The ACCU-CHEK Combo System is indicated for the treatment of insulin-requiring diabetes and for the quantitative measurement of glucose in fresh capillary whole blood from the finger. The ACCU-CHEK Spirit Combo Insulin Pump is intended for the subcutaneous continuous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin as prescribed by a physician. The ACCU-CHEK 360° Insulin Pump Configuration software facilitates the monitoring and programming of the pump settings. The ACCU-CHEK Aviva Combo blood glucose system is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitoring the effectiveness of diabetes control. The ACCU-CHEK Aviva Combo blood glucose monitoring system should not be used for the diagnosis or screening of diabetes or for neonatal use. Alternative site testing should NOT be used with the ACCU-CHEK Aviva Combo blood glucose monitoring system. The ACCU-CHEK Aviva Plus test strips are for use with the ACCU-CHEK Aviva Combo blood glucose meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. The ACCU-CHEK Aviva Combo meter can also be used to interface with, and remotely control, the ACCU-CHEK Spirit Combo insulin infusion pumps via radio frequency communication. The ACCU-CHEK Aviva Combo meter is also indicated for the management of diabetes by calculating an insulin dose or carbohydrate intake based on user-entered data. For in vitro diagnostic use. The ACCU-CHEK Aviva Combo system is intended for single patient use and should not be shared.
Device Story
System comprises ACCU-CHEK Spirit Combo insulin pump and ACCU-CHEK Aviva Combo blood glucose meter. Pump delivers subcutaneous insulin at set/variable rates via syringe mechanism (brushless DC motor, piston rod). Meter measures glucose via electrochemical reaction on test strips; provides results in 5 seconds from 0.6 µL capillary blood. Meter interfaces with pump via radio frequency (RF) to enable remote control and insulin dose/carbohydrate intake calculations based on user-entered data. Used by patients at home for diabetes management. Healthcare providers use pump configuration software for programming. System benefits include integrated glucose monitoring and insulin delivery control, facilitating diabetes management.
Clinical Evidence
Laboratory and human factors studies conducted. No clinical trial data provided. Bench testing verified delivery accuracy (±5% at 1.0 I.U./h; ±30% at 0.05 I.U./h) and occlusion detection performance.
Technological Characteristics
External portable insulin pump with syringe mechanism (brushless DC motor, piston rod). Powered by AA alkaline, lithium, or NiMH battery. Glucose meter uses electrochemical biosensor technology. Connectivity via RF for pump-meter communication. Software used for pump configuration.
Indications for Use
Indicated for insulin-requiring diabetes patients for subcutaneous insulin delivery and quantitative capillary blood glucose measurement. Contraindicated for neonatal use, diabetes screening/diagnosis, and alternative site testing.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
K080639 — SYMPHONY GLUCOSE MANAGEMENT SYSTEM · Animas Corp. · Jun 24, 2008
K131029 — ACCU-CHEK AVIVA PLUS TEST STRIP · Roche Diagnostics Operations, Inc. · Aug 9, 2013
K082590 — SYMPHONY METER REMOTE · Lifescan, Inc. · Dec 17, 2008
K110587 — CONTOUR LINK BLOOD GLUCOSE METER, CONTOUR BLOOD GLUCOSE TEST STRIPS, CONTOUR CONTROL SOLUTION (HIGH, NORMAL, LOW) · Bayer Healthcare, LLC · Mar 28, 2012
Submission Summary (Full Text)
{0}------------------------------------------------
K111353
### JUL 1 7 2012 510(k) Summary of Safety and Effectiveness
| Introduction | According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence. |
|-------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitted By | Roche Diabetes Care, AG<br>Kirchbergstrass 190, Postfach<br>CH-3401 Burgdorf, Switzerland |
| | USA Contact Person:<br>Scott Thiel<br>Anson Group<br>9001 Wesleyan Road, Suite 200<br>Indianapolis, IN 46268 |
| Device Name | Proprietary name: ACCU-CHEK Combo System<br>Common name: Insulin infusion pump, glucose test system, and accessories<br>Classification name:<br><ul><li>Pump, infusion, insulin (21 CFR 880.5725); Class II; Product Code: LZG,</li><li>Glucose test system (21 CFR 862.1345); Class II; Product Code: NBW,</li><li>Drug dosing calculator (21 CFR 868.1890); Class II; Product Code: NDC</li></ul> |
| Predicate Devices | ACCU-CHEK Spirit Insulin Pump - K060876<br>ACCU-CHEK Aviva Plus blood glucose monitoring system - K101299<br>ACCU-CHEK Pocket Compass Diabetes Management Software - K082595 |
Continued on next page
政
. *
:
ון י
{1}------------------------------------------------
Device Description The ACCU-CHEK Combo System includes:
- An ACCU-CHEK Spirit Combo insulin infusion pump and optional . accessory software (ACCU-CHEK 360° Insulin Pump Configuration software), and
- . An ACCU-CHEK Aviva Combo blood glucose monitor. Both devices utilize disposables that are already legally marketed.
