SONICISION STERILIZATION TRAY

K131170 · Covidien, Formerly Valleylab, A Division of Tyco H · KCT · Aug 28, 2013 · General Hospital

Device Facts

Record IDK131170
Device NameSONICISION STERILIZATION TRAY
ApplicantCovidien, Formerly Valleylab, A Division of Tyco H
Product CodeKCT · General Hospital
Decision DateAug 28, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6850
Device ClassClass 2

Indications for Use

The Sonicision Sterilization Tray is intended to be used to encase and protect reusable battery packs and generators of the Sonicision system during sterilization and storage. The tray containing up to one generator and one battery pack (maximum load) can be used with the following models and pre-set (non-adjustable) cycles. STERRAD® 100S (Standard), STERRAD® NX® (Standard), STERRAD® 100NX® (Standard), STERIS Amsco® V-PRO® 1 (Lumen), STERIS Amsco® V-PRO® 1 Plus (Lumen), STERIS Amsco® V-PRO® maX (Lumen). Maintenance of sterility of the content within the tray is based on demonstrated performance of the sterilization wrap or pouch used to encase the tray.

Device Story

Sonicision Sterilization Tray is a reusable, two-part (lid/base) container designed to hold and protect Sonicision generators and battery packs during sterilization, storage, and transport. Composed of polysulfone with silicone inserts, the tray features latches, handles, and surface perforations to allow sterilant penetration. It does not provide a sterile barrier; it must be wrapped in an FDA-cleared sterilization wrap or pouched in a Tyvek pouch. Used in clinical settings, the tray is operated by healthcare staff to facilitate the decontamination and sterilization of surgical equipment. By securing the generator and battery pack, the tray ensures the integrity of the devices during the sterilization process, ultimately supporting the availability of sterile equipment for surgical procedures.

Clinical Evidence

No clinical data. Substantial equivalence is supported by non-clinical bench testing, including sterilization efficacy (six-log reduction of Geobacillus stearothermophilus), hydrogen peroxide residual testing via UV/VIS spectroscopy, material compatibility (340 reprocessing cycles), biocompatibility (ISO 10993-5/12), and mechanical performance per AAMI ST77:2006.

Technological Characteristics

Materials: Polysulfone (lid/base), silicone (inserts). Design: Two-part system with latches, handles, and surface perforations (9.7%). Energy: Passive device; relies on external sterilization systems (STERRAD/V-PRO). Connectivity: None. Sterilization: Reusable; compatible with hydrogen peroxide gas plasma sterilization. Conformance: AAMI ST77:2006, ISO 10993, ISO 14937.

Indications for Use

Indicated for encasing and protecting reusable Sonicision system battery packs and generators during sterilization and storage. Compatible with specific STERRAD and STERIS V-PRO sterilization cycles. No specific patient population or contraindications described.

