# KIMGUARD ONE-STEP STERILIZATION WRAP (K082177)

_Kimberly-Clark Corp. · FRG · Mar 27, 2009 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/FRG/K082177

## Device Facts

- **Applicant:** Kimberly-Clark Corp.
- **Product Code:** [FRG](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/FRG.md)
- **Decision Date:** Mar 27, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6850
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

KIMGUARD ONE-STEP* Sterilization Wrap is intended to be used to enclose another medical device that is to be sterilized by a health care provider by pre-vacuum steam at 270°F/132°C for 4 minutes or by 100% ethylene oxide (EtO) with a concentration of 725-735 mg/L at 131°F/ 55°C and 40% - 80% relative humidity for 60 minutes. The wrap is intended to allow sterilization of the enclosed medical device(s) and also to maintain sterility of the enclosed device(s) until opened. The wrap was validated for aeration times for EO sterilization of 8 hours at 55 °C or 12 hours at 43.3 °C. The wrap was validated for dry times for pre-vacuum steam sterilization of 20 minutes for Models 100 and 200, and for 30 minutes for Models 400, 500, and 600.

## Device Story

Sterilization wrap system; consists of two sheets of polypropylene fabric ultrasonically seamed on two sides; enables simultaneous two-sheet wrapping of medical devices. Used in healthcare facilities by sterile processing staff. Input: medical devices/trays requiring sterilization. Process: device enclosed in wrap; subjected to pre-vacuum steam or EtO sterilization; wrap maintains sterility of contents post-sterilization until opened. Output: sterile, wrapped medical device. Benefits: provides barrier protection to maintain sterility of instruments until point of use.

## Clinical Evidence

Bench testing only. Testing performed per FDA guidance for medical sterilization packaging systems. Included biocompatibility (ISO 10993 irritation/sensitization), sterilant penetration, dry time, and physical integrity. Sterility maintenance validated for up to 30 days post-sterilization. All results met acceptance criteria.

## Technological Characteristics

Three-layer SMS (spunbond-meltblown-spunbond) polypropylene laminate. Two sheets ultrasonically seamed. Models KC100-KC600 vary by weight capacity (3 lbs to 25 lbs).

## Regulatory Identification

A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.

## Predicate Devices

- KIMGUARD* Regular and Heavy Duty Sterile Wraps ([K881471](/device/K881471.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K082107

# 510(k) Summary for the Kimberly-Clark* Corporation KIMGUARD ONE-STEP* Sterilization Wrap (Models KC100, KC200, KC300, KC400, KC500, and KC600)

| Date Summary was Prepared:                  | March 26, 2009<br>MAR 2 7 2009                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Submitter:                           | Thomas Kozma<br>Director, Regulatory Affairs<br>Kimberly-Clark Health Care<br>1400 Holcomb Bridge Road<br>Roswell, GA 30076<br>Ph: 770.587.8393<br>FAX: 920.225.3408<br>Email: thomas.kozma@kcc.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Primary Contact for this 510(k) Submission: | Lisa Peacock, Consultant to Kimberly-Clark Health Care<br>SciMed, Inc.<br>Ph: 706.216.3413<br>FAX: 800.713.7754<br>Email: lisa.peacock@kcc.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Device Common Name:                         | Sterilization Wrap                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Classification Name:                        | Sterilization Wrap (21 CFR 880.6850)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Product Code:                               | FRG                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Intended Use:                               | KIMGUARD ONE-STEP* Sterilization Wrap is intended to be used to<br>enclose another medical device that is to be sterilized by a health care<br>provider by pre-vacuum steam at 270°F/132°C for 4 minutes or by<br>100% ethylene oxide (EtO) with a concentration of 725-735 mg/L at<br>131°F/ 55°C and 40% - 80% relative humidity for 60 minutes. The wrap<br>is intended to allow sterilization of the enclosed medical device(s) and<br>also to maintain sterility of the enclosed device(s) until opened. The<br>wrap was validated for aeration times for EO sterilization of 8 hours at<br>55 °C or 12 hours at 43.3 °C. The wrap was validated for dry times for<br>pre-vacuum steam sterilization of 20 minutes for Models 100 and 200,<br>and for 30 minutes for Models 400, 500, and 600. |
|                                             | KIMGUARD ONE-STEP* Sterilization Wrap is not indicated for use for gravity steam sterilization.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                             | *The KC300 Model KIMGUARD ONE-STEP* Sterilization Wrap is not indicated for use for pre-vacuum steam sterilization.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                             | See Wrap Model Recommendations on Page 3.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |

Page 1 of 3 - 510(k) Summary

:

{1}------------------------------------------------

| Predicate<br>Devices:       | The KIMGUARD ONE-STEP* Sterilization Wrap<br>(Models KC100, KC200, KC300, KC400, KC500, and KC600) are<br>substantially equivalent to the predicate KIMGUARD* Regular and<br>Heavy Duty Sterile Wraps (K881471)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Substantial<br>Equivalence: | The KIMGUARD ONE-STEP* Sterilization Wrap is substantially<br>equivalent to the predicates in intended use, design, and materials.<br>The predicate devices were constructed of a three-layer laminate<br>composed of a layer of meltblown polypropylene bonded on both<br>surfaces with a layer of spunbonded polypropylene. The KIMGUARD<br>ONE-STEP* Sterilization Wrap is comprised of two sheets of<br>KIMGUARD* Sterilization Wrap ultrasonically seamed on two sides.<br>This allows for convenient wrapping with two sheets simultaneously.<br>The sheets of sterilization wrap are square or rectangular fabric<br>produced using a polypropylene three-layer SMS (spunbond-<br>meltblown-spunbond) process. |
| Summary of<br>Testing:      | KIMGUARD ONE-STEP* Sterilization Wrap performance has been<br>tested in accordance with the applicable requirements recommended in<br>Premarket Notification [510(k)] Submissions for Medical Sterilization<br>Packaging Systems in Health Care Facilities; Draft Guidance for<br>Industry and FDA (March 7, 2002). Testing included biocompatibility<br>(i.e., irritation and sensitization) in compliance with the methods of ISO<br>10993, sterilant penetration, dry time, and physical integrity. The Wrap<br>has also been tested for the ability to maintain sterility of pack contents<br>after sterilization for up to 30 days under standard conditions. All<br>results of testing met acceptance criteria. |

and the comments of the country

・

:

.

:

{2}------------------------------------------------

| Wrap Model Recommendations<br>KIMGUARD<br>ONE-STEP*<br>Sterilization Wrap<br>Models | Intended Loads                                                                           | Maximum<br>Wrapped<br>Package<br>Content<br>Weights Used<br>in Sterility<br>Maintenance<br>Validation<br>Study2 | Descriptions of Loads Used in<br>Sterility Maintenance Validation<br>Study2                                                                                   |
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| KC100                                                                               | Very Light Weight<br>Package (for<br>example: towel<br>packs)                            | 3 lbs                                                                                                           | 16 huck towels (17"x 29")                                                                                                                                     |
| KC200                                                                               | Light Weight<br>Package (for<br>example: standard<br>linen packs)                        | 6 lbs                                                                                                           | 2 huck towels (17"x 29")<br>2 fluid resistant U-drape (68"x109")<br>1 fluid resistant universal bar drape<br>(70" x 108")                                     |
| KC3003                                                                              | Light to Moderate<br>Weight Package (for<br>example: general use<br>medical instruments) | 9 lbs                                                                                                           | FOR EO:<br>16 huck towels<br>2 fluid resistant large drapes (76"x100")<br>1 fluid resistant small drape (76"x60")<br>1 fluid resistant table cover (60"x 90") |
| KC4004                                                                              | Moderate to Heavy<br>Weight Package (for<br>example: general use<br>medical instruments) | 13 lbs                                                                                                          | 4 tray liners 20" x 25" stacked<br>10" x 10" x 3 ½ " tray containing 11 lbs<br>of metal mass                                                                  |
| KC5004                                                                              | Heavyweight<br>Package (for<br>example: general use<br>medical instruments)              | 17 lbs                                                                                                          | 4 tray liners 20" x 25" stacked<br>10" x 10" x 3 ½ " tray containing<br>15 lbs of metal mass                                                                  |
| KC6004                                                                              | Very Heavy Weight<br>Package (for<br>example: general<br>use medical<br>instruments)     | 25 lbs                                                                                                          | 4 tray liners 20" x 25" stacked<br>10" x 10" x 3 ½ " tray containing<br>23 lbs of metal mass                                                                  |

y individual results may differ the time in handling practices wraping techniques, and folding
 individual results and the transferenced costents, which may nut added stress Individual results may differ due to firegularly handle proceded to the may all and and the may and

Results may also differ due to the on irregulary shaped content, which intended use.
facility should determine for itself which wrap model is most appropriated for on 
first recommended to not exceed the most a most appoprate for each micriced do.
 I it is recommended to not except persize content weights included for the KMGUARD lices for mended to not exceed the number individual content includes hat were video hour model house house house house house house house hour mass. is recommended to not exceed the number and size of the weight of the weigh mass).
Sterilization Wraps (i.e.: the number and size of the weight of the weigh mass).
Sterilized

Sterilization Wraps (i.e.: the number and size of the
pletion Wran is not indicated for use for pre-vacuum steam sterilization.

Sterlization Wraps (i.e. the number and indicated for use for use for use for use for use includination
 Sterlization West Medication Wes is not indicated for use for use for "The KC300 Model KiMGUARD Waps were validated to the Piction Mult S bsorne find heatmal inen, and i
"The KC400, KC500, and KC600 model was were validation will 3 lbs models. The KC50, and KC600 modelwaps were validated to stelling syles with helt below in the recommended that is is ecommended to not exeed 3 the find resistant intenting of the min these much in the common
the user not include fluit not been evaluated with these models.

{3}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized representation of an eagle.

#### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

### MAR 2 7 2009

Kimberly-Clark Corporation C/o Ms. Lisa Peacock, RAC SciMed, Incorporated 172 Conductor Drive Dawsonville, Georgia 30534

Re: K082177

> Trade/Device Name: KIMGUARD ONE-STEP* Sterilization Wrap (Models KC100, KC200, KC300, KC400, KC500, and KC600)

Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: FRG Dated: March 9, 2009 Received: March 10, 2009

Dear Ms. Peacock:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

#### Page 2 - Ms. Peacock

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Antony D. morton bon

Ginette Y. Michaud, M.D. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

### Indications for Use

510(k) Number (if known): _K082177

#### KIMGUARD ONE-STEP* Sterilization Wrap (Models KC100, KC200, Device Name: KC300, KC400, KC500, and KC600)

Indications for Use:

KIMGUARD ONE-STEP* Sterilization Wrap is intended to be used to enclose another medical device that is to be sterilized by a health care provider by pre-vacuum steam at 2709F/132°C for 4 minutes or by 100% ethylene oxide (EtO) with a concentration of 725-735 mg/L at 131°F/ 55°C and 40% - 80% relative humidity for 60 minutes. The wrap is intended to allow sterilization of the enclosed medical device(s) and also to maintain interility of the enclosed device(s) until opened. The wrap was validated for aeration sterilized the oncliebed as not (9) or 12 hours at 43.3 ℃. The wrap was times for 20 cry times for pre-vacuum steam sterilization of 20 minutes for Models 100 and 200, and for 30 minutes for Models 400, 500, and 600.

KIMGUARD ONE-STEP* Sterilization Wrap is not indicated for use for gravity steam sterilization.

*The KC300 Model KIMGUARD ONE-STEP* Sterilization Wrap is not indicated for use for pre-vacuum steam sterilization.

See Wrap Model Recommendations on Page 2.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

× Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Shiela H. Murphy, D

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K0821

{6}------------------------------------------------

#### Indications for Use

| KIMGUARD<br>ONE-STEP*<br>Sterilization Wrap<br>Models | Intended Loads1                                                                          | Maximum<br>Wrapped<br>Package<br>Content<br>Weights Used<br>in Sterility<br>Maintenance<br>Validation<br>Study2 | Descriptions of Loads Used in<br>Sterility Maintenance Validation<br>Study2                                                                                   |
|-------------------------------------------------------|------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| KC100                                                 | Very Light Weight<br>Package (for<br>example: towel<br>packs)                            | 3 lbs                                                                                                           | 16 huck towels (17"x 29")                                                                                                                                     |
| KC200                                                 | Light Weight<br>Package (for<br>example: standard<br>linen packs)                        | 6 lbs                                                                                                           | 2 huck towels (17"x 29")<br>2 fluid resistant U-drape (68"x109")<br>1 fluid resistant universal bar drape<br>(70" x 108")                                     |
| KC3003                                                | Light to Moderate<br>Weight Package (for<br>example: general use<br>medical instruments) | 9 lbs                                                                                                           | FOR EO:<br>16 huck towels<br>2 fluid resistant large drapes (76"x100")<br>1 fluid resistant small drape (76"x60")<br>1 fluid resistant table cover (60"x 90") |
| KC4004                                                | Moderate to Heavy<br>Weight Package (for<br>example: general use<br>medical instruments) | 13 lbs                                                                                                          | 4 tray liners 20" x 25" stacked<br>10" x 10" x 3 ½ " tray containing 11 lbs<br>of metal mass                                                                  |
| KC5004                                                | Heavyweight<br>Package (for<br>example: general use<br>medical instruments)              | 17 lbs                                                                                                          | 4 tray liners 20" x 25" stacked<br>10" x 10" x 3 ½ " tray containing<br>15 lbs of metal mass                                                                  |
| KC6004                                                | Very Heavy Weight<br>Package (for<br>example: general use medical<br>instruments)        | 25 lbs                                                                                                          | 4 tray liners 20" x 25" stacked<br>10" x 10" x 3 ½ " tray containing<br>23 lbs of metal mass                                                                  |

Individual results may differ due to factors such as variations in handling practices, wrapping techniques, and folling methods individual results may differ due its valiation minding phavos, magend syres on the wrap. Each healthcare facility should determine for itself which wrap model is most appropriate for each intended use.

 it is recommended to not exceed the may model is nice layed in cached for rach wrap model. Futhermore, it It is recommended to not exceed the maximum wiapped passis of individual ontent types fible weets in the KIMGUARD
is recommended to not exceed the number, weight, and size of is recommended to fiel exceed the number and size of the fluid resistant intense or the weight of the weight of the mediated for use yours to

 Stemization Waps (t.e. nie number and the senior of the endony stem sterillzation.
The KC300 Model KMGUARD Were valiaties to the for endrum that resisten in en and in The KC400, KC500, and NC800 mlade resident in stellization cycles with his or notel. is recommended to fot exceed 3 bs of for-libre in stamization your his and model wraps, as use of such filled resistant materials has not been evaluated with these models.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/FRG/K082177](https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/FRG/K082177)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).