K202462 · Sigma Medical Supplies Corporation · FRG · Apr 23, 2021 · General Hospital
Device Facts
Record ID
K202462
Device Name
SIGMA Sterilization Pouch and Roll
Applicant
Sigma Medical Supplies Corporation
Product Code
FRG · General Hospital
Decision Date
Apr 23, 2021
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The SIGMA sterilization pouch and roll are intended to provide health care workers with an effective method to enclose devices intended for sterilization in the Steam Sterilizer and via Ethylene Oxide (EO). The recommended sterilization cycles are as follows: - Gravity steam at 121°C (250°F) for 30 minutes; Drying time of 30 minutes. - Gravity steam at 132°C (270°F) for 15 minutes; Drying time of 30 minutes. - Gravity steam at 135°C (275°F) for 10 minutes; Drying time of 30 minutes. - Pre-vacuum steam at 132°C (270°F) for 4 minutes; Drying time of 20 minutes. - Pre-vacuum steam at 135°C (275°F) for 3 minutes; Drying time of 16 minutes. - EO sterilization cycle is 4 hours at 55°C (131°F) with a relative humidity between 50%-85% and a sterilant concentration of 600 mg/L. Furthermore, the sterilization pouch and roll maintains the enclosed devices up until 3 years post EO gas sterilization and maintains the enclosed devices up until 6 months post Steam sterilization. Lastly, the pouch's external chemical ink indicators are designed to indicate to the pouch has undergone either a steam or EO sterilization process.
Device Story
Device consists of sterilization pouches and rolls used by healthcare workers to enclose medical instruments for steam or EO sterilization. Pouches/rolls are constructed from medical-grade paper and plastic film (CPP/PET) with PU adhesive. External chemical ink indicators printed on the paper change color (Blue to Greenish Black for steam; Red to Yellow for EO) to confirm exposure to the sterilization process. The device maintains sterility of enclosed instruments for up to 3 years post-EO and 6 months post-steam. The device is used in clinical settings; instruments are placed inside, sealed (self-sealing or heat-sealed), and processed in a sterilizer. The color-changing indicator provides visual verification of exposure, aiding in workflow management, though it does not replace biological indicators. Benefits include maintaining instrument sterility and providing a clear visual signal of processing status.
Clinical Evidence
No clinical data. Bench testing only. Validation included sterilant penetration (SAL 10^-6), drying time, aeration, biocompatibility (ISO 10993-10), package integrity (ASTM F2251, F1140, F1929, F88, F2096), and chemical indicator efficacy (ISO 11140-1). Accelerated aging studies (ASTM F1980) confirmed 3-year shelf life for EO and 6-month shelf life for steam.
Technological Characteristics
Materials: Medical grade paper, CPP, PET, PU adhesive. Sensing: Chemical indicator ink (color change). Connectivity: None. Sterilization: Steam and EO gas. Form factor: Pouches (self-sealing, flat, gusseted, multi) and rolls (flat, gusseted). Standards: ASTM F2251, D882, D1004, F1140, F1929, F88, F1980, F2096, F1608; ISO 11607-1, 1924-2, 5636-3, 10993-10, 11140-1; DIN 58953-6.
Indications for Use
Indicated for healthcare workers to enclose medical devices for sterilization via steam or ethylene oxide (EO) gas. Suitable for various instrument types within specified load limits. Not for use as a primary sterilization indicator.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Sigma Medical Supplies Corporation % Uta Shih Regulatory Affairs Manager Sen Mu Technology Co., LTD. I5F-2, No.I, Lane 26, Mincyuan 1st Rd., Lingya District, Kaohsiung City, 802 Taiwan
Re: K202462
Trade/Device Name: SIGMA Sterilization Pouch and Roll Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: FRG Dated: August 21, 2020 Received: August 27, 2020
Dear Uta Shih:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Clarence W. Murray, III, Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202462
#### Device Name
SIGMA sterilization pouch and roll
### Indications for Use (Describe)
The SIGMA sterilization pouch and roll are intended to provide health care workers with an effective method to enclose devices intended for sterilization in the Steam Sterilizer and via Ethylene Oxide (EO). The recommended sterilization cycles are as follows:
- · Gravity steam at 121°C (250°F) for 30 minutes; Drying time of 30 minutes.
