← Product Code [FRG](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/FRG) · K242898

# Sterilization Pouches and Reels (K242898)

_Shanghai Jianzhong Medical Packaging Co., Ltd. · FRG · Feb 21, 2025 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/FRG/K242898

## Device Facts

- **Applicant:** Shanghai Jianzhong Medical Packaging Co., Ltd.
- **Product Code:** [FRG](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/FRG.md)
- **Decision Date:** Feb 21, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6850
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The Sterilization Pouches and Reels is intended to provide health care workers with an effective method to enclosed devices in a single or double pouch configuration by steam or ethylene oxide sterilization. The recommended sterilization cycle parameters are as follows: a) Steam sterilization at 132°C (270°F) for 4 minutes; Drying time of 20 minutes. b) Ethylene Oxide (EO) with a concentration of 800 mg/L at 55 ℃ (131 °F) and 40%-80% relative humidity for 6 hours. Aeration time of 7 days at 20℃ (68 °F). The Sterilization Pouches and Reels is composed of medical compound film, maintains the sterility of the enclosed devices for up to 12 months post Steam or EO gas sterilization, and before sterilization has a maximum shelf life of 5 years from the date of manufacture. The pouch's external chemical indicators are designed to indicate to the user that the pouch has undergone either a steam or EO Sterilization indicators will change color according to the sterilization method (Steam, Ethylene oxide). The color of steam sterilization indicator changes from Pink before sterilization to Cocoa after sterilization. The color of Ethylene oxide sterilization indicator changes from Blue before sterilization to Golden Brown after sterilization. The maximum validated pouch load is 1.87 lbs. (850g); The maximum validated reel load is 1.87 lbs. (850 g)

## Device Story

Device consists of medical compound film pouches and reels used by healthcare workers to contain medical instruments during steam or EO sterilization. Input: medical devices placed inside pouch/reel. Operation: heat-sealed by user; subjected to validated steam (132°C) or EO (800 mg/L) cycles. Output: sterilized, sealed package maintaining sterility for 12 months. External chemical indicators (printed on film) change color (Pink-to-Cocoa for steam; Blue-to-Golden Brown for EO) to confirm exposure to sterilization process. Benefits: provides microbial barrier for sterile storage of instruments. Used in clinical settings; operator is healthcare worker. Output confirms process exposure, not sterility; used alongside biological indicators.

## Clinical Evidence

No clinical data. Bench testing only, including seal strength (ASTM F88), leak detection (ASTM F1929, F2096), tensile properties (ISO 1924-2), thickness (ASTM F2251), microbial barrier (DIN 58953-6), and biocompatibility (ISO 10993 series).

## Technological Characteristics

Materials: medical grade paper and medical compound film. Sensing: chemical indicator bars (ISO 11140-1). Dimensions: 72 variations of pouches and reels. Sterilization: steam (132°C) and EO (800 mg/L). Biocompatibility: non-cytotoxic, non-irritative, non-sensitizing (ISO 10993).

## Regulatory Identification

A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.

## Predicate Devices

- Sterilization Pouch and Roll ([K221875](/device/K221875.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format. The logo is simple and professional, and it is easily recognizable.

February 21, 2025

Shanghai Jianzhong Medical Packaging Co., Ltd. Xinhua Wei Medical Device Regulatory Affairs Specialist Bldg 16, 789# Puxing Road, Minhang District Shanghai. 201114 China

Re: K242898

Trade/Device Name: Sterilization Pouches and Reels Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: FRG, JOJ Dated: September 24, 2024 Received: January 24, 2025

Dear Xinhua Wei:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Digitally signed by Stephen A. Stephen A. Anisko -S Date: 2025.02.21 Anisko -S 15:53:10 -05'00'

for: Christopher K. Dugard, M.S. Division Director DHT4C: Division of Infection Control Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K242898

Device Name Sterilization Pouches and Reels

#### Indications for Use (Describe)

The Sterilization Pouches and Reels is intended to provide health care workers with an effective method to enclosed devices in a single or double pouch configuration by steam or ethylene oxide sterilization. The recommended sterilization cycle parameters are as follows:

a) Steam sterilization at 132°C (270°F) for 4 minutes; Drying time of 20 minutes.

b) Ethylene Oxide (EO) with a concentration of 800 mg/L at 55 ℃ (131 °F) and 40%-80% relative humidity for 6 hours. Aeration time of 7 days at 20℃ (68 °F).