The ACCU-CHEK Spirit Combo utilizes disposables that are already legally marketed:
- ACCU-CHEK FlexLink Plus .
- ACCU-CHEK Ultraflex ●
- ACCU-CHEK Tender .
- ACCU-CHEK Rapid-D .
- ACCU-CHEK 3.15 ml Plastic Cartridge .
The ACCU-CHEK Aviva Combo utilizes disposables that are already legally marketed:
- ACCU-CHEK Aviva Plus test strips .
- · ACCU-CHEK Aviva control solutions
The ACCU-CHEK Combo System is meant only for single-patient use; the device may not be used for screening or diagnosis of diabetes or for multiplepatient use.
Continued on next page
{2}------------------------------------------------
The ACCU-CHEK Combo System is indicated for the treatment of insulin-Intended Use requiring diabetes and for the quantitative measurement of glucose in fresh capillary whole blood from the finger.
> The ACCU-CHEK Spirit Combo Insulin Pump is intended for the subcutaneous continuous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin as prescribed by a physician. The ACCU-CHEK 360° Insulin Pump Configuration software facilitates the monitoring and programming of the pump settings.
> The ACCU-CHEK Aviva Combo blood glucose system is intended for selftesting outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitoring the effectiveness of diabetes control. The ACCU-CHEK Aviva Combo blood glucose monitoring system should not be used for the diagnosis or screening of diabetes or for neonatal use. Alternative site testing should NOT be used with the
> ACCU-CHEK Aviva Combo blood glucose monitoring system. The ACCU-CHEK Aviva Plus test strips are for use with the ACCU-CHEK Aviva Combo blood glucose meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.
The ACCU-CHEK Aviva Combo meter can also be used to interface with, and remotely control, the
ACCU-CHEK Spirit Combo insulin infusion pumps via radio frequency communication. The ACCUCHEK Aviva Combo meter is also indicated for the management of diabetes by calculating an insulin dose or carbohydrate intake based on user-entered data.
For in vitro diagnostic use
The ACCU-CHEK Aviva Combo system is intended for single patient use and should not be shared.
Continued on next page
{3}------------------------------------------------
#### Technical Characteristics
The ACCU-CHEK Spirit Combo Insulin Pump is an external, portable insulin pump designed for continuous delivery of insulin. The design allows the delivery of 0.05 to 25.0 units of U100 insulin per hour in basal rates and 0.1 up to 25.0 units of U100 insulin per meal or meal bolus. Factory settings default to 0.05 to 25.0 units of U100 insulin per hour in basal rates and 0.1 up to 25.0 units of U100 insulin per meal or meal bolus. ACCU-CHEK 360° Insulin Pump Configuration Software is needed to adjust the hourly basal rates in the basal rate profiles that are set to zero by the factory.
The physical method used to accomplish the pumping mechanism action is a state-of-the-art syringe mechanism.
This mechanism consists of:
- Brushless DC motor to drive the piston rod .
- Telescoping piston rod connected with a thread to the cartridge plunger . (plunger retention) to prevent free flow and siphoning
- Force sensor to detect occlusions in the fluid path .
