PADDED MITTS/PADDED-PLUS MITTS/FINGER SEPARATOR PADDED MITTS/RIGID PALM PADDED MITTS

K963458 · Skil-Care Corp. · FMQ · Oct 18, 1996 · General Hospital

Device Facts

Record IDK963458
Device NamePADDED MITTS/PADDED-PLUS MITTS/FINGER SEPARATOR PADDED MITTS/RIGID PALM PADDED MITTS
ApplicantSkil-Care Corp.
Product CodeFMQ · General Hospital
Decision DateOct 18, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6760
Device ClassClass 1
AttributesTherapeutic

Indications for Use

Padded Mitts, Padded-Plus Mitts, Rigid Palm Padded Mitts, and Finger Separator Padded Mitts are intended for patients who remove bandages and dressings, pick and scratch at wounds, grasp and pull out feeding and naso-gastric tubes. Rigid Palm Padded Mitts have a rigid insert and are intended for patients who require an additional barrier to prevent grasping. Finger-Separator Padded Mitts have finger channels that help to control hand contracture.

Device Story

Protective hand restraints designed for clinical/nursing home use. Device consists of woven polyester mitts with foam palm padding, polyester mesh top for ventilation, and polyester wrist straps with metal buckles. Variations include: Padded-Plus (thicker foam), Finger Separator (stitched channels for contracture control), and Rigid-Palm (removable 6mm PVC insert to prevent grasping). Operated by healthcare staff to restrict patient hand/arm range of motion. Prevents self-injury, wound interference, and accidental removal of medical tubes. Benefits include patient safety and wound healing support. Used under professional supervision per facility and regulatory guidelines.

Clinical Evidence

No clinical data. The devices are established products with approximately 18 years of commercial distribution history. Safety and effectiveness are supported by historical use and adherence to HCFA regulations and facility accreditation standards regarding the appropriate use of patient restraints.

Technological Characteristics

Materials: Woven polyester, polyester mesh, polyester webbing, foam padding, metal buckles, 6mm PVC insert (Rigid-Palm model). Form factor: Mitt with wrist strap and optional finger channels or rigid palm insert. Mechanical restraint principle. No energy source or software.

Indications for Use

Indicated for patients requiring protective restraint to prevent removal of bandages, dressings, or medical tubes (feeding/naso-gastric), or to control hand contracture. Contraindicated for use as a means of discipline or convenience.

Regulatory Classification

Identification

A protective restraint is a device, including but not limited to a wristlet, anklet, vest, mitt, straight jacket, body/limb holder, or other type of strap, that is intended for medical purposes and that limits the patient's movements to the extent necessary for treatment, examination, or protection of the patient or others.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} SkillCare CORPORATION 167 Saw Mill River Road, Yonkers, New York 10701 (914) 963-2040 (800) 431-2972 OCT 18 1996 FAX (914) 963-2567 K963458 TAB F: 510(k) Summary of Safety and Effectiveness Name, address, phone and fax numbers for person submitting the 510(k) notification: Arnold Silverman, President Skil-Care Corporation 167 Saw Mill River Road Yonkers, NY 10701 Phone: 1-914-963-2040 Fax: 1-914-963-2567 Contact person: Arnold Silverman Date summary was prepared: August 27, 1996 Device name: Trade name: Padded Mitts, Finger Separator Padded Mitts, Rigid Palm Padded Mitts Common name: Same Classification name: Protective Restraint Predicate device: Padded Mitts, Padded-Plus Mitts, Finger Separator Padded Mitts, and Rigid Palm Padded Mitts marketed by Skil-Care Corporation. Device Description: Padded Mitts are constructed from woven polyester that encloses a 3/4-inch-thick foam palm pad. A top piece of polyester mesh is sewn on the mitts to prevent heat build-up. The mitts have a 1 1/4-inch-wide polyester wrist strap that secures by means of a metal buckle. The underside of the mitt cuff has a loop through which a 44-inch-long looped strap passes. This strap is used to secure the patient's hand/arm range of motion. The outer edges are finished with a bias cut binding. All straps are made of polyester webbing. Padded-Plus Mitts are identical to the above except that the foam used for the palm pad is two-inches-thick. 31 {1} Finger Separator Padded Mitts are identical to Padded-Plus Mitts except that the polyester mesh top piece is stitched to the polyester woven fabric in a such a way as to form finger channels. Finger Separator Rigid-Palm Padded Mitts are identical to the Padded Mitts except that they have a pocket on the bottom side which contains a 6mm pvc insert. This rigid insert, which removes for laundering, is used for patients who require an additional barrier to prevent against grasping at feeding and naso-gastric tubes. **Indications for use:** Padded Mitts, Padded-Plus Mitts, Rigid Palm Padded Mitts, and Finger Separator Padded Mitts are intended for patients who remove bandages and dressings, pick and scratch at wounds, grasp and pull out feeding and naso-gastric tubes. Rigid Palm Padded Mitts have a rigid insert and are intended for patients who require an additional barrier to prevent grasping. Finger-Separator Padded Mitts have finger channels that help to control hand contracture. **Comparative information:** The device (devices) used for comparative purposes is (are) currently marketed as described in this submission. Device (devices) is (are): Padded Mitts, Padded-Plus Mitts, Rigid-Palm Padded Mitts, and Finger Separator Padded Mitts. These devices are currently exempt from 510(k) Premarket Notification Procedures and Good Manufacturing Practice Regulations and are legally marketed by Skil-Care Corporation as of the date of this submission. Skil-Care Corporation has been marketing and commercially distributing these devices for approximately 18 years. The difference from our currently marketed devices are that the labeling will be changed to incorporate many of the suggestions in FDA’s draft document, "Guidance on the Content of Premarket Notification [510(k)] Submissions for Protective Restraints. The use of all patient restraints in nursing homes are subject to Health Care Financing Administration’s Regulations which prohibit the use of any restraint, physical or chemical, imposed for the purpose of discipline or convenience. Further, most health care facilities are accredited. HCFA rules governing appropriate use and accreditation standards for device use and personnel training provide the control necessary to ensure that the devices are used correctly. The application of these standards along with public awareness and health care provider training have contributed significantly to 32 {2} ensuring that the least restrictive restraint is used, that restraints are used only when needed for proper medical treatment, and that their use is under appropriate supervision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...