The Sterilization Pouches and Reels is composed of medical compound film, maintains the sterility of the enclosed devices for up to 12 months post Steam or EO gas sterilization, and before sterilization has a maximum shelf life of 5 years from the date of manufacture. The pouch's external chemical indicators are designed to indicate to the user that the pouch has undergone either a steam or EO Sterilization indicators will change color according to the sterilization method (Steam, Ethylene oxide). The color of steam sterilization indicator changes from Pink before sterilization to Cocoa after sterilization. The color of Ethylene oxide sterilization indicator changes from Blue before sterilization to Golden Brown after sterilization. The maximum validated pouch load is 1.87 lbs. (850g); The maximum validated reel load is 1.87 lbs. (850 g)

The following tables (Table 1~5) identifies the size of subject device and its Maximum validated load. Table 1:Sterilization Pouches, self seal

| Size of device |                  | Content/Max Load (Ibs.) |         |              |
|----------------|------------------|-------------------------|---------|--------------|
| Inch           | Metric           | Metal                   | Plastic | Gauze/Linens |
| 2.25"×4"       | 57× (100+30) mm  | 0.004                   | 0.0088  | 0.005        |
| 2.25"×2.75"    | 57× (70+30) mm   | 0.004                   | 0.0088  | 0.004        |
| 2.75"×9"       | 70× (230+30) mm  | 0.020                   | 0.04    | 0.02         |
| 3.5"×6.5"      | 90× (135+30) mm  | 0.016                   | 0.032   | 0.016        |
| 3.5"×9"        | 90× (230+30) mm  | 0.024                   | 0.12    | 0.024        |
| 3.5"×10.5"     | 90× (270+30) mm  | 0.032                   | 0.15    | 0.032        |
| 3.5"×22"       | 90× (560+30) mm  | 0.072                   | 0.32    | 0.072        |
| 5"×15"         | 127× (380+30) mm | 0.062                   | 0.30    | 0.062        |
| 5.25"×10"      | 135× (254+29) mm | 0.045                   | 0.22    | 0.045        |
| 5.25"×10"      | 135× (255+30) mm | 0.045                   | 0.22    | 0.045        |
| 5.25"×12"      | 135× (305+30) mm | 0.052                   | 0.26    | 0.05         |
| 5.25"×13"      | 140× (330+30) mm | 0.072                   | 0.36    | 0.07         |
| 6"×9"          | 150× (230+30) mm | 0.055                   | 0.27    | 0.05         |
| 7.5"×13"       | 190× (330+30) mm | 0.102                   | 0.48    | 0.10         |
| 8"×12.5"       | 200× (320+30) mm | 0.12                    | 0.53    | 0.10         |
| 10"×14.5"      | 250× (370+30) mm | 0.21                    | 0.65    | 0.19         |
| 10.6"×17"      | 270× (400+30) mm | 0.38                    | 0.78    | 0.22         |
| 10.5"×17"      | 270× (430+30) mm | 0.42                    | 0.82    | 0.24         |
| 12"×15"        | 300× (380+30) mm | 0.52                    | 0.90    | 0.23         |
| 12"×16.5"      | 300× (420+30) mm | 0.58                    | 1.02    | 0.25         |
| 12"×18"        | 300× (460+30) mm | 0.68                    | 1.10    | 0.28         |
| 14"×19"        | 355× (480+30) mm | 0.92                    | 1.10    | 0.34         |

{4}------------------------------------------------

| 15"×25" | 380× (635+30) mm | 1.30 | 1.10 | 0.50 |
|---------|------------------|------|------|------|
| 20"×22" | 500× (560+40) mm | 1.87 | 1.10 | 0.60 |

Table 2 Sterilization Pouches, flat, heat seal

| Size of device |           |
|----------------|-----------|
| Inch           | Metric    |
| 3.5"×22"       | 90×560mm  |
| 4"×12"         | 100×300mm |
| 5"×15"         | 127×380mm |
| 6"×10"         | 150×254mm |
| 6"×22"         | 150×560mm |
| 7.5"×13"       | 190×330mm |
| 8"×20"         | 200×500mm |
| 12"×15"        | 300×380mm |
| 12"×18"        | 300×460mm |
| 16.5"×22.5"    | 420×570mm |

