DIB INFUSOR-RA
Device Facts
| Record ID | K960318 |
|---|---|
| Device Name | DIB INFUSOR-RA |
| Applicant | Novacon Corp. |
| Product Code | MEB · General Hospital |
| Decision Date | Aug 5, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 880.5725 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
The intended use of the DIB-RA Infusor is similar to that of other legally marketed elastomeric infusion pumps.
Device Story
DIB-RA Infusor; elastomeric infusion pump; utilizes medical grade silicone shaft and silicone balloon. Device operates via internal pressure to deliver medication; balloon tested to 110±5 mmHg; clinical operating pressure ~70 mmHg. Used in clinical settings for pain management and IV infusions. Healthcare providers manage infusion; device provides continuous delivery to benefit patients requiring sustained medication administration.
Clinical Evidence
Evidence includes in-vitro flow studies and clinical literature. In-vitro testing conducted by University of Toronto and University of Utah. Clinical experience cited from international studies (Japan) covering continuous epidural block, postoperative pain relief, and cervical epidural anesthesia for Ramsey Hunt Syndrome.
Technological Characteristics
Materials: medical grade silicone shaft and balloon. Sensing/Actuation: elastomeric balloon pressure (110±5 mmHg test pressure; ~70 mmHg clinical operating pressure). Standards: AAMI Draft Infusion Device Standard. Manufacturing: Class 10,000 clean room, GMP compliant.
Indications for Use
Indicated for patients requiring continuous infusion of medications, including applications for obstetrical analgesia, post-operative pain management, chronic pain management, and intravenous infusions.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Reference Devices
- DIB Catheter (K930404)
Related Devices
- K955849 — DIB INFUSOR · Novacon Corp. · Aug 5, 1996