K142109 · Sterilucent, Inc. · KCT · Jan 20, 2015 · General Hospital
Device Facts
Record ID
K142109
Device Name
STERILUCENT STERILIZATION CONTAINER SYSTEM
Applicant
Sterilucent, Inc.
Product Code
KCT · General Hospital
Decision Date
Jan 20, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The Sterilucent Sterilization Container System is a device intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used. The container system is intended to be used in the Sterilucent PSD-85 Hydrogen Peroxide Sterilizer Lumen and Non-Lumen Cycles. The sterility maintenance period has been validated for a period of up to 180 days. Reusable baskets and accessory items (clips, posts, pins, dividers, brackets, bars and mats) are intended to organize and secure enclosed medical devices during sterilization, transport, and storage of the container. Consumable accessory items (filter media, data card and tamper evident arrows) provide a range of functions and are indicated as single-use devices. Filter media allows ingress and egress of sterilant while providing a microbial barrier. Data cards are used to record information regarding a specific sterilization process load. Tamper evident arrows provide a visual indication that the container system has not been inadvertently opened prior to use. Each arrow contains an external process indicator that serves as a visual indication that the system has been exposed to the Sterilucent VHP sterilization process.
Device Story
Rigid, reusable, stackable container system; houses surgical instrumentation in baskets during sterilization, transport, and storage. Input: medical devices placed inside container; container processed in Sterilucent PSD-85 Hydrogen Peroxide Sterilizer. Transformation: container design allows sterilant ingress/egress via filter media while maintaining microbial barrier; tamper-evident arrows and process indicators provide visual confirmation of sterilization exposure. Output: sterilized, protected medical devices. Used in clinical settings by healthcare staff. Benefits: ensures sterility maintenance of instruments for up to 180 days post-processing; provides organized, secure transport and storage.
Clinical Evidence
No clinical data. Bench testing only: 12-log reduction and SAL of 10^-6 achieved via BI overkill method; 180-day shelf-life validation; aerosol microbial challenge (1x10^6 CFU Bacillus atrophaeus) confirmed barrier integrity.
Technological Characteristics
Materials: 5000/1100 series anodized aluminum, 300/400 series stainless steel, silicone elastomer gaskets/mats, SMS polypropylene filters, HDPE data cards. Principle: Rigid container with filter-based gas exchange for VHP sterilization. Dimensions: Various sizes (Mini to Full Length). Connectivity: None. Sterilization: Vaporized Hydrogen Peroxide (VHP).
Indications for Use
Indicated for use by healthcare providers to enclose medical devices for sterilization in the Sterilucent PSD-85 Hydrogen Peroxide Sterilizer (Lumen and Non-Lumen cycles) and to maintain sterility of the enclosed devices for up to 180 days.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Predicate Devices
Genesis™ Reusable Rigid Container System (K112535)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of three human profiles facing right, arranged in a way that they appear to be interconnected. The profiles are rendered in black and have a flowing, abstract design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the profile graphic.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 20, 2015
Sterilucent, Inc. Peter Kalkbrenner Director of Engineering 1400 Marshall Street NE Minneapolis, MN 55413
Re: K142109
Trade/Device Name: Sterilucent Sterilization Container System Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Container Regulatory Class: II Product Code: KCT Dated: December 19, 2014 Received: December 22, 2014
Dear Mr. Kalkbrenner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K142109
Device Name Sterilucent Sterilization Container System
#### Indications for Use (Describe)
The Sterilucent Sterilization Container System is a device intended to enclose another medical device that is to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used. This container system is intended to be used in the Sterilucent PSD-85 Hydrogen Peroxide Sterilizer Lumen Cycles. The sterility maintenance period has been validated for a period of up to 180 days.
Reusable baskets and accessory items (clips, posts, pins, dividers, bars and mats) are intended to organize and secure enclosed medical devices during sterilization, transport, and storage of the container.
Consumable accessory items (filter media, data card and tamper evident arrows) provide a range of functions and are indicated as single-use devices. Filter media allows ingress of sterilant while providing a microbial barrier. Data cards are used to record information regarding a specific sterilization process load. Tamper evident arrows provide a visual indication that the container system has not been inadvertently opened prior to use. Each arrow contains an external process indicator that serves as a visual indication that the system has been exposed to the Sterilization process.
