K072087 · Belimed Sauter AG · FLE · Oct 5, 2007 · General Hospital
Device Facts
Record ID
K072087
Device Name
BELIMED STEAM STERILIZER MODEL TOP 5000 SERIES 24
Applicant
Belimed Sauter AG
Product Code
FLE · General Hospital
Decision Date
Oct 5, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6880
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The Belimed Steam Sterilizer TOP 5000, series 24, is designed for sterilization of non porous and porous heat and moisture-stable materials used in healthcare facilities. The Belimed Steam Sterilizer TOP 5000, series 24, is equipped with the following factory-programmed Sterilization cycles and cycle values (Table 1). The Belimed Steam Sterilizer TOP5000, series 24, is available as a single door prevacuum or a double door prevacuum version.
Device Story
Steam sterilizer for healthcare facilities; processes porous/non-porous heat-stable materials (textiles, glassware, instruments). Device utilizes steam under pressure; factory-programmed cycles (PREVAC, LIQUID, EXPRESS, FLASH) manage temperature, exposure, and drying times. Larger chamber volume (2480L) compared to predicates; increased vacuum-pump power. Operated by healthcare staff; microprocessor controller monitors cycle parameters, providing alerts/aborts if conditions fail. Output includes sterilized loads verified by biological/chemical indicators. Benefits include efficient terminal sterilization of medical equipment, reducing cross-contamination risk.
Clinical Evidence
No clinical data required. Bench testing only. Validation performed per AAMI/ANSI ST-8:2001. PREVAC cycles demonstrated SAL of at least 10^-6, F0 >= 12, moisture retention <3% (fabric) and <20% (towels). FLASH cycles validated per AAMI/ANSI ST37:1996. LIQUID cycles validated for 121°C maintenance for 12 minutes. Bowie-Dick test confirmed uniform steam penetration.
Indicated for sterilization of non-porous and porous heat and moisture-stable materials in healthcare facilities. Includes textiles, glassware, and wrapped/unwrapped instrument trays.
Regulatory Classification
Identification
A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.
Predicate Devices
Belimed Steam Sterilizer TOP 5000 Series 9 - 18 (K021223)
Belimed Steam Sterilizer TOP 5000 Series 4-8 (K033538)
Submission Summary (Full Text)
{0}------------------------------------------------
OCT 5 2007
Image /page/0/Picture/1 description: The image contains a sequence of handwritten digits and letters. The sequence appears to be "k072087". The characters are written in a simple, slightly irregular style, typical of handwriting.
510(k) Summary
| Date: | June, 19, 2007 |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter's Name / Address: | Belimed Sauter AG<br>Zelgstrasse 8<br>CH- 8583 Sulgen / Switzerland |
| Contact Person: | Hans Stadler, Head of Product Development,<br>Email: hans.stadler@belimed-sauter.ch<br>Phone: +41 71 644 85 97<br>Fax: +41 71 644 86 00 |
| Trade Name: | Belimed Steam Sterilizer TOP 5000<br>Series 24 |
| Classification: | Steam Sterilizer -- Class II, as listed per 21 CFR 880.6880 |
| Predicate Devices: | Belimed Steam Sterilizer TOP 5000<br>Series 9 - 18 (K021223) and<br>Series 4-8 (K033538) |
## DEVICE DESCRIPTION:
The Belimed Steam Sterilizers TOP 5000 Series 24 is intended for use in hospital and health care facilities and is intended to be used in an identical manner as the Belimed Steam Sterilizer TOP 5000 series 9 - 18.
The larger chamber size (higher chamber) incorporates additional flexibility, and allows operating the sterilizers in an economical way (less floor space required).
# NONCLINICAL COMPARISON TO THE PREDICATE DEVICE:
The Belimed Steam Sterilizer TOP 5000 series 24, are very similar to the predicate device. Modifications made from the predicate device include:
- Larger chamber size (chamber height increased) .
- Vacuum-pump power was increased according to larger chamber volume. .
- Improved functionallity (as done for series 4-8 K033538) .
## CLINICAL DATA
No clinical data is required for this device classification submission
## INDICATIONS FOR USE:
The Belimed Steam Sterilizer TOP 5000 series 24 is designed for sterilization of non porous and porous heat and moisture-stable materials used in healthcare facilities.