## Table 3 Sterilization Pouches, gusseted, heat seal

| Size of device |               |
|----------------|---------------|
| Inch           | Metric        |
| 4"×12"         | 100×50×300mm  |
| 4"×14"         | 100×50×360mm  |
| 6"×16"         | 150×50×400mm  |
| 6"×18"         | 150×50×460mm  |
| 8"×20"         | 200×55×500mm  |
| 8"×16"         | 200×55×400mm  |
| 8"×19"         | 250×65×480mm  |
| 12"×22"        | 300×80×550mm  |
| 16.5"×22.5"    | 420×100×570mm |

## Table 4 Sterilization Reels, flat, heat seal

|  | Size of device |
|--|----------------|
|--|----------------|

| Inch  | Metric       |
|-------|--------------|
| 3"    | 75mm×30.5m   |
| 4"    | 100mm×30.5m  |
| 6"    | 150mm×30.5m  |
| 8"    | 200mm×30.5m  |
| 10"   | 250mm×30.5m  |
| 2"    | 50mm × 200m  |
| 2.15" | 55mm × 200m  |
| 3"    | 75mm × 200m  |
| 4"    | 100mm × 200m |
| 4.7"  | 120mm × 200m |
| 6"    | 150mm × 200m |
| 8"    | 200mm × 200m |
| 10"   | 250mm×200m   |
| 12"   | 300mm × 200m |
| 14"   | 350mm × 200m |
| 16"   | 400mm × 200m |
| 18"   | 450mm × 200m |
| 20"   | 500mm × 200m |

|  | Table 5 Sterilization Reels, gusseted, heat seal |  |  |  |  |
|--|--------------------------------------------------|--|--|--|--|
|--|--------------------------------------------------|--|--|--|--|

| Content/Max Load (Ibs.) | Metal | Plastic | Gauze/Linens |
|-------------------------|-------|---------|--------------|
|                         | 0.005 | 0.01    | 0.020        |
|                         | 0.004 | 0.0088  | 0.0176       |
|                         | 0.07  | 0.20    | 0.055        |
|                         | 0.06  | 0.19    | 0.05         |
|                         | 0.12  | 0.32    | 0.10         |
|                         | 0.10  | 0.25    | 0.09         |
|                         | 0.12  | 0.53    | 0.10         |
|                         | 0.44  | 0.54    | 0.38         |
|                         | 0.58  | 0.74    | 0.52         |
|                         | 1.87  | 1.10    | 0.60         |

| Content/Max Load (Ibs.) |         |              |
|-------------------------|---------|--------------|
| Metal                   | Plastic | Gauze/Linens |
| 0.004                   | 0.0088  | 0.0176       |
| 0.02                    | 0.04    | 0.025        |
| 0.05                    | 0.12    | 0.04         |
| 0.06                    | 0.18    | 0.06         |
| 0.12                    | 0.53    | 0.10         |
| 0.12                    | 0.48    | 0.10         |
| 0.30                    | 0.74    | 0.26         |
| 1.15                    | 0.98    | 0.40         |
| 1.87                    | 1.10    | 0.60         |

| Content/Max Load (Ibs.) |         |              |
|-------------------------|---------|--------------|
| Metal                   | Plastic | Gauze/Linens |
| 0.02                    | 0.04    | 0.02         |
| 0.025                   | 0.15    | 0.032        |
| 0.06                    | 0.27    | 0.07         |
| 0.10                    | 0.53    | 0.10         |
| 0.12                    | 0.53    | 0.10         |
| 0.004                   | 0.0088  | 0.004        |
| 0.004                   | 0.0088  | 0.005        |
| 0.02                    | 0.04    | 0.002        |
| 0.025                   | 0.08    | 0.035        |
| 0.04                    | 0.10    | 0.04         |
| 0.06                    | 0.20    | 0.06         |
| 0.12                    | 0.38    | 0.08         |
| 0.12                    | 0.53    | 0.10         |
| 0.52                    | 0.53    | 0.25         |
| 0.92                    | 1.05    | 0.34         |
| 1.08                    | 1.07    | 0.52         |
| 1.30                    | 1.10    | 0.58         |
| 1.87                    | 1.10    | 0.60         |

