K973550 · Tuttnauer U.S.A Co, Ltd. · FLE · Oct 29, 1997 · General Hospital
Device Facts
Record ID
K973550
Device Name
TUTTNAUER TABLE-TOP AUTOCLAVE
Applicant
Tuttnauer U.S.A Co, Ltd.
Product Code
FLE · General Hospital
Decision Date
Oct 29, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6880
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Model M is intended to provide sterilization of medical and dental instruments and to sterilize liquids for non-clinical applications.
Device Story
Table-top steam sterilizer; utilizes multi-valve control assembly, timer, water reservoir, heat source, and sealed chamber. Operates by subjecting instruments/liquids to specific temperature/pressure cycles: unwrapped (134°C, 3 min, 29 psi), wrapped (134°C, 7 min, 29 psi), and liquids (121°C, 30 min, 16 psi). M-series requires manual parameter setting; E-series uses pre-programmed cycles. Used in medical/dental clinics and labs; operated by clinical staff. Modification involves replacing ball-type thrust bearing with solid-type thrust bearing in door closure mechanism to ensure safe operation. Output is sterilized medical/dental equipment/liquids, reducing risk of infection transmission.
Clinical Evidence
Bench testing only. Pressure vessels tested for conformance to ASME standards. No clinical data provided.
Technological Characteristics
Stainless steel pressure chamber, door, and locking device. Conforms to ASME code for pressure vessels. Manual or electronic control of sterilization cycles. Complies with UL 544, CSA, TUV, and Japanese safety standards. Solid thrust bearing in door closure.
Indications for Use
Indicated for sterilization of solid metal instruments, solid metal hinged instruments, and other materials capable of withstanding sterilization temperatures and pressures. Intended for use in ophthalmic, dental, and medical clinics, first aid rooms, and small laboratories.
Regulatory Classification
Identification
A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.
{0}------------------------------------------------
Special 510(k) - Corrective Action Being Taken Door Closure Modification
510(K) Summarv
K973550
OCT 2 9 1997
#### Submitter:
Tuttnauer USA Co. LTD. 33 Comac Loop Equi Park Ronkonkoma, New York 11779
Phone (516) 737-4850 Fax (516) 737-0720
Contact Name: Robert R. Basile, V.P.
Date Prepared: September 18, 1997
Common Name: Autoclave, table-top steam sterilizer
Trade Name: Tuttnauer Manual Autoclave Models (M, MK, MP).
Classification Name:
Steam Sterilizer Class II Device - 21 C.F.R. § 880.6880
#### Substantial Equivalence:
The Modified Tuttnauer Model M series autoclave is claimed to be substantially equivalent in safety and effectiveness to the Tuttnauer Model M series autoclave that is currently cleared and legally marketed under 510(k) number K833837. With the exception of the change from a ball-type thrust bearing to a solid type thrust bearing in the door closure mechanism, the modified devices are identical to the Model M cleared under K833837.
### General Description:
The Tuttnauer Model M series autoclave is a table-top steam sterilizer that includes as its main components a multi-valve control assembly & timer, water reservoir, heat source and sealed chamber. The M can sterilize unwrapped instruments (134°C at 3 minutes), wrapped instruments (134°C at 7 minutes) and liquids (121°C at 30 minutes).
The Model M autoclave is produced in four sizes, the 1730 with a chamber size of 7" x 12"; the 2340 with a chamber size of 9" x 18.5" and the 2540 with a chamber size of 10" x 18.7". The 3870, which has a chamber size of 15" x 27", is not subject to this corrective action.
{1}------------------------------------------------
# Design and Materials:
The pressure chamber, door and locking device are constructed from stainless steel. The pressure chamber, and components are designed, tested and stamped for conformance to ASME code.
## Intended Use:
The Model M is intended to provide sterilization of medical and dental instruments and to sterilize liquids for non-clinical applications.
# Technology Considerations:
The modified Model M exhibits the same technological characteristics as the Model M, its predicate device. With the exception of a solid thrust bearing, all of the components that may be found in the unmodified Model M are identical to those in the modified version.
## Safety and Effectiveness:
Because modification of the thrust bearing on the door closure mechanism does not effect the autoclave's sterilization parameters, the modified Model M series presents an identical sterilization profile.
The Tuttnauer Model M complies with domestic and international safety standards including Underwriter's Laboratories UL 544, Canadian Standards Association (CSA) medical device standards, European TUV standards and Japanese safety standards. The pressure chamber complies with ASME standards for pressure vessels and Canadian CRN requirements. Additionally, each pressure vessel is individually tested for conformance prior to release.
## Conclusion:
It is Tuttnauer USA Co. Ltd.'s conclusion that the modified Model M series autoclave is substantially equivalent to its predicate device, the Tuttnauer Model M series autoclave, 510(k) number K833837. Based upon the engineering calculations and ASME testing, the modification to the door closure assembly provides safe operation for these devices.
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN-SERVICES --------
Image /page/2/Picture/2 description: The image is a logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it, often used as a symbol of medicine. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the caduceus symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 2 9 1997
Tuttnauer USA Company, Ltd. C/O Mr. Mark M. Yacura Akin, Gump, Strauss, Hauer & Feld, L.L.P. 1333 New Hampshire Avenue, N.W., Suite 400 Washington, DC 20036
Re : K973550 Tuttnauer Table-Top Autoclaves, Models 1730, Trade Name: 2340 and 2540 "M" and "E" Series Regulatory Class: II Product Code: FLE == ========================================================================================================================================================= September 18, 1997 Dated: September 18, -1997 --------------------Received:
Dear Mr. Yacura:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਬੋ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obliqation you miqht have under sections 531
{3}------------------------------------------------
Page 2 - Mr. Yacura
through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely Yours,
Ky Celatrust
Timot Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radioloqical Health
.Enclosure
{4}------------------------------------------------
PAGE 002 OF 002
510(k) Number: K973550
Tuttnauer Models 1730, 2340 and 2540 "M" and "E" Series Table-top Device Name: Autoclave
Indications For Use: Tuttnauer electronic and manual controlled autoclaves are intended to be used for sterilization of solid metal instruments, solid metal hinged instruments and other materials that will withstand temperatures and pressure required for sterilization.
> The sterilizers have 3 cycles: unwrapped instruments (134°C at 3 minutes and 29 psi); wrapped instruments (134°C at 7 minutes and 29 psi); and liquids (121°C at 30 minutes and 16 psi). In the "E" Series; these cycles are pre-programmed, whereas in the "M" Series the user must manually set the parameters, however, these parameters are recommended in the labeling.
These autoclaves are intended for use in ophthalmic, dental and medical clinics, first aid rooms, and small laboratories in conformity with 21 C.F.R. § 880.6880.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) …………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………… Concurrence of CDRH, Office of Device Evaluation (ODE)
Qhin S. Lin
(Division Sign-510(k) Num
Prescription Use
OR Over-The-Counter Use X
(Per 21 C.F.R. § 801.109)
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.