K973225 · Getinge/Castle, Inc. · FLE · Feb 10, 1998 · General Hospital
Device Facts
Record ID
K973225
Device Name
CASTLE SERIES 200 STEAM STERILIZER (POWERCLAVE)
Applicant
Getinge/Castle, Inc.
Product Code
FLE · General Hospital
Decision Date
Feb 10, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6880
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The Castle Model 222 Gravity Steam Sterilizer is intended for steam sterilization of hospital supplies. Typical clinical (Hospital) applications include wrapped hard goods and linens. Wrapped-goods cycles are provided for processing wrapped linen packs, as well as wrapped hard goods such as instruments and utensils. Unwrapped goods cycles (flash) are used to process unwrapped hard goods such as instruments and utensils. Steam stenlization by the unwrapped (flash) method is employed when time does not permit the use of the preferable, wrapped sterilization procedure. Implantables should never be sterilized by the unwrapped (flash) method. All cycles employ gravity or downward displacement with positive pulsing conditioning for dynamic air removal. Processing temperatures range from 250°F to 275°F (121°C to 135°C). The Castle Model 233 Vacuum Steam Sterilizer is intended for steam sterilization of hospital supplies. Typical clinical (Hospital) applications include wrapped hard goods and linens. Wrapped-goods (Hi Vac) cycles are provided for processing wrapped linen packs, as well as wrapped hard goods such as instruments and utensils. A gravity cycle is provided for processing wrapped dry goods and/or hard goods. Additionally, a Vacuum Leak Test cycle, not intended for sterilization, and an Air-Removal-Test-Pack (Bowie-Dick) cycle, not intended for sterilization is provided. The wrapped-goods (Hi Vac) cycles employ vacuum-and-positive pulsing to condition a load before processing at the selected exposure temperature. The gravity cycle employs the gravity, or downward displacement method of air removal. Processing temperatures for wrapped goods cycles range from 250°F (121°C to 135°C). The Castle Model 223 Gravity Steam Sterilizer is intended for steam sterilization of hospital supplies. Typical clinical (Hospital) applications include wrapped and unwrapped hard goods and linens. Wrapped-goods cycles are provided for processing wrapped linen packs, as well as wrapped hard goods such as instruments and utensils. The cycles employ gravity or downward displacement with positive pulsing conditioning for dynamic air removal. Processing temperatures for wrapped goods cycles range from 250°F to 275°F (121°C to 135°C). Castle Series 200 Steam Sterilizers are programmed with factory recommended sterilization parameters (cycle settings) which have been verified and validated for efficacy. By using a supervisory password, the controls allow for flexibility in selecting exposure temperature, exposure time, and drying time. It is possible, with that password, to select cycle settings that will not necessarily achieve the desired sterlity assurance level. Users with password access are responsible for the efficacy of any cycle settings other than the factory recommended settings.
Device Story
Steam sterilizer (Powerclave) for hospital use; processes wrapped/unwrapped instruments, hard goods, linens. Inputs: user-selected cycle parameters (temp, time, drying). Operation: microprocessor-controlled steam sterilization; gravity or prevacuum air removal; positive pulsing conditioning. Outputs: sterilized goods; status/cycle data via interactive operator interface. Used in clinical environments by hospital staff. Benefits: effective sterilization of surgical supplies. Features chamber drain float switch and thermostatic drain discharge temperature control. Users can modify factory settings via password, assuming responsibility for efficacy of non-standard cycles.
Clinical Evidence
Bench testing only. Validation performed using half-cycle analysis per AAMI ST8-1994. Results consistent with sterility assurance level (SAL) of at least 10^-6. Testing included fabric and instrument loads (wrapped/unwrapped) across various cycles (Gravity, Prevacuum, Flash). No clinical data required.
Indicated for steam sterilization of hospital supplies including wrapped/unwrapped hard goods, instruments, utensils, and linens in healthcare facilities. Contraindicated for implantables in unwrapped (flash) cycles.
Regulatory Classification
Identification
A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.
Predicate Devices
Castle Microcomputer Controls and Sterilizers (K820783)
Submission Summary (Full Text)
{0}------------------------------------------------
FEB 1 0 1998
K973225
# 510(k) SAFETY AND EFECTIVENESS SUMMARY
for
## Castle Series 200 Steam Sterilizers (Powerclave)
Submitted by
Getinge/Castle, Inc. 1777 East Henrietta Road Rochester, NY 14623
Phone (716) 475-1400 (716) 272-5033 Fax
November 24, 1997
Oz Osborne Senior Quality Assurance Engineer Quality Assurance/Regulatory Affairs Phone 716-272-5129 FAX 716-272-5033
Michael J. Palladino Senior Quality Assurance Engineer Quality Assurance/Regulatory Affairs Phone 716-272-5016 FAX 716-272-5033
Per 21 CFR 807.92(a)(1) & (a)
{1}------------------------------------------------
This summary is for the:
| Trade Name | Castle Series 200 Steam Sterilizers (Powerclave) |
|---------------------|--------------------------------------------------|
| Common Name | Steam Sterilizer (greater than 2 cubic feet) |
| Classification Name | Steam Sterilizer (per 21 CFR 880.6880) |
Per 21CFR 807.92(a)(3)
We believe this product to be substantially equivalent to our Castle Microcomputer Controls and Sterilizers K820783.
