K200500 · Molekule, Inc. · FRA · Apr 15, 2020 · General Hospital
Device Facts
Record ID
K200500
Device Name
Molekule Air Pro RX
Applicant
Molekule, Inc.
Product Code
FRA · General Hospital
Decision Date
Apr 15, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Molekule Air Pro RX air purifier is a device intended for medical purposes that is used to destroy bacteria and viruses in the air by exposure to ultraviolet radiation. The core technology components of the Molekule Air Pro RX air purifier have been demonstrated to destroy the following MS2 bacteriophage bioaerosol entrained on the filter of the subject device under the following exposure conditions: Organism Name Average Maximum log reduction /exposure time (hours) Room temperature test Virus MS2 bacteriophage 5.21 / 24 hours
Device Story
Freestanding air purifier for medical facilities; operated by healthcare professionals. Uses centrifugal fan to draw air through pre-filter and catalytic filter. Employs photo-electrochemical oxidation (PECO) technology; UV-A light (320-400 nm) activates photocatalyst on filter surface; generates hydroxyl radicals to oxidize captured bacteria, viruses, and mold. Output is purified air. Safety features include door interlocks to prevent UV exposure and protective fan grilles. Benefits patient by reducing airborne microbiological contaminants in clinical environments.
Clinical Evidence
No clinical data. Bench testing only. Performance characterized by MS2 bacteriophage log reduction on filter media: 3.41 log reduction after 1 hour; 5.21 log reduction after 24 hours at room temperature.
Technological Characteristics
PECO air purification; UV-A (320-400 nm) LED light source; catalytic filter media. Centrifugal fan; 4-speed flow control (300-800 CFM). Safety: UL 507 (fans), IEC 60601-1-2 (EMC). Dimensions: 22" x 22" x 52". Power: 120V, 3.72A, 450W max.
Indications for Use
Indicated for use in medical facilities, including hospital and general surgery settings, to destroy bacteria and viruses in the air via ultraviolet radiation.
Regulatory Classification
Identification
A medical ultraviolet air purifier is a device intended for medical purposes that is used to destroy bacteria in the air by exposure to ultraviolet radiation.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
April 15, 2020
Molekule, Inc. % Adrienne Lenz Senior Medical Device Regulation Expert Hyman, Phelps, & McNamara, P.C. 700 Thirteenth Street N.W., Suite 1200 Washington, District of Columbia 20005
Re: K200500
Trade/Device Name: Molekule Air Pro RX Regulation Number: 21 CFR 880.6500 Regulation Name: Medical Ultraviolet Air Purifier Regulatory Class: Class II Product Code: FRA Dated: February 27, 2020 Received: February 28, 2020
Dear Adrienne Lenz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
CAPT Elizabeth F. Claverie, MS Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
#### K200500
Device Name Molekule Air Pro RX
#### Indications for Use (Describe)
The Molekule Air Pro RX air purifier is a device intended for medical purposes that is used to destrov bacteria and viruses in the air by exposure to ultraviolet radiation.
The core technology components of the Molekule Air Pro RX air purifier have been demonstrated to destroy the following MS2 bacteriophage bioaerosol entrained on the filter of the subject device under the following exposure conditions:
| Organism | Name | Average Maximum log reduction<br>/exposure time (hours) |
|----------|-------------------|---------------------------------------------------------|
| | | Room temperature test |
| Virus | MS2 bacteriophage | 5.21 / 24 hours |
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
[X] Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
# MOLEKULE AIR PRO RX
{4}------------------------------------------------
## 510(k) Summary
In accordance with 21 C.F.R. § 807.92 the following summary of information is provided:
DATE: April 15, 2020
## SUBMITTER:
Molekule, Inc. 1301 Folsom St San Francisco, CA 94103 T 855-999-9069
### PRIMARY CONTACT PERSON:
Adrienne R. Lenz Senior Medical Device Regulation Expert Hyman, Phelps, & McNamara, P.C. T 202-737-4292
## SECONDARY CONTACT PERSON:
Frank Bianco FDA Compliance Molekule, Inc. T 925-404-7724
### DEVICE:
TRADE NAME: Molekule Air Pro RX COMMON/USUAL NAME: Air Purifier CLASSIFICATION NAMES: Purifier, Air, Ultraviolet, Medical REVIEW PANEL: General Hospital PRODUCT CODE: FRA
PREDICATE DEVICE(S):
Transformair Indoor Air Purifier, K161468
{5}------------------------------------------------
Molekule Inc. is the current holder of the Transformair 510(k). This predicate has not been subject to a design-related recall. The ABRACAIR (K052732) is used as a reference device as, like the Molekule Air Pro RX, it is a freestanding device.
