K112985 · Catheter Connections, Inc. · QBP · Nov 9, 2011 · General Hospital
Device Facts
Record ID
K112985
Device Name
CATHETER CONNECTIONS' DUALCAP DUO
Applicant
Catheter Connections, Inc.
Product Code
QBP · General Hospital
Decision Date
Nov 9, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
DualCap Duo™ is intended for use on Luer access valves. DualCap Duo™ will disinfect and decontaminate the valve and act as a barrier to contamination between IV administration line accesses. DualCap Duo™ will disinfect the connections within five (5) minutes after application and act as a physical barrier to contamination up to ninety-six (96) hours under normal conditions if not removed.
Device Story
DualCap Duo is a single-use, sterile, Luer access valve cap containing 70% isopropyl alcohol. Designed for clinical use on IV administration sets to disinfect valve surfaces and prevent contamination. The device is applied by a clinician to the Luer valve; the alcohol provides disinfection within five minutes; the cap remains in place to serve as a physical barrier for up to 96 hours between line accesses. This passive disinfection and protection mechanism reduces the risk of catheter-related bloodstream infections by maintaining a sterile environment on the valve surface.
Clinical Evidence
Bench testing only. Verification testing performed according to ISO 11137 (sterilization) and ISO 10993 (biocompatibility) protocols demonstrated that the device meets all established acceptance criteria for performance and is substantially equivalent to the predicate.
Technological Characteristics
Single-use Luer access valve cap containing 70% isopropyl alcohol. Materials are latex-free, non-pyrogenic, preservative-free, and DEHP-free. Sterilization via radiation per ISO 11137. Biocompatibility evaluated per ISO 10993.
Indications for Use
Indicated for disinfection of needleless Luer access valves via 5-minute application, providing a physical barrier to contamination for up to 96 hours if left in place.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
K113842 — CATHETER CONNECTIONS' DUALCAP SOLO · Catheter Connections, Inc. · Jan 27, 2012
K123065 — CATHETER CONNECTIONS DUALCAP SOLO · Catheter Connections, Inc. · Dec 14, 2012
K142806 — DualCap · Catheter Connections, Inc. · May 22, 2015
K142399 — DualCap IV Pole Strips(Disinfectant Caps for Male Luers) · Catheter Connections, Inc. · Nov 19, 2014
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left, and the FDA logo on the right. The FDA logo is in blue and includes the acronym "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
Catheter Connections, Inc. Donald D. Solomon President and COO 615 Arapeen Drive, Suite 302a Salt Lake City, Utah 84108
March 11, 2022
Re: K112985
Trade/Device Name: Catheter Connections' Dualcap Duo Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: QBP
Dear Donald D. Solomon:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated November 9, 2011 and the correction letter dated December 14, 2018. Specifically, FDA is updating this SE Letter because FDA has better categorized your device technology under regulation 880.5440.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Payal Patel, OHT3: Office of GastroRenal, Ob-Gyn, General Hospital and Urology Devices, 240-402-6029, Payal.Patel(@fda.hhs.gov.
Sincerely,
FDA
Payal Patel Assistant Director for General Hospital Devices DHT3C: Division of Drug Delivery and General Hospital Devices and Human Factors OHT3: Office of GastroRenal, Ob-Gyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym along with the full name of the agency on the right. The symbol on the left is a stylized representation of a caduceus, a traditional symbol of medicine. To the right of the symbol is a black square containing the acronym "FDA" in white letters, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in bold, sans-serif font.
December 14, 2018
Catheter Connections, Inc. Donald Solomon President and COO 2455 E Parleys Way - Suite 150 Salt Lake City, Utah 84109
Re: K112985
Trade/Device Name: Catheter Connections' Dualcap Duo Regulatory Class: Unclassified Product Code: OBP Dated: September 30, 2011 Received: October 6, 2011
Dear Donald Solomon:
This letter corrects our substantially equivalent letter of November 9, 2011.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration, Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
# K112985 - Donald Solomon
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html: good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
FDA
Tina Kiang
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{3}------------------------------------------------
## Indications For Use
510(k) Number (if known):
Device Name: Catheter Connections' DualCap Duo™
Indications For Use:
When left in place for five (5) minutes, DualCap Duo™ disinfects needleless Luer access valves; thereafter the caps provide a physical barrier to contamination up to ninety-six (96) hours under normal conditions if not removed.