### Per 21CFR 807.92(a)(4), (5) & (6)
The Castle Series 200 Sterilizers (Powerclave) is a control system upgrade of an existing microcomputer control system intended to be used with hospital steam sterilizers. Specifically, PACS 2000 controls applied to the Series 200 (Powerclave) Steam Sterilizer are to be used with "Castle" brand sterilizers manufactured by Getinge/Castle, Inc for use in health care facility environments. These sterilizers are intended to be used to sterilize wrapped and unwrapped surgical instruments, hard goods, and linens.
The difference between the predicate sterilizer controls and the PACS 2000 controls is that the PACS 2000 controls are provided with a more modern microprocessor, software, and an interactive operator interface. The piping remains relatively unchanged, with the addition of a chamber drain float switch to signal possible water-in-chamber conditions and a thermostatic drain discharge temperature control. A minor change on each chamber vessel size has changed the interior chamber width from 24 inches to 26 inches; the length on the mid-sized vessel has changed from 49 inches to 50 inches.
Per 21CFR 807.92(b)(1), (2) & (3)
The following standards were used to establish the minimum construction and performance requirements for this product:
ANSI/AAMI ST-8 - 1994 American National Standard for Hospital Steam Sterilizers,
CAN/CSA-Z314.7-M91 Effective Sterilization in Hospitals by the Steam Process (ref. AAMI ST8-1988), GGS-1340A November 24, 1975 Federal Specification Sterilizer, Surgical Instrument and Supply Gravity Air Removal, Non-Portable (Heat and Moisture Stable),
GGS-1343A November 26, 1975 Federal Specification Sterilizer, Surgical Instrument and Supply Mechanical Air Removal, Non-Portable (Heat and Moisture Stable)
{2}------------------------------------------------
Evaluation studies consisting of fabric and instrument loads (wrapped and unwrapped) have been completed to validate the safety and effectiveness of the Series 200 Steam Sterilizer (Powerclave). The plan for this was derived from AAMI ST8 1994 and the accepted industrial standards listed above. Validation testing was performed using accepted half cycle analysis and results consistent with a sterility assurance level of at least 10 were obtained.
| Sterilizer<br>Model | Cycle Type | Exposure<br>Time<br>Minutes* | Exposure<br>Temperature | Drying<br>Time<br>Minutes | Loads |
|---------------------|-----------------|------------------------------|-------------------------|---------------------------|-----------------------------------------|
| 222 | Gravity | 30 | 250 °F | 30 | Wrapped Linen Packs |
| 222 | Gravity | 10 | 275 °F | 30 | Wrapped Hard Goods |
| 222 | Gravity/Flash | 3 | 275 °F | 0 | Unwrapped Nonporous Instruments |
| 222 | Gravity/Flash | 10 | 275 °F | 0 | Unwrapped Porous Instruments |
| 223 | Gravity | 30 | 250 °F | 30 | Wrapped Linen Packs and Hard Goods |
| 223 | Gravity | 10 | 275 °F | 30 | Wrapped Hard and Dry Goods |
| 233 | Prevacuum | 3 | 275 °F | 16 | Wrapped Hard and Dry Goods |
| 233 | Prevacuum | 3 | 275 °F | 3 | Linen Packs & Single Wrapped Hard Goods |
| 233 | Gravity | 30 | 250 °F | 30 | Wrapped Linen Packs and Hard Goods |
| 233 | Bowie-Dick Test | 3.5 | 273 °F | 0 | Air Removal Test Pack |
| 233 | Air leak Test | 3 | 268 °F | 15 | No Load - Vacuum Leak test |
The cycles to be cleared for each of the models are listed in the following table:
* Exposure times listed are actual times of the cycles to be cleared and are not half times.
This product has been also designed to meet the requirements of UL544 and CSA C22.2 Nº 151 product safety standards for medical devices. The vessels are designed and constructed to Section VIII of the ASME pressure vessel code, and each vessel is so certified.
We believe the product to conform to the above requirements.