(Continued on separate page.)
Type of Use (Select one or both, as applicable)
- Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
Indications for Use Continuation page (Form FDA 3881 (1/14)
| Container | Basket | Contents/Configuration | Sterilization Cycle | |
|-----------|-----------|-----------------------------------------------------|-------------------------|-------------------------|
| Catalog # | Catalog # | | Lumen<br>Cycle | Non-Lumen<br>Cycle |
| SL065 | SL011 | Baskets | YES | YES |
| SL066 | SL012 | | | |
| SL067 | SL013 | | | |
| SL068 | SL014 | Stacking Baskets | YES | YES |
| SL069 | SL015 | Lumen: 1mm (ID) or larger x 60mm (L) or<br>shorter | YES<br>Qty. 10 | NO |
| SL070 | SL016 | | | |
| SL071 | SL017 | | | |
| SL072 | SL018 | | | |
| SL073 | SL019 | Lumen: 2mm (ID) or larger x 250mm (L) or<br>shorter | YES<br>Qty. 10 | NO |
| SL074 | SL020 | | | |
| SL075 | SL021 | | | |
| SL076 | SL022 | | | |
| SL076 | SL023 | Lumen: 3mm (ID) or larger x 350mm (L) or<br>shorter | YES<br>Qty. 10 | NO |
| SL076 | SL023 | | | |
| SL077 | SL024 | Occluded/Mated Challenge | YES | YES |
| SL078 | SL025 | | | |
| SL098 | SL026 | | | |
| SL098 | SL027 | Silicone Support Bars | YES | YES |
| SL098 | SL028 | Dividers | YES | YES |
| SL098 | SL029 | | | |
| SL098 | SL030 | | | |
| SL098 | SL031 | Brackets | YES | YES |
| SL098 | SL032 | | | |
| SL098 | SL033 | | | |
| SL098 | SL034 | Silicone Mat | YES | YES |
| SL098 | SL035 | | | |
| SL098 | SL036 | | | |
| SL098 | SL037 | Filter | SL082<br>SL083<br>SL084 | SL082<br>SL083<br>SL084 |
| SL098 | SL038 | | | |
| SL098 | SL039 | | | |
| SL098 | SL040 | Data Card | SL081 | SL081 |
| SL098 | SL041 | | | |
| SL098 | SL042 | | | |
| SL098 | SL043 | Tamper-Evident Arrow | SL085 | SL085 |
| SL098 | SL044 | | | |
| SL098 | SL045 | | | |
| SL098 | SL046 | Stack Height | NO STACKING | NO STACKING |
| SL098 | SL047 | | | |
| SL098 | SL048 | | | |
| SL098 | SL049 | Max. Total Container System Weight | 10 lbs<br>(4.5 kg) | 25 lbs<br>(11.3 kg) |
| SL098 | SL050 | | | |
## Container and Accessory Device Challenges
{4}------------------------------------------------
| Catalog<br>Code | Container Description | External Dimensions | Total Container Plus Load<br>Weight | |
|-----------------|-----------------------|----------------------|-------------------------------------|---------------------|
| | | | Lumen Cycle¹ | Non-Lumen² |
| SL065 | Mini | 10.2" x 7.2" x 3.2" | 5 lbs<br>(2.3 kg) | 5 lbs<br>(2.3 kg) |
| SL066 | Quarter Length | 9.5" x 12.4" x 3.8" | 8 lbs<br>(3.3 kg) | 8 lbs<br>(3.3 kg) |
| SL067 | Half Length, 4" Deep | 11.8" x 12.4" x 4.5" | 10 lbs<br>(4.5 kg) | 12 lbs³<br>(5.4 kg) |
| SL068 | Half Length, 5" Deep | 11.8" x 12.4" x 5.3" | 10 lbs<br>(4.5 kg) | 12 lbs³<br>(5.4 kg) |
| SL069 | Half Length, 6" Deep | 11.8" x 12.4" x 6.1" | 10 lbs<br>(4.5 kg) | 12 lbs³<br>(5.4 kg) |
| SL070 | Mid Length, 4" Deep | 19.2" x 12.4" x 4.5" | 10 lbs<br>(4.5 kg) | 20 lbs<br>(9.1 kg) |
| SL071 | Mid Length, 5" Deep | 19.2" x 12.4" x 5.3" | 10 lbs<br>(4.5 kg) | 20 lbs<br>(9.1 kg) |
| SL098 | Mid Length, 6" Deep | 19.2" x 12.4" x 6.1" | 10 lbs<br>(4.5 kg) | 20 lbs<br>(9.1 kg) |
| SL072 | Mid Length, 8" Deep | 19.2" x 12.4" x 7.8" | 10 lbs<br>(4.5 kg) | 20 lbs<br>(9.1 kg) |
