K072563 · Miltex, Inc. · KCT · Feb 12, 2008 · General Hospital
Device Facts
Record ID
K072563
Device Name
MILTEX RIGID CONTAINER SYSTEMS
Applicant
Miltex, Inc.
Product Code
KCT · General Hospital
Decision Date
Feb 12, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The Miltex Rigid Sterilization Container System is intended for use in hospitals and health care facilities to contain other medical devices that are to be sterilized and to allow sterilization of the enclosed medical devices, including surfaces and lumens, using gravity steam sterilizers. The recommended gravity steam sterilization cycle parameters are 45 minutes exposure at 121 deg C and a 20 minute dry time. 2 metal lumens 3-mm in diameter and up to 200-mm in length were validated. The container is intended for use with non-porous materials such as stainless steel surgical instruments. For effective sterilization and drying, the weight of the single container, basket and basket contents should not exceed the following: 16 lbs for the one-half size container, 20 lbs for the three-quarter size container, and 25 lbs for the full size container. Sterilized devices may be stored and transported within the container. The container is intended to maintain sterility for a maximum of 60 days as long as the integrity of the container has not been compromised. Containers should not be stacked when used within gravity steam sterilizers.
Device Story
System of rigid, reusable, sealed containers; includes perforated bottoms, lids, and baskets. Used in hospitals/healthcare facilities to house surgical instruments during gravity steam sterilization. Sterilant penetrates through lid/bottom perforations; filter system prevents microbial migration during transport. Operated by healthcare staff; instruments placed in baskets within containers. Output is a sterile, protected environment for instruments. Benefits include effective sterilization, sterility maintenance for up to 60 days, and safe transport/storage. Not for stacking in sterilizers.
Clinical Evidence
Bench testing only. Validation included gravity steam efficacy testing, thermal profile studies, and sterility maintenance (30-day real-time). Tested with inoculated 3-mm I.D. x 200-mm metal lumens and stainless steel instruments (hinges/knurled surfaces).
Technological Characteristics
Materials: Aluminum alloy, stainless steel, silicone. Design: Rigid, reusable, sealed container with perforated lid/bottom and internal filter system. Energy: Gravity steam sterilization. Connectivity: None. Sterilization: Gravity steam (121°C, 45 min exposure, 20 min dry). Standards: Conforms to AAMI ST 77 (Draft).
Indications for Use
Indicated for use in hospitals and healthcare facilities to contain and sterilize non-porous medical devices (e.g., stainless steel instruments) via gravity steam sterilization. Supports metal lumens up to 3mm diameter and 200mm length. Weight limits: 16 lbs (1/2 size), 20 lbs (3/4 size), 25 lbs (full size).
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Predicate Devices
SteriTite® Rigid Sterilization Container System with MediTray Products (K022978)
Submission Summary (Full Text)
{0}------------------------------------------------
# FEB 1 2 2008
### 510(k) Summary [21 CFR §807.92]
Prepared: August 15, 2007
Device Trade Name: Miltex Rigid Sterilization Container System.
Device Common Name: Rigid Sterilization Container.
Classification Name: Sterilization wrap containers, trays, cassettes, and other accessories.
Class of Device: Class II device, product code KCT
Predicate Device: SteriTite® Rigid Sterilization Container System with MediTray Products- Case Medical, Incorporated- K022978
Official Contact: Audie Margrave, Senior Director QA/RA
#### Device Description:
The Miltex Rigid Sterilization Container System consists of a family of rigid, re-usable, sealed containers that provides an effective sterilization packaging method for medical devices. Container bottoms and lids, within a given size, are interchangeable. The system is composed of the following components:
- Container bottoms (perforated versions) .
- . Container baskets,
- . Container lids (perforated only)
The container system is designed for sterilant penetration through perforations in the lid and container bottom models that are perforated.
#### Intended Use:
The Miltex Rigid Sterilization Container System is intended for use in hospitals and health care facilities to contain other medical devices that are to be sterilized and to allow sterilization of the enclosed medical devices, including surfaces and lumens, using gravity steam sterilizers. The recommended gravity steam sterilization cycle parameters are 45 minutes exposure at 121 deg C and a 20 minute dry time. 2 metal lumens 3-mm in diameter and up to 200-mm in length were validated. The container is intended for use with non-porous materials such as stainless steel surgical instruments.