{1}------------------------------------------------
The Belimed Steam Sterilizer TOP 5000 series 24 is equipped with the following factory-The belined Steam Sterilizer For Sales, which correspond with Series 9-18 (Table 1).
| CYCLES | STERILIZE TEMP | STERILIZE TIME(min) | DRY TIME (min) | RECOMMENDED LOAD |
|---------------------------|-------------------|---------------------|----------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| PREVAC<br>270° F (132°C) | 270° F<br>(132°C) | 4 | 20 | Double-wrapped Instrument Trays, max.<br>weight of 20 lbs (9 kg) each<br>or Fabric packs.<br>Refer to Table 2 for recommended quantities. |
| PREVAC<br>270° F (132°C) | 270° F<br>(132°C) | 4 | 5 | Fabric Packs.<br>Refer to Table 2 for recommended quantities<br>Packs are to place horizontally on shelves |
| LIQUID<br>250° F (121°C) | 250° F<br>(121°C) | 45 | - | Liquids not intended for direct patient<br>contact!<br>Refer to Table 3 for Guidelines |
| EXPRESS<br>270° F (132°C) | 270° F<br>(132°C) | 4 | 3 | Single Wrapped Instrument Tray with non<br>porous single instrument |
| FLASH<br>270° F (132°C) | 270° F<br>(132°C) | 3 | 1 | Unwrapped Instrument Tray with a single<br>Instrument |
| FLASH 270° F<br>(132°C) | 270° F<br>(132°C) | 10 | 1 | Unwrapped Instrument Tray with non porous<br>multiple instruments (max. weight of 20 lbs) |
Table 1: Factory programmed sterilization cycles
The following table (Table 2) shows BELIMED SAUTER AG´s recommended load by sterilizer sizes: Do not sterilize a mix load of instrument trays and Fabric packs.
| Series | Model<br>single door<br><i>double door</i> | Sterilizer<br>Chamber Size<br>(HxWxD) | Wrapped Instrument<br>Trays 20"x10"<br>max. 20 lbs each | Fabric Packs<br>11"x11"x9"<br>max.6.6 lbs each | Fabric Packs<br>23"x11"x11"<br>max. 17 lbs each |
|--------|--------------------------------------------|---------------------------------------|---------------------------------------------------------|------------------------------------------------|-------------------------------------------------|
| 24 | GR15-6-18 HS1<br>GR15-6-18 HS2 | 63"x26"x79" | 24 | 56 | 24 |
#### Table 2: Recommended Loads
The following table (Table 3) is BELIMED SAUTER AG's guidelines for liquid cycle processing for models:
| Series | Model<br>single door<br>double door | Sterilizer<br>Chamber Size<br>(HxWxD) | Volume of Liquid in one<br>bottle | Max. number of<br>bottles |
|--------|-------------------------------------|---------------------------------------|-----------------------------------|---------------------------|
| 24 | GR15-6-18 HS1<br>GR15-6-18 HS2 | 63"x26"x79" | 1000 ml | 336 |
Table 3: Guidelines for liquid 250°F cycle processing
{2}------------------------------------------------
| Model | Con-<br>figuration | Chamber size<br>(H x W x D)<br>(inch) | (mm) | Overall Dimensions<br>(H x W x D)<br>(inch) | (mm) |
|----------------|-----------------------------------|---------------------------------------|-----------------|---------------------------------------------|------------------|
| GR 15-6-18 HS1 | Floor flush<br>design,<br>1 door | 63"x 26"x 79" | 1600x 660x 2000 | 79"x 75"x 98" | 2000x 1900x 2500 |
| GR 15-6-18 HS2 | Floor flush<br>design,<br>2 doors | 63"x 26"x 79" | 1600x 660x 2000 | 79"x 75"x 98" | 2000x 1900x 2500 |
The Belimed Steam Sterilizer TOP 5000 Series 24 are offered in the following size configurations:
Table 4: Dimensions
The Belimed Steam Sterilizer TOP 5000 Series 24 is designed to be used for the terminal Sterilization of porous and non porous, heat and moisture stabile materials in the healthcare facilities.
Depending of the chosen cycle materials as different as textiles, glassware, unwrapped or wrapped instrument trays with single or multiple instruments may be sterilized.
The Belimed Steam Sterilizer TOP 5000 Series 24 is factory equipped with cycles which has been tested in accordance with AAMI/ANSI ST-8:2001 under defined load conditions. The predicate devices with a chamber volume between 752 l (series 9) and 1900 l (series 18) have been tested in 2002. The new models have a chamber volume of 2480 1. The sterilizer has been validated.
#### EFFECTIVENESS:
Efficacy of sterilizer function and exposure time recommendations are ultimately shown by complete kill of biological indicators and verifying an appropriate safety factor or sterility assurance level (SAL) of at least 106 reduction. BELIMED SAUTER AG validates its sterilization cycles by recommended practices, standards and guidelines developed by various independent organizations such as the Association for Advancement of Medical Instrumentation (AAMI). Prior to release, Belimed Steam Sterilizer TOP 5000 series 24 was validated to meet the requirements of AAMI/ANSI-ST8-2001.