| SL073 | Full Length, 4" Deep | 23.1" x 12.4" x 4.5" | 10 lbs<br>(4.5 kg) | 25 lbs<br>(11.3kg) |
| SL074 | Full Length, 5" Deep | 23.1" x 12.4" x 5.3" | 10 lbs<br>(4.5 kg) | 25 lbs<br>(11.3 kg) |
| SL075 | Full Length, 6" Deep | 23.1" x 12.4" x 6.1" | 10 lbs<br>(4.5 kg) | 25 lbs<br>(11.3kg) |
| SL076 | Full Length, 7" Deep | 23.1" x 12.4" x 7.0" | 10 lbs<br>(4.5 kg) | 25 lbs<br>(11.3kg) |
| SL077 | Small Narrow, 3" Deep | 20.8" x 7.3" x 3.9" | 10 lbs<br>(4.5 kg) | 10 lbs<br>(4.5 kg) |
| SL078 | Small Narrow, 5" Deep | 20.8" x 7.3" x 5.2" | 10 lbs<br>(4.5 kg) | 10 lbs<br>(4.5 kg) |
Maximum Container Loads
1 The Lumen Cycle validation testing was conducted using a maximum of ten (10) lumens per load. The validation studies were performed using the following device models and weights:
| Model | SL065 | SL066 | SL069 | SL072 | SL076 | SL078 |
|-------------------------------|-------|-------|-------|-------|-------|-------|
| Validation Load Weight (lbs.) | 5.1 | 8.0 | 10.3 | 14.0 | 17.1 | 10.1 |
2 The Non-Lumen Cycle validation studies were performed using the following device models and weights:
| | Model | SL065 | SL066 | SL069 | SL072 | SL076 | SL078 |
|-------------------------------|-------|-------|-------|-------|-------|-------|-------|
| Validation Load Weight (lbs.) | 5.0 | 8.0 | 12.1 | 20.1 | 25.1 | 10.1 | |
3 Two (2) containers, each containing 12 lbs. (5.4 kg), for a total chamber load weight of 24 lbs. (10.8 kg), has also been validated for use in the PSD-85 Non-Lumen Cycle.
{5}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Sterilucent. The logo consists of a blue abstract symbol to the left of the word "sterilucent" in blue, sans-serif font. Below the word "sterilucent" is the phrase "scientific sterilization solutions" in a smaller, blue, sans-serif font. The background of the image is light blue.
# 510(k) Summary for the Sterilucent Sterilization Container System K142109
| Owner: | |
|----------|--|
| Address: | |
Telephone:
Sterilucent, Inc. 1400 Marshall Street NE Minneapolis, MN 55413 612-767-3260 612-767-3261
Contact:
Fax:
Peter R. Kalkbrenner Director of Engineering
612-767-3253
612-767-3261
Telephone: Fax:
Summary Date: 16 January 2015
{7}------------------------------------------------
## 1. Device Name and Classification
| Trade Name: | Sterilucent Sterilization Container System |
|----------------------|-------------------------------------------------------------------------|
| Common/Usual Name: | Sterilization Container |
| Classification Name: | Sterilization Wrap, Containers, Trays, Cassettes & Other<br>Accessories |
| Product Code: | KCT (21 CFR 880.6850) |
| Class: | II |
## 2. Predicate Device
Genesis™ Reusable Rigid Container System (K112535)
## 3. Device Description
The Sterilucent Sterilization Container System is an assortment of rigid, reusable, stackable containers that are used to sterilize other medical devices and maintain sterility of those devices until used. The container system is comprised of a lid, bottom, filter, tamper evident arrows, and data cards.
The container system houses reusable baskets of varying depths and organizing accessory items that are used to organize and secure surgical instrumentation and/or other medical devices.