{5}------------------------------------------------

| Size of device |              | Content/Max Load (Ibs.) |         |              |
|----------------|--------------|-------------------------|---------|--------------|
| Inch           | Metric       | Metal                   | Plastic | Gauze/Linens |
| 4.15"          | 55mm × 100m  | 0.004                   | 0.0088  | 0.004        |
| 3"             | 75mm × 100m  | 0.02                    | 0.04    | 0.04         |
| 4"             | 100mm × 100m | 0.05                    | 0.15    | 0.08         |
| 6"             | 150mm × 100m | 0.07                    | 0.20    | 0.10         |
| 8"             | 200mm x 100m | 0.12                    | 0.53    | 0.10         |
| 10"            | 250mm x 100m | 0.12                    | 0.53    | 0.10         |
| 12"            | 300mm × 100m | 1.15                    | 0.98    | 0.40         |
| 14"            | 350mm x 100m | 1.35                    | 1.0     | 0.50         |
| 16"            | 400mm × 100m | 1.65                    | 1.03    | 0.58         |
| 18"            | 450mm x 100m | 1.87                    | 1.10    | 0.60         |
| 20"            | 500mm x 100m | 1.87                    | 1.10    | 0.60         |

#### Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{6}------------------------------------------------

# 510(k) Summary

This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.The

assigned 510(k) Number: K242898

- 1. Date of Preparation: February 20, 2025
- 2. Sponsor Identification

### Shanghai Jianzhong Medical Packaging Co., Ltd.

Bldg 16, 789# Puxing Road, Minhang District, Shanghai 201114, P.R. China Registration Number: 3005030944. Contact Person: Xinhua Wei Position: Medical Device Regulatory Affairs Specialist Tel: +86-21-54315666-5201 Fax: +86-21-54315801 Email: weixinhua@mpackchina.com

- 3. Identification of Proposed Device
Trade Name: Sterilization Pouches and Reels

Common Name: Sterilization Pouches and Reels

Regulatory Information

Classification Name: Wrap, Sterilization

Classification: II;

Product Code: FRG;

Subsequent Product Code: JOJ;

Regulation Number: 21 CFR 880.6850

Review Panel: General Hospital;

- 4. Identification of Predicate Device 510(k) Number: K221875 Product Name: Sterilization Pouch and Roll Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II

{7}------------------------------------------------

Product Code: FRG, Subsequent Product Code: JOJ Dated: January 17, 2023 Received: January 17, 2023

#### 5. Device Description

The Sterilization Pouches and Reels is composed of medical compound film. It is intended to be used to contain medical devices to be terminally sterilized by Steam or EO sterilization process. The recommended sterilization cycle parameters are as follows:

- Steam sterilization at 132°C (270°F) for 4 minutes; Drying time of 20 minutes. a)
- b) Ethylene Oxide (EO) with a concentration of 800 mg/L at 55 °C (131 °F) and 40%-80% relative humidity for 6 hours. Aeration time of 7 days at 20℃ (68 °F).

The medical devices are inserted into the Pouches/Reels, and then sterilized for the EO or Steam Sterilization Process. The heat-sealed Pouches/Reels are heat sealed prior to sterilization processing. After completion of the sterilization process, the Pouches/Reels maintain sterility of the enclosed medical devices until the seal is opened. The Sterilization Pouches and Reels maintains the sterility of the enclosed devices for up to 12 months post Steam or EO gas sterilization, and before sterilization has a maximum shelf life of 5 years from the date of manufacture.

The Sterilization Pouches and Reels is printed with chemical indicator bars that changed from Pink to Cocoa(Steam) or Blue to Golden Brown (EO) when exposed to steam vapor or EO gas during process. The steam and EO chemical indicator offer an addition way to verify processing in the sterilization cycle. The chemical indicator should be used in addition to, not in place of, the biological indicator. The steam and EO chemical indicators do not signify sterilization; they only indicate that the indicators have been exposed to the Steam or EO.

- 6. Indication for Use Statement
The Sterilization Pouches and Reels is intended to provide health care workers with an effective method to enclosed devices in a single or double pouch configuration by steam or ethylene oxide sterilization. The recommended sterilization cycle parameters are as follows:

a) Steam sterilization at 132°C (270°F) for 4 minutes; Drying time of 20 minutes.

b) Ethylene Oxide (EQ) with a concentration of 800 mg/L at 55 ℃ (131 °F) and 40%-80% relative humidity for 6 hours. Aeration time of 7 days at 20°C (68 °F).