For effective sterilization and drying, the weight of the single container, basket and basket contents should not exceed the following: 16 lbs for the one-half size container, 20 lbs for the three-quarter size container, and 25 lbs for the full size container.
Sterilized devices may be stored and transported within the container. The container is intended to maintain sterility for a maximum of 60 days as long as the integrity of the container has not been compromised.
Containers should not be stacked when used within gravity steam sterilizers.
The device models that are the subject of this pre-market notification on listed on page 2 of this Indication for Use statement.
{1}------------------------------------------------
| Part Number | Description | UM |
|-------------|---------------------------------------------------------------------------------------|----|
| 3-5110-10 | STERI CONTAINER ½ (11 ¼ in x 11 in) x 4 inches in height PERFORATED<br>BOTTOM, SILVER | EA |
| 3-5110-13 | STERI CONTAINER ½ (11 ¼ in x 11 in) x 5 inches in height PERFORATED<br>BOTTOM, SILVER | EA |
| 3-5110-15 | STERI CONTAINER ½ (11 ¼ in x 11 in) x 6 inches in height PERFORATED<br>BOTTOM, SILVER | EA |
| 3-5310-10 | STERI CONTAINER ¾ (18 ¼ in x 11 in) x 4 inches in height PERFORATED<br>BOTTOM, SILVER | EA |
| 3-5310-13 | STERI CONTAINER ¾ (18 ¼ in x 11 in) x 5 inches in height PERFORATED<br>BOTTOM, SILVER | EA |
| 3-5310-15 | STERI CONTAINER ¾ (18 ¼ in x 11 in) x 6 inches in height PERFORATED<br>BOTTOM, SILVER | EA |
| 3-5510-10 | STERI CONTAINER 1 (23 in x 11 in)x 4 inches in height PERFORATED<br>BOTTOM, SILVER | EA |
| 3-5510-13 | STERI CONTAINER 1 (23 in x 11 in)x 5 inches in height PERFORATED<br>BOTTOM, SILVER | EA |
| 3-5510-15 | STERI CONTAINER 1 (23 in x 11 in) x 6 inches in height PERFORATED<br>BOTTOM, SILVER | EA |
| 3-6100-00 | STERI CONTAINER LID ½ (11 ¼ in x 11 in) PERFORATED, SILVER | EA |
| 3-6300-00 | STERI CONTAINER LID ¾ (18 ¼ in x 11 in) PERFORATED, SILVER | EA |
| 3-6500-00 | STERI CONTAINER LID 1/1 (23 in x 11 in) PERFORATED, SILVER | EA |
Key - ½ indicates one-half size or 11 ¼ inches x 11 inches % indicates three-quarter size or 18 ¼ inches x 11 inches l indicates full size or 23 inches x 11 inches
## The following container lids can be used for each container base:
| Lid Size | Perforated Base to be used with Lid |
|--------------------------------------------------|---------------------------------------------------------------------------------------------|
| 1/2 size Lid (11 1/4 in x 11 in)<br>PN 3-6100-00 | PN 3-5110-10 (4 in height)<br>PN 3-5110-13 (15 1/3 in height)<br>PN 3-5110-15 (6 in height) |
| 3/4 size Lid (18 1/4 in x 11 in)<br>PN 3-6300-00 | PN 3-5310-10 (4 in height)<br>PN 3-5310-13 (5 1/3 in height)<br>PN 3-5310-15 (6 in height) |
| Full size Lid (23 in x 11 in)<br>PN 3-6500-00 | PN 3-5510-10 (4 in height)<br>PN 3-5510-13 (5 1/3 in height)<br>PN 3-5510-15 (6 in height) |
### Technological Characteristics:
A comparison of the technology characteristics of the Miltex Rigid Sterilization Containers to the predicate device.