## 4. Statement of Intended Use:
The Sterilucent Sterilization Container System is a device intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used. The container system is intended to be used in the Sterilucent PSD-85 Hydrogen Peroxide Sterilizer Lumen and Non-Lumen Cycles. The sterility maintenance period has been validated for a period of up to 180 days.
Reusable baskets and accessory items (clips, posts, pins, dividers, brackets, bars and mats) are intended to organize and secure enclosed medical devices during sterilization, transport, and storage of the container.
Consumable accessory items (filter media, data card and tamper evident arrows) provide a range of functions and are indicated as single-use devices. Filter media allows ingress and egress of sterilant while providing a microbial barrier. Data cards are used to record information regarding a specific sterilization process load. Tamper evident arrows provide a
{8}------------------------------------------------
visual indication that the container system has not been inadvertently opened prior to use. Each arrow contains an external process indicator that serves as a visual indication that the system has been exposed to the Sterilucent VHP sterilization process.
| Container | Basket | | Sterilization Cycle | | |
|-------------------------|-------------------------------------------------------------------------------------------------|------------------------------------------------------|---------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|--|
| Catalog # | Catalog # | Contents/Configuration | Lumen | Non-Lumen | |
| | | | Cycle | Cycle | |
| ട് സ്കാരം<br>SLOE6 | SL011<br>SL012 | Baskets | YES | YES | |
| SL067<br>SL068 | SL013<br>SL014 | Stacking Baskets | YES | YES | |
| SLO69<br>SL070<br>SL071 | SL015<br>SL016<br>SL017 | Lumen: 1mm (ID) or larger x 60mm (L) or<br>shorter | YES<br>Qty. 10 | NO | |
| SL072<br>SL073 | SL018<br>SL019<br>SL020 | Lumen: 2mm (ID) or larger x 250mm (L) or<br>shorter | YES<br>Qty. 10 | NO | |
| SL074<br>SL075<br>SL076 | SL021<br>SL022<br>SL023 | Lumen: 3mm (ID) or larger x 350mm (L) or<br>shorter | YES<br>Qty. 10 | NO | |
| SL077<br>SL078 | SL024<br>SL025 | Occluded/Mated Challenge | YES | YES | |
| SL098 | SL026<br>SL027 | Silicone Support Bars | YES | YES | |
| | SL028<br>SL029 | Dividers | YES | YES | |
| | SL030<br>SL031 | Brackets | YES | YES | |
| | SL032<br>SL033 | Silicone Mat | YES | YES | |
| | SL034<br>SL035<br>SL036<br>SL037 | Filter | SL082<br>SL083<br>SL084 | SL082<br>SL083<br>SL084 | |
| | SL038<br>SL039 | Data Card | SL081 | SL081 | |
| | SL040<br>SL041<br>SL042<br>SL043<br>SL044<br>SL045<br>SL046<br>SL047<br>SL048<br>SL049<br>SL050 | Tamper-Evident Arrow | ട് സ്റ്റാറ്റ് ട്രോഗ്യമായി സ്റ്റാമിക്കുന്നു. അവലംബം കോട്ടിക്കുന്നു വിവ്യാവിക്കുന്നു. അവലംബം കേരള | ട് സ്റ്റാറ്റ് ട്രോഗ്യമായി സ്റ്റാമിക്കുന്നു. അവലംബം കോട്ടിക്കുന്നു കേരള | |