The Sterilization Pouches and Reels is composed of medical compound film, maintains the sterility of the enclosed devices for up to 12 months post Steam or EO gas sterilization, and before sterilization has a maximum shelf life of 5 years from the date of manufacture. The pouch's external chemical indicators are designed to indicate to the user that the pouch has undergone either a steam or EO Sterilization process. Sterilization indicators will change color according to the sterilization method (Steam, Ethylene oxide). The color of steam sterilization indicator changes from Pink before sterilization to Cocoa after sterilization. The color of

{8}------------------------------------------------

Ethylene oxide sterilization indicator changes from Blue before sterilization to Golden Brown after sterilization. The maximum validated pouch load is 1.87 lbs. (850g); The maximum validated reel load is 1.87 lbs. (850 g)

| Item                                                           | Size of Device |                | Content/Max Load(Ibs) |         |              |
|----------------------------------------------------------------|----------------|----------------|-----------------------|---------|--------------|
|                                                                | Inch           | Metric         | Metal                 | Plastic | Gauze/Linens |
| Sterilization<br>Pouches,<br>self-seal                         | 2.25"x4"       | 57x(100+30)mm  | 0.004                 | 0.0088  | 0.005        |
|                                                                | 2.25"x2.75"    | 57x(70+30)mm   | 0.004                 | 0.0088  | 0.004        |
|                                                                | 2.75"x9"       | 70x(230+30)mm  | 0.020                 | 0.04    | 0.02         |
|                                                                | 3.5"x6.5"      | 90x(135+30)mm  | 0.016                 | 0.032   | 0.016        |
|                                                                | 3.5"x9"        | 90x(230+30)mm  | 0.024                 | 0.12    | 0.024        |
|                                                                | 3.5"x10.5"     | 90x(270+30)mm  | 0.032                 | 0.15    | 0.032        |
|                                                                | 3.5"x22"       | 90x(560+30)mm  | 0.072                 | 0.32    | 0.072        |
|                                                                | 5"x15"         | 127x(380+30)mm | 0.062                 | 0.30    | 0.062        |
|                                                                | 5.25"x10"      | 135x(254+29)mm | 0.045                 | 0.22    | 0.045        |
|                                                                | 5.25"x10"      | 135x(255+30)mm | 0.045                 | 0.22    | 0.045        |
|                                                                | 5.25"x12"      | 135x(305+30)mm | 0.052                 | 0.26    | 0.05         |
|                                                                | 5.25"x13"      | 140x(330+30)mm | 0.072                 | 0.36    | 0.07         |
| es,<br>self-<br>seal                                           | 6"x9"          | 150x(230+30)mm | 0.055                 | 0.27    | 0.05         |
|                                                                | 7.5"x13"       | 190x(330+30)mm | 0.102                 | 0.48    | 0.10         |
|                                                                | 8"x12.5"       | 200x(320+30)mm | 0.12                  | 0.53    | 0.10         |
|                                                                | 10"x14.5"      | 250x(370+30)mm | 0.21                  | 0.65    | 0.19         |
|                                                                | 10.6"x17"      | 270x(400+30)mm | 0.38                  | 0.78    | 0.22         |
|                                                                | 10.5"x17"      | 270x(430+30)mm | 0.42                  | 0.82    | 0.24         |
|                                                                | 12"x15"        | 300x(380+30)mm | 0.52                  | 0.90    | 0.23         |
|                                                                | 12"x16.5"      | 300x(420+30)mm | 0.58                  | 1.02    | 0.25         |
|                                                                | 12"x18"        | 300x(460+30)mm | 0.68                  | 1.10    | 0.28         |
|                                                                | 14"x19"        | 355x(480+30)mm | 0.92                  | 1.10    | 0.34         |
|                                                                | 15"x25"        | 380x(635+30)mm | 1.30                  | 1.10    | 0.50         |
|                                                                | 20"x22"        | 500x(560+40)mm | 1.87                  | 1.10    | 0.60         |
| Sterilization<br>Pouches,<br>flat,<br>heat<br>seal             | 3.5"x22"       | 90x560mm       | 0.005                 | 0.01    | 0.020        |
|                                                                | 4"x12"         | 100x300mm      | 0.004                 | 0.0088  | 0.0176       |
|                                                                | 5"x15"         | 127x380mm      | 0.07                  | 0.20    | 0.055        |
|                                                                | 6"x10"         | 150x254mm      | 0.06                  | 0.19    | 0.05         |
|                                                                | 6"x22"         | 150x560mm      | 0.12                  | 0.32    | 0.10         |
|                                                                | 7.5"x13"       | 190x330mm      | 0.10                  | 0.25    | 0.09         |
|                                                                | 8"x20"         | 200x500mm      | 0.12                  | 0.53    | 0.10         |
