Enbio S

K210279 · Enbio Group AG · FLE · Oct 29, 2021 · General Hospital

Device Facts

Record IDK210279
Device NameEnbio S
ApplicantEnbio Group AG
Product CodeFLE · General Hospital
Decision DateOct 29, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6880
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Enbio S is an air-removal (pre-vacuum) table-top steam sterilizer intended for use by a health care provider to sterilize medical products by means of pressurized steam. It is suitable for the sterilization of dental and medical instruments that are validated to be sterilized by steam. The Enbio S has not been designed to sterilize liquid loads, biomedical waste or materials not compatible with sterilization. The processing of such loads may result in incomplete sterilization and/or damage to the autoclave.

Device Story

Enbio S is a table-top, pre-vacuum steam sterilizer for medical/dental instruments. User places load in aluminum chamber; selects program via TFT touch screen. Aluminum steam generator produces superheated steam; steam penetrates instruments at 121°C or 134°C for specified exposure times. Device pumps out steam; performs drying cycle; displays process completion to user. Used by healthcare providers in clinical settings. Output confirms sterilization status, ensuring instrument safety for patient use.

Clinical Evidence

No clinical data. Bench testing only. Performance verified via ANSI AAMI ST55:2016 standards, including vacuum tests, Bowie-Dick tests, empty/full/half-cycle thermal/pressure studies, and biological indicator testing (10^-6 SAL). Electrical safety and EMC verified per IEC 61010-1, 61010-2, IEC 60601-1-2, and IEC 61326-1.

Technological Characteristics

Table-top pre-vacuum steam sterilizer. Aluminum chamber with integrated heaters. 2.7L chamber volume. External water tank. TFT touch screen interface. USB port for data. Power: 110-120V, 60Hz, 15A. Complies with ANSI AAMI ST55:2016, IEC 61010-1/2, IEC 60601-1-2, IEC 61326-1.

Indications for Use

Indicated for sterilization of dental and medical instruments validated for steam sterilization. Suitable for solid, small porous, simple recessed, narrow-clearance items, dental handpieces, textiles, and plastics (121°C cycle). Not for liquid loads, biomedical waste, or incompatible materials.