| | | Stack Height | NO STACKING | NO STACKING | |
| | | Max. Total Container System Weight | 10 lbs<br>(4.5 kg) | 25 lbs<br>(11.3 kg) | |
# Container and Accessory Device Challenges
{9}------------------------------------------------
| Catalog<br>Code | Container Description | External Dimensions | Total Container Plus Load<br>Weight | |
|-----------------|-----------------------|----------------------|-------------------------------------|---------------------|
| | | | Lumen Cycle¹ | Non-Lumen² |
| SL065 | Mini | 10.2" x 7.2" x 3.2" | 5 lbs<br>(2.3 kg) | 5 lbs<br>(2.3 kg) |
| SL066 | Quarter Length | 9.5" x 12.4" x 3.8" | 8 lbs<br>(3.3 kg) | 8 lbs<br>(3.3 kg) |
| SL067 | Half Length, 4" Deep | 11.8" x 12.4" x 4.5" | 10 lbs<br>(4.5 kg) | 12 lbs³<br>(5.4 kg) |
| SL068 | Half Length, 5" Deep | 11.8" x 12.4" x 5.3" | 10 lbs<br>(4.5 kg) | 12 lbs³<br>(5.4 kg) |
| SL069 | Half Length, 6" Deep | 11.8" x 12.4" x 6.1" | 10 lbs<br>(4.5 kg) | 12 lbs³<br>(5.4 kg) |
| SL070 | Mid Length, 4" Deep | 19.2" x 12.4" x 4.5" | 10 lbs<br>(4.5 kg) | 20 lbs<br>(9.1 kg) |
| SL071 | Mid Length, 5" Deep | 19.2" x 12.4" x 5.3" | 10 lbs<br>(4.5 kg) | 20 lbs<br>(9.1 kg) |
| SL098 | Mid Length, 6" Deep | 19.2" x 12.4" x 6.1" | 10 lbs<br>(4.5 kg) | 20 lbs<br>(9.1 kg) |
| SL072 | Mid Length, 8" Deep | 19.2" x 12.4" x 7.8" | 10 lbs<br>(4.5 kg) | 20 lbs<br>(9.1 kg) |
| SL073 | Full Length, 4" Deep | 23.1" x 12.4" x 4.5" | 10 lbs<br>(4.5 kg) | 25 lbs<br>(11.3 kg) |
| SL074 | Full Length, 5" Deep | 23.1" x 12.4" x 5.3" | 10 lbs<br>(4.5 kg) | 25 lbs<br>(11.3 kg) |
| SL075 | Full Length, 6" Deep | 23.1" x 12.4" x 6.1" | 10 lbs<br>(4.5 kg) | 25 lbs<br>(11.3 kg) |
| SL076 | Full Length, 7" Deep | 23.1" x 12.4" x 7.0" | 10 lbs<br>(4.5 kg) | 25 lbs<br>(11.3 kg) |
| SL077 | Small Narrow, 3" Deep | 20.8" x 7.3" x 3.9" | 10 lbs<br>(4.5 kg) | 10 lbs<br>(4.5 kg) |
| SL078 | Small Narrow, 5" Deep | 20.8" x 7.3" x 5.2" | 10 lbs<br>(4.5 kg) | 10 lbs<br>(4.5 kg) |
Maximum Container Loads
1 The Lumen Cycle validation testing was conducted using a maximum of ten (10) lumens per load. The validation studies were performed using the following device models and weights:
| Model | SL065 | SL066 | SL069 | SL072 | SL076 | SL078 |
|-------------------------------|-------|-------|-------|-------|-------|-------|
| Validation Load Weight (lbs.) | 5.1 | 8.0 | 10.3 | 14.0 | 17.1 | 10.1 |
2 The Non-Lumen Cycle validation studies were performed using the following device models and weights:
| Model | SL065 | SL066 | SL069 | SL072 | SL076 | SL078 |
|-------------------------------|-------|-------|-------|-------|-------|-------|
| Validation Load Weight (lbs.) | 5.0 | 8.0 | 12.1 | 20.1 | 25.1 | 10.1 |
3 Two (2) containers, each containing 12 lbs. (5.4 kg), for a total chamber load weight of 24 lbs. (10.8 kg), has also been validated for use in the PSD-85 Non-Lumen Cycle.