The ACCU-CHEK Spirit Combo Insulin pump utilizes a single AA alkaline or lithium battery, or rechargeable NiMH accumulator.
The ACCU-CHEK Spirit Combo Insulin Pump accuracy has been verified for the following delivery rates and operational conditions.
| Area | Description |
|----------------------------------|-----------------------------------------------------------------------|
| Delivery Rates: | 1.0 I.U./h Accuracy ± 5% |
| | 0.05 I.U./h Accuracy ± 30% |
| Occlusion bolus volume: U100 | Maximum Occlusion Volume with a<br>feed rate of 1.0 IU/h”<br>• 1.5 IU |
| Occlusion bolus delay time: U100 | • 2 h with 1.0 IU/h<br>• 24 h with 0.05 IU/h |
The ACCU-CHEK Aviva Combo meter was developed to utilize the ACCUCHEK Aviva Plus test system's technology and performance characteristics. These test strips are calibrated to deliver plasmalike results. Blood from your fingertip reacts with the chemicals in the test strip to create a harmless electrical current in the test strip. The meter reads the current and gives you the blood glucose result. The ACCU-CHEK Aviva Combo meter and ACCU-CHEK Aviva Plus test strip are designed to be simple and easy to use. It provides accurate blood glucose test results in 5 second while requiring a very small (0.6 µ) sample.
Continued on next page
{4}------------------------------------------------
Laboratory and Human Factors studies demonstrate that the ACCU-CHEK® Performance Summary Combo System performed in an equivalent manner to the predicate devices and is suitable for its intended use.
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle with three stripes representing the three levels of government: federal, state, and local. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Roche Diagnostics C/O Mr. Scott Thiel Senior Regulatory Consultant Anson Group 9001 Wesleyan Road, Suite 200 Indianapolis. Indiana 46268
JUL 1 7 .2012
Re: K111353
Trade/Device Name: ACCU-CHEK® Combo System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: LZG, NBW & NDC Dated: July 9, 2012 Received: July 10, 2012
Dear Mr. Thiel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
#### Page 2 - Mr. Thiel
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements modi that I Dr mas made statutes and regulations administered by other Federal agencies. or the Act of any I ederal the Act's requirements, including, but not limited to: registration 1 od intest compty with the mart 807); labeling (21 CFR Part 801); medical device reporting and houng (21 er readical device-related adverse events) (21 CFR 803); good manufacturing (reporting of measur as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), nlease go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Touriner of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Fw. D. Stetson
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{7}------------------------------------------------
### Indications for Use
510(k) Number (if known): _
Device Name: ACCU-CHEK® Combo System
Indications for Use:
The ACCU-CHEK Combo System is indicated for the treatment of insulin-requiring diabetes and for the quantitative measurement of glucose in fresh capillary whole blood from the finger.
The ACCU-CHEK Spirit Combo Insulin Pump is intended for the subcutaneous continuous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin as prescribed by a physician. The ACCU-CHEK 360° Insulin Pump Configuration software facilitates the monitoring and programming of the pump settings.
The ACCU-CHEK Aviva Combo blood glucose system is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitoring the effectiveness of diabetes control. The ACCU-CHEK Aviva Combo blood glucose monitoring system should not be used for the diagnosis or screening of diabetes or for neonatal use. Alternative site testing should NOT be used with the ACCU-CHEK Aviva Combo blood glucose monitoring system. The ACCU-CHEK Aviva Plus test strips are for use with the ACCU-CHEK Aviva Combo blood glucose meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.
The ACCU-CHEK Aviva Combo meter can also be used to interface with, and remotely control, the ACCU-CHEK Spirit Combo insulin infusion pumps via radio frequency communication. The ACCU-CHEK Aviva Combo meter is also indicated for the management of diabetes by calculating an insulin dose or carbohydrate intake based on user-entered data.
For in vitro diagnostic use
The ACCU-CHEK Aviva Combo system is intended for single patient use and should not be shared.
Prescription Use XX (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Lola for RSE Hielm
Page 1 of 1
Division Sign-Uff) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K11130
VOL 007 page000003
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.