Prescription Use × AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
11/9/11
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K112985
Confidential
{4}------------------------------------------------
NOV - 9 2011
CATHETER CONNECTIONS. INC. 510(k) Premarket Notification Submission: Catheter Connections' DualCap Duo™ Page 20 of 80
#### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
(21 CFR 807.92)
### for Catheter Connections' DualCap Duo™
#### SUBMITTER:
Catheter Connections, Inc. 615 Arapeen Drive, Suite 302a Salt Lake City, UT 84108
## ESTABLISHMENT REGISTRATION NUMBER: 3009141010
CONTACT: Donald D. Solomon, Ph.D. President and COO Telephone: (801) 209-1269 Fax: (888) 862-2693 Email: dsolomon@cathconn.com
### DATE PREPARED:
September 30, 2011
#### MODIFIED DEVICE (Submission Device):
Trade Name: DualCap Duo™ Regulation Number: Unclassified Regulation Classification Name: Pad, Alcohol, Device Disinfectant Regulatory Class: Unclassified Classification Product Code: LKB General Hospital Classification Advisory Panel:
#### SPONSOR'S CLEARED DEVICE - DualCap™ (K093229):
510(k) Holder of CLEARED DEVICE (K093229): Catheter Connections, Inc. Regulation Number: Unclassified Regulation Classification Name: Pad, Alcohol, Device Disinfectant Unclassified Regulatory Class: Classification Product Code: LKB Classification Advisory Panel: General Hospital
Confidential
{5}------------------------------------------------
### DEVICE DESCRIPTION:
The DualCap Duo™ is designed to fit securely on Luer access valves. The cap contains 70% isopropyl alcohol. The product is intended for single-use and is provided sterile, latex free, non-pyrogenic, preservative free and DEHP free.
#### INTENDED USE:
DualCap Duo™ is intended for use on Luer access valves. DualCap Duo™ will disinfect and decontaminate the valve and act as a barrier to contamination between IV administration line accesses.
DualCap Duo™ will disinfect the connections within five (5) minutes after application and act as a physical barrier to contamination up to ninety-six (96) hours under normal conditions if not removed.
#### INDICATIONS FOR USE:
When left in place for five (5) minutes DualCap Duo™ disinfect needleless Luer access valves; thereafter the caps provide a physical barrier to contamination up to ninety-six (96) hours under normal conditions if not removed.
#### TECHNOLOGICAL COMPARISON TO PREDICATE DEVICES:
New device is compared to Marketed Device? Yes. It is compared to legally marketed predicates (Sponsor's Cleared Device).
Does the new device have the same indication statements? Yes.
Do the differences alter the intended therapeutic/diagnostic/etc. effect (i.e. deciding may consider impact on safety and effectiveness)? No, the differences do not alter the intended use of the device.
Does the new device have the same technological characteristics, e.g. design, material, etc.? Yes. The Catheter Connections' DualCap Duo™ is substantially in equivalent design, materials, packaging, sterilization method of operation. The basic fundamental scientific technology of the device has not changed.
Could the new characteristics affect safety or effectiveness? No.
Do the new characteristics raise new types of safety and effectiveness questions? No. There are no new types of safety and effectiveness questions.
Do accepted scientific methods exist for assessing effects of the new characteristics? Yes.
Confidential
{6}------------------------------------------------
Sterilization requirements of ISO 11137: 2006, Sterilization of health care products -Radiation.
Biocompatibility requirements according to of ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing.
These and other standards were used to determine the appropriate methods for evaluating the device's performance.
Are performance data available to assess effects of new characteristics? Yes. Verification testing was performed according to protocols based on the above-referenced guidance document recommendations and additional standards and protocols.
Do performance data demonstrate equivalence? Yes. Performance data gathered demonstrated that the Catheter Connections' DualCap Duo™ is substantially equivalent to the noted predicate (Sponsor's Cleared Device - DualCap™).
## CONCLUSION
The Catheter Connections' DualCap Duo™ will meet all established acceptance criteria for performance testing. This testing demonstrated that the Catheter Connections' DualCap Duo™ is safe and effective for its intended use, and based on FDA's decision tree is substantially equivalent to the above noted Sponsor's Cleared Device (DualCap™ - K093229).
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.