K110867 · Cpac, Inc. · FRG · Jul 20, 2011 · General Hospital
Device Facts
Record ID
K110867
Device Name
STERI-DENT, STERI-SURE
Applicant
Cpac, Inc.
Product Code
FRG · General Hospital
Decision Date
Jul 20, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The Steri-Dent Self-Seal Nylon Pouch is intended to be used to allow sterilization of enclosed medical instruments by a dry heat sterilizer operating at 160°C (320°F) for 1 hour (60 minutes) and also to maintain the sterility of the enclosed instruments for 30 days after sterilization.
Device Story
Single-use nylon pouches; designed for dry heat sterilization of metal instruments. Healthcare providers place instruments inside; seal using integrated adhesive strip. Pouches allow thermal energy transfer during 160°C (320°F) dry heat cycle (e.g., Steri-Dent 2100/3100); provide microbial barrier post-sterilization. Maintains sterility for 30 days. If unused within 30 days, instruments require reprocessing. Benefits include instrument protection and sterility maintenance until point of use.
Clinical Evidence
Bench testing only. Half-cycle efficacy studies on Steri-Dent 2100/3100 sterilizers confirmed heat penetration and sterilization. Aerosol challenge and 30-day shelf-life testing confirmed microbial barrier integrity. Biocompatibility testing confirmed non-toxicity.
Indicated for use by healthcare providers to enclose medical instruments for sterilization in a dry heat sterilizer (160°C for 60 minutes) and to maintain sterility for up to 30 days post-sterilization.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
# Section 5 - 510(k) Summary
# K 110867
JUL 20 2011
# Device Owner and Manufacturer
CPAC, Inc 2364 Leicester Rd Leicester, NY 14481
## Contact person
Jeffrey Crane 2364 Leicester Rd Leicester, NY 14481 Phone: 585-382-3223 585-382-3736 Fax:
Summary prepared on June 3, 2011
{1}------------------------------------------------
Device
Steri-Dent Self-Seal Nylon Pouch . Trade name:
Common name: Sterilization Wrap
Classification name: Wrap, Sterilization (21 CFR 880.6850, Product Code FRG)
### Predicate Device
The Steri-Dent Self-Seal Nylon Pouches are substantially equivalent to the Winner Self Seal Sterilization Pouch both of which are single use sterilization wraps. The predicate device is legally marketed under 510(k) K051242.
# Device Description
The Steri-Dent Self-Seal Nylon Pouches are single use sterilization wraps that provide a means for health care providers to enclose metal instruments, which are sterilized in a dry heat sterilizer, and to maintain sterility of the instruments until they are used. The pouches can be used in dry heat sterilizers with a sterilization cycle of 1 hour (60 minutes) operating at 160°C (320°F) such as the Steri-Dent Model 2100 and 3100. The nylon pouches will maintain the sterility of the instruments for a maximum of 30 days. If the instruments inside the pouch are not used within 30 days after being sterilized, they should be removed, placed in a new bag and resterilized.
#### Indications for Use
The Steri-Dent Self-Seal Nylon Pouch is intended to be used to allow sterilization of enclosed medical instruments by a dry heat sterilizer operating at 160°C (320°F) for 1 hour (60 minutes) and also to maintain the sterility of the enclosed instruments for 30 days after sterilization.