The results of the Belimed Steam Sterilizer TOP 5000 verification studies demonstrate that the sterilizer performs as intended and are summarized as follows:
- All PREVAC cycles verified using the fabric test pack, as described in Section 5.5.2 AAMI/ANSI-. ST8:2001 were qualified according to section 5.5.2.5 ANSI/AAMI-ST8. These cycles demonstrated a sterility assurance level of at least 10° through achievement of a time-attemperature sufficient to produce an FO of at least 12 by half cycle, moisture retention of fess than 3% increase in pre-sterilization test pack weight, and exhibited no wet spots.
- All PREVAC cycles verified using full load instruments trays were qualified according to section t 5.5.4 of ANSI/AAMI-ST8: 2001. These cycles demonstrated a sterility assurance level of at least 10 * through achievement of a time-at-temperature sufficient to produce an F0 of at least 12 by half cycle, moisture retention of less than 20% increase in pre-sterilization weight of the towel, and exhibited no wet spots on the outer wrapper.
{3}------------------------------------------------
- All FLASH cycles verified using the unwrapped instrument tray were qualified according to . All 1 LAJT Cycles Vernico asing the Lind ANSI/AAMI ST37:1996 section 7.7.3. These cycles Section 3.5.3.1 ANMARKSTOFS.COM of at least 10 * through achievement of a time at demonstated '6' stemly '450, and east 12 by half cycle and exhibited no wet spots.
- tellipendial sameline to produce three 1'000 ml flasks, as described in section 5.5.3 of . All ElQUD Cycles Venned asing to section 5.5.3.5. These cycles demonstrated a AxiMirANST 516.2017 16:26 10° through achievement of a time-at-temperature sufficient stenlity assurance ferer of at 12 by half cycle, a water loss not exceeding 50 ml, and automatic to produce and of at least to bperature of 121°C was achieved and maintained in the center of the liquid for at least 12 minutes.
- of the ngore for at was verified using the Bowie-Dick Test Pack were qualified according to section . 5.6 of AAMI/ANSI-ST8, and demonstrated a uniform color change throughout the test sheet.
- 5.0 of halfaller validation for the cycle operation was performed according to FDA's moderate . The Software Nandation Total of Torided in the document "Guidance for the Content for Prelever of Colbmissions for Software Contained in Medical Devices (May 2005)".
#### SAFETY:
BELIMED SAUTER AG's sterilizers including the Belimed Steam Sterilizer TOP 5000 have been designed, constructed and tested to meet the safety and performance requirements of various national safety codes and standards. The Belimed Steam Sterilizer TOP 5000 complies with the following requirements:
- ing requirences:
Underwriter Laboratory (UL) Code UL 61010A-1:2002 and UL 61010A-2-041:2002 1. and IEC 61010-1:2001 and IEC 61010-2-041:1995
- IEC / EN 61326:2001 2.
- AAMI ST8:2001 3.
- AAMI ST37:1996 4.
- American Society of Mechanical Engineers (ASME), Section Vill, Division 1 for unfired pressure 5. vessels: 2004.
# HAZARDS-FAILURE OF PERFORMANCES
Failure of the sterilization process can lead to incidence of cross contamination, the transmission of i unar of the stemlednon prisess from one infected person to another who was not otherwise infected prior to the incident.
To avoid failure, the user must ensure the materials, instruments and devices to be sterilized are thoroughly cleaned, that the manufacturer's instructions for use are followed., that the cycle to be thoroughly cleaned, that will as been validated, that the sterilizer has been maintained in ascordance with the sterilizer's manufacturer's maintenance schedule and is operating properly, and that each sterilizer load is monitored with available and validated biological and chemical sterilization process indicators.
stemization process managers in daily use in hospitals throughout the United States. The incident of sterilizer malfunction or sterilization process failure is relatively rare considering the wide medies of steam sterilizers. Further, there are no known reports in the literature of patient spread ass - thave resulted from steam sterilizer failure. The technology designed in Belimed Steam Sterilizer TOP 5000 provides microprocessor controller safeguard that aborts the cycle and give appropriate signals, alerts and warnings when required conditions have not been met or when a malfunction occurs.
{4}------------------------------------------------
## USER INFORMATION
BELIMED SAUTER AG provides information to the user that is intended to insure SatureR AG BELIMED JAOTER / AS provides Information of the International and other labeling. BELIMED SAUTER AG use of stemsedian in to accomment the AAMI steam sterilization standards to enclude asso and recomments the asse and point offective use of steam sterilization equipment in health care facilities.
