400mm          | 2                |                     |  |
| IN-6105                                                                                                                                                                                                                                                                                                                                                                                                          | 3mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 200mm          | 1                |                     |  |
| IN-2681                                                                                                                                                                                                                                                                                                                                                                                                          | 1mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 64mm           | 1                |                     |  |
| The worst case validated load by vent-to-volume calculation is the IN-2681 tray. The intended use of the subject device includes the Amsco V-PRO Low Temperature Sterilization Systems. Performance testing has been completed for the Amsco V-PRO Low Temperature Sterilization Systems. These new sterilization cycles do not affect safety and effectiveness of the Instru-Safe Instrument Protection System. |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                |                  |                     |  |
| Comparison of Characteristics to Predicate Device:                                                                                                                                                                                                                                                                                                                                                               | Based on a comparison of the design, technology, materials, manufacturing, performance, specifications and methods of use, the Instru-Safe Instrument Protection System is equivalent to the identified 510(k) cleared predicate device.                                                                                                                                                                                                                                                                                                                                                                                                                         |                |                  |                     |  |
| Element                                                                                                                                                                                                                                                                                                                                                                                                          | New Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                |                  | Predicate (K133015) |  |

Summit Medical, Inc. | 815 Northwest Parkway, Suite 100 | St. Paul, MN 55121 USA phone: 651.789.3939 | fax: 651.789.3979 www.summitmedicalusa.com

Page 2 of 11

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for Summit Medical. The logo consists of the word "SUMMIT" in a stylized font, with the word "medical" in a smaller font below it. Above the word "SUMMIT" are three blue diamonds. The logo is simple and modern, and the use of blue gives it a professional look.

| Intended Use                                                                                                                                                                                                                                                             | Instru-Safe® Instrument Protection System cassettes are used to organize and protect other medical devices that are sterilized by a healthcare provider. Instru-Safe Instrument Protection System cassettes are intended to allow sterilization of the enclosed medical devices during Amsco V-PRO Low Temperature Sterilization Cycles. The Instru-Safe Instrument Protection System cassettes are intended to be used in conjunction with legally marketed wrap or Aesculap rigid container. The Instru-Safe Instrument Protection System cassettes are not intended on their own to maintain sterility. A full list of device models is provided in table 1. |                                                                                                                                                                                                                                                         |                 |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|
|                                                                                                                                                                                                                                                                          | Instru-Safe Instrument Protection System cassettes are used to organize and protect other medical devices that are sterilized by a healthcare provider. Instru-Safe System cassettes are intended to allow sterilization of the enclosed medical devices during a pre-vacuum steam sterilization cycle. The Instru-Safe System cassettes are intended to be used in conjunction with legally marketed wrap or Aesculap rigid container. The Instru-Safe System cassettes are not intended on their own to maintain sterility.                                                                                                                                   |                                                                                                                                                                                                                                                         |                 |
|                                                                                                                                                                                                                                                                          | Sterilization methods and configurations                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Sterilization methods and configurations                                                                                                                                                                                                                |                 |
|                                                                                                                                                                                                                                                                          | Amsco V-PRO Low Temperature Sterilization Systems                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | • Autoclave Sterilization Parameter:<br>Cycle: Pre-vacuum<br>Temperature: 270°F (132°C)<br>Exposure Time: 4 minutes<br>Minimum Dry Time: 30 minutes                                                                                                     |                 |
| Sterilizer                                                                                                                                                                                                                                                               | Standard Cycle                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Lumen Cycle                                                                                                                                                                                                                                             | Non Lumen Cycle |
| V-PRO 1                                                                                                                                                                                                                                                                  | X                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | N/A                                                                                                                                                                                                                                                     | N/A             |
| V-PRO 1 Plus                                                                                                                                                                                                                                                             | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | X                                                                                                                                                                                                                                                       | X               |
| V-PRO MAX                                                                                                                                                                                                                                                                | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | X                                                                                                                                                                                                                                                       | X               |
| Summit Cassette Model                                                                                                                                                                                                                                                    | Aesculap Container Model                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                         |                 |
| IN-8823-AE                                                                                                                                                                                                                                                               | *JN444                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                         |                 |
| IN-2880                                                                                                                                                                                                                                                                  | *JK444                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                         |                 |
| IN-6105                                                                                                                                                                                                                                                                  | *JN742                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                         |                 |
| IN-8823                                                                                                                                                                                                                                                                  | *JM444                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                         |                 |
| IN-6105                                                                                                                                                                                                                                                                  | *JM742                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                         |                 |
| *Validated by Summit Medical for use in Amsco V-PRO Low Temperature Sterilization Systems ONLY. When using the Aesculap container as a sterile barrier, the load (Summit tray and contents), should not exceed the load claims for the container in weight or load type. |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | *Validated by Summit Medical for use in steam prevacuum sterilizers ONLY operating at 270°F (132°C) for 4 minutes exposure time. Consult container instructions to ensure that contents do not exceed the sterilization container intended load claims. |                 |