Regulatory Classification

Identification

A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. October 29, 2021 Enbio Group AG Lukasz Rogowski Corporate Quality Manager Eichengasse 3 Oensingen, CH-4702 Switzerland Re: K210279 Trade/Device Name: Enbio S Regulation Number: 21 CFR 880.6880 Regulation Name: Steam sterilizer Regulatory Class: Class II Product Code: FLE Dated: September 24, 2021 Received: September 27, 2021 Dear Lukasz Rogowski: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {2}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Clarence W. Murray, III, Ph.D. Assistant Director THT4B1:Sterility Devices Team DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use 510(k) Number (if known) ## K210279 Device Name Enbio S ### Indications for Use (Describe) The Enbio S is an air-removal (pre-vacuum) table-top steam sterilizer intended for use by a health care provider to sterilize medical products by means of pressurized steam. It is suitable for the sterilization of dental and medical instruments that are validated to be sterilized by steam. The Enbio S has not been designed to sterilize liquid loads, biomedical waste or materials not compatible with sterilization. The processing of such loads may result in incomplete sterilization and/or damage to the autoclave. Please refer to the table below for program name, load description, sterilization temperature, exposure time, drying time and maximum load. | Program<br>Name | Load Description | Sterilization<br>Temperature | Sterilization<br>Time | Drying<br>Time | Maximum<br>Load | |-----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|-----------------------|----------------|--------------------| | 134°C | solid objects, small porous<br>objects, simple objects<br>recessed, narrow-clearance<br>items, dental handpieces,<br>and textiles; wrapped and<br>unwrapped | 134°C (273°F) | 4 minutes | 3 minutes | 0.5 Kg/<br>1.1 lbs | | 121°C | solid objects, small porous<br>objects, simple objects<br>recessed, narrow-clearance<br>items, dental handpieces,<br>textiles, and plastics;<br>wrapped and unwrapped | 121°C (250°F) | 30 minutes | 5 minutes | 0.5 Kg/<br>1.1 lbs | Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510(k) Summary K210279 #### Sponsor/ Applicant 1. Enbio Group AG Eichengasse 3 CH-4702 Oensingen, Switzerland Mr. Lukasz Rogowski Corporate Quality Manager Email: lukasz.rogowski@enbio.com Phone: +48 605 058 629 Summary Preparation Date: October 29, 2021 #### 2. Device | Trade Name | Enbio S | |---------------------|------------------| | Classification | Class 2 | | Classification Name | Steam Sterilizer | | Product Code | FLE | | Regulation Number | 21 CFR 880.6880 | | Review Panel | General Hospital | #### 3. Predicate Device Statclave G4 Chamber Autoclave (K190062) #### 4. Device Description The Enbio S is an air-removal (pre-vacuum) table-top steam sterilizer intended for use by a health care provider to sterilize medical products by means of pressurized steam. It has a hermetically sealed, heated chamber made from aluminum, with two heaters sink inside chamber walls. Inside this chamber, the sterilized load is placed on a special perforated tray. After closing the chamber, the user selects the appropriate sterilization program through the TFT touch screen. {5}------------------------------------------------ The actual sterilization phase starts after the pre-vacuum phase. The aluminum steam generator produces superheated steam and applies it inside the chamber. That steam penetrates the sterilized instruments. The set temperature is maintained inside the chamber depending on the selected sterilization cycle (121C° or 134°C), during a specified time (30 minutes or 4 minutes). After that time all the steam accumulated inside the chamber is pumped out and the drying cycle begins. When sterilization is finished, device displays to the user that process is completed, and that the load is sterile. #### 5. Indications for Use The Enbio S is an air-removal (pre-vacuum) table-top steam sterilizer intended for use by a health care provider to sterilize medical products by means of pressurized steam. It is suitable for the sterilization of dental and medical instruments that are validated to be sterilized by steam. The Enbio S has not been designed to sterilize liquid loads, bio-medical waste or materials not compatible with steam sterilization. The processing of such loads may result in incomplete sterilization and/or damage to the autoclave. Please refer to the table below for program name, load description, sterilization temperature, exposure time, drying time and maximum load. | Program<br>Name | Load Description | Sterilization<br>Temperature | Sterilization<br>Time | Drying<br>Time | Maximum<br>Load | |-----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|-----------------------|----------------|--------------------| | 134°C | solid objects, small porous<br>objects, simple objects<br>recessed, narrow-clearance<br>items, dental handpieces,<br>and textiles; wrapped and<br>unwrapped | 134°C (273°F) | 4 minutes | 3 minutes | 0.5 Kg/<br>1.1 lbs | | 121°C | solid objects, small porous<br>objects, simple objects<br>recessed, narrow-clearance<br>items, dental handpieces,<br>textiles, and plastics;<br>wrapped and unwrapped | 121°C (250°F) | 30 minutes | 5 minutes | 0.5 Kg/<br>1.1 lbs | {6}------------------------------------------------ #### Technological Characteristics Comparision Table 6. Provided below is a technological comparison of the subject device with the predicate device. | | Subject Device | Predicate Device | Comparison | |----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------| | Trade Name | Enbio S | STATCLAVE G4 Chamber<br>Autoclave (K190062) | | | Submitter | Enbio Group AG | SciCan Ltd. | - | | Product Code | FLE | FLE | Same | | Regulation Number | 21 CFR 880.6880 | 21 