|                                                                | 12"x15"        | 300x380mm      | 0.44                  | 0.54    | 0.38         |
|                                                                | 12"x18"        | 300x460mm      | 0.58                  | 0.74    | 0.52         |
|                                                                | 16.5"x22.5"    | 420x570mm      | 1.87                  | 1.10    | 0.60         |
|                                                                | 4"x12"         | 100x50x300mm   | 0.004                 | 0.0088  | 0.0176       |
|                                                                | 4"x14"         | 100x50x360mm   | 0.02                  | 0.04    | 0.025        |
| Sterilization<br>Pouch<br>es,<br>gusset<br>ed,<br>heat<br>seal | 6"x16"         | 150x50x400mm   | 0.05                  | 0.12    | 0.04         |
|                                                                | 6"x18"         | 150x50x460mm   | 0.06                  | 0.18    | 0.06         |
|                                                                | 8"x20"         | 200x55x500mm   | 0.12                  | 0.53    | 0.10         |
|                                                                | 8"x16"         | 200x55x400mm   | 0.12                  | 0.48    | 0.10         |
|                                                                | 8"x19"         | 250x65x480mm   | 0.30                  | 0.74    | 0.26         |
|                                                                | 12"x22"        | 300x80x550mm   | 1.15                  | 0.98    | 0.40         |
|                                                                | 16.5"x22.5"    | 420x100x570mm  | 1.87                  | 1.10    | 0.60         |
|                                                                | 3"             | 75mmx30.5m     | 0.02                  | 0.04    | 0.02         |
|                                                                | 4"             | 100mmx30.5m    | 0.025                 | 0.15    | 0.032        |
|                                                                | 6"             | 150mmx30.5m    | 0.06                  | 0.27    | 0.07         |
|                                                                | 8"             | 200mmx30.5m    | 0.10                  | 0.53    | 0.10         |
|                                                                | 10"            | 250mmx30.5m    | 0.12                  | 0.53    | 0.10         |
|                                                                | 2"             | 50mm x 200m    | 0.004                 | 0.0088  | 0.004        |
|                                                                | 2.15"          | 55mm x 200m    | 0.004                 | 0.0088  | 0.005        |
| Sterilization<br>Reels,<br>flat,<br>heat<br>seal               | 3"             | 75mm x 200m    | 0.02                  | 0.04    | 0.002        |
|                                                                | 4"             | 100mm x 200m   | 0.025                 | 0.08    | 0.035        |
|                                                                | 4.7"           | 120mm x 200m   | 0.04                  | 0.10    | 0.04         |
|                                                                | 6"             | 150mm x 200m   | 0.06                  | 0.20    | 0.06         |
|                                                                | 8"             | 200mm x 200m   | 0.12                  | 0.38    | 0.08         |
|                                                                | 10"            | 250mmx200m     | 0.12                  | 0.53    | 0.10         |
|                                                                | 12"            | 300mm x 200m   | 0.52                  | 0.53    | 0.25         |
|                                                                | 14"            | 350mm x 200m   | 0.92                  | 1.05    | 0.34         |
|                                                                | 16"            | 400mm x 200m   | 1.08                  | 1.07    | 0.52         |
|                                                                | 18"            | 450mm x 200m   | 1.30                  | 1.10    | 0.58         |
|                                                                | 20"            | 500mm x 200m   | 1.87                  | 1.10    | 0.60         |
|                                                                | 4.15"          | 55mm x 100m    | 0.004                 | 0.0088  | 0.004        |
| Sterilization<br>Reels,<br>gusset<br>ed,                       | 3"             | 75mm x 100m    | 0.02                  | 0.04    | 0.04         |
|                                                                | 4"             | 100mm x 100m   | 0.05                  | 0.15    | 0.08         |
|                                                                | 6"             | 150mm x 100m   | 0.07                  | 0.20    | 0.10         |
|                                                                | 8"             | 200mm x 100m   | 0.12                  | 0.53    | 0.10         |
| heat<br>seal                                                   | 10"            | 250mm x 100m   | 0.12                  | 0.53    | 0.10         |
|                                                                | 12"            | 300mm x 100m   | 1.15                  | 0.98    | 0.40         |
|                                                                | 14"            | 350mm x 100m   | 1.35                  | 1.0     | 0.50         |
|                                                                | 16"            | 400mm x 100m   | 1.65                  | 1.03    | 0.58         |
|                                                                | 18"            | 450mm x 100m   | 1.87                  | 1.10    | 0.60         |
|                                                                | 20"            | 500mm x 100m   | 1.87                  | 1.10    | 0.60         |

