OMI RETRACTABLE SAFETY SYRINGE

K060109 · Vitalcare Group, Inc. · MEG · May 5, 2006 · General Hospital

Device Facts

Record IDK060109
Device NameOMI RETRACTABLE SAFETY SYRINGE
ApplicantVitalcare Group, Inc.
Product CodeMEG · General Hospital
Decision DateMay 5, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5860
Device ClassClass 2
AttributesTherapeutic

Indications for Use

This device is a safety hypodermic syringe for intramuscular and subcutaneous injection of medication to patients. This device aids in prevention of needlestick injuries. EO sterilized, non-pyrogenic, and latex free.

Device Story

OMI Retractable Safety Syringe is a single-use, EO-sterilized hypodermic syringe for intramuscular/subcutaneous injections. Device functions like a conventional syringe during fluid aspiration and injection. Safety mechanism is integrated: upon full plunger depression, additional force fractures the needle hub web, allowing a coiled spring to retract the needle into the hollow plunger rod. Retraction is confirmed by an audible click and visual separation of the hub. Designed for single-handed operation to minimize needlestick injury risk. Minimal dead space design reduces fluid spatter. Intended for use by healthcare professionals in clinical settings.

Clinical Evidence

Clinical investigation performed in accordance with FDA's 'Supplementary Guidance on Premarket Notifications for Medical Devices with Sharps Injury Prevention Features'. Study compared OMI Retractable Safety Syringe to the predicate device. Results demonstrated the device is clinically acceptable for its intended use.

Technological Characteristics

Materials: Polypropylene (barrel), thermoplastic rubber (gasket), Medipeel film/Ethypel paper (packaging). Sterilization: Ethylene oxide (SAL 10^-6). Sizes: 1, 3, 5, 10 ml. Mechanism: Spring-loaded needle retraction triggered by plunger depression. Standards: Barrel markings conform to ISO 7886-1:1993(E); Biocompatibility conforms to ISO 10993-1.

Indications for Use

Indicated for intramuscular and subcutaneous medication injection in patients; aids in prevention of needlestick injuries.

Regulatory Classification

Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Kobolog MAY - 5 2006 ## EXHIBIT 2 510(k) Summary VitalCare Group, Inc. 8935 N.W. 27th Street Miami, Florida 33172 USA Telephone 305-620-4007 Fax 305-620-5220 Contact: Ramzi Abulhaj, President November 3, 2005 - 1) Identification of the Device: Proprietary-Trade Name: OMI Retractable Safety Syringe Classification Name: Piston Syringe Common/Usual Name: Safety Syringe - 2) Equivalent legally marketed device: SEZ Safety Syringe, K031163 - 3) Indications for Use (intended use) . This device is a safety hypodermic syringe for intramuscular and subcutaneous injection of medication to patients. This device aids in prevention of needlestick injuries. EO sterilized, non-pyrogenic, and latex free. - 4) Description of the Device: The OMI Retractable Safety Syringe works like this: STEP 1 After cap removal, fluid draw-up and aspiration is unchanged from conventional syringes. STEP 2 As the plunger rod is depressed and the plunger stopper makes contact with the needle hub, the elastic membrane conforms to the conical shape of the needle hub. The plunger stopper membrane stretches until the plunger rod is fully depressed and maximum fluid is expelled from the svringe. STEP 3 At the point of full plunger rod depression, the application of the injection force overcomes the strength of the web joining the inner and outer portions of the needle hub. As this web fractures and the outer portion of the needle hub moves forward, the plunger stopper membrane is simultaneously pierced by the conical end of the hub, releasing the inner portion of the hub carrying the needle to be retracted into the hollow stem of the plunger rod by the coiled spring. The activation force is controlled. The extra pressure required at the end of the injection stroke to retract the needle is minimal. The entire procedure designed for singlehanded use, with no change to common habits. STEP 4 The fracturing of the needle hub and the piercing of the plunger stopper allow the needle to be forced into the plunger rod by the coiled spring. The speed of this retraction and the minimal dead space avoids blood or fluid spatter. Retraction is accompanied by an audible "click" and the highly visible separation of the needle hub. The syringe is EO sterilized, non-pyrogenic, and latex free. {1}------------------------------------------------ | Device Name | SEZ safety syringe<br>[K031163] | OMI Retractable Safety Syringe | |---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | This device is a Safety hypodermic<br>syringe for Intramuscular and<br>Subcutaneous injection.<br>This device aids in prevention of<br>needlestick injuries. EO sterilized.<br>non-pyrogenic, and latex free. | Identical | | Principle operation | Activation of safety feature consists<br>of two steps :<br>1) Disassemble needle assembly<br>from the barrel by turning the<br>plunger<br>2) Retract needle into barrel and<br>confine it in the barrel by<br>pushing the plunger forward<br>before disposal | Activation of safety feature consists<br>of the following single step:<br>After injecting the medicine, the<br>plunger is pushed in just a bit harder<br>causing the retraction mechanism to<br>pull the needle permanently inside<br>the barrel. | | Volume (ml/cc) | 3 and 5 | 1, 3, 5 and 10 ml sizes | | Nozzle type | Female conical lock fitting with<br>rotatable internally threaded neck | Needle and hub are integral to the<br>syringe, not separable. | | Barrel Marking | Scale: confirms to ISO7886-<br>1:1993(E) | Identical | | Reuse | Non-reusable | Identical | | Biocompatibility | Conform to ISO 10993-1 | Identical | | Materials | 1 ) Plastic parts : polypropylene<br>(homo type) | Identical | | | 2) Gasket : thermoplastic rubber | Identical | | | 3) Packing film : Medipeel film | Identical | | | 4) Packing paper : Ethypel<br>paper | Identical | | Sterility | Sterilized by ethylene oxide gas<br>SAL = 10⁶ | Identical | - 5) Safety and Effectiveness, comparison to predicate device: - 6) Conclusion: In all material respects, the OMI Retractable Safety Syringe is substantially equivalent to the predicate device. The conclusion is based on biocompatibility testing, clinical testing, compliance with voluntary standards, and comparison to the previous device. A clinical investigation was performed, and test for the comparison between OMI Retractable Safety Syringe and the legally marketed predicate device was performed in accordance with "Supplementary Guidance on Premarket Notifications for Medical Devices with Sharps Injury Prevention Features; Guidance for Industry and FDA". The results of the investigation showed that the OMI Retractable Safety Syringe is clinically acceptable. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of a bird-like figure, possibly representing an eagle, with its wings spread. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAY - 5 2006 VitalCare Group, Incorporated C/O Mr. Daniel Kamm, P.E. Principal Engineer Kamm & Associates P.O. Box 7007 Deerfield, Illinois 60015 Re: K060109 Trade/Device Name: OMI Retractable Safety Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: MEG Dated: January 11, 2006 Received: February 8, 2006 Dear Mr. Kamm: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 8010 tartoon. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use 510(k) Number (if known): | 060 i 09 Device Name: OMI Retractable Safety Syringe Indications For Use: This device is a safety hypodermic syringe for intramuscular and subcutaneous injection of medication to patients. This device aids in prevention of needlestick injuries. EQ sterliized non-pyrogenic, and latex free. Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Quro Ls k) Number: trion Sign-Off) ( ്ട്രിാന of Anesthoundential D Page 1 of 1
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...