# DEVICE DESCRIPTION:
The Molekule Air Pro RX air purifier employs a photo-electrochemical oxidation (PECO) ultraviolet air purification technology that destroys bacteria, viruses and mold in air in medical facilities. The Molekule Air Pro RX air purifier includes a pre-filter and low energy ultraviolet lights (UV-A 320-400 nm), and a catalytic filter.
## INTENDED USE:
The Molekule Air Pro RX air purifier is a device intended for medical purposes that is used to destroy bacteria and viruses in the air by exposure to ultraviolet radiation.
The core technology components of the Molekule Air Pro RX air purifier have been demonstrated to destroy the following MS2 bacteriophage bioaerosol entrained on the filter of the subject device under the following exposure conditions:
| Organism | Name | Average Maximum log reduction<br>/exposure time (hours)<br>Room temperature test |
|----------|-------------------|----------------------------------------------------------------------------------|
| Virus | MS2 bacteriophage | 5.21 / 24 hours |
# TECHNOLOGY:
The Molekule Air Pro RX air purifier employs a photo-electrochemical oxidation (PECO) ultraviolet air purification technology that destroys bacteria in air in medical facilities. It is designed using the technology platform from the Transformair Air Purifier. The Molekule Air Pro RX air purifier includes a pre-filter, low energy ultraviolet lights (UV-A 320-400 nm), and a catalytic filter.
PECO is an air purification technology that fully oxidizes bacteria, viruses, and mold. PECO works by shining UV-A light on the surface of the catalytic filter to initiate a chemical reaction that generates hydroxyl radicals. These radicals combine with microbiological contaminants, such as bacteria, viruses, and mold spores that are captured on the filter. Once combined, a chemical reaction takes place killing the contaminants.
{6}------------------------------------------------
Unlike the Transformair Air Purifier which is installed in a vent, the Molekule Air Pro RX air purifier is a freestanding device. The following table summarizes the similarities and differences between the subject and predicate devices.
| | Transformair air purifier (K161468) | Molekule Air Pro RX air purifier (K200500) | | | | | | | | | | | |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|--|--|--|--|--|--|--|--|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Product Name | Transformair Air purifier | Molekule Air Pro RX Air purifier | | | | | | | | | | | |
| Device Type | Medical Ultraviolet Air purifier | Medical Ultraviolet Air purifier | | | | | | | | | | | |
| Product Code | FRA | FRA | | | | | | | | | | | |
| Classification Regulation | 21 C.F.R. § 880.6500 | 21 C.F.R. § 880.6500 | | | | | | | | | | | |
| Class | II | II | | | | | | | | | | | |
| Patient Population | General Hospital | General Hospital | | | | | | | | | | | |
| User | Healthcare Professional | Healthcare Professional | | | | | | | | | | | |
| Indications for Use | The Transformair Indoor Air Purifier, In Duct Model 16108 is a device intended for medical purposes that is used to destroy bacteria in the air by exposure to ultraviolet radiation.<br><br>Transformair Indoor Air Purifier, In Duct Model 16108 has been demonstrated to destroy Staphylococcus epidermidis, Escherichia coli, MS2, Phi-X174, Aspergillus Niger and Bacillus globigii entrained on the filter of the subject device under the following exposure conditions:<br><br>Average Maximum log reduction /exposure time (hours)<br>Test temperature<br>45 °F 72 °F 110 °F Virus, MS2 bacteriophage<br>4.13 / 24 hours 4.25 / 24 hours 5.51 / 24 hours Bacteria, Staphylococcus epidermidis<br>4.88 / 24 hours 4.02 / 0.33 hours 4.20 / 0.33 hours | | | | | | | | | | The Molekule Air Pro RX air purifier is a device intended for medical purposes that is used to destroy bacteria and viruses in the air by exposure to ultraviolet radiation.