Regulatory Classification

Identification

A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the word "COVIDIEN" in bold, black letters. To the left of the word is a black and white logo that resembles a plus sign. The logo is made up of black and white squares. The word "COVIDIEN" is in all capital letters. # 510(k) Summary Date summary prepared: August 23, 2013 #### 510(k) Submitter/Holder Covidien 5920 Longbow Drive Boulder. CO 80301 #### Contact **AUG 2 8 2013** David M. Horton Product Manager, Regulatory Affairs Telephone: 303-530-6391 Fax: 303-530-6313 david.m.horton@covidien.com #### Name of (Subject) Device Trade Name: Sonicision™ Sterilization Tray Common Name: Sterilization Tray Classification Name: Sterilization wrap containers, trays, cassettes & other accessories Device Class: II Classification Regulation: 21 CFR 880.6850 Product Code: KCT #### Predicate Devices The sterilization tray described in this submission was compared and found to be substantially equivalent to the following sterilization trays in commercial distribution: Trade Name: Sonicision™ Sterilization Tray 510(k) Number: K112536, cleared 4/19/2012 Manufacturer: Covidien Trade Name: V-PRO® Sterilization Tray 510(k) Number: K070769, cleared 10/5/2007 Manufacturer: STERIS Corporation Trade Name: FinESS™ Endoscope Sterilization Tray 510(k) Number: K103213, cleared 2/2/2011 Manufacturer: Entellus Medical, Inc. #### Device Description The Sonicision™ Sterilization Tray is intended to hold and protect Sonicision generators and battery packs during sterilization and storage, as indicated. It is comprised of a polysulfone base containing silicone inserts and a removable/latching polysulfone lid. The tray is reusable and provided to the customer in a non-sterile condition. Perforations on the lid and base permit {1}------------------------------------------------ K131170 510(k) Summary exposure of the sterilant. Handles on the sides of the base facilitate transfer of the sterilized components, and the contoured inserts secure the contents within the tray. The Sonicision™ Sterilization Tray does not contain gaskets, filters, or valves to act as a barrier to microorganisms during storage, handling, and transport. The tray containing its intended content (load) must be wrapped with an FDA-cleared STERRAD or STERIS V-PRO-compatible polypropylene sterilization wrap or pouched with an FDA-cleared STERRAD or STERIS V-PRO-compatible Tyvek® pouch (note: the wrap and pouch are not supplied by Covidien) and sterilized as instructed by the labeling. ### Intended Use The Sonicision Sterilization Tray is intended to be used to encase and protect reusable battery packs and generators of the Sonicision system during sterilization and storage. The tray containing up to one generator and one battery pack (maximum load) can be used with the following models and pre-set (non-adjustable) cycles. | Model | Cycle | |-----------------------------|----------| | STERRAD® 100S | Standard | | STERRAD® NX® | Standard | | STERRAD® 100NX® | Standard | | STERIS Amsco® V-PRO® 1 | Lumen | | STERIS Amsco® V-PRO® 1 Plus | Lumen | | STERIS Amsco® V-PRO® maX | Lumen | Maintenance of sterility of the content within the tray is based on demonstrated performance of the sterilization wrap or pouch used to encase the tray. #### Technological and Performance Characteristics The proposed Sonicision tray was found to be similar to the predicate trays in several ways. Three fundamental similarities are identified and discussed below: - (1) Basic design All trays are reusable accessories designed as two-part (lid/base) systems with latches, handles, perforations, and contoured inserts intended to contain items for sterilization, storage, transportation, aseptic presentation of contents, and return of contaminated items to the decontamination area. Notably other design aspects, such as the general shape, size, weight, and materials, though not identical, are similar. - (2) Role in the sterile barrier system Neither the proposed tray nor the predicate trays contain gaskets, valves, or filters - they must all be enclosed with a qualified FDAcleared sterilization wrap or pouch to maintain sterility. - (3) Fundamental technology All trays rely on surface perforations (holes or other openings) to allow the sterilant to penetrate and render its content sterile. The primary differences between the predicate and proposed trays are variances in shape, size, weight, intended content, and surface perforations. Although a small difference in percent of surface perforations is noted, performance data demonstrate that the difference does not adversely affect safety and effectiveness. A comparison of the proposed device to the predicate devices is provided on the next page. {2}------------------------------------------------ | K131170 | 510(k) Summary | |---------|----------------| |---------|----------------| | | Proposed: Sonicision™<br>Sterilization Tray | Predicate: Sonicision™<br>Sterilization Tray,<br>K112536 | Predicate: V-PRO®<br>Sterilization Tray, K070769 | Predicate: FinESS™ Endoscope<br>Sterilization Tray, K103213 | |------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | To encase and protect reusable<br>batteries and generators of the<br>Sonicision system during<br>sterilization and storage.