HEELBO WRAP AROUND VEST

K963294 · Heelbo, Inc. · FMQ · Oct 11, 1996 · General Hospital

Device Facts

Record IDK963294
Device NameHEELBO WRAP AROUND VEST
ApplicantHeelbo, Inc.
Product CodeFMQ · General Hospital
Decision DateOct 11, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6760
Device ClassClass 1
AttributesTherapeutic

Indications for Use

The Heelbo Wrap Around Vest is intended to help support and help prevent a patient from slumping or sliding forward while seated in a wheelchair or geriatric chair. The Heelbo Wrap Around Vest is also intended to limit the patient’s movement while lying in bed or on a gurney to prevent the patient from falling out of bed or climbing over the side rails.

Device Story

Crisscross-style safety vest; secures to wheelchair, geriatric chair, bed, or gurney frames via two straps; optional knotless quick-disconnect system for caregiver use; available in poly/cotton blend or polyester mesh; machine washable; color-coded for size identification. Used in clinical or home settings to support posture and prevent patient falls or unauthorized exit from beds/chairs.

Clinical Evidence

No clinical data. Biocompatibility assessment based on history of safe use of materials in the clothing and garment industry and absence of reported skin irritation complaints.

Technological Characteristics

Crisscross-style safety vest; materials: poly/cotton blend or polyester mesh; mechanical attachment via straps; optional knotless quick-disconnect system; machine washable; five sizes; color-coded.

Indications for Use

Indicated for patients requiring support to prevent slumping or sliding while seated in wheelchairs or geriatric chairs, and for patients requiring movement limitation while in bed or on a gurney to prevent falls or climbing over side rails.

Regulatory Classification

Identification

A protective restraint is a device, including but not limited to a wristlet, anklet, vest, mitt, straight jacket, body/limb holder, or other type of strap, that is intended for medical purposes and that limits the patient's movements to the extent necessary for treatment, examination, or protection of the patient or others.

Predicate Devices

Submission Summary (Full Text)

{0} Heelbo, Inc. 1134 N. Homan Ave. Chicago, IL 60651 312/489-7711 Telex RCA 297-248 800/323-5444 Fax 312/489-7744 K963294 OCT 11 1996 Heelbo, Inc. Wrap Around Vest # Safety and Effectiveness Summary ## 1. Submitter's name, Address and Contact Person | Submitter | Contact Person | | --- | --- | | Heelbo, Inc. 1134 N. Homan Ave. Chicago, IL 60651 | Joseph S. Tokarz; Manager, Regulatory Affairs Hollister Incorporated 2000 Hollister Drive Libertyville, IL 60048 Ph: (847)680-2849, Fax: (847)918-3860 | Date Summary Prepared - August 19, 1996 ## 2. Name of Device: Heelbo Wrap Around Vest ## 3. Name of Predicate Device(s) Heelbo Wrap Around Vest ## 4. Description of Device The Heelbo Wrap Around Vest is a crisscross-style safety vest. The Heelbo Wrap Around Vest is intended to help support and help prevent a patient from slumping or sliding forward while seated in a wheelchair or geriatric chair. The Heelbo Wrap Around Vest is also intended to limit the patient’s movement while lying on a bed or gurney to prevent the patient from falling out of bed or climbing over the side rails. The vest is secured to the bed or wheelchair frame with two straps. There are also loops sewn to the shoulder area of the vest in the event extra support is needed. The Wrap Around Vest is available with a knotless system (quick disconnect option). This option allows the caregiver to quickly and easily disconnect the restraint from the chair or bed frame by eliminating time consuming untying of knots (a picture of the loop accessory is attached to this section). The Wrap Around Vest is available in a poly/cotton blend or a polyester mesh material. Five sizes accommodate a wide range of patients. The Wrap Around Vest is made of durable machine washable materials and is color coded for easy size identification. Page 22 {1} Heelbo, Inc. 1134 N. Homan Ave. Chicago, IL 60651 312/489-7711 Telex RCA 297-248 800/323-5444 Fax 312/489-7744 Heelbo, Inc. Wrap Around Vest The following chart shows product availability: The Wrap Around Vest Is Available in the following Sizes and Configurations | Series | Material | Size of Vest | Color | | --- | --- | --- | --- | | 9420 9420K* | Poly/Cotton blend | x-small | White | | | | small | Yellow | | | | medium | Blue | | | | large | Green | | | | x-large | Beige | | 9420B* | Polyester Mesh material | x-small | White bias on white breezewear | | | | small | Yellow bias on white breezewear | | | | medium | Blue bias on white breezewear | | | | large | Green bias on white breezewear | | | | x-large | Beige bias on white breezewear | * K=Knotless, B=Breezewear 5. Statement of Intended Use The Heelbo Wrap Around Vest is intended to help support and help prevent a patient from slumping or sliding forward while seated in a wheelchair or geriatric chair. The Heelbo Wrap Around Vest is also intended to limit the patient’s movement while lying in bed or on a gurney to prevent the patient from falling out of bed or climbing over the side rails. 6. Statement of Technological Characteristics of the Device The Heelbo Wrap Around Vest is a crisscross-style safety vest. The Heelbo Wrap Around Vest is intended to help support and help prevent a patient from slumping or sliding forward while seated in a wheelchair or geriatric chair. The Heelbo Wrap Around Vest is also intended to limit the patient’s movement while lying on a bed or gurney to prevent the patient from falling out of bed or climbing over the side rails. The vest is secured to the bed or wheelchair frame with two straps. There are also loops sewn to the shoulder area of the vest in the event extra support is needed. Page 23 {2} Heelbo, Inc. 1134 N. Homan Ave. Chicago, IL 60651 312/489-7711 Telex RCA 297-248 800/323-5444 Fax 312/489-7744 Heelbo, Inc. Wrap Around Vest The Wrap Around Vest is available with a knotless system (quick disconnect option). This option allows the caregiver to quickly and easily disconnect the restraint from the chair or bed frame by eliminating time consuming untying of knots The Wrap Around Vest is available in a poly/cotton blend or a polyester mesh material. Five sizes accommodate a wide range of patients. The Wrap Around Vest is made of durable machine washable materials and is color coded for easy size identification. The subject devices are identical in intended use, design, materials, manufacturing process, physical and mechanical specifications and issues of safety and effectiveness to the devices prior to the submission of this notification. The only difference is that the product labeling has been revised to comply with the Agency's labeling requirements set forth in the draft "Guidance on the Content of Premarket Notification [501(k)] Submissions for Protective Restraints" dated December 1995. 7. Biocompatibility The subject devices are identical in component materials to the predicate devices. The suppliers of the materials use in the fabrication of these devices have stated that there is a history of safe use of their materials in the clothing and garment industry. In addition, Heelbo, Inc., is not aware of any reports or complaints of skin irritation associated with these materials. 8. Conclusion Based upon the information presented above it is concluded that the proposed Heelbo Wrap Around Vest is safe and effective for its intended use and is substantially equivalent to the predicate device. Page 24
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...