Oryum and Ovem Epidermal Deri Prick Test Applicator
Applicant
Allergy & Applicator Depot, LLC
Product Code
SCL · General Hospital
Decision Date
Apr 10, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Indications for Use
Oryum and Ovem Epidermal Skin Prick Test Applicator is used for the percutaneous administration of diagnostic allergenic extracts.
Device Story
Oryum and Ovem Epidermal Skin Prick Test Applicator is a sterile, disposable, multi-head applicator for percutaneous delivery of diagnostic allergenic extracts. Device features 1 to 12 test heads with protruding tines arranged in an asymmetrical design. Operation involves loading allergen extract from a matching dipwell tray onto the tines via capillary action; applying the applicator to the patient's skin surface to create indentations for testing. Used by clinicians in a medical setting to assess hypersensitivity reactions. Device does not pierce the skin. Benefits include standardized, efficient administration of multiple allergen tests simultaneously.
Clinical Evidence
Bench testing only. Biocompatibility testing per ISO 10993-5 (cytotoxicity), ISO 10993-10 (sensitization and irritation), and ISO 10993-11 (acute systemic toxicity). Sterility validated per ISO 11737-2. Performance testing included shipping container validation (ASTM D4169) and internal comparison of allergen collection and delivery efficiency against the predicate device.
Technological Characteristics
Acrylic polymer construction. Multi-head applicator (1-12 heads) with capillary-action tines. Asymmetrical design for dipwell tray alignment. Sterilized via Ethylene Oxide (SAL 10^-6). Dimensions: 1.2-1.6mm prick size, 2.0-2.5cm lancet intervals. Shelf life: 3 years. Packaging: PET plastic sterilization pouch.
Indications for Use
Indicated for percutaneous administration of diagnostic allergenic extracts in individuals suspected of having allergies to perform skin tests for hypersensitivity reactions.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
K172062 — AllergiEnd ST-9 Multiple Skin Test Applicator, AllergiEnd ST-9 Multiple Well Test Tray · Medscience Science Research Group, Inc. · Mar 5, 2018
K003195 — GREERTRACK SKIN TESTING SYSTEM, MODEL GT-IT, GT-IC · Greer Laboratories, Inc. · Dec 13, 2000
K000299 — DERMAPIK II SKIN TEST SYSTEM, DERMAPIK II (SINGLE TEST DEVICE), POLYPIK II (MULTIPLE SKIN TEST DEVICE), DERMAPIK TRAY, D · Biomedex, Inc. · Apr 28, 2000
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 1, 2024
Allergy & Applicator Depot, LLC % Marc Sanchez Contract In-House Counsel and Consultants, LLC (FDA Atty) 53516 Bickett Drive Chapel Hill, North Carolina 27517
Re: K182582
Trade/Device Name: Oryum and Ovem Epidermal Skin Prick Test Applicator Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic single lumen needle Regulatory Class: Class II Product Code: SCL
Dear Marc Sanchez:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated April 10, 2019. Specifically, FDA is updating this SE Letter because FDA has better categorized your device technology under product code SCL.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact David Wolloscheck, OHT3: Office of GastroRenal, OB-Gyn, General Hospital and Urology Devices, 301-796-1480, David.Wolloscheck(@fda.hhs.gov.
Sincerely,
# David Wolloscheck -S
David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices and Human Factors OHT3: Office of GastroRenal, OB-Gyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 10, 2019
Allergy & Applicator Depot, LLC % Marc Sanchez Contract In-House Counsel and Consultants, LLC (FDA Atty) 53516 Bickett Drive Chapel Hill, North Carolina 27517
Re: K182582
Trade/Device Name: Oryum and Ovem Epidermal Skin Prick Test Applicator Regulatory Class: Unclassified Product Code: LDH Dated: March 6, 2019 Received: March 8, 2019
Dear Marc Sanchez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{2}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Geeta K. Pamidimukkala -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K182582
Device Name
Oryum and Ovem Epidermal Skin Prick Test Applicator
Indications for Use (Describe)
Oryum and Ovem Epidermal Skin Prick Test Applicator is used for the percutaneous administration of diagnostic allergenic extracts.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### Allergy & Applicator Depot, LLC
Traditional 510k Submis sion Oryum and OvemEpidermal Skin Prick Test Applicator
### K182582
# 510(K) Summary
The following information is provided as required by 21 CFR 807.92 for the Oryum and Ovem Epidermal Skin Prick Test Applicator 510(k) premarket notification.