No clinical testing is required for this submittal.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, facing to the right.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 1 0 1998 Gettinge/Castle, Incorporated C/O Mr. Gordon Gillerman Underwriters Laboratories, Incorporated 333 Pfingsten Road Northbrook, Illinois 60062
Re : K973225 Castle Series 200 Steam Sterilizer Trade Name: (Straightline) Regulatory Class: II Product Code: FLE Dated: February 2, 1998 February 3, 1998 Received:
Dear Mr. Gillerman:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A ------substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531
{4}------------------------------------------------
Page 2 - Mr. Gillerman
through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
Timothy A Directo Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if Known):
#### Device Name: Castle Series 200 Steam Sterilizer (Powerclave)
Indications For Use:
#### Model 222:
The Castle Model 222 Gravity Steam Sterilizer is intended for steam sterilization of hospital supplies. Typical clinical (Hospital) applications include wrapped hard goods and linens.
Wrapped-goods cycles are provided for processing wrapped linen packs, as well as wrapped hard goods such as instruments and utensils. Unwrapped goods cycles (flash) are used to process unwrapped hard goods such as instruments and utensils. Steam stenlization by the unwrapped (flash) method is employed when time does not permit the use of the preferable, wrapped sterilization procedure. Implantables should never be sterilized by the unwrapped (flash) method. All cycles employ gravity or downward displacement with positive pulsing conditioning for dynamic air removal. Processing temperatures range from 250°F to 275°F (121°C to 135°C)
The cycles to be cleared for this model (Model 222) are:
| Cycle | Name | Type | Exposure Temp | Minutes | Dry Minutes |
|-------|-------------|--------------------|---------------|---------|-------------|
| P1 | GRAV WRAP 1 | Gravity 250°F | 250°F | 30 | 30 |
| P2 | GRAV WRAP 2 | Gravity 275°F | 275°F | 10 | 30 |
| P3 | FLASH 3 + | Gravity 3 Minutes | 275°F | 3 | 0 |
| P4 | FLASH 10 + | Gravity 10 Minutes | 275°F | 10 | 0 |
#### Model 233:
The Castle Model 233 Vacuum Steam Sterilizer is intended for steam sterilization of hospital supplies. Typical clinical (Hospital) applications include wrapped hard goods and linens.
Wrapped-goods (Hi Vac) cycles are provided for processing wrapped linen packs, as well as wrapped hard goods such as instruments and utensils. A gravity cycle is provided for processing wrapped dry goods and/or hard goods. Additionally, a Vacuum Leak Test cycle, not intended for sterilization, and an Air-Removal-Test-Pack (Bowie-Dick) cycle, not intended for sterilization is provided. The wrapped-goods (Hi Vac) cycles employ vacuum-and-positive pulsing to condition a load before processing at the selected exposure temperature. The gravity cycle employs the gravity, or downward displacement method of air removal. Processing temperatures for wrapped goods cycles range from 250°F (121°C to 135°C).
The cycles to be cleared for this model (Model 233) are:
| Cycle | Name | Type | Exposure<br>Temp | Minutes | Dry<br>Minutes |
|-------|--------------|---------------|------------------|---------|---------------------------|
| P1 | PREVAC ONE | Prevac 275°F | 275°F | 3 | 16 |
| P2 | PREVAC TWO | Prevac 275°F | 275°F | 3 | 3 |
| P3 | GRAVITY WRAP | Gravity 250°F | 250°F | 30 | 30 |
| P5 | B-D TEST | Air Removal | 273°F | 3.5 | 0 |
| P6 | VAC LEAK TST | Vacuum Leak | 268°F | 3 | 15(dry)+5(dwell)+15(test) |
{6}------------------------------------------------
#### Model 223:
The Castle Model 223 Gravity Steam Sterilizer is intended for steam sterilization of hospital supplies. Typical clinical (Hospital) applications include wrapped and unwrapped hard goods and linens.
Wrapped-goods cycles are provided for processing wrapped linen packs, as well as wrapped hard goods such as instruments and utensils. The cycles employ gravity or downward displacement with positive pulsing conditioning for dynamic air removal. Processing temperatures for wrapped goods cycles range from 250°F to 275°F (121°C to 135°C).
The cycles to be cleared for this model (Model 223) are:
| Cycle | Name | Type | Exposure<br>Temp | Minutes | Dry<br>Minutes |
|-------|-------------|---------------|------------------|---------|----------------|
| P1 | GRAV WRAP 1 | Gravity 250°F | 250°F | 30 | 30 |
| P2 | GRAV WRAP 2 | Gravity 275°F | 275°F | 10 | 30 |
#### All Models:
Castle Series 200 Steam Sterilizers are programmed with factory recommended sterilization parameters (cycle settings) which have been verified and validated for efficacy. By using a supervisory password, the controls allow for flexibility in selecting exposure temperature, exposure time, and drying time. It is possible, with that password, to select cycle settings that will not necessarily achieve the desired sterlity assurance level. Users with password access are responsible for the efficacy of any cycle settings other than the factory recommended settings.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Lillian
(Division Sign-Off)
Division of Dental Infection Control, and General Hospital Devic 510(k) Number
Prescription Use (Per 21 CFR 801.109)
OR
Over-The Counter Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.