Page 3 of 11

{12}------------------------------------------------

| Lumen size of instrumentation validated includes:                                    |                                                                                                                                                                                                                                                              |                   |                                                                                                                                                                                                       |
|--------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Summit<br>Cassette<br>Model                                                          | Minimum<br>Inside<br>Diameter                                                                                                                                                                                                                                | Maximum<br>Length | Number of<br>Lumens                                                                                                                                                                                   |
| IN-8823                                                                              | 3mm                                                                                                                                                                                                                                                          | 400mm             | 2                                                                                                                                                                                                     |
| IN-6105                                                                              | 3mm                                                                                                                                                                                                                                                          | 200mm             | 1                                                                                                                                                                                                     |
| IN-2681                                                                              | 1mm                                                                                                                                                                                                                                                          | 64mm              | 1                                                                                                                                                                                                     |
| The worst case validated load by vent-to- volume calculation<br>is the IN-2681 tray. |                                                                                                                                                                                                                                                              |                   |                                                                                                                                                                                                       |
| Material<br>Composition                                                              | No changes from predicate device                                                                                                                                                                                                                             |                   | The cassette contains<br>components made<br>of anodized aluminum,<br>stainless steel, blue                                                                                                            |
| Physical<br>Properties                                                               | Instru-Safe Instrument Protection<br>System cassettes include<br>- perforated base<br>- perforated cover<br>- silicone inserts (hold-it / hold down)<br>- Handles<br>- Latches<br>- Feet<br>- Posts (optional)<br>- Divider (optional)<br>- Shelf (optional) |                   | Instru-Safe Instrument<br>Protection<br>System cassettes include<br>- perforated base<br>- perforated cover<br>- silicone<br>inserts (hold-<br>it / hold<br>down)<br>- Handles<br>- Latches<br>- Feet |
| Chemical<br>Properties                                                               | Not Applicable                                                                                                                                                                                                                                               |                   | Not Applicable                                                                                                                                                                                        |
| Configurations<br>/ Dimensions                                                       | Various configurations / dimensions                                                                                                                                                                                                                          |                   | See table located in<br>predicate device                                                                                                                                                              |
| Air permeance                                                                        | Not Applicable                                                                                                                                                                                                                                               |                   | Not Applicable                                                                                                                                                                                        |

Summit Medical, Inc. | 815 Northwest Parkway, Suite 100 | St. Paul, MN 55121 USA phone: 651.789.3939 | fax: 651.789.3979 www.summitmedicalusa.com

Page 4 of 11

{13}------------------------------------------------

| Percent of<br>surface<br>performations | Not Applicable                         | Not Applicable            |
|----------------------------------------|----------------------------------------|---------------------------|
| Performance                            | New Device                             | Predicate                 |
| Sterilant                              | Amsco V-PRO Low Temperature            | Pre-Vacuum Steam          |
| Penetration                            | Sterilization Systems.                 | Cycle: Pre-               |
|                                        |                                        | vacuum                    |
|…

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/KCT/K150540](https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/KCT/K150540)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
