SAFE'N'SOUND PASSIVE DELIVERY SYSTEM

K101233 · Rexam Pharma · MEG · Feb 18, 2011 · General Hospital

Device Facts

Record IDK101233
Device NameSAFE'N'SOUND PASSIVE DELIVERY SYSTEM
ApplicantRexam Pharma
Product CodeMEG · General Hospital
Decision DateFeb 18, 2011
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 880.5860
Device ClassClass 2
AttributesPediatric

Indications for Use

Single use devices that are indicated for use as an accessory with sterile 1 mL long staked needle prefilled ISO Standard glass syringes to aid in the protection of healthcare professionals, patients who self-inject doctor prescribed medications, and individuals that assist selfinjecting patients, from accidental needlesticks. The devices can be used on a wide range of patients including children and adults.

Device Story

Safe'n'Sound Passive Delivery System is a single-use, antineedlestick accessory for 1 mL long staked needle prefilled ISO Standard glass syringes. Device consists of a molded plastic body assembly, sleeve, plunger rod, and spring. During injection, the device is used by healthcare professionals, patients, or caregivers; upon injection completion, the spring-loaded mechanism automatically activates to fully contain the needle, shielding the user from accidental sticks. Device is used in clinical or home settings to enhance safety during medication administration.

Clinical Evidence

Simulated clinical use testing confirmed the device safely and effectively shields needles post-injection. Bench testing confirmed functional performance. Biocompatibility testing met ISO 10993-5 and ISO 10993-10 requirements.

Technological Characteristics

Molded plastic assembly (body, sleeve, plunger rod, spring). Passive spring-actuated needle containment mechanism. Designed for 1 mL long staked needle prefilled ISO Standard glass syringes. Biocompatibility per ISO 10993-5 and ISO 10993-10.

Indications for Use

Indicated for use as an accessory with 1 mL long staked needle prefilled ISO Standard glass syringes to prevent accidental needlesticks. Population includes healthcare professionals, patients (children and adults) self-injecting prescribed medications, and caregivers assisting with injections.

Regulatory Classification

Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Abbreviated 510(k) Safe'n'Sound Passive Delivery System # FEB 1 8 201 ### Section 5: 510(k) Summary Assigned 510(k) number: K101233 Company: Rexam Healthcare 600 Deerfield Parkway Buffalo Grove, IL 60089 Phone: 317-346-5178 Fax: 317-736-9016 - Contact: Jeffrey Burris Date Prepared: March 23, 2010 Proprietary Names: Safe'n'Sound Passive Delivery System Classification Name: Piston syringe accessory 21 CFR 880.5860, Class II, Product Code MEG Classification: K060743 UltraSafe® Passive Delivery System by Predicate Device: Safety Syringes, Inc. Device Description: The Safe'n'Sound Passive Delivery · System is an antineedlestick accessory for use with sterile prefilled ISO Standard glass syringes. It fits 1 mL long staked needle {1}------------------------------------------------ syringes, and consists of a body assembly and a loose plunger rod. Intended Use: Single use devices that are indicated for use as an accessory with sterile 1 mL long staked needle prefilled ISO Standard glass syringes to aid in the protection of healthcare professionals, patients who self-inject doctor prescribed medications, and individuals that assist selfinjecting patients, from accidental needlesticks. The devices can be used on a wide range of patients including children and adults. ## Technological Characteristic Comparison Summary to Predicate Device: The Safe'n'Sound Passive Delivery System is similar to the predicate device in general technological features and principle of operation. Both are molded plastic assemblies consisting of a body, sleeve, plunger rod, and spring that activates upon injection completion to fully contain the Minor differences between the devices in needle. technological features and performance have been demonstrated to be insignificant based upon bench testing and simulated clinical use studies performed. - Performance Testing: Bench testing has been performed on the Safe'n'Sound Passive Delivery System. It confirmed the product functions as intended and is substantially equivalent to the predicate device. Biocompatibility testing has been performed demonstrating that the product meets ISO 10993-5 and ISO 10993-10 requirements. - Clinical Testing: Simulated clinical use testing has been performed. It confirmed that the Safe'n'Sound Passive Delivery System could be used safely and effectively to shield needles inside the protection device after use. Conclusion: Based upon the design, technology, performance, functional testing, and intended use, the Safe'n'Sound Passive Delivery System is substantially equivalent to the predicate device currently marketed under the Food, Drug and Cosmetic Act. The Safe'n'Sound Passive Delivery System raises no new issues of safety or effectiveness. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized symbol that resembles three abstract human figures or waves, stacked on top of each other. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD_20993-0002 Mr. Jeffrey Burris Global Regulatory Affairs Manager Rexam Pharma 600 Deerfield Parkway Buffalo Grove, Illinois 60089 ## . FEB 18 201 Re: K101233 Trade/Device Name: Safe'n'Sound Passive Delivery System Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Accessory Regulatory Class: II Product Code: MEG Dated: February 9, 2011 Received: February 11, 2011 Dear Mr. Burris: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If vour device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2- Mr. Burris Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.goy/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, J. Kosol Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): K101233 Device Name: Safe'n'Sound Passive Delivery System Indications for Use: Single use devices that are indicated for use as an accessory with sterile 1 mL long staked needle prefilled ISO Standard glass syringes to aid in the protection of healthcare professionals, patients who self-inject doctor prescribed medications, and individuals that assist self-injecting patients, from accidental needlesticks. The devices can be used on a wide range of patients including children and adults. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) | Concurrence of CDRH, Office of Device Evaluation (ODE) | |--------------------------------------------------------| | (Division Sign-Off) | | 2/18/14 | Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: k/01233
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...