CFR 880.6880 | Same | | Device Class | Class 2 | Class 2 | Same | | Prescription / Over-The-Counter Use | Over-The-Counter | Over-The-Counter | Same | | Intended Use | The Enbio S is an air-removal (pre-vacuum) table-top steam sterilizer intended for use by a health care provider to sterilize medical products by means of pressurized steam. It is suitable for the sterilization of dental and medical instruments that are validated to be sterilized by steam. The Enbio S has not been designed to sterilize liquid loads, bio-medical waste or materials not compatible with steam sterilization. The processing of such loads may result in incomplete sterilization and/or damage to the autoclave. | The STATCLAVE G4 is a dynamic-air-removal (pre-vacuum) table-top steam sterilizer intended for use by a health care provider to sterilize medical products by means of pressurized steam. It is suitable for the sterilization of dental and medical instruments that are validated to be sterilized by steam. The STATCLAVE G4 has not been designed to sterilize liquid loads, bio-medical waste or materials not compatible with steam sterilization. The processing of such loads may result in incomplete sterilization and / or damage to the autoclave. | Same | | Water tank | External | Internal reservoir | Different | | Sterilization Chamber Volume | 2.7 L | 26 L | Different | | Sterilization Chamber Dimensions | 292 x 192 x 45 mm<br>(L x W x H) | Diameter: 280 mm<br>Depth: 381 mm | Different | | Device Dimensions (L x W x H) | 561 x 252 x 162 mm | 635 x 450 x 495 mm | Different | | Weight | 15 kg (approximately) | 62.7kg (approximately) | Different | | Power Rating | 110-120 V, 60Hz, 15A | 120V, 60Hz, 12 A | Similar | | Wireless<br>Transmission<br>Capability | No | Yes | Different | | USB Port | Yes | Yes | Same | | Sterility and Shelf-<br>life | Not provided sterile.<br>No shelf-life claimed | Not provided sterile.<br>No shelf-life claimed | Same | {7}------------------------------------------------ #### Non-clinical Bench (Performance) testing 7. Provided below is the non-clinical testing to demonstrated the subject device meets the specification and accpetance criteria found in the Standards and test method listed below. | Test Method | Purpose | Acceptance criteria | Result | |---------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------| | ANSI AAMI ST55:2016<br>Vacuum Test | Verify air removal<br>performance | Average leak rate of 1<br>millimeter of mercury<br>(mmHg) (0.13 kPa) (0.019<br>psia) per min or less | Pass | | Bowie Dick Test | Verify air removal<br>performance | The Bowie-Dick test<br>indicator sheet shall show<br>a uniform color change;<br>i.e., the color in the center<br>should be the same as<br>that at the outer edges. | Pass | | ANSI AAMI ST55:2016<br>Empty Chamber Study<br>• 134°C (273°F) in 4 minutes | Verify pressure and<br>temperature – to ensure<br>that the sterilizer is<br>capable of providing<br>steady-state thermal<br>and pressure conditions<br>within the chamber | +3°C/-0°C<br>±0.3 bar | Pass | | ANSI AAMI ST55:2016<br>Empty Chamber Study<br>• 121°C (250°F) in 30 minutes | Verify pressure and<br>temperature - to ensure<br>that the sterilizer is<br>capable of providing<br>steady-state thermal<br>and pressure conditions<br>within the chamber | | Pass | | ANSI AAMI ST55:2016<br>Full Cycle Study<br>• 134°C (273°F) in 4<br>minutes | Verify pressure and<br>temperature - to ensure<br>that the sterilizer is<br>capable of providing<br>steady-state thermal<br>and pressure conditions | +3°C/-0°C<br>±0.3 bar | Pass | | ANSI AAMI ST55:2016<br>Full Cycle Study<br>• 121°C (250°F) in 30<br>minutes | Verify pressure and<br>temperature - to ensure<br>that the sterilizer is<br>capable of providing<br>steady-state thermal<br>and pressure conditions<br>during the cycle | | Pass | | ANSI AAMI ST55:2016<br>Half Cycle Study<br>• 134°C (273°F) in 4<br>minutes | Verify pressure and<br>temperature – to ensure<br>that the sterilizer is<br>capable of providing | +3°C/-0°C<br>±0.3 bar | Pass | | ANSI AAMI ST55:2016<br>Half Cycle Study<br>• 121°C (250°F) in 30 minutes | steady-state thermal<br>and pressure conditions<br>during the cycle | | Pass | | ANSI AAMI ST55:2016<br>Full Cycle Biological<br>Indicators<br>• 134°C (273°F) in 4 minutes | To verify biological<br>sterilization performance | The tested cycle has a<br>10-6 SAL | Pass | | ANSI AAMI ST55:2016<br>Full Cycle Biological<br>Indicators<br>• 121°C (250°F) in 30 minutes | | | Pass | | ANSI AAMI ST55:2016<br>Half Cycle Biological<br>Indicators<br>• 134°C (273°F) in 2 minutes | To ensure the efficacy of<br>the equipment and the<br>lethality of the<br>recommended<br>processing parameters<br>by biological challenge | The tested cycle has a<br>10-6 SAL | Pass | | ANSI AAMI ST55:2016<br>Half Cycle Biological<br>Indicators<br>• 121°C (250°F) in 15 minutes | | | Pass | | ANSI AAMI ST55:2016<br>Half Cycle Study /<br>Textile PCD<br>• 134°C (273°F) in 2 minutes | Verify half-cycle<br>sterilization of BI inside<br>textile PCD | The tested cycle has a<br>10-6 SAL | Pass | | ANSI AAMI ST55:2016<br>Half Cycle Study /<br>Textile PCD<br>• 121°C (250°F) in 15 minutes | | | Pass | | IEC 61010-1, 61010-2 | Verify electrical safety | Meets specifications of<br>standard | Pass | | IEC 60601-1-2,<br>IEC 61326-1,<br>47 CFR 15 Subpart B, | Verify electromagnetic<br>compatibility | Meets specifications of<br>standards | Pass | {8}------------------------------------------------ #### 8. Clinical Testing The submission does not contain any data from clinical testing. #### 9. Conclusion: The conclusions drawn from the nonclinical tests demonstrate that the Enbio S is as safe, as effective, and performs as well as or better than the legally marketed predicate device, STATCLAVE G4 Chamber Autoclave (K190062).
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...