| Sponsor: | Allergy & Applicator Depot, LLC<br>5018 Expressway Dr. S.<br>Suite LL5<br>Ronkonkoma, NY 11779<br>Establishment Registration: t/b/d |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | Yılmaz Medikal Mustafa Nazlier<br>Budak Mah. Hasirci Sami Bey Cad. No: 12<br>Sehitkamil, Gaziantep, TURKEY<br>Ph.: +90 342 321 74 77<br>E-mail: info@yilmazmedikal.com<br>Establishment Registration: t/b/d |
| Contact: | Marc C. Sanchez, Esq.<br>Contract In-House Counsel and Consultants, LLC<br>(d/b/a FDA Atty)<br>1717 Pennsylvania Ave. Suite 1025<br>Washington, D.C. 20006<br>Ph: 202.765.4491<br>E-mail: msanchez@fdaatty.com |
| Date of Submission: | August 23, 2018 |
| Proprietary Name: | Oryum and Ovem Epidermal Skin Prick Test Applicator |
| Common Name: | System, Delivery, Allergen And Vaccine |
| Regulation Number: | Pre-Amendment |
| Regulatory Class: | Unclassified |
| Product Code: | LDH |
| Predicate Device(s): | Multi-Test II (K961918) |
{5}------------------------------------------------
# Allergy & Applicator Depot. LLC
Traditional 510k Submission Oryum and OvemEpidermal Skin Prick Test Applicator
Device Description: Oryum and Ovem Epidermal Skin Prick Test Applicator is a sterile, disposable, multiple test head applicator used to administer skin test substances to the surface of the skin.
The Oryum and Ovem Epidermal Skin Prick Test Applicator are used for the conventional percutaneous application of substance directly onto the surface of the skin of diagnostic allergen extracts for performing skin tests for hypersensitivity reactions in individuals suspected of having allergies.
The Oryum and Ovem Epidermal Skin Prick Test Applicator is offered in several configurations with 1 to 12 test heads arranged in an asymmetrical design
Each of the test heads have a "leg." At the tip of each leg is an array of protruding test points (tines). The tines utilize capillary action to hold the allergenic material for percutaneous delivery when the applicator is applied to the skin. This is an industry wide standard design.
The test heads are designed to fit into the matching asymmetrical well design of a dipwell tray.
The applicator loads the allergen from each well in the tray on to each test head. When properly applied to the skin, the applicator will leave visible indentations in the patient's skin corresponding to the test heads of each applicator is not intended to pierce the skin.
#### Intended Use :
Oryum and Ovem Epidermal Skin Prick Test Applicator is used for the percutaneous administration of diagnostic allergenic extracts.
#### Summary of Non-Clinical Testing
Oryum and Ovem Epidermal Skin Prick Test Applicator has completed non-clinical tests to identify the substantial equivalence from the predicate device. The tests include the concerning of the biocompatibility, sterility and performance, which are demonstrated in the table below.
| Biocompatibility | ISO10993-5:2009 | In Vitro Cytotoxicity Test |
|------------------|------------------|----------------------------|
| | ISO10993-10:2010 | Skin Sensitization Test |
| | ISO10993-10:2010 | Skin Irritation Test |
| | ISO10993-11:2017 | Acute systemic toxicity |
Table 1 Summary of Non-Clinical Performance testing
{6}------------------------------------------------
# Allergy & Applicator Depot, LLC
Traditional 510k Submission Oryum and OvemEpidermal Skin Prick Test Applicator
| Sterility | ISO11737-2:2009; ISO 11737-<br>2:1998 Sterilization of Medical<br>Devices | Sterilization of Medical Devices |
|-----------------------------------------------------------|---------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|
| Shelf-Life Validation | ISO 11737-2:2009 | Sterilization of Medical Devices |
| Performance Testing of Shipping<br>Containers and Systems | ASTM D4169 | Standard Practice for Performance<br>Testing of Shipping Containers and<br>Systems |
| Performance comparison with<br>predicate device | Internal method | Comparison of ability to collect<br>sufficient allergen extract from<br>dipwell and deliver to the patient |
### Summary of Technological Similarities/Differences:
The Oryum and Ovem Epidermal Skin Prick Test Applicator and the predicate are both acrylic plastic model applicators use to load allergenic extracts on the points of the applicator and apply to the skin. They share the same intended use and near identical design. Any differences in technology are minor and do not raise new questions of safety or efficacy.
| Parameter/application | Oryum and Ovem<br>Epidermal | Multi-Test II (K961918) |
|---------------------------------------|-------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| Product code | LDH | LDH |
| Indications for use | Percutaneous<br>administration of<br>diagnostic allergenic<br>extracts. | Percutaneous administration<br>of diagnostic allergenic<br>extracts. |
| Applicator for<br>Allergenic Extracts | YES | YES |
| Acrylic Polymer<br>Construction | YES | YES |
| SAL | 10-6 | 10-6 |
| Sterilization Method | Ethylene Oxide<br>(applicator); AAMI<br>Guidelines | Ethylene Oxide (applicator);<br>Gamma Radiation (tray<br>with lid); AAMI Guidelines |
| Use Area | Skin Surface | Skin Surface |
| Prick Size | 1.2-1.6mm | 2.0mm |
| Lancet Intervals | 2.0-2.5cm | 2.0-2.4cm |
| Packaging Method | Sterilization Pouch | Sterilization Pouch |
| Packaging Materials | PET Plastic | PET Plastic |
| Shelf Life | 3 years | 3 years |
Table 2 Technological Characteristics
{7}------------------------------------------------
## Allergy & Applicator Depot, LLC
Traditional 510k Submission Oryum and OvemEpidermal Skin Prick Test Applicator
#### Conclusion
Therefore, taking into consideration Table 1 for Substantial Equivalence, an analysis of safety, indications, intended uses, performance, and technological properties, the Oryum and Ovem Epidermal Skin Prick Test Applicator raises no different questions of safety or effectiveness and has been found to be substantially equivalent to the predicate devices.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.