{10}------------------------------------------------
# 5. Technological Characteristics Summary Comparison
The Sterilucent Sterilization Container System has similar technological characteristics as the predicate device:
| Summary of Technological Characteristics of the Device Compared to the | | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Predicate Device | | |
| Characteristic | New Device<br>Sterilucent Sterilization<br>Container System<br>K142109 | Predicate Device<br>Genesis™ Reusable Rigid<br>Container System<br>K112535 |
| Container | 5000 and 1100 Series Anodized<br>Aluminum;<br>300 Series Stainless Steel | Same |
| Container Gasket | Closed Cell Silicone Foam | Same |
| Reusable Baskets | 304 Electropolished Stainless Steel | Same |
| Clips, Posts, Pins | 300 and 400 Series Stainless Steel | Same |
| Dividers, Brackets | 5000 Series Aluminum | Same |
| Bars, Mats | Silicone Elastomer | Same |
| Filter Material | SMS Polypropylene | Same |
| Data Card | High-Density Polyethylene (Tyvek®) | Same |
| Tamper Evident Arrow | Polypropylene | Same |
| Arrow Process Indicator | Synthetic Substrate printed with Reactive<br>Ink | Same |
| Sterilization Modality | Vaporized Hydrogen Peroxide (VHP) | Same (some models also<br>approved for steam and<br>ETO) |
| Indications for Use | The Sterilucent Sterilization Container<br>System is a device intended to be used to<br>enclose another medical device that is to<br>be sterilized by a healthcare provider. It<br>is intended to allow sterilization of the<br>enclosed medical device and also to<br>maintain sterility of the enclosed device<br>until used. The container system is<br>intended to be used in the Sterilucent<br>PSD-85 Hydrogen Peroxide Sterilizer<br>Lumen and Non-Lumen Cycles.<br>Reusable baskets and accessory items<br>(clips, posts, pins, dividers, brackets, bars<br>and mats) are intended to organize and<br>secure enclosed medical devices during<br>sterilization, transport, and storage of the<br>container. | The Genesis Reusable<br>Rigid Sterilization<br>Container System is a<br>device intended to be used<br>to enclose another medical<br>device that is to be<br>sterilized by a healthcare<br>provider. It allows<br>sterilization of the<br>enclosed medical device<br>and maintains sterility of<br>the enclosed device until<br>used for a maximum of<br>180 days.<br>Containers are suitable for<br>dynamic air removal (pre-<br>vacuum) steam |
| | | |
| Consumable accessory items (e.g. filter<br>media, data card and tamper evident<br>arrows) provide a range of functions and<br>are indicated as single-use devices. Filter<br>media allows ingress and egress of<br>sterilant while providing a microbial<br>barrier. Data cards are used to record<br>information regarding a specific<br>sterilization process load. Tamper evident<br>arrows provide a visual indication that<br>the container system has not been<br>inadvertently opened prior to use. Each<br>arrow contains an external process<br>indicator that serves as a visual indication<br>that the system has been exposed to the<br>Sterilucent VHP sterilization process. | use pre-vacuum steam<br>sterilization and 100%<br>ethylene oxide sterilization<br>when used as described in<br>the instructions for use.<br>Reusable baskets and<br>accessory items (pins,<br>dividers, mats, etc) are<br>intended to organize and<br>secure enclosed medical<br>devices during<br>sterilization, transport, and<br>storage of the container.<br>Data cards are used to<br>record information<br>regarding a specific<br>sterilization process load.<br>Filter media allows ingress<br>and egress of sterilant<br>while providing a<br>microbial barrier. Tamper<br>evident arrows provide a<br>visual indication that the<br>container system has not<br>been inadvertently opened<br>prior to use. Each arrow<br>contains a modality-<br>specific external process<br>indicator that serves as a | |
| | visual indication that the<br>system has been exposed<br>to a specific sterilization<br>parameter. Data cards,<br>filters and tamper evident | |
| | arrows are single use only. | |
{11}------------------------------------------------
# 6. Summary of Non-Clinical Performance Data
Sterilization performance studies were conducted for the Sterilization Container System and all acceptance criteria were met. Sterilization efficacy testing demonstrated a 12 log reduction and a sterility assurance level (SAL) of 10 ° using the biological indicator (BI) overkill method and half-cycle validation under worst case conditions. Real time event related shelf life studies demonstrated sterility maintenance for a 180 day time period. Whole
{12}------------------------------------------------
package microbial challenge testing, exposing a container to a minimum of 1 x 106 Bacillus atrophaeus colony forming units (CFU) via an aerosol challenge, demonstrated appropriate microbial barrier properties following exposure to the Sterilucent hydrogen peroxide sterilization processes under worst case conditions.
# 7. Summary of Clinical Performance Data
N/A - No clinical tests were conducted for this submission.
# 8. Overall Performance Conclusion Statement
The non-clinical studies demonstrated that the Sterilization Container System is as safe, as effective, and performs as well as the predicate device for the sterilization of the enclosed medical devices and maintains sterility of those devices for a period of up to 180 days. The proposed device and the predicate device are composed of the same designs, materials, and manufacturing characteristics. The proposed device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.