<br>Compatible sterilization systems are<br>indicated as follows:<br>-STERRAD® 100S<br>-STERRAD® NX®<br>-STERRAD® 100NX®<br>-STERIS Amsco® V-PRO® 1<br>-STERIS Amsco® V-PRO® 1 Plus<br>-STERIS Amsco® V-PRO® maX | To encase and protect<br>reusable batteries and<br>generators of the Sonicision<br>system during sterilization<br>and storage. Compatible<br>sterilization systems are<br>indicated as follows:<br>-STERRAD® 100S | To contain, transport, and store<br>reusable medical devices for<br>sterilization. Compatible<br>sterilization systems are indicated<br>as follows:<br>-STERIS Amsco® V-PRO® 1<br>-STERIS Amsco® V-PRO 1 Plus<br>-STERIS Amsco® V-PRO® maX | To encase and protect the FinESS<br>Endoscope for sterilization.<br>Compatible sterilization systems<br>are indicated as follows:<br>-STERRAD® NX®<br>-STERRAD® 100NX® | | Design<br>Characteristics | | | | | | Composition | Base and lid | Identical to proposed tray | Base and lid | Base and lid | | Intended Content<br>(maximum load) | One Sonicision generator and one<br>battery pack | Identical to proposed tray | Reusable medical instrument(s),<br>as indicated by size and weight | One FinESS™ Endoscope | | Inserts | Yes. Designed to secure intended<br>content | Identical to proposed tray | Yes. Various. Designed to secure<br>intended content | Yes. Designed to secure intended<br>content | | Handles | Yes | Identical to proposed tray | Yes | No | | Latches | Yes | Identical to proposed tray | Yes | Yes | | Reusable | Yes | Identical to proposed tray | Yes | Yes | | Materials of<br>Construction | | | | | | Lid/Base | Polysulfone | Identical to proposed tray | Noryl resin | Polyetherimide | | Inserts | Silicone | Identical to proposed tray | Polypropylene | Silicone | | Latch | Polysulfone | Identical to proposed tray | 304 Stainless Steel | Polyetherimide | 510(k) Summary - Comparison Tabl {3}------------------------------------------------ | 70 | ummary<br>S | |----|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | I | (1) 0 11 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1<br>l<br>ച്ച് ഗ | | | Proposed: Sonicision™<br>Sterilization Tray | Predicate: Sonicision™<br>Sterilization Tray, K112536 | Predicate: V-PRO®<br>Sterilization Tray, K070769 | Predicate: FinESS™ Endoscope<br>Sterilization Tray, K103213 | | |---------------------------|----------------------------------------------|-------------------------------------------------------|--------------------------------------------------|-------------------------------------------------------------|-----------------------------| | Dimensions | Assembled (L×W) | 5.9 in. × 8.1 in. | Identical to proposed tray | 10 in. × 14 in. | 7.6 in × 4.1 in. | | Weight | Containing<br>intended content<br>(Max Load) | 1.58 lbs<br>(1 battery and 1 generator) | Identical to proposed tray | 5.16 lbs<br>(various reusable medical<br>devices) | 11.61 oz<br>(one endoscope) | | Fundamental<br>Technology | Percent of surface<br>perforations | 9.7% | Identical to proposed tray | 5% | 1% | . 1 Page 4 of 6 {4}------------------------------------------------ K 131170 510(k) Summary #### Non-Clinical Performance Studies Objective test evidence supporting sterilization efficacy, hydrogen peroxide residuals, material compatibility, biocompatibility, and mechanical performance were provided in the submission. A brief description of all tests used to support the conclusion of substantial equivalence with the predicate devices is provided as follows: Sterilization efficacy: Sonicision Trays containing the intended maximum load (one reusable battery pack and one reusable generator) were inoculated with geobacillus stearothermophilus. The biological indictor containing the indicator organism was placed in areas of the battery pack, generator, and sterilization trav determined to be the most difficult (worstcase) for the sterilant to penetrate. Following inoculation, the sterilization trays were either wrapped with an FDA-cleared STERRAD or STERIS V-PRO-compatible polypropylene wrap or enclosed in an FDA-cleared STERRAD or STERIS V-PRO-compatible Tyvek® Pouch. The loaded, inoculated, and wrapped or pouched trays were then exposed to the Standard cycles of the STERRAD® NX® and 100NX® systems, or the Lumen cycle of the STERIS Amsco® V-PRO® system. The results show that the indicated STERRAD and STERIS "half cycles" provided a sixlog reduction of the indicator organism. Sterilant Residuals: Residuals testing by ultraviolet/visible (UV/VIS) spectroscopy compared the measured results of the Sonicision tray's load with those obtained for a reference predicate tray's load. The study consisted of three runs of loaded (one battery pack and one generator), wrapped, or pouched Sonicision trays placed in STERRAD or STERIS systems. Testing of the reference predicate tray consisted of at least one run of loaded placed in STERRAD or STERIS systems. The products were exposed to Standard cycles of each STERRAD NX® and 100NX®, and Lumen cycle of the STERIS Amsco V-PRO® model. The results show that the Sonicision Sterilization Tray permits adequate dissipation of hydrogen peroxide