{2}------------------------------------------------
# Technological Characteristics Comparison
# Below is a comparison of the technological characteristics of the Winner Self Seal Sterilization Pouch and the Steri-Dent Self-Seal Nylon Pouch
| Characteristic | Predicate Device - Winner® | Device - Steri-Dent |
|---------------------------|--------------------------------------------------|--------------------------|
| Single Use or Reuse | Single Use | Single Use |
| Pouch Material | Medical Grade Paper and<br>plastic film | Nylon |
| Seal Method | Self Seal adhesive strip | Self Seal adhesive strip |
| Sterilization Method | Steam or ethylene oxide (EtO) | Dry Heat |
| Sterilization Temperature | Steam - 121°C (250°F)<br>EtO - no recommendation | 160°C (320°F) |
| Sterilization Time | Steam - 15 minutes<br>EtO - no recommendation | 1 hr (60 minutes) |
| Smallest pouch size | 3" x 8" | 2" x 10" |
| Largest pouch size | 18" x 22" | 9.5" x 13" |
The technological difference between the Steri-Dent device and the predicate device is material differences due to the type of sterilization method used. The predicate device is designed to be used for steam or gas (EtO) sterilization and so must provide a means for the steam and gas to penetrate in and out of the pouch. This is accomplished by using a medical grade paper on one side of the pouch to allow this transfer of air. The Steri-Dent nylon pouch is designed to be used for dry heat sterilization and so does not require air to penetrate the pouch, only thermal energy. This technological difference allows the new device to be made of entirely of nylon which allows for the transfer of heat in and out of the pouch while providing an excellent microbial barrier.
# Non-Clinical Performance Data
# Safety
The biocompatibility tests demonstrated that the pouch materials are not toxic to the health care provider or the patient.
# Effectiveness of Sterilization of Enclosed Instruments
Half cycle efficacy studies performed on both the model 2100 and 3100 dry heat sterilizers using the nylon pouches demonstrated that the heat penetrated through the pouch and sterilized the instruments inside.
# Effectiveness of Maintaining Sterility of Enclosed Instruments
The aerosol challenge test, 30 day shelf life test and material compatibility tests demonstrated that the pouches maintained the sterility of the enclosed instruments.
{3}------------------------------------------------
#### Conclusion
The non-clinical tests demonstrate that the nylon pouch is as safe and effective as the predicate device in providing health care providers a single use pouch to keep sterilized medical instruments sterile until their use. The pouch is made entirely of nylon with an adhesive seal tape which the biocompatibility tests prove to be safe for direct patient contact. The half cycle tests performed using a dry heat sterilizer demonstrates that the pouch allows for the transfer of heat to enable sterilization of the enclosed instruments. The ability of the nylon pouch to maintain the sterility of the enclosed instruments after sterilization is proven by a number of tests. The material compatibility shows the high strength and puncture resistance of the nylon material. The 30 day shelf life and aerosol challenge tests show the nylon material and adhesive seal tape provide the microbial barrier required to keep the instruments sterile.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo for the Department of Health & Human Services. The logo features the department's name in a circular arrangement on the left side. To the right of the text, there is a stylized symbol consisting of three overlapping, curved shapes that resemble abstract human figures or waves. The logo is presented in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Jeffery S. Crane Engineering Project Manager CPAC, Incorporated 2364 Leicester Road P.O. Box 175 Leicester, New York 14481
# JUL 20 2011
Re: K110867
Trade/Device Name: Steri-Dent / Steri-Sure Self Seal Nylon Pouches Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: FRG Dated: July 5, 2011 Received: July 6, 2011
Dear Mr. Crane:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
# Page 2 – Mr. Crane
Please be advised that FDA's issuance of a substantial equivalence determination does not · mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of . the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutF DA/CentersOffices/CDR H/CDRHOffices/uc m115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony O. Winter
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
501(k) Number:
# K110867
Device Names: Steri-Dent / Steri-Sure Self-Seal Nylon Pouches
# Indications for Use
The Steri-Dent / Steri-Sure Self-Seal Nylon Pouch is intended to be used to allow sterilization of enclosed medical instruments by a dry heat sterilizer operating at 160°C (320°F) for 1 hour (60 minutes) and also to maintain the sterility of the enclosed instruments for 30 days after sterilization.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use XX (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY),
Concurrence of CDRH, Office of Device Evaluation (ODE)
Cohie Langston
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Sec. 4 -1
Confidential Document CPAC, Inc.
510(k) Number: K110867