(Model and Dimension and Content/Max. Load)

{9}------------------------------------------------

{10}------------------------------------------------

#### 7. Non-Clinical Test Conclusion

Performance testing and Biocompatibility assessment were conducted to verify that the subject device met all design specifications. The test results demonstrated that the subject device meet the following standards (included but not fully described):

- > ASTM F88/F88M-2015, Standard Test Method for Seal Strength of Flexible Barrier Materials.(Sterility)
- > ASTM F1929-2015 Standard Test Method for Detecting Seal Leaks in Porous Medical Package by Dye Penetration
- > ASTM F 2096-2011 (2019) Standard Test Method for Detecting Gross Leaks in Packaging by Internal Pressurization (Bubble Test)
- A ISO 1924-2:2008 Paper and board -- Determination of tensile properties.
- A ASTM F 2251-13 Standard Test Method for Thickness Measurement of Flexible Packaging Material
- > ISO 11140-1:2014 Sterilization of health care products. Chemical indicators. General requirements
- > ISO 10993-1: Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process
- > ISO 10993-5:2009 Biological evaluation of medical devices- Part 5: Tests for in vitro cytotoxicity
- > ISO 10993-7:2008 Biological Evaluation of Medical Device- Part 7: Ethylene Oxide Sterilization Residuals
- > ISO 10993-10:2010 Biological evaluation of medical device- Part 10: Tests for irritation and skin sensitization
- > ISO 10993-12: Biological evaluation of medical devices Part 12: Sample preparation and reference materials
- > ISO 11135: 2014 Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices.
- > DIN58953-6-2010 Sterilization - Sterile supply - Part 6: Microbial barrier testing of packaging materials for medical devices which are to be sterilized.
- > ISO17665-1:2006 Sterilization of health care products-Moist heat-Part1: requirements for the development, validation and routine control of a sterilization process for medical devices.