on the load when compared to the predicate devices used in the study. Use Life (Material Compatibility): Life testing required all key functional/performance specifications of the tray to be evaluated after being subjected to complete reprocessing cycles using a worst-case methodology. The worst-case sterilization systems and cycles chosen for testing were the STERRAD 100NX with Standard cycle, and the STERIS V-Pro with Lumen cycle. The protocol required the tray to be assembled, cleaned, disinfected, dried and sterilized repeatedly (340 cycles) using the methods identified in the labeling. The results show the tray meets its performance specifications when reprocessed repeatedly in accordance with the labeling. Biocompatibility: Biocompatibility testing was performed using a worst-case methodology by exposing the Sonicision Sterilization Tray to two full cycles (as indicated) of the STERRAD and STERIS systems. Test samples were prepared in accordance with ISO 10993-12:2007, Part 12: Sample Preparation and Reference Materials. All biocompatibility assays were performed on neat extracts. A quantitative assessment was made in accordance with applicable requirements of ISO 10993-5:2009, Part 5: Tests for In Vitro Cytotoxicity. The results show that the Sonicision Sterilization Tray passes all applicable biocompatibility requirements and is, therefore, considered non-cytotoxic following exposure to the intended sterilization systems. Mechanical/Bench: The Sonicision Sterilization Tray was evaluated in accordance with applicable clauses/criteria specified in AAMI ST77:2006. The results show that all requirements were met. {5}------------------------------------------------ K131170 510(k) Summary #### Conformance to Standards and Guidance The Sonicision Sterilization Tray conforms to applicable clauses of FDA-recognized consensus standard, AAMI ST77:2006, and draft FDA Guidance Document, "Premarket Notification [510(k)] Submissions for Medical Sterilization Packaging Systems in Health Care Facilities; Draft Guidance for Industry and FDA" (March 7, 2002). Notably, conformance to this standard and guidance document rely on adherence to the requirements and procedures of the following national and international standards: ISO 10993-1:2003 (Biocompatibility), AAMI ST81:2004 (Labeling), ISO 14937:2009 (Sterilization). #### Clinical Performance Studies This premarket notification report does not rely on the assessment of clinical performance data to demonstrate substantial equivalence. #### Conclusion The Sonicision™ Sterilization Tray subject to this submission is substantially equivalent to the Sonicision™ Sterilization Tray (K112536), V-PRO® Sterilization Tray (K070769), and the FinESS™ Endoscope Sterilization Tray (K103213). None of the differences between the Sonicision tray and the predicate trays change the intended use or raise new questions of safety or effectiveness. {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows a circular logo with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the top half of the circle. Inside the circle is a stylized image of an eagle or bird-like figure with outstretched wings. The logo is black and white and appears to be a seal or emblem. ## DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 August 28. 2013 Covidien Mr. David M. Horton Products Manager, Regulatory Affairs 5920 Longbow Drive BOULDER CO 80301 Rc: K131170 > Trade/Device Name: Sonicision™ Sterilization Tray Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Tray Regulatory Class: II Product Code: KCT Dated: August 2. 2013 Received: August 6, 2013 Dear Mr. Horton: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into cither class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {7}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Mary S. Runner -S Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {8}------------------------------------------------ ## Indications for Use Statement 510(k) Number (if known): K131170 Device Name: Sonicision™ Sterilization Tray Indications for Use: The Sonicision Sterilization Tray is intended to be used to encase and protect reusable battery packs and generators of the Sonicision system during sterilization and storage. The tray containing up to one generator and one battery pack (maximum load) can be used with the following models and pre-set (non-adjustable) cycles. | Model | Cycle | |-----------------------------|----------| | STERRAD® 100S | Standard | | STERRAD® NX® | Standard | | STERRAD® 100NX® | Standard | | STERIS Amsco® V-PRO® 1 | Lumen | | STERIS Amsco® V-PRO® 1 Plus | Lumen | | STERIS Amsco® V-PRO® maX | Lumen | Maintenance of sterility of the content within the tray is based on demonstrated performance of the sterilization wrap or pouch used to encase the tray. Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use ✔ AND/OR (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) . Concurrence of CDRH, Office of Device Evaluation (ODE) # Sreekanth Gutala -S 2013.08.28 13:57:43 -04'00' (Division Sign-Off) Division of Anesthesiology, General Hospital Respiratory, Infection Control and Dental Devices 510 (k) Number:
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...