<br><br>The core technology components of the Molekule Air Pro RX air purifier have been demonstrated to destroy the following MS2 bacteriophage bioaerosol entrained on the filter of the subject device under the following exposure conditions:<br><br>Average Maximum log reduction /exposure time (hours)<br>Room temperature Virus, MS2 bacteriophage<br>5.21 / 24 hours | | |
{7}------------------------------------------------
| | Transformair air purifier<br>(K161468) | Molekule Air Pro RX air<br>purifier (K200500) |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Bacteria, Escherichia coli<br>4.31 / 24 4.79 / 24 4.40 / 0.33<br>hours hours hours | |
| | Virus, Phi-X174 bacteriophage<br>4.37 / 24 4.37 / 24 4.37 / 24<br>hours hours hours | |
| | Mold endospore, Aspergillus Niger<br>3.91 / 72 3.99 / 72 4.22 / 72<br>hours hours hours | |
| | Bacterial endospore, Bacillus globigii<br>4.11 / 72 4.41 / 72 4.41 / 72<br>hours hours hours | |
| Environment of Use | Hospital and general surgery<br>setting | Hospital and general surgery<br>setting |
| User Control | HVAC fan speed controls the<br>air flow | One knob controls the four-<br>speed fan setting<br>One button turns the unit on<br>and off. |
| Software | None. The device is on<br>whenever power is connected. | Basic Firmware, used to turn<br>the unit on, off, and change fan<br>speed. |
| Mechanism of Action | UV light of sufficient energy<br>(UV-A) activates photocatalyst<br>that destroys microorganisms<br>entrained on the filter through a<br>photochemical reaction. | UV light of sufficient energy<br>(UV-A) activates photocatalyst<br>that destroys microorganisms<br>entrained on the filter through a<br>photochemical reaction. |
| Installation | In-duct | Free standing |
| Pre-Filter(s) | Synthetic Media for<br>mechanical filtration<br>upstream of the PECO<br>filter. Dimensions 21.25 in. x 26<br>in. x 2 in. Pleats per inch 1.25 Total Filter surface area<br>2800 sq. in. | Synthetic Media for<br>mechanical filtration<br>upstream of the PECO<br>filter. Dimensions 20 in. x 20 in.<br>x 4 in. Pleats per inch 1.5 Total Filter surface area<br>4800 sq. in. |
| | Transformair air purifier<br>(K161468) | Molekule Air Pro RX air<br>purifier (K200500) |
| Catalytic Filter | Proprietary filter media Dimensions 21.25 in. x 26 in. x 2 in. Pleats per inch 1.25 Total Filter surface area 2800 sq. in. Filter coated with proprietary photocatalyst and a wire-mesh | Proprietary filter media Dimensions 20 in. x 20 in. x 4 in. Pleats per inch 1.5 Total Filter surface area 4800 sq. in. Filter coated with the proprietary photocatalyst and a wire-mesh |
| Photocatalyst | Proprietary catalyst | Same proprietary catalyst |
| Light Source | UV Light Source, black light fluorescent bulbs Wavelength, 320-400nm (See Figure 13.3) Power per Lamp/ String, 17W Number of Lamps/ String, 6 Total UV Power, 102W Filter Irradiance (Minimum), 25W/m² | UV Light Source, LED Wavelength, 320-400nm Power per Lamp/ String, 11.4W Number of Lamps/ String, 6 Total UV Power, 68.4W Filter Irradiance (Minimum), 30W/m² |
| Air Source | HVAC fan | Centrifugal Fan |
| Flow Control | Building HVAC system | Four speeds (low, medium, high, boost) provide 300 - 800 CFM flow. |
| Air Changes Per<br>Hour | Standard building HVAC of 2-3 ACH (air changes per hour) when used in building volume of 45600 cubic feet with flowrates of 1300-1950 CFM. | 6-9 ACH when used in room with volume of 4000 cubic feet (a typical Operating Room Volume) with flowrates of 450-650 CFM at settings 2 and 3. |