- · Gravity steam at 132°C (270°F) for 15 minutes; Drying time of 30 minutes.
- · Gravity steam at 135°C (275°F) for 10 minutes; Drying time of 30 minutes.
- · Pre-vacuum steam at 132°C (270°F) for 4 minutes; Drying time of 20 minutes.
- · Pre-vacuum steam at 135°C (275°F) for 3 minutes; Drying time of 16 minutes.
· EO sterilization cycle is 4 hours at 55°C (131°F) with a relative humidity between 50%-85% and a sterilant concentration of 600 mg/L.
Furthermore, the sterilization pouch and roll maintains the enclosed devices up until 3 years post EO gas sterilization and maintains the enclosed devices up until 6 months post Steam sterilization. Lastly, the pouch's external chemical ink indicators are designed to indicate to the pouch has undergone either a steam or EO sterilization process.
The SIGMA sterilization pouch and roll is offered in the following 6 types:
- · Self-sealing sterilization pouches
- · Self-sealing sterilization multi pouches
- · Sterilization pouches. Flat
- · Sterilization pouches, Gusseted
- · Sterilization rolls, Flat
- · Sterilization rolls, Gusseted
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Type | Size | Model |
|------------------------------------------|--------------------------------------------------------------|-------|
| Self-sealing sterilization pouches | L: 133 mm ~ 610 mm<br>W: 57 mm ~ 460 mm | SMSE |
| Self-sealing sterilization multi pouches | L: 283 mm ~ 335 mm<br>W: 90 mm ~ 240 mm | SMSE |
| Sterilization pouches, Flat | L: 150 mm ~ 600 mm<br>W: 75 mm ~ 420 mm | SMFP |
| Sterilization pouches, Gusseted | L: 300 mm ~ 500 mm<br>W: 100 mm ~ 300 mm<br>H: 40 mm ~ 70 mm | SMGP |
| Sterilization rolls, Flat | L: 30.5M~200 M<br>W: 50 mm ~ 400 mm | SMFR |
| Sterilization rolls, Gusseted | L: ~100 M<br>W: 75 mm ~ 400 mm<br>H: 35 mm ~ 80 mm | SMGR |
The maximum validated pouch load is 3.53 pounds (1.6 kg).
{4}------------------------------------------------
## Submitter's Information
| Name: | SIGMA Medical Supplies Corp. |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | No.34, Ding-Ping Road, Rui Fang Industrial Park, Rui Fang Dist., New Taipei City 224, Taiwan (R.O.C.) |
| Establishment Registration Number: | 3004970050 |
| Contact: | Isabel Tsai<br>RA<br>Legal & Regulatory Affairs Dept.<br>886-3-3748880 ext. 3561<br>email: Isabel.Tsai@Benqmaterials.com<br>Or<br>Ethan Lin<br>Sr. Specialist<br>Market Planning Sec.<br>886-3-3648880 ext. 3572<br>email: Ethan.Lin @sigma-medical.com.tw |
| Date Prepared: | August 21, 2020 |
| Device Name | |
| Trade Name: | SIGMA Sterilization Pouch and Roll |
| Common/usual Name: | SIGMA Sterilization Pouch and Roll |
| Device Classification Names: | 1) Wrap, Sterilization.<br>2) Indicator, Physical/Chemical Sterilization Process |
| Panel: | General Hospital |
| Classification Product Code: | 1) FRG |
| Subsequent Product Code: | 2) JOJ |
| Device Classification: | Class II |
| Regulation Number: | 1) 21 CFR 880.6850<br>2) 21 CFR 880.2800 |
## Predicate Devices
The predicate device [510(k) Notification K180661, cleared June 05, 2018] is the SIGMA Sterilization Pouch and Roll.
Manufacturer: SIGMA Medical Supplies Corp.
{5}------------------------------------------------
Product Code: 1) FRG 2) JOJ 510(k) number: K180661
# Indications for Use
The SIGMA sterilization pouch and roll are intended to provide health care workers with an effective method to enclose devices intended for sterilization in the Steam Sterilizer and via Ethylene Oxide (EO). The recommended sterilization cycles are as follows:
- Gravity steam at 121℃ (250°F) for 30 minutes; Drying time of 30 minutes. .