{11}------------------------------------------------

#### 8. Technological Characteristic Comparison Table

| ITEM                                                  | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Predicate Device K221875                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Comparison<br>analysis                                                                                                                                                                          |
|-------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code                                          | FRG, JOJ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | FRG, JOJ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | same                                                                                                                                                                                            |
| Regulation No.                                        | 21 CFR 880.6850                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 21 CFR 880.6850                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | same                                                                                                                                                                                            |
| Class                                                 | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | same                                                                                                                                                                                            |
| Indication for Use                                    | The Sterilization Pouches and Reels is intended to provide<br>health care workers with an effective method to enclosed<br>devices in a single or double pouch configuration by steam or<br>ethylene oxide sterilization. The recommended sterilization<br>cycle parameters are as follows:<br>a)  Steam sterilization at 132°C (270°F) for 4 minutes;<br>Drying time of 20 minutes.<br>b) Ethylene Oxide (EO) with a concentration of 800 mg/L<br>at 55 °C (131 °F) and 40%-80% relative humidity for 6<br>hours. Aeration time of 7 days at 20℃ (68 °F).<br>The Sterilization Pouches and Reels is composed of medical<br>grade paper and medical compound film, maintains the sterility<br>of the enclosed devices for up to 12 months post Steam or EO<br>gas sterilization, and before sterilization has a maximum shelf<br>life of 5 years from the date of manufacture.<br>The pouch's external chemical indicators are designed to<br>indicate to the user that the pouch has undergone either a steam<br>or EO Sterilization process. Sterilization indicators will change<br>color according to the sterilization method (Steam, Ethylene<br>oxide). The color of steam sterilization indicator changes from<br>Pink before sterilization to Cocoa after sterilization. The color<br>of Ethylene oxide sterilization indicator changes from Blue<br>before sterilization to Golden Brown after sterilization.<br>The maximum validated pouch load is 1.87 lbs. (850g); The<br>maximum validated reel load is 1.87 lbs. (850 g) | The Sterilization Pouch and Roll are intended to provide health<br>care workers with an effective method to enclose devices<br>intended for sterilization in the Steam or via Ethylene Oxide<br>(EO). The recommended sterilization cycles are as follows:<br>• Steam Sterilization at 132°C<br>(270°F) for 4 minutes; Drying time of 20 minutes.<br>• Ethylene Oxide (EO) with a concentration of 800 mg/L at<br>55°C (131 °F) and 40% to 90% relative humidity for 6 hours.<br>Aeration time of 7 days at 20℃).<br>The sterilization pouch and roll are made with medical grade<br>paper and medical compound film. The sterilization pouch and<br>roll maintains the sterility of the enclosed devices for up to 24<br>months post Steam or EO gas sterilization, and before<br>sterilization has a maximum shelf life of 5 years from the date<br>of manufacture. The pouch's external chemical ink indicators<br>are designed to indicate to the user that the pouch has<br>undergone either a steam or EO sterilization process.<br>Sterilization indicator will change color according to the<br>sterilization method (ethylene oxide, steam). Ethylene oxide<br>sterilization indicator color change from Pink before<br>sterilization to Yellow after sterilization. The steam<br>sterilization indicator color change from Blue before<br>sterilization to Dark Grey after sterilization. | Same                                                                                                                                                                                            |
|                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | K242898                                                                                                                                                                                         |
| Material Composition                                  | Composed of medical grade paper and medical compound film<br>and printed with a chemical indicator bar that changed from<br>Pink to Cocoa (Steam) or Blue to Golden Brown (EO) when<br>exposed to Steam vapor or EO gas during process.<br>"Sterilization Pouches, flat, self seal" type contains tape.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Composed of medical grade paper, medical compound film,<br>EO and Steam Process Indicator.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Similar. The<br>Pouches with tape<br>were validated for<br>their effectiveness<br>and safety. The test<br>results show that<br>the pouches are as<br>effective and safe<br>as predicate device. |
| Sterilization<br>Cycles                               | Steam sterilization at 132°C (270°F) for 4 minutes; Drying time<br>of 20 minutes.<br>Ethylene Oxide (EO) with a concentration of 800 mg/L at 55 °C<br>(131 °F) and 40%-80% relative humidity for 6 hours. Aeration<br>time of 7 days at 20°C (68 °F).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Steam Sterilization at 132°C (270°F) for 4 minutes; Drying time<br>of 20 minutes.<br>Ethylene Oxide (EO) with a concentration of 800 mg/L at 55°C<br>(131 °F) and 40% to 90% relative humidity for 6 hours. Aeration<br>time of 7 days at 20°C.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | same                                                                                                                                                                                            |
| Dimensions                                            | Length of Pouches ranges from 2.75" to 25";<br>Width of Pouches ranges from 2.25" to 20";<br>Length of Reels ranges from 30.5m (=1200.78") to 200m<br>(=7874.01");<br>Width of Reels ranges from 2" to 20";<br>There are 72 sizes in total.<br>Please refer to table of Comparison of Dimension for detailed<br>information.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Length of Pouches ranges from 3.5" to 17";<br>Width of Pouches ranges from 2.25" to 12";<br>Length of Reels (or Rolls) ranges from 100m (=3937") to<br>200m (=7874.01");<br>Width of Reels ranges from 2" to 17.7"<br>There are 32 sizes in total.<br>Please refer to table of Comparison of Dimension for detailed<br>information.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Similar                                                                                                                                                                                         |
| Sterilant Penetration Efficiency                      | It meets the requirement of SAL10-6                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | It meets the requirement of SAL10-6                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | same                                                                                                                                                                                            |
| EO、ECH residue                                        | Referring to ISO 10993-7:2008, EO<4mg, ECH<9mg                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Referring to ISO 10993-7:2008, EO<4mg, ECH<9mg                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | same                                                                                                                                                                                            |
|                                                       | Performance Testing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                 |
| Chemical Indicator (CI)<br>Functionality and Endpoint | The sterilant penetrated through the pouch configuration and<br>rendered the CI color the endpoint color.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | The sterilant penetrated through the pouch configuration and<br>rendered the CI color the endpoint color.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | same                                                                                                                                                                                            |
| Device Design of Steam CI                             | It changes from Pink to Cocoa when exposed to Steam vapor.…

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/FRG/K242898](https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/FRG/K242898)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