| | Transformair air purifier<br>(K161468) | Molekule Air Pro RX air<br>purifier (K200500) |
| UV Light Exposure<br>Safety Features | Installed in duct away from<br>user, none required. | Safety switches exist in the<br>following locations: PECO<br>filter door, pre filter door,<br>PECO filter compartment, and<br>pre filter compartment. If any<br>door is open or if a filter is<br>missing, the unit will not<br>operate. The purpose of these<br>switches is to protect the user<br>from any possibility of<br>exposure to direct contact with<br>UV light. Safety feature<br>confirmed by UL507. |
| Fan Exposure Safety<br>Features | Installed in duct without fan,<br>none required. | Grill at outlet and inlet of fan<br>with small enough grating to<br>block user from accessing<br>spinning fan without tools.<br>Safety feature confirmed by<br>UL507. |
| Voltage | 120 Volt<br>(plugs into standard outlet: no<br>switch, always on) | 120 Volt<br>(plugs into standard single<br>phase 120 Volt outlet) |
| Current | Up to 1.25 amps | Up to 3.72 amps |
| Power Consumption | Up to 240 Watts | Up to 450 Watts |
| Dimensions | Outer frame dimensions:<br>28.5" x 21" x 12.25"<br>Filter dimensions :<br>Filter: 7-20" x 7-26" x 5"<br>Pre-Filter: 7-20" x 7-26" x 2" | Outer frame dimensions:<br>22" x 22" x 52"<br>Filter dimensions:<br>Filter: 20" x 20" x 4"<br>Pre-Filter: 20" x 20" x 4" or<br>20″ x 20″ x 2″ |
| | Transformair air purifier<br>(K161468) | Molekule Air Pro RX air<br>purifier (K200500) |
| Standards | ANSI/UL 1995-2011 & CAN-<br>CSA C22.2 No. 236-11,<br>Heating and Cooling<br>Equipment and UL 2043, Heat<br>and Visible Smoke Release For<br>Discrete Products and Their<br>Accessories Installed in Air-<br>Handling Spaces. | UL 507 Standard for Electrical<br>Fans<br>IEC 60601-1-2 EMC. EMC for<br>Medical Devices. |
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
# SUMMARY OF NON-CLINICAL TESTS:
The Molekule Air Pro RX air purifier complies with voluntary standards for electrical safety and electromagnetic compatibility. The following were provided in support of the substantial equivalence determination:
- Risk Analysis ●
- Software verification and validation testing and software information ● recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"
- Electrical safety and electromagnetic compatibility testing per UL 507:2017 . Electric Fans and IEC 60601-1-2: 2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests, respectively
- Performance tests for UV light leakage and light intensity ●
- Fluid modeling to confirm the system does not impact laminar flow systems ●
- Testing to demonstrate filtration performance and destruction of bacteria ●
- Third party testing was performed to characterize the kill rate kinetics for MS2 ● bacteriophage deposited onto the surface of Air Pro RX's proprietary photocatalytic coated filter media after exposure to Air Pro RX's UV-A light. The Air Pro RX Catalytic media effectively kills viable MS2 bacteriophage at room temperature. After one hour of UV exposure, an average of 3.41 LOG reduction was observed. After 24 hours of exposure, a 5.21 LOG reduction was observed.
SUMMARY OF CLINICAL TESTS: Not applicable.
{11}------------------------------------------------
## CONCLUSION:
Based on the intended uses, technological characteristics and non-clinical performance data, the Molekule Air Pro RX is as safe, as effective, and performs at least as well as the predicate,Transformair air purifier, cleared under K161468.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.