- Gravity steam at 132℃ (270°F) for 15 minutes; Drying time of 30 minutes. .
- · Gravity steam at 135℃ (275°F) for 10 minutes; Drying time of 30 minutes.
- Pre-vacuum steam at 132℃ (270°F) for 4 minutes; Drying time of 20 minutes. .
- · Pre-vacuum steam at 135°C (275°F) for 3 minutes; Drying time of 16 minutes.
- EO sterilization cycle is 4 hours at 55℃ (131 °F) with a relative humidity between . 50%-85% and a sterilant concentration of 600 mg/L.
Furthermore, the sterilization pouch and roll maintains the enclosed devices up until 3 years post EO gas sterilization and maintains the enclosed devices up until 6 months post Steam sterilization. Lastly, the pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EO sterilization process.
The SIGMA sterilization pouch and roll is offered in the following 6 types:
- Self-sealing sterilization pouches ●
- Self-sealing sterilization multi pouches
- . Sterilization pouches. Flat
- Sterilization pouches, Gusseted
- Sterilization rolls,
- Flat Sterilization rolls, Gusseted
The defining characteristics of the 6 types are as follows:
- · Self-sealing sterilization pouches: These pouches are made from a medical grade plastic film that is heat sealed on three sides. The forth side has an adhesive strip that is used to seal the pouch. Release paper used in the pouch is a laminated sheet with composing structure of PE/paper/PE. It is a strip to cover the adhesive area and is released before seal the pouch. The medical grade paper conforms to recognized material standards and can be sterilized by steam or ethylene oxide gas. The Process Indicators Ink printed on the medical grade paper will exhibit a color change after the pouch is exposed to steam or ethylene oxide gas.
- · Self-sealing sterilization multi pouches: These pouches have the same components as the Self-sealing sterilization pouches, except that the interior of the pouch is divided into several compartments for packaging multiple instruments separately.
{6}------------------------------------------------
- · Sterilization pouches, Flat: These pouches have the same components with the Self-sealing sterilization pouches, except the forth side is left opened instead of an adhesive strip and will be heat-sealed when using.
- · Sterilization pouches, Gusseted: These pouches are the same with the Sterilization pouches, flat, except that the plastic film is folded on both longest sides instead of flat. This design is convenient to enclose the medical devices with certain height.
- · Sterilization rolls, Flat: These rolls are made from a medical grade paper and plastic film that are heat sealed on opposite two sides. It will be cut into the suitable length and the opened sides will be heat-sealed. The indicators printed on the medical grade paper are the same with the self-sealing sterilization pouches.
- · Sterilization rolls, Gusseted: These rolls are the same with the flat sterilization roll, except that the plastic film is folded on both longest sides instead of flat. This design is convenient to enclose the medical devices with certain height.
The following table (Table 1) lists the model numbers of the SIGMA sterilization pouch and roll by type, model, dimensions, and content/max. load (lbs.):