## CONCLUSION
The Belimed Steam Sterilizer Top 5000 Series 24 is a substanially equivalent device to that of the predicate device. There have been no substantial changes in technology and intended use of this device. The control system and the hardware interface remain unchanged to predicate device series 4-8 (K033538). This steam sterilizer meets the appicable requirements of the applicable standards. Based on the provided information in this premarket notification, it can be concluded that the subject device is substantial equivalent to the predicate device and is safe and effective when used. as intended.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Belimed Sauter AG C/O Mr. Stefan Preiss Responsible Third Party Official TUV SUD America, Incorporated 1775 Old Highway 8 NW New Brighton, Minnesota 55112-1891
OCT 5 ี 2007
Re: K072087
Trade/Device Name: Belimed Steam Sterilizer TOP 5000 Regulation Number: 880.6880 Regulation Name: Steam Sterilizer Regulatory Class: II Product Code: FLE Dated: September 20, 2007 Received: September 24, 2007
Dear Mr. Preiss:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
#### Page 2 - Mr. Preiss
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: Belimed Steam Sterilizer TOP 5000
Indications for Use:
The Belimed Steam Sterilizer TOP 5000, series 24, is designed for sterilization of non porous and porous heat and moisture-stable materials used in healthcare facilities.
The Belimed Steam Sterilizer TOP 5000, series 24, is equipped with the following factory-programmed Sterilization cycles and cycle values (Table 1).
The Belimed Steam Sterilizer TOP5000, series 24, is available as a single door prevacuum or a double door prevacuum version.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use x OTC (21 CFR 807 Subpart C)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Clus
(Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices
510(k) Number: K072087
Page 1 of 2
{8}------------------------------------------------
| CYCLES | STERILIZE<br>TEMP | STERILIZE<br>TIME(min) | DRY TIME<br>(min) | RECOMMENDED LOAD |
|--------------------------|-------------------|------------------------|-------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| PREVAC<br>270° F (132°C) | 270° F<br>(132°C) | 4 | 20 | Double-wrapped Instrument Trays, max.<br>weight of 20 lbs (9 kg) each or Fabric packs.<br>Refer to Table 2 for recommended quantities. |
| PREVAC<br>270°F (132°C) | 270° F<br>(132°C) | 4 | 5 | Fabric Packs,<br>Refer to Table 2 for recommended quantities<br>Packs are to place horizontally on shelves. |
| LIQUID<br>250°F (121°C) | 250° F<br>(121°C) | 45 | 0 | Liquids not intended for direct patient contact!<br>Refer to Table 3 for Guidelines |
| EXPRESS<br>270°F (132°C) | 270° F<br>(132°C) | 4 | 3 | Single Wrapped Instrument Tray with non<br>porous single instrument |
| FLASH<br>270° F (132°C) | 270° F<br>(132°C) | 3 | 1 | Unwrapped Instrument Tray with a single<br>Instrument |
| FLASH 270°F<br>(132°C) | 270° F<br>(132°C) | 10 | 1 | Unwrapped Instrument Trays with non porous<br>multiple instruments (max. weight of 20 lbs) |
Table 1: Factory programmed Sterilization Cycles
The following table (Table 2) shows Belimed Sauter AG's recommended load by sterilizer size:
Do not sterilize a mix load of instrument trays and fabric packs.
| ( " " " " " " " " " " " " " " "<br>148 - 28860. Bu | vode.<br>1341 114 11<br>single door<br>1201-1-1 2 22 - 11 - 35: 4<br>「テートが意書から出版を<br>年本月の行きを読む 流行し<br>ouble door | Sterilizer Sterilizer<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>イ・アイテル・アイテム、アルター、アルキーストーム・アー<br>An "E" 1 " 1 " 1 " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " "<br>"HxWxD) == | Wrapped Instrument Fabric Packs<br>1 2 3 3 - 5 - 5 - 1 - 8 6 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1<br> Chamber Size Trays 20"x10" = " = "<br>o gos "Thesels rices in not<br>- 14 malder = " =<br>nilhe pa<br> | :<br>11*x11"x9" ===================================================================================================================================================================<br>The Fact "Fact . I be de .<br><br>max 6 6 lbs | abrical<br>*92"x11"x11"<br>-23-11-12<br>and Extraction in the<br>rach Emar = |
|----------------------------------------------------|------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| | | | | | |
| | GR15-6-18<br>r | | | | |
Table 2: Recommended Loads
The following table (Table 3) is Belimed Sauter AG's guidelines for liquid cycle:
| Series --<br>.<br>Comments of the count | Model Model<br>i single goor<br>double door | Chamber Size ==== bottle<br>(HxWxD) i ----- | Sterilizer Volume of Liquid in one Max. number of<br>17 15 4 5 4 5 4 3 4 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1<br>1 - 1 - 1 7 - 1 1 3 6 - 1 - 1 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2<br>"+ }<br><br> | bottles<br>r - 1 - 1 |
|-----------------------------------------|---------------------------------------------|---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| | . GR15-6-18 HS1<br>GR15-6-18 HS2 | ، 63"x26"x79" | 1000 ml | 250 |
Table 3: Guidelines for liquid 250°F cycle processing
ASON SIGN-Off)
vision of Anesthesiology, General Holsiger, 2 of 2 Infection Control, Dental Devices
KD72087
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.