{2}------------------------------------------------
| Properties | Miltex System | SteriTite System |
|--------------------------------------------------------------------------------------|------------------------------------------------|------------------------------------------------|
| Intended use to contain instruments being<br>sterilized in gravity steam sterilizers | Yes | Yes |
| Intended to be re-used | Yes | Yes |
| Closed System | Yes | Yes |
| Sealed | Yes | Yes |
| Design | | |
| Incorporates a filter system to permit entry of<br>sterilant agent | Yes | Yes |
| Incorporates a filter system to prevent<br>microbial migration during transport. | Yes | Yes |
| Materials | | |
| Container | Aluminum alloy, Stainless<br>Steel, & Silicone | Aluminum alloy, Stainless<br>Steel, & Silicone |
#### Performance Data:
A comparison of the non-clinical performance of the Miltex Rigid Sterilization Containers to the predicate device.
| Properties | Miltex System | SteriTite System |
|--------------------------------------------|--------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance Standards | | |
| Conformance to appropriate AAMI standards | Yes, conforms to AAMI ST<br>77 Draft- Containment<br>Devices for Reusable<br>Medical Device<br>Sterilization | Yes, conforms to AAMI ST-33 -<br>Guidelines for the Selection and Use<br>of Reusable Rigid Sterilization<br>Container Systems for ETO<br>Sterilization and Steam Sterilization<br>in Healthcare Facilities |
| Validation Testing | | |
| Gravity Steam Efficacy Testing | Yes | Yes |
| Thermal Profile Study | Sterilant penetration can be<br>achieved in a single layer<br>configuration | Sterilant penetration can be achieved<br>in a stacked configuration |
| Sterility Maintenance | 30-day real time w/ weekly<br>handling events (perforated<br>bottom). | 90-day real time w/ weekly handling<br>events. 30-day real time w/ daily<br>handling events. |
| Load | Up to 16-lbs. (small)<br>Up to 20-lbs. (med.)<br>Up to 25-lbs. (large) | Up to 22-lbs. |
| Test Organisms/ Inoculated Product | | |
| Inoculated Lumens | Yes--3-mm I.D. x 200-<br>mm, metal | Yes |
| Inoculated Stainless Steel Medical Devices | Yes (hinge area of medical<br>pliers and knurled<br>instruments) | Yes |
### Conclusion:
The Miltex Rigid Sterilization Containers is substantially equivalent to the SteriTite Container (K022978).
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image is a black and white circular seal for the Department of Health and Human Services - USA. The seal features the department's logo, which is a stylized human figure embracing a globe. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the perimeter of the circle.
# FEB 1 2 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Patricia Kihn D.D.S. Manager Quality Systems Compliance & Regulatory Affairs Miltex, Incorporated 589 Davies Drive York, Pennsylvania 17402
Re: K072563
Trade/Device Name: Miltex Rigid Containers Regulation Number: 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: KCT Dated: January 25, 2008 Received: January 28, 2008
Dear Dr. Patricia Kihn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 -- Dr. Patricia Kihn
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chih-Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K072563
Device Name: Miltex Rigid Containers
Indications For Use:
The Miltex Rigid Sterilization Container System is intended for use in hospitals and health care facilities to contain other medical devices that are to be sterilized and to allow sterilization of the enclosed medical devices, including surfaces and lumens, using gravity steam sterilizers. The recommended gravity steam sterilization cycle parameters are 45 minutes exposure at 121 deg C and a 20 minute dry time. 2 metal lumens 3-mm in diameter and up to 200-mm in length were validated. The container is intended for use with non-porous materials such as stainless steel surgical instruments.
For effective sterilization and drying, the weight of the single container, basket and basket contents should not exceed the following: 16 lbs for the one-half size container, 20 Ibs for the three-quarter size container, and 25 Ibs for the full size container.
Sterilized devices may be stored and transported within the container. The container is intended to maintain sterility for a maximum of 60 days as long as the integrity of the container has not been compromised.
| Lid Size | Perforated Base to be used with Lid |
|------------------------------------------------|-------------------------------------------------------------------------------------------|
| 1/2 size Lid (11 ¼ in x 11 in)<br>PN 3-6100-00 | PN 3-5110-10 (4 in height)<br>PN 3-5110-13 (5 in height)<br>PN 3-5110-15-15 (6 in height) |
| 3/4 size Lid (18 ¼ in x 11 in)<br>PN 3-6300-00 | PN 3-5310-10 (4 in height)<br>PN 3-5310-13 (5 in height)<br>PN 3-5310-15 (6 in height) |
| Full size Lid (23 in x 11 in)<br>PN 3-6500-00 | PN 3-5510-10 (4 in height)<br>PN 3-5510-13 (5 in height)<br>PN 3-5510-15 (6 in height) |
Containers should not be stacked when used within gravity steam sterilizers.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use X (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IFNEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K072563
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.