{7}------------------------------------------------
| Type | Model | Dimension in S.I. | *Content / Max Load (lbs.) | | |
|------------------------------------------------|------------|-------------------------|----------------------------|---------|---------------------|
| | | | Metal | Plastic | Linens and<br>Gauze |
| | SMSE057133 | 57 mm x 133 mm | 0.11 | 0.03 | 0.03 |
| | SMSE070257 | 70 mm × 257 mm | 0.39 | 0.11 | 0.10 |
| | SMSE090162 | 90 mm x 162 mm | 0.30 | 0.09 | 0.08 |
| | SMSE090168 | 90 mm × 168 mm | 0.32 | 0.09 | 0.08 |
| | SMSE090257 | 90 mm × 257 mm | 0.57 | 0.16 | 0.14 |
| | SMSE090265 | 90 mm x 265 mm | 0.59 | 0.17 | 0.15 |
| | SMSE090594 | 90 mm x 594 mm | 1.35 | 0.45 | 0.39 |
| | SMSE133289 | 133 mm × 289 mm | 1.00 | 0.33 | 0.28 |
| | SMSE133391 | 133 mm x 391 mm | 1.43 | 0.48 | 0.41 |
| | SMSE135193 | 135 mm x 193 mm | 0.67 | 0.19 | 0.17 |
| | SMSE135283 | 135 mm × 283 mm | 0.99 | 0.33 | 0.28 |
| Self-Sealing<br>Sterilization<br>Pouches | SMSE180335 | 135 mm x 335 mm | 1.20 | 0.40 | 0.34 |
| | SMSE150610 | 150 mm x 610 mm | 2.08 | 0.83 | 0.69 |
| | SMSE190358 | 190 mm x 358 mm | 1.75 | 0.64 | 0.54 |
| | SMSE190365 | 190 mm x 365 mm | 1.79 | 0.65 | 0.55 |
| | SMSE200435 | 200 mm x 435 mm | 2.00 | 0.80 | 0.67 |
| | SMSE300380 | 300 mm x 380 mm | 2.48 | 1.10 | 0.90 |
| | SMSE300474 | 300 mm x 474 mm | 2.63 | 1.50 | 1.17 |
| | SMSE300485 | 300 mm x 485 mm | 2.70 | 1.54 | 1.20 |
| | SMSE303474 | 303 mm x 474 mm | 2.66 | 1.52 | 1.18 |
| | SMSE305416 | 305 mm x 416 mm | 2.56 | 1.28 | 1.02 |
| | SMSE356483 | 356 mm x 483 mm | 2.88 | 1.92 | 1.44 |
| | SMSE380635 | 380 mm x 635 mm | 3.16 | 2.53 | 1.80 |
| | SMSE460610 | 460 mm x 610 mm | 3.40 | 2.83 | 2.00 |
| | SMSE090283 | 90 mm × 283 mm | 0.24 | 0.07 | 0.06 |
| Self-sealing<br>Sterilization<br>Multi Pouches | SMSE097314 | 97 mm x 314 mm | 0.31 | 0.09 | 0.08 |
| | SMSE152314 | 152 mm x 314 mm | 0.41 | 0.12 | 0.10 |
| | SMSE228314 | 228 mm × 314 mm | 0.64 | 0.18 | 0.16 |
| | SMSE240335 | 240 mm x 335mm | 0.78 | 0.22 | 0.20 |
| Type | Model | Dimension in S.I. | *Content / Max Load (lbs.) | | |
| | | | Metal | Plastic | Linens and<br>Gauze |
| | SMFP075150 | 75 mm × 150 mm | 0.25 | 0.07 | 0.06 |
| | SMFP075200 | 75 mm × 200 mm | 0.36 | 0.10 | 0.09 |
| | SMFP075300 | 75 mm × 300 mm | 0.57 | 0.16 | 0.14 |
| | SMFP100200 | 100 mm × 200 mm | 0.55 | 0.16 | 0.14 |
| | SMFP100300 | 100 mm × 300 mm | 0.78 | 0.24 | 0.21 |
| | SMFP100550 | 100 mm × 550 mm | 1.32 | 0.44 | 0.33 |
| | SMFP120250 | 120 mm × 250 mm | 0.81 | 0.25 | 0.22 |
| | SMFP120280 | 120 mm × 280 mm | 0.92 | 0.28 | 0.25 |
| | SMFP150200 | 150 mm × 200 mm | 0.82 | 0.25 | 0.22 |
| | SMFP150250 | 150 mm × 250 mm | 1.07 | 0.33 | 0.29 |
| | SMFP150300 | 150 mm × 300 mm | 1.16 | 0.39 | 0.33 |
| | SMFP150380 | 150 mm × 380 mm | 1.51 | 0.50 | 0.43 |
| Sterilization<br>Pouches, Flat | SMFP150550 | 150 mm × 550 mm | 1.94 | 0.70 | 0.60 |
| | SMFP180300 | 180 mm × 300 mm | 1.44 | 0.48 | 0.41 |
| | SMFP200330 | 200 mm × 330 mm | 1.56 | 0.57 | 0.48 |
| | SMFP200400 | 200 mm × 400 mm | 1.94 | 0.70 | 0.60 |
| | SMFP200600 | 200 mm × 600 mm | 2.75 | 1.10 | 0.92 |
| | SMFP205400 | 205 mm × 400 mm | 2.48 | 1.10 | 0.90 |
| | SMFP250300 | 250 mm × 300 mm | 2.26 | 1.00 | 0.82 |
| | SMFP250450 | 250 mm × 450 mm | 2.61 | 1.04 | 0.87 |
| | SMFP300380 | 300 mm × 380 mm | 2.65 | 1.06 | 0.88 |
| | SMFP300460 | 300 mm × 460 mm | 2.98 | 1.49 | 1.19 |
| | SMFP300500 | 300 mm × 500 mm | 3.26 | 1.63 | 1.30 |
| | SMFP420400 | 420 mm × 400 mm | 3.32 | 2.21 | 1.66 |
| | SMFP420600 | 420 mm × 600 mm | 3.44 | 2.86 | 2.02 |
| Sterilization<br>Pouches,<br>Gusseted | SMGP100300 | 100 mm × 40 mm × 300 mm | 2.26 | 1.88 | 1.33 |
| | SMGP150400 | 150 mm × 50 mm × 400 mm | 2.55 | 2.13 | 1.50 |
| | SMGP200400 | 200 mm × 50 mm × 400 mm | 2.68 | 2.23 | 1.58 |
| | SMGP250480 | 250 mm × 60 mm × 480 mm | 3.06 | 2.55 | 1.80 |
| | SMGP300500 | 300 mm × 70 mm × 500 mm | 3.45 | 2.88 | 2.03 |
| Type | Model | Dimension in S.I. | *Content / Max Load (lbs.) | | |
| | | | Metal | Plastic | Linens and<br>Gauze |
| Sterilization<br>Rolls, Flat | SMFR 022 | 50 mm × 200 M | 0.26 | 0.08 | 0.07 |
| | SMFR902 | 50.8 mm × 200 M | 0.27 | 0.08 | 0.07 |
| | SMFR 032 | 75 mm × 200 M | 0.57 | 0.16 | 0.14 |
| | SMFR912 | 76.2 mm × 200 M | 0.58 | 0.17 | 0.15 |
| | SMFR 042 | 100 mm × 200 M | 0.87 | 0.25 | 0.22 |
| | SMFR922 | 101.6 mm × 200 M | 0.89 | 0.26 | 0.22 |
| | SMFR 062 | 150 mm × 200 M | 1.40 | 0.47 | 0.40 |
| | SMFR932 | 152.4 mm × 200 M | 1.43 | 0.48 | 0.41 |
| | SMFR 082 | 200 mm × 200 M | 1.86 | 0.75 | 0.62 |
| | SMFR942 | 203.2 mm × 200 M | 1.90 | 0.76 | 0.63 |
| | SMFR 102 | 250 mm × 200 M | 2.41 | 0.96 | 0.80 |
| | SMFR952 | 254 mm × 200 M | 2.45 | 0.98 | 0.82 |
| | SMFR 122 | 300 mm × 200 M | 2.77 | 1.38 | 1.11 |
| | SMFR962 | 304.8 mm × 200 M | 2.82 | 1.41 | 1.13 |
| | SMFR 142 | 350 mm × 200 M | 3.28 | 1.64 | 1.31 |
| | SMFR 162 | 400 mm × 200 M | 3.53 | 2.36 | 1.77 |
| | SMFR02-1 | 50.8 mm × 30.5 M | 0.27 | 0.08 | 0.07 |
| | SMFR03-1 | 76.2 mm × 30.5 M | 0.58 | 0.17 | 0.15 |
| | SMFR04-1 | 101.6 mm × 30.5 M | 0.89 | 0.26 | 0.22 |
| | SMFR06-1 | 152.4 mm × 30.5 M | 1.43 | 0.48 | 0.41 |
| | SMFR08-1 | 203.2 mm × 30.5 M | 1.90 | 0.76 | 0.63 |
| | SMFR10-1 | 254 mm × 30.5 M | 2.45 | 0.98 | 0.82 |
| Sterilization<br>Rolls,<br>Gusseted | SMGR 031 | 75 mm × 35 mm × 100 M | 2.21 | 1.84 | 1.30 |
| | SMGR 041 | 100 mm × 40 mm × 100M | 2.26 | 1.88 | 1.33 |
| | SMGR 061 | 150 mm × 50 mm × 100 M | 2.47 | 2.06 | 1.45 |
| | SMGR 081 | 200 mm × 50 mm × 100 M | 2.56 | 2.13 | 1.51 |
| | SMGR 101 | 250 mm × 60 mm × 100 M | 2.73 | 2.27 | 1.61 |
| | SMGR 121 | 300 mm × 70 mm × 100 M | 2.93 | 2.44 | 1.73 |
| | SMGR 141 | 350 mm × 80 mm × 100 M | 3.17 | 2.64 | 1.87 |
| | SMGR 161 | 400 mm × 80 mm × 100 M | 3.32 | 2.76 | 1.95 |
Table 1. The model numbers of SIGMA sterilization pouch and roll (Type, Model and Dimension and Content/Max. Load)
{8}------------------------------------------------
Table 1. (Continued) The model numbers of SIGMA sterilization pouch and roll (Type, Model and Dimension and Content/Max Load)
{9}------------------------------------------------
Table 1. (Continued) The model numbers of SIGMA sterilization pouch and roll (Type, Model and Dimension and Content/Max Load)
{10}------------------------------------------------
## Device Description
The SIGMA sterilization pouches and rolls are inserted into the Pouch/Roll, sealed, and then sterilized in the Sterilization System. After completion of the sterilization process, the Pouch/Roll maintain sterility of the enclosed medical devices until the seal is opened. The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices until used up to 3 years post EO gas sterilization and maintains the enclosed devices up until 6 months post Steam sterilization.
The Self-seal pouch permits sealing of the pouch without need of heat- sealing equipment, while the heat sealed pouches and rolls are heat sealed prior to processing in the steam/or EO Sterilization cycles.
The chemical indicators ink printed on the medical grade paper will exhibit a color change after the pouch is exposed to steam or ethylene oxide gas. The color of Chemical Indicator changes from Blue to Greenish Black, when exposed to Steam. And the color changes from red to yellow, when exposed to EO gas.
The Chemical Indicator offers an addition way to verity processing in the sterilization cycle. The Chemical Indicator should be used in addition to, not in place of, the biological indicator. The Chemical Indicators do not signify sterilization; they only indicate that the indicator has been exposed to Steam/or EO gas.
## Description of Comparison and Substantial Equivalence
A summary of the technological characteristics of the device subject of this premarket notification in comparison to those of the predicate devices is included in Table 2.
| Feature | Proposed device | Predicate device | | |
|-------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|
| Device name | SIGMA sterilization pouch and roll<br>(K202462) | SIGMA sterilization pouch and<br>roll (K180661) | Comparison | |
| Material<br>Composition | Medical Grade Paper, CPP, PET, PU adhesive, EO and Steam<br>Process Indicator, Print Ink | Medical Grade Paper, CPP, PET, PU<br>adhesive, EO and Steam Process<br>Indicator, Print Ink | Same | |
| Intended use | The SIGMA sterilization pouch and roll are intended to<br>provide health care workers with an effective method to<br>enclose devices intended for sterilization in the Steam<br>Sterilizer and via Ethylene Oxide (EO). The<br>recommended sterilization cycles are as follows:<br>• Gravity steam at 121 °C for 30 min; Drying time of 30 min.<br>• Gravity steam at 132°C for 15 min; Drying time of 30 min.<br>• Gravity steam at 135°C for 10 min; Drying time of 30 min.<br>• Pre-vacuum steam at 132°C for 4 min; Drying time of 20 min.<br>• Pre-vacuum steam at 135°C for 3 min; Drying time of 16 min.<br>• EO sterilization cycle is 4 hours at 55°C with a relative<br>humidity between 50%-85% and a sterilant concentration of<br>600 mg/L.<br>Furthermore, the sterilization pouch and roll maintains<br>the enclosed devices up until 3 years post EO gas<br>sterilization and maintains the enclosed devices up until<br>6 months post Steam sterilization. Lastly, the pouch's<br>external chemical ink indicators are designed to<br>indicate to the user that the pouch has undergone either<br>a steam or EO sterilization process. | The sterilization pouch and roll are<br>intended to provide health care<br>workers with an effective method to<br>enclose devices intended for<br>sterilization in steam auto claves and<br>via Ethylene Oxide (EO). The<br>recommended steam sterilization<br>cycle parameters are 30 minutes at<br>121°C. The recommended EO<br>sterilization cycle is 4 hours at 55°C<br>with a relative humidity between<br>50%-85% and a sterilant<br>concentration of 600 mg/L.<br>Furthermore, the sterilization pouch<br>and roll maintains the enclosed<br>devices up until 3 years post<br>sterilization. Lastly, the pouch's<br>external chemical ink indicators are<br>designed to indicate to the user that<br>the pouch has undergone either a<br>steam or EO sterilization process. | Only the Steam<br>recommended<br>sterilization cycles<br>are different.<br>The recommended<br>Steam sterilization<br>cycles of proposed<br>device increase 4<br>types of<br>sterilization cycle,<br>which are Gravity<br>132°C/15 min,<br>Gravity 135°C/10<br>min, Pre-vacuum<br>132°C/ 4 min and<br>Pre-vacuum<br>135°C/ 3 min. | |
| Feature | Subject device: SIGMA sterilization pouch and roll | | Predicate devices: SIGMA sterilization pouch and roll (K180661) | Comparison |
| Pouch Types | Self-sealing sterilization pouches Self-sealing sterilization multi pouches Sterilization pouches, Flat Sterilization pouches, Gusseted Sterilization rolls, Flat Sterilization rolls, Gusseted | | Self-sealing sterilization pouches Sterilization pouches, Flat Sterilization pouches, Gusseted Sterilization rolls, Flat Sterilization rolls, Gusseted | The proposed device increase one type of the sterilization pouch, which is "Self-sealing sterilization multi pouches". |
| Device models (Configurations/Dimensions) | Self-Sealing Sterilization Pouches | | Self-Sealing Sterilization Pouches | |
| | Size<br>L: 133 mm ~ 610 mm<br>W: 57 mm ~ 460 mm | Model<br>SMSE | Size<br>L: 133 mm ~ 610 mm<br>W: 57 mm ~ 460 mm<br>Model<br>SMSE | Same |
| | Self-Sealing Sterilization Multi Pouches | | N/A | The proposed device increase one type and the size is different. |
| | Size<br>L: 283 mm ~ 335 mm<br>W: 90 mm ~ 240 mm | Model<br>SMSE | | |
| | Sterilization Pouches, Flat | | Sterilization Pouches, Flat | Same |
| | Size<br>L: 150 mm ~ 600 mm<br>W: 75 mm ~ 420 mm | Model<br>SMFP | Size<br>L: 150 mm ~ 600 mm<br>W: 75 mm ~ 420 mm<br>Model<br>SMFP | |
| | Sterilization pouches, Gusseted | | Sterilization pouches, Gusseted | Same |
| | Size<br>L: 300 mm ~ 500 mm<br>W: 100 mm ~ 300 mm<br>H: 40 mm ~ 70 mm | Model<br>SMGP | Size<br>L: 300 mm ~ 500 mm<br>W: 100 mm ~ 300 mm<br>H: 40 mm ~ 70 mm<br>Model<br>SMGP | |
| | Sterilization rolls, Flat | | Sterilization rolls, Flat | Same |
| | Size<br>L: 30.5M ~ 200 M<br>W: 50 mm ~ 400 mm | Model<br>SMFR | Size<br>L: 30.5M ~ 200 M<br>W: 50 mm ~ 400 mm<br>Model<br>SMGP | |
| | Sterilization Rolls, Gusseted | | Sterilization Rolls, Gusseted | Same |
| | Size<br>L: ~100 M<br>W: 75 mm ~ 400 mm<br>H: 35 mm ~ 80 mm | Model<br>SMGR | Size<br>L: ~100 M<br>W: 75 mm ~ 400 mm<br>H: 35 mm ~ 80 mm<br>Model<br>SMGR | |
Table 2. Summary of the Proposed and Predicate Devices Technological Characteristics
